Certification of a new selenized yeast reference material (SELM-1) for methionine, selenomethinone and total selenium content and its use in an intercomparison exercise for quantifying these analytes

  • Original Paper
  • Published:
Analytical and Bioanalytical Chemistry Aims and scope Submit manuscript

Abstract

A new selenized yeast reference material (SELM-1) produced by the Institute for National Measurement Standards, National Research Council of Canada (INMS, NRC) certified for total selenium (2,059±64 mg kg−1), methionine (Met, 5,758±277 mg kg−1) and selenomethionine (SeMet, 3,431±157 mg kg−1) content is described. The ±value represents an expanded uncertainty with a coverage factor of 2. SeMet and Met amount contents were established following a methanesulfonic acid digestion of the yeast using GC-MS and LC-MS quantitation. Isotope dilution (ID) calibration was used for both compounds, using 13C-labelled SeMet and Met. Total Se was determined after complete microwave acid digestion based on ID ICP-MS using a 82Se spike or ICP-OES spectrometry using external calibration. An international intercomparison exercise was piloted by NRC to assess the state-of-the-art of measurement of selenomethione in SELM-1. Determination of total Se and methionine was also attempted. Seven laboratories submitted results (2 National Metrology Institutes (NMIs) and 5 university/government laboratories). For SeMet, ten independent mean values were generated. Various acid digestion and enzymatic procedures followed by LC ICP-MS, LC AFS or GC-MS quantitation were used. Four values were based on species-specific ID calibration, one on non-species-specific ID with the remainder using standard addition (SA) or external calibration (EC). For total selenium, laboratories employed various acid digestion procedures followed by ICP-MS, AFS or GC-MS quantitation. Four laboratories employed ID calibration, the remaining used SA or EC. A total of seven independent results were submitted. Results for methionine were reported by only three laboratories, all of which used various acid digestion protocols combined with determination by GC-MS and LC UV. The majority of participants submitted values within the certified range for SeMet and total Se, whereas the intercomparison was judged unsuccessful for Met because only two external laboratories provided values, both of which were outside the certified range.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Yang L, Sturgeon RE, McSheehy S, Mester Z (2004) J Chromatogr A 1055:177–184

    Article  CAS  Google Scholar 

  2. Nelms SM, Quetel CR, Prohaska T, Vogl J, Taylor PDP (2001) J Anal At Spectrom 16:333–338

    Article  CAS  Google Scholar 

  3. Wrobel K, Kannamkumarath SS, Caruso JA (2003) Anal Bioanal Chem 375:133–138

    CAS  Google Scholar 

  4. Haberhauer Troyer C, Alvarez Llamas G, Zitting E, Rodriguez Gonzalez P, Rosenberg E, Sanz Medel A (2003) J Chromatogr A 1015:1–10

    Article  CAS  Google Scholar 

  5. Yang L, Mester Z, Sturgeon RE (2004) Anal Chem 76:5149–5156

    Article  CAS  Google Scholar 

  6. McSheehy S, Yang L, Sturgeon R, Mester Z (2005) Anal Chem 77:344–349

    Article  CAS  Google Scholar 

  7. Yang L, Bancon Montigny C, Mester Z, Sturgeon RE, Willie SN, Boyko VJ (2003) Anal Bioanal Chem 376:85–91

    CAS  Google Scholar 

  8. Bodo ET, Stefanka Z, Ipolyi I, Soros C, Dernovics M, Fodor P (2003) Anal Bioanal Chem 377:32–38

    Article  CAS  Google Scholar 

  9. Infante HG, G O C, Rayman M, Wahlen R, Entwisle J, Norris P, Hearn R, Catterick T (2004) J Anal At Spectrom 19:1529–1538

    Article  Google Scholar 

  10. Polatajko A, Sliwka Kaszynska M, Dernovics M, Ruzik R, Encinar JR, Szpunar J (2004) J Anal At Spectrom 19:114–120

    Article  CAS  Google Scholar 

  11. Reyes LH, Sanz FM, Espilez PH, Marchante Gayon JM, Alonso JIG, Sanz Medel A (2004) J Anal At Spectrom 19:1230–1235

    Article  Google Scholar 

  12. McSheehy S, Yang L, Mester Z (2006) Microchimica Acta, http://www.springerlink.com/openurl.asp?genre=article&id=doi:10.1007/s00604-006-0520-2

  13. Ellison SLR, Burke S, Walker RF, Heydorn K, Mansson M, Pauwels J, Wegscheider W, Nijenhuis BT (2001) Accred Qual Assur 6:274–277

    Article  Google Scholar 

  14. http://inms-ienm.nrc-cnrc.gc.ca/en/calserv/crm_files_e/SELM1_certificate.pdf(2005)

Download references

Acknowledgements

The NRC thanks V. Boyko, M. McCooeye and V. Clancy for preparation of the yeast standard and G. Gardner and J. Lam for the Karl Fischer measurements. The NRC is grateful to Institute Rosell-Lallemand for supporting this research and providing the yeast sample used in this study.

BCU thank the Department of Biochemistry and Food Technology at BUTE for the amino acid analysis.

LGC acknowledges support under contract by the UK Department of Trade and Industry as part of the National Measurement System Valid Analytical Measurement (VAM) program.

OVIE acknowledges support from project BQU-2003-04671 and thanks Dr. Vanessa Díaz Huerta for her instrumental work.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Zoltán Mester.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Mester, Z., Willie, S., Yang, L. et al. Certification of a new selenized yeast reference material (SELM-1) for methionine, selenomethinone and total selenium content and its use in an intercomparison exercise for quantifying these analytes. Anal Bioanal Chem 385, 168–180 (2006). https://doi.org/10.1007/s00216-006-0338-0

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00216-006-0338-0

Keywords

Navigation