Abstract.
A recently described thermospray flame furnace atomic absorption spectrometric (TS-FF-AAS) system has been modified in order to extend the applicability of the method for the determination of elemental traces in very small sample volumes (microliter or submicroliter). As an easily available, effective thermospray vaporizer, a fused silica capillary was used and the liquid sample was transported by 1 MPa (10 bar) gas pressure delivered by a standard gas cylinder. A 0.3 µL sample volume can be analyzed with a higher power of detection than using 3 orders of magnitude larger sample volumes with conventional flame atomic absorption spectrometry. The relative standard deviations (N=12) for 0.3 µL volumes and 5 µg/mL Pb samples amount to 3.1% and 3.8% in signal height and signal area, respectively. The detection limit was found to be 69 ng/mL. Initial experiments with other elements (Cd, Hg, Tl, Zn) led to similar results.
Similar content being viewed by others
Author information
Authors and Affiliations
Additional information
Electronic Publication
Rights and permissions
About this article
Cite this article
Gáspár, A., Széles, É. & Berndt, H. Analysis of submicroliter samples using micro thermospray flame furnace atomic absorption spectrometry. Anal Bioanal Chem 372, 136–140 (2002). https://doi.org/10.1007/s00216-001-1136-3