Abstract
A longitudinal study was conducted to assess the value of quantitative ultrasound (QUS) measurement in predicting the risk of fracture and to evaluate how QUS parameters change with ageing and the climacteric. A group of 211 female subjects underwent assessment by QUS at the distal metaphysis of the first phalanx of the last four fingers of the hand on two occasions 3 years apart. The subjects were selected from outpatients attending the orthopaedic clinic, provided they were not affected by metabolic disease or under treatment with drugs known to interfere with bone metabolism. In vivo the coefficient of variation and the standardized coefficient of variation of the QUS device were respectively 0.5% and 3.5%. The correlation between the values of the amplitude-dependent speed of sound (AD-SoS) in the two measurements wasr=0.92. In 77.3% of the subjects during the observation period we recorded a reduction in AD-SoS. During the study 22 fractures were observed in peripheral sites, 8 of which were associated with ‘low-energy trauma’. By multiple logistic regression analysis we found that the relative risk of fracture for a 1 SD reduction in AD-SoS was 1.5 (95% CI 1.1–1.7) (p<0.03). The percentage of low-energy fractures significantly increased among those subjects with an AD-SoS value lower than 1850 m/s (T-score <−3.5) at the first examination (p<0.0001). QUS investigation proved to be especially sensitive to hormonal changes associated with the climacteric: we observed a mean decrease of 56 m/s in the AD-SoS for women who entered the menopause between the first and the second QUS test (average time since menopause 2 years), as against 10 m/s in subjects remaining premenopausal. In a group of 146 subjects with ‘normal’ AD-SoS at the first examination, we observed a significant reduction in AD-SoS only after 40 years of age. This study demonstrates that measurement of the AD-SoS at the phalanx is reproducible, can be employed to assess the risk of fracture, and is able to detect age-related alterations in bone tissue.
Similar content being viewed by others
References
Langton CM, Palmer SB, Porter RW. The measurement of broadband ultrasound attenuation in cancellous bone. Eng Med 1984;13:89–91.
Glüer CC, Wu CY, Jergas M, Goldstein SA, Genant HK. Three quantitative ultrasound parameters reflect bone structure. Calcif Tissue Int 1994;55:46–52.
Hans D, Arlot ME, Schott AM, Roux JP, Kotzki PO, Meunier PJ. Do ultrasound measurement of the os calcis reflect more the bone architecture than the bone mass? A two dimensional histomorphometric study. Bone 1995;16:295–300.
Cadossi R, Canè V. Pathways of transmission of ultrasound energy through the distal metaphysis of the second phalanx of pigs: an in vitro study. Osteoporos Int 1996;6:196–206.
Baran DT, Kelly AM Karellas A, Gionet M, Price M, Leahey D, et al. Ultrasound attenuation of the os calcis in women with osteoporosis and hip fractures. Calcif Tissue Int 1988;43:138–42.
Heaney RP, Avioli LV, Chestunt CH, Lappe J, Reicher RR, Brandenburger GH. Osteoporotic bone fragility: detection by ultrasound transmission velocity. JAMA 1989;261:2986–90.
Fumck C, Wüster C, Alenfeld FE, Pereira-Lima JFS, Fritz T, Meeder PJ, et al. Ultrasound velocity of the tibia in normal German women and hip fracture patients. Calcif Tissue Int 1996;58:390–4.
Sili Scavalli A, Marini M, Spadaro A, Riccieri V, Cremona A, Zoppini A. Comparison of ultrasound transmission velocity with computed metacarpal radiogrammetry and dual-photon absorptiometry. Eur Radiol 1996;6:192–5.
Wüster C, Pereira-Lima J, Beck C, Götz M, Paetzold W, Brandt K, et al. Quantitative Ultrashall-Densitometrie (QUS) zur Osteoporose-Risiko-Beurteilung: Referenzdaten für verschiedene Messtellen-Grenzen und Einsatzmöglichkeiten. Frauen Artzt 1995;11:1304–13.
Duboeuf F, Hans D, Schott AM, Giraud S, Delmas PD, Meunier PJ. Ultrasound velocity measured at the proximal phalanges: precision and age-related changes in normal females. Rev Rhum 1996;63:427–34.
Hans D, Dargent-Molina P, Schott AM, Sebert JL, Cormier C, Kotski PO, et al. Ultrasonographic heel measurements to predict hip fracture in elderly women: the EPIDOS prospective study. Lancet 1996;348:511–4.
Bauer DC, Glüer CC, Cauley JA, Vogt TM, Ensrud KE, Genant HK, Black DM. Broad-band ultrasound attenuation predicts fractures strongly and independently of densitometry in older women: a prospective study. Arch Intern Med 1997;157:629–34.
Gonnelli S, Cepollaro C, Pondrelli C, Martini S, Rossi S, Gennari C. Ultrasound parameters in osteoporotic patients treated with salmon calcitonin: a longitudinal study. Osteoporos Int 1996;6:303–7.
Paltrinieri F, Ventura V, Mauloni M, Mura M, de Aloysio D. Hormone replacement therapy monitoring by means of quantitative ultrasound studies: preliminary study [abstract]. Osteoporos Int 1996;6(Suppl 1):229.
Mele R, Passari L, Traina GC. Metodica ad ultrasuoni in biomeccanica ortopedica. Min Ort Traum 1987;38:801–5.
Jergas M, Uffmann M, Muller P, Koster O. Ultraschallgeschwindigkeitsmessung zur Diagnose dere postmenopausalen Osteoporose. Fortschr Rontgenstr 1993;158:207–11.
Buckwalter JA, Glimcher MJ, Cooper RR, Recker R. Bone biology. J Bone Joint Surg Am 1995;77:1256–89.
Meema EH, Meindok H. Advantages of peripheral radiogrammetry over dual-photon absorptiometry of the spine in the assessment of prevalence of osteoporotic vertebral fractures in women. J Bone Miner Res 1992;7:897–903.
Kleerekoper M, Nelson DA, Flynn MJ, Pawluska AS, Jacobsen G, Peterson EL. Comparison of radiographic absorptiometry with dual-energy X-ray absorptiometry and quantitative computed tomography in normal older white and black women. J Bone Miner Res 1994;9:1745–9.
Ventura V, Mauloni M, Mura M, Paltrinieri F, de Aloysio D. Ultrasound velocity changes at the proximal phalanxes of the hand in pre-, peri-, and postmenopausal women. Osteoporos Int 1996;6:368–75.
Alenfeld FE, Wüster C, Beck C, Ziegler R. Validity of ultrasound measurement of bone mineral density on the phalanges of the hand. Perth International Bone Meeting, 10–13 February 1995. Perth: Scott Wilson & Ivan Price, 1995:61.
Theiler R, Altermatt M, Görres G, Tyndall A. Ultrasound assessment in patients with and without vertebral fractures [abstract]. Osteoporos Int 1996;6(Suppl 1):179.
Mussolino M, Looker A, Madans J, Edelstein D, Walker R, Lydick E. Epstein R. Phalangeal bone density and hip fracture risk [abstract]. J Bone Miner Res 1995;10:360.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Mele, R., Masci, G., Ventura, V. et al. Three-year longitudinal study with quantitative ultrasound at the hand phalanx in a female population. Osteoporosis Int 7, 550–557 (1997). https://doi.org/10.1007/BF02652561
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02652561