Many variables are involved in hydroxyapatite coating of metals by plasma-spray techniques. The authors have investigated the biological response to some of the most relevant variables in a controlled in vivo trial. The bone response in the rabbit towards hydroxyapatite coated cylinders was studied keeping the following variables fixed: (a) crystallinity of coating (greater than 90% and between 70% and 60%); (b) thickness of coating (50 and 100 μm); (c) metallic substrate (titanium alloy and duplex stainless stell). Analysis of the results highlights the importance of defining the crystallinity of the coating to forecast its in vivo behaviour: highly crystalline coating is more stable in time but can give rise to fragmented bulky particles; a less crystalline coating is subject to slow degradation in the long term but facilitates its substitution by newly formed bone. Furthermore, it has been found that no relevant differences can be ascribed to a variation in coating thickness between 50 and 100 μm. It has, also, been observed that there are no differences when duplex stainless steel is used instead of titanium alloy as metallic substrate, confirming that bone responds primarily to the coating.
Similar content being viewed by others
References
J. E. LEMONS, Clin. Orthop. 235 (1988) 220.
R. X. LEGEROS, I. ORLY, M. GREGOIRE, Trans. Soc. Biomater. 10 (1987) 84.
C. GABBI, A. MEROLLI and A. CACCHIOLI, Biomateriali 1/2 (1993) 17.
S. D. COOK, K. A. THOMAS, J. F. KAY, Clin. Orthop. 232 (1988) 225.
K. DE GROOT, R. GEESINK, C. P. KLEIN, J. Biomed. Mater. Res. 21 (1987) 1375.
B. KOCH, J. G. WOLKE and K. DE GROOT, J. Biomed. Mater. Res. 21 (1990)655.
T. G. ABRAHAMS and O. D. CROTHERS, Invest.-Radiol. 27 (1992) 779.
J. A. D'ANTONIO, W. N. CAPELLO, O. D. CROTHERS, W. L. JAFFE and M. T. MANLEY, J. Bone Joint Surg. 74 (1992) 995.
R. G. GEESINK, K. DE GROOT and C. P. KLEIN, Clin. Orthop. 225 (1987) 147.
R. G. GEESINK, Orthopedics 12 (1989) 1239.
R. G. GEESINK, Clin. Orthop. 261 (1990) 39.
K. GOMI, B. LOWENBERG, G. SHAPIRO and J. E. DAVIES, Biomaterials 14 (1993) 91.
N. TAKESHITA, T. AKAGI, M. YAMASAKI, T. OZEKI, T. NOJIMA, Y. HIRAMATSU and N. NAGAI, J. Electron Microsc. 41 (1992) 141.
G. J. WANG, T. E. BOMAN, K. C. CHUNG, Trans. Soc. Biomater. 12 (1989) 20.
S. D. COOK, J. F. KAY, K. A. THOMAS, Int. J. Oral Maxillofac. Implants 2 (1987) 15
J. D. DE BRUIJN in “Bioeeramics 5” (edited by T. Kokubo, Butterworth-Heinemann, London, 1992) p. 291.
J. G. C. WOLKE et al., J. Thermal Spray Technol. 1 (1992) 75.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Tranquilli, P.L., Merolli, A., Palmacci, O. et al. Evaluation of different preparations of plasma-spray hydroxyapatite coating on titanium alloy and duplex stainless steel in the rabbit. J Mater Sci: Mater Med 5, 345–349 (1994). https://doi.org/10.1007/BF00058960
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF00058960