Abstract
The 90° pulse calibration is essential in NMR spectroscopy to prevent artefacts in the liquid state or to enhance cross-polarization efficiency in the solid state. We verified pulse-angle (PA) errors due to circuit impedances in solid-state NMR and suggested a possible solution to prevent the inconvenience of PA errors. The classic pulse sequences used to calibrate 1H 90° pulse lengths by direct detection of protons or by cross-polarization were modified in order to replace single 1H pulses with 1H pulse trains. Pulse trains were found to decrease the effect of PA imperfections in the calibration of basic pulses (i.e. 90° and 180°) for a number of organic substrates. The modified sequences are especially important to rapidly obtain pulse calibration of complex and heterogeneous molecular systems such as humic substances, which usually require a long time when using single 1H pulses.
Similar content being viewed by others
References
Duer MJ (ed) (2002) Solid-state NMR spectroscopy: principles and applications. Blackwell, Oxford
Taylor RE (2004) Concepts Magn Reson Part A 22:37–49
Goez M (1992) Magn Reson Chem 30:958–961
Berger S, Braun S (eds) (2004) 200 and more NMR experiments. Wiley-VCH, Weinheim
Piccolo A, Conte P, Spaccini R, Mbagwu JSC (2005) Eur J Soil Sci 56:343–352
Conte P, Spaccini R, Piccolo A (2004) Prog Nucl Magn Reson Spectrosc 44:215–223
Brown DE (1995) J Magn Reson 114:268–270
Hafner S, Spiess HW (1997) Solid State Nucl Magn Reson 8:17–24
Lewitt MW, Lowe AJ (1997) Thermochim Acta 294:13–21
Hartmann SR, Hahn EI (1962) Phys Rev 128:2042–2053
Wilson MA (1987) NMR techniques and applications in geochemistry and soil chemistry, 1st edn. Pergamon, Oxford
Acknowledgements
The authors gratefully acknowledge a fruitful discussion with Francesca Benevelli from Bruker Biospin, Italy, and the help of Daniela Šmejkalová (Università di Napoli Federico II) in evaluation of spectra.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Conte, P., Piccolo, A. Precise measurement of 1H 90° pulse in solid-state NMR spectroscopy for complex and heterogeneous molecular systems. Anal Bioanal Chem 387, 2903–2909 (2007). https://doi.org/10.1007/s00216-007-1156-8
Received:
Revised:
Accepted:
Published:
Issue Date: