Summary
An instrument to track the sedimentation properties of red blood cells quickly and automatically and to evaluate the hematocrit as well as the plasma density of a blood sample in the same operation is described. Erythrocyte-sedimentation basically is due to the density difference between the red blood cells and the plasma and the new method makes use of the mechanical-oscillator-technique, where the resonant frequency of the bending-type oscillations of a U-shaped glass tube is related to the density of the fluid with which the tube is filled. The oscillating U-tube is part of a sedimentation-tube of special dimensions and particular spatial orientation. The tube is tilted at 60°, the tip of the U-tube being raised above the horizontal plane. Filling the sedimentation-tube with anticoagulated blood first permits recording of the density of the homogeneous suspension. Then, after a few minutes, the onset of erythrocyte-sedimentation produces a decrease in recorded density. The final value is related to the density of the supernatant plasma-fluid. Hematocrit, blood and plasma density of the undiluted blood sample as well as the sedimentation-rate of the anticoagulated sample are automatically calculated from the recorded sedimentation-curve. The particular arrangement of the tilted sedimentation-tube facilitates a determination of the maximum sedimentation-rate within 15 min at most.
Similar content being viewed by others
References
Acrivos A, Herbholzheimer E (1979) Enhanced Sedimentation in Settling Tanks with Inclined Walls. J Fluid Mech 92: 435–457
Boycott AE (1920) Sedimentation of blood corpuscles. Nature 105: 532
Bull BS (1981) Clinical and laboratory implications of present ESR methodology. Clin Lab Haematol 3: 283–298
Fåhræus R (1921) The suspension stability of the blood. Acta Med Scand 55: 1–228
Fourcade C, Belleville J, Lenormand D, Descotes J, Jossinet J (1973) Description d'un VS-mètre automatique. Ann Phys Biol Med 7: 29–38
Gabel JC, Drake RE (1984) Plasma Proteins and Protein Osmotic Pressure. In: Staub NC, Taylor AE (eds.) Edema. Raven Press, New York, pp 371–381
Gregersen MI, Rawson RA (1959) Blood Volume. Physiol Rev 39: 307–342
Hinghofer-Szalkay H, Holzer H (1979) The calculation of hemoglobin concentration from blood and plasma densities, measured by the mechanical oscillator technique. J Clin Chem Clin Biochem 17: 613–618
Hinghofer-Szalkay H, Leopold H, Kenner T, Holzer H (1980) On the coefficient of thermal expansion of blood and its constituents. Biomed Technik 25: 151–157
Holzer H, Leopold H, Hinghofer-Szalkay H, Stübchen-Kirchner H, Maurer E (1978) Determination of the serum protein concentration based on the measurement of density by the mechanical oscillator technique. J Clin Chem Clin Biochem 16: 391–395
Kratky O, Leopold H, Stabinger H (1969) Dichtemessung an Flüssigkeiten und Gasen auf 10−6 g/cm3 bei 0.6 cm3 Präparatvolumen. Z Angew Physik 27: 273–277
Kratky O, Leopold H, Stabinger H (1972) Vorrichtung zur Messung der Dichte. AT-Patent No. 331539 Österr. Patentamt, Wien
Kratky O, Leopold H, Stabinger H (1978) Vorrichtung zur Bestimmung der Dichte von Flüssigkeiten und Gasen. AT-Patent No. 356943 Österr. Patentamt Wien
Leopold H, Hinghofer-Szalkay H, Kenner T, Holzer H (1978) Schnelle, gravitationsunabhängige Bestimmung der Sedimentationsrate des Blutes. Biomed Techn 23 (5): 99–103
Linzenmeier G, Rodecourt M (1931) Versuch einer Schnellmethode zur Bestimmung der Blutsenkungsgeschwindigkeit. Zentralbl Gynäkol 11: 736–741
Llory J, Bénézech C (1958) Cinétique de la sédimentation globulaire par procédé photoélectrique. Montpellier médical 53: 131–145
Misiaszek E, Williams R, Stasiw DM, Cerny LC (1977) An Automatic Sedimentimeter. Biorheology 14: 145–149
Moser M (1980) Die Anwendbarkeit von Blut- und Plasmadichtemessungen mittels der Biegeschwingermethode auf Fragen des Flüssigkeitsaustausches in der Mikrozirkulation. PhD thesis, Graz
Moser M, Hinghofer-Szalkay H, Kenner T, Holzer H (1980) Measurement of the colloid osmotic pressure from plasma density by means of the mechanical oscillator technique. J Clin Chem Clin Biochem 18: 233–236
JP-Patent No. 59145964, Denkikagaku Keiki K (1984)
ES-Patent No. 8402938, Lopez FV (1984)
EP-Patent No. 102764, Diesse Diagnostica (1984)
FR-Patent No. 2528578, Université de Rennes (1983)
US-Patent No. 3952579, Konan Electric KK (1976
Prandtl L, Oswatitsch K, Wieghardt K (1984) Führer durch die Strömungslehre. 8th edn. Vieweg, Braunschweig-Wiesbaden, pp 404–407
Ruhenstroth-Bauer G (1966) Mechanismus und Bedeutung der beschleunigten Erythrocytensenkung. Klin Wochenschr 44: 533–539
Starling EH (1896) On the absorption of fluids from the connective tissue spaces. J Physiol (London) 19: 312–326
Westergren A (1921) Studies of the Suspension Stability of the Blood in Pulmonary Tuberculosis. Acta Med Scand 54: 247–281
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Schneditz, D., Kenner, T., Gallasch, E. et al. Quick measurement of hematocrit and erythrocyte sedimentation-rate by means of a density tracking method. Blut 55, 153–163 (1987). https://doi.org/10.1007/BF00320569
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF00320569