To the Editor: In early studies, the use of continuous subcutaneous insulin infusion (CSII) was associated with an increased rate of diabetic ketoacidosis, caused mainly by infusion system malfunction [1–3]. However, these findings were not confirmed in most subsequent studies [4, 5]. It has been suggested that technological improvements and/or patient and physician vigilance has lowered this risk. Nevertheless, the reliability of pumps has not been re-evaluated during the last two decades.
The aim of this study was to determine the failure rate of pumps from four manufacturers and to assess the clinical consequences in type 1 diabetic patients using new generation insulin pumps.
Between April 2001 and November 2007, we prospectively studied insulin pump failures with 640 consecutive new pumps provided and maintained by the same institution (AIR Bretagne, Rennes, France). The pumps were used in ambulatory CSII treatment by 252 adult type 1 diabetic patients (mean age 40.7 ± 13.1 years; 169 women, 83 men). Patients were given a choice between pumps manufactured by MiniMed Medtronic (MiniMed508, Paradigm) (Northridge, CA, USA) and Roche Disetronic (H-Tron plus, Accu-Chek D-Tronplus, Accu-Chek spirit) (Burgdorf, Switzerland) in 2001 to 2007, and by Smiths Medical (Deltec Cozmo) (St Paul, MN, USA) and Animas (Animas IR 1100, Animas IR 1200) (West Chester, PA, USA) after 2003. The same insulin pump was used by several successive patients once it had been checked and was working correctly. CSII treatment was initiated in an inpatient setting with a specific education programme to prevent and enable correction of CSII malfunctions. A professional healthcare service (on-call 24 h per day) was available to assist pump users. Patients’ knowledge was checked by at least four contacts per year and with an annual treatment evaluation in the initiating centre, according to French legislation [6].
We focus on malfunctions directly related to the pump per se and not related to shock or incorrect use, which were treated as censored observations. All pump failures were documented and managed by AIR Bretagne; patients could not contact the manufacturer to exchange the pump directly. Clinical consequences of pump malfunctions were registered: hyperglycaemia (≥13.2 mmol/l), presence of ketones, hypoglycaemia and the need for hospitalisation.
During the study, 232 pumps (36%) broke down after a median time of 15.1 (range 0.1–64.4) months. The rate of pump malfunction was 25 per 100 pump-years. The characteristics of the pumps are shown in Electronic supplementary material [ESM] Table 1. There were 21 minor defects, 103 complete failures (pumps immediately unusable), 66 alarms set off and 42 mechanical malfunctions that required pump replacement, as recommended by the manufacturer (Table 1).
In total, 97 women (57%) and 45 men (54%) experienced a pump failure (χ 2, p = 0.63). In univariate analysis, the total duration of CSII during the follow-up (median 58, range 2–79 months) was significantly associated with an increased risk of pump failure (OR 1.04, 95% CI 1.03–1.05; p < 0.0001). The age of patients was not related to the risk of pump failure, after adjusting for the duration of CSII (OR 1.00, 95% CI 0.98–1.03; p = 0.77).
The pump Kaplan–Meier survival curves (pump malfunction-free) for the four pump manufacturers were significantly different (logrank test, p < 0.0001; Fig. 1). As the survival curves did not differ for the models of MiniMed Medtronic, Roche Disetronic and Smiths Medical (ESM Fig. 1), we regrouped the pump models for each of these manufacturers. For Animas, the survival curves differed significantly, and we used only the pump with the better survival curve (Animas IR 1200). Using Cox proportional hazards models to compare the failure rates for manufacturers, MiniMed Medtronic pumps (n = 380) had a lower risk of pump failure than Roche Disetronic (n = 163) (HR 0.59, 95% CI 0.40–0.88; p = 0.009), Smiths Medical (n = 54) (HR 0.40 95% CI 0.27–0.59; p < 0.0001) or Animas IR1200 (n = 31) (HR 0.25, 95% CI 0.14–0.47; p < 0.0001) pumps. Roche Disetronic pumps had a lower risk of failure than Smiths Medical (HR 0.52 95% CI 0.32–0.86; p = 0.009) or Animas IR1200 (HR 0.42, 95% CI 0.22–0.82; p = 0.01) pumps. There was no difference between Smiths Medical and Animas pumps.
For the 232 pump failures, there were no metabolic consequences in 161 cases, hyperglycaemia in 40 (17%) (with the presence of ketones in ten of these cases), one severe instance of hypoglycaemia, possibly due to an overdose by the pump, and consequences were not determined in 30 cases. No hospitalisation was necessary and no patient discontinued CSII therapy after insulin pump failure.
The main finding of this study is that pump failure still occurs frequently, with substantial differences between manufacturers. This study is the first to provide data on the rate of pump failure for new generation devices. The majority of reports on technical features or difficulties with CSII were published before 1990 [1, 2, 7].
Importantly, our findings do not suggest that pump malfunctions are less frequent or even less severe than those observed in early reports: we recorded substantial rates of severe pump failures, with a complete failure in 44% of cases. However, the majority of patients did not experience serious adverse metabolic consequences. We emphasise the importance of patient education and the 24 h on-call service to limit the consequences of pump failures. The advanced security features of modern pumps probably also play a role.
The strengths of the study are the large number of insulin pumps, the prospective follow-up and the fact that all pump users received the same technical education, according to French legislation [6]. The main limitation of our study is the absence of randomised allocation of the different pumps.
These findings show that there are still important technical issues with new generation insulin pumps, more than 25 years after the introduction of CSII. Randomised studies are now needed to further compare the reliability of different models of pumps.
Abbreviations
- CSII:
-
Continuous subcutaneous insulin infusion
References
Peden NR, Braaten JT, McKendry JB (1984) Diabetic ketoacidosis during long-term treatment with continuous subcutaneous insulin infusion. Diabetes Care 7:1–5
Mecklenburg RS, Guinn TS, Sannar CA, Blumenstein BA (1986) Malfunction of continuous subcutaneous insulin infusion systems: a one-year prospective study of 127 patients. Diabetes Care 9:351–355
Chantelau E, Spraul M, Muhlhauser I, Gause R, Berger M (1989) Long-term safety, efficacy and side-effects of continuous subcutaneous insulin infusion treatment for type 1 (insulin-dependent) diabetes mellitus: a one centre experience. Diabetologia 32:421–426
Lenhard MJ, Reeves GD (2001) Continuous subcutaneous insulin infusion: a comprehensive review of insulin pump therapy. Arch Intern Med 161:2293–2300
Pickup J, Keen H (2002) Continuous subcutaneous insulin infusion at 25 years: evidence base for the expanding use of insulin pump therapy in type 1 diabetes. Diabetes Care 25:593–598
Hanaire H, Lassmann-Vague V, Jeandidier N et al (2008) Treatment of diabetes mellitus using an external insulin pump: the state of the art. Diabetes Metab 34:401–423
Tschope D, Schumacher M, Gries A (1987) Technical monitoring of miniaturized insulin infusion pumps for increased therapeutical security in the treatment of type-I diabetics. Med Prog Technol 13:47–52
Acknowledgements
We thank V. Derennes, V. Texier-Lethuillier, C. Colmar-Montiel and A. Colobert for their assistance with data collection.
Duality of interest
The authors declare that there is no duality of interest associated with this manuscript.
Author information
Authors and Affiliations
Corresponding author
Electronic supplementary material
Below is the link to the electronic supplementary material.
ESM Table 1
Characteristics of the 640 insulin pumps used with median time before failure (PDF 11 kb)
ESM Fig. 1
Curves (failure-free) for the different models of the four pump manufacturers. a MiniMed Medtronic pumps: model Paradigm (black triangles) vs MiniMed 508 (white circles) (logrank test, p=0.91); (b) Roche Disetronic pumps: model H-Tron plus (white triangles) vs Accu-Chek D-Tronplus (white squares) and vs Accu-Chek spirit (black diamonds) (p=0.37); (c) Smiths Medical pump: Deltec Cozmo; (d) Animas pumps: model IR 1100 (black squares) vs IR 1200 (black circles) (p=0.0006) (PDF 201 kb)
Rights and permissions
About this article
Cite this article
Guilhem, I., Balkau, B., Lecordier, F. et al. Insulin pump failures are still frequent: a prospective study over 6 years from 2001 to 2007. Diabetologia 52, 2662–2664 (2009). https://doi.org/10.1007/s00125-009-1549-7