Abstract
An analytical method, SSMS, for the determination of trace impurities in cobalt is described. This element is used as raw material for the production of60Co in Embalse Nuclear Power Station. Tin, aluminum, lead, silicon, molybdenum, calcium, vanadium, boron, chromium, zirconium, tungsten, nickel, magnesium, iron, zinc and copper were determined. Manganese, analyzed by AAS, was used as an internal standard. The measurements were semiquantitative. The accuracy of the results was compared with that of emission spectrography and atomic absorption spectrometry, the agreement was found to be satisfactory.
Similar content being viewed by others
References
R. W. Andrews, Cobalt, H. M. Stationary Office, London, 1962.
G. Glavin, Ch. Kopetskii, R. Nikolaev, G. Sikharulidze,J. Less Common Metals, 26 /1972/ 1.
R. Lukaszew, R. Cretella, CNEA Internal Report, D/Q/QA-92, 1985.
A. Ahearn, Mass Spectrometry Analysis of Solids, Elsevier Publ. Co., 1986, p. 84.
A. Ahearn, Trace Analysis by Mass Spectrometry, Academic Press, 1972, p. 194.
J. Bacon, A. Ure,Analyst, 109 /1984/ 1229.
A. Ure, J. Bacon,Analyst, 103 /1978/ 807.
X. Liu, J. Verlinden, F. Adams,Bull. Soc. Chim. Belg., 95, Nr. 5–6 1986.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Gautier, E., Marrero, J. & Servant, R. Determination of trace impurities in cobalt by spark source mass spectrometry. Journal of Radioanalytical and Nuclear Chemistry Letters 135, 333–340 (1989). https://doi.org/10.1007/BF02163426