Abstract
The principle of a CQM system in electrosurgical units (ESU) is shown and various such systems are discussed. Theoretical calculations of the current distribution inside the patient’s skin showed the influence of the size and shape of a neutral electrode (NE).
Measurements of the temperature increase under a NE on a suitable test device showed the relation between the change of the CQM impedance for a detached NE, the skin properties, and the temperature increase with a reduced contact area of the NE.
It could be shown that CQM systems that activate an alarm simply when the CQM impedance exceeds a fixed upper limit are not safe in all circumstances.
Systems taking into account the skin properties of the patient by measuring an initial CQM impedance and reacting to a relative impedance increase proved to be safe under all tested circumstances.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2009 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Nessler, N.H., Salchner, M. (2009). CQM - the Unknown Risk in Electrosurgery. In: Dössel, O., Schlegel, W.C. (eds) World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany. IFMBE Proceedings, vol 25/6. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-03906-5_5
Download citation
DOI: https://doi.org/10.1007/978-3-642-03906-5_5
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-03905-8
Online ISBN: 978-3-642-03906-5
eBook Packages: EngineeringEngineering (R0)