Abstract
A pulsed Doppler ultrasound technique was used for mapping two-dimensional blood velocity profiles in the human ascending aorta during open-heart surgery. An electronic position-sensitive device was constructed and linked to an intraluminal 10 MHz Doppler ultrasound probe. From a plane perpendicular to the central direction of blood flow, velocity mapping was performed covering the entire cross-section of the ascending aorta 6–7 cm above the valve. This method is based on a sequential sampling of velocity from continuously changing locations during a stable haemodynamic period; typically velocity points are recorded from 150–300 beats. Further processing transformed data to suit a previously developed velocity distribution model for normal blood flow in the human ascending aorta, based on multiregression analyses. In this model, the time series of data from consecutive beats were computed into an average two-dimensional profile described through one cardiac cycle. This method allows high spatial resolution (1.5 mm), in addition to the high-frequency response (200 Hz) of the modified ultrasound Doppler meter. Together with the advantage of velocity directionality and minimal time interventions, this makes the method well suited for studies on normal flow conditions as well as flow velocity distribution distal to different heart valve prostheses.
Similar content being viewed by others
References
Angelsen, B. A. J., andKristoffersen, K. (1979): ‘On ultrasonic MTI measurement of velocity profiles in blood flow,’IEEE Trans.,BME-26, 12, pp. 665–671
Gill, R. W. (1979): ‘Performance of the mean frequency Doppler modulator,’Ultrasound Med. Biol.,5, pp. 237–247
Hasenkam, J. M., Westphal, D., Reul, H., Gormsen, J., Giersiepen, M., Stodkilde-Jorgensen, H., andPaulsen, P. K. (1987): ‘Three-dimensional visualization of axial velocity profiles downstream of six different mechanical aortic valve prostheses, measured with a hot-film anemometer in a steady state flow model,’J. Biomech.,20, pp. 353–364
Hasenkam, J. M., Westphal, D., Nygaard, M., Reul, H., Giersiepen, M., andStodkilde-Jorgensen, H. (1988). ‘In vitro stress measurement in the vicinity of six mechanical aoric valves using hot-film anemometry in steady flow’,J. Biomech.,21, pp. 235–247
Hatle, L., andAngelsen, B. A. J. (1982): ‘Doppler ultrasound in cardiology’ (Lea & Febiger, Philadelphia)
Jenni, R., Vieli, A., Ruffmann, K., Krayenbuehl, H. P., andAnliker, M. (1984): ‘A comparison between single gate and multigate ultrasonic Doppler measurements for the assessment of the velocity pattern in the human ascending aorta,’Eur. Heart J.,5, pp. 948–953
Kilner, P. J., Firmin, D. N., Rees, R. S. O., Martinez, J., Pennell, D. J., Mohiaddin, R. H., Underwood, S. R., andLongmore, D. B. (1991): ‘Valve and great vessel stonosis: assessment with MR jet velocity mapping,’Radiol.,178, pp. 229–235
Klipstein, R. H., Firmin, D. N., Underwood, S. R., Rees, R. S. O., andLongmore, D. B. (1987): ‘Blood flow patterns in the human aorta studied by magnetic resonance,’Br. Heart J.,58, pp. 316–323
Lucas, C. L., Keagy, B. B. A., Hsiao, H. S., Johnson T. A., Henry, G. W., andWilcox, B. R. (1984): ‘The velocity profile in the canine ascending aorta and its effect on the accuracy of pulsed Doppler determination of mean blood velocity,’Cardiovasc. Res.,18, pp. 282–293
Mathison, M., Furuse, A., andAsano, K. (1988): ‘Doppler analyses of flow velocity profile at the aortic root,’J. Amer. Coll. Cardiol.,12, pp. 947–954
Mitchell, D. G. (1990): ‘Color doppler imagining: principles, limitations, and artifacts,’Radiolog.,177, pp. 1–10
Nicols, W. W., andO’Rourke, M. F. (1990): ‘McDonal’s blood flow in arteries’ (Edward Arnold, London) 3rd edn., pp. 35–46
Nygaard, H., Giersiepen, M., Hasenkam, J. M., Reul, H., Paulsen, P. K., Rovsing P. E., andWestphal, D. (1992): ‘Two-dimensional color-mapping of turbulent shear stress distribution downstream of two aortic bioprosthetic valuesin vitro,’J. Biomech.,4, pp. 429–440
Paulsen, P. K. (1980): ‘The hot film anemometer—a method for blood velocity determination. I.In vivo comparison with the electromagnetic blood flowmeter,’Eur. Surg. Res.,12, pp. 149–158
Paulsen, P. K., andHasenkam, J. M. (1983): ‘Three-dimensional visualization of velocity profiles in the ascending aorta in dogs, measured with a hot-film anemometer,’J. Biomech.,16, pp. 201–210
Paulsen, P. K., Nygaard, H., Hasenkam, J. M., Gormsen, J., Stodkilde Jorgensen, H., andAlbrechtsen, O. (1988): ‘Analysis of velocity in the ascending aorta in humans: a comparative study among normal aortic valves, S. Jude Medical and Starr-Edwards Silastic Ball values,’Int. J. Artif. Organs.,11, (4), pp. 293–302
Rossvoll, O., Samstad, S. O., Torp, H. G., Linker, D. T., Skjaerpe, T., Angelsen, B. A. J., andHatle, L. (1991): ‘The velocity distribution in the aortic anulus in normal subjects: a quantitative analyses of two-dimensional Doppler flow maps,’J. Amer. Soc. Echocardiogr.,4, pp. 367–378
Samstad, S. O., Torp, H., Matre, K., Rossvoll, O., Segadal, L., andPiene, H. (1990): ‘Instantaneous crosssectional flow velocity profiles: a comparative study of two ultrasound Doppler methods applied to anin vitro pulsatile flow model,’J. Amer. Soc. Echocardiogr., 1990,3, pp. 451–464
Segadal, L., andMatre, K. (1987): ‘Blood velocity distribution in the human ascending aorta,’Circ.,76, pp. 90–100
Segadal, L. (1991): ‘Velocity distribution model for normal blood flow in the human ascending aorta,’Med. Biol. Eng. Comput.,29, pp. 489–492
Vieli, A., Jenni, R., andAnliker, M. (1986): ‘Spatial velocity distribution in the ascending aorta of health humans and cardiac patients,’IEEE Trans.,BME-33, pp. 28–34
Woo, Y.-R., andYoganathan, A. P. (1985): ‘In vitro pulsatile flow velocity and turbulent shear stress measurement in the vicinity of mechanical aortic heart valve prostheses,’Life Suppl. Syst.,3, pp. 283–312
Woo, Y.-R., Sung, H.-W., Williams, F. P., andYoganathan, A. P. (1986): ‘In vitro fluid dynamic characteristics of aortic Bioprostheses: old versus new,’Life Supp. Syst.,4, pp. 63–85
Yang, S. S., Bentivoglio, L G., Maranho, V., andGoldberg, H. (1972): ‘From cardiac catheterization data to hemodynamic parameters’ (F. A. Davis Company, Philadelphia) p. 292
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Hessevik, I., Matre, K., Kvltting, P. et al. Intraluminal recording of cross-sectional blood velocity distribution of human ascending aorta by ultrasound Doppler technique. Med. Biol. Eng. Comput. 32 (Suppl 1), S171–S177 (1994). https://doi.org/10.1007/BF02523344
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02523344