Validation of Cone Beam CT Scan for Measurement of Palatal Depth in Study Casts

  • Conference paper
4th Kuala Lumpur International Conference on Biomedical Engineering 2008

Abstract

The purpose of this study was to validate the use of digital dental study casts obtained from Cone beam CT Scan (CBCT) against gold standard; that is traditional dental study cast measured with digital caliper.

Thirty-four study casts of Malaysian Aborigines were selected from Centre of Malaysian Pribumi Studies Databank. Palatal depth of each study casts were measured using Mitutoyo digital caliper. Subsequently, all the study models were scanned using i-CAT CBCT and the palatal depth were measured using i-CAT software. Data were analysed using SPSS version 12.0 where Intraclass correlation and paired T-test were employed.

There was no significant difference in palatal depth between CBCT and Mitutoyo digital caliper (p>0.05). Average measures for Intraclass Correlation is 0.753 indicate that it is consistently good.

Within the limitation of this study palatal depth measurement produced by CBCT was as accurate as the digital caliper. However, more samples and parameters should be added to further substantiate the result of this study.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 259.00
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 329.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Similar content being viewed by others

References

  1. Mitchell L. An introduction to Orthodontics. Third edition. Oxford. Oxford University Press. 2007. 49–60

    Google Scholar 

  2. Asquith J, Gillgrass T, Mossey P. Three-dimensional imaging of orthodontic models: a pilot study. Eur J Orthod. 2007; 29: 517–522

    Article  Google Scholar 

  3. Ferracane JL. Materials in Dentistry. Principles and Applications. Second Edition. Philadelphia. Lippincott Williams & Wilkins. 2001. 173–222

    Google Scholar 

  4. Machen DE. Legal aspects of orthodontic practice: risk management concepts. British Association of Orthodontists Newsletter, 1991. September: 4.

    Google Scholar 

  5. McGuinness NJ, Stephens CD. Storage of orthodontic study models in hospital units in the U.K. 1992. 19:227–232

    Google Scholar 

  6. Champagne M. Reliability of measurements from photocopies of study models. J Clin Orthod. 1992; 26:648–650

    Google Scholar 

  7. Nollet PJPM, Katsaros C, van’ t Hof MA, Bongaarts CA, Semb G, Shaw WC, Kuijpers-Jagtmann AM. Photographs of study casts: an alternative medium for rating dental arch relationships in unilateral cleft lip and palate. Cleft Palate-Craniofac J. 2004; 41:646–650

    Article  Google Scholar 

  8. Schwaninger B, Schmidt RL, Hurst RVV. Holography in dentistry. J Am Dent Assoc. 1977; 95:814–817

    Google Scholar 

  9. Keating PJ, Parker RA, Keane D, Wright L. The holographic storage of study models. Br J Orthod. 1984; 11:119–125

    Google Scholar 

  10. Buschang PH, Ceen RF, Schroeder JN. Holographic storage of dental casts. J Clin Orthod. 1990; 24:308–311.

    Google Scholar 

  11. Ayoub AF, Wray D, Moos KF, Jin J, Niblett TB, Urquhart C, Mourfourth P, Siebert P. A three-dimensional imaging system for archiving dental study casts: a preliminary report. Int J Adult Orthod Orthognathic Surg.1997;12:79–84.

    Google Scholar 

  12. Bell A, Ayoub AF, Siebert P. Assessment of the accuracy of a threedimensional imaging system for archiving dental study models. J Orthod. 2003; 30:219–223

    Article  Google Scholar 

  13. Joffe L. OrthoCAD™: digital models for a digital era. J Orthod. 2004; 31:344–347.

    Article  Google Scholar 

  14. Sukovic P. (2003) Cone beam computed tomography in craniofacial imaging. Orthod. Craniofacial Res. 6(Suppl.1): 31–36

    Article  Google Scholar 

  15. Koudi MS and Patil SP. Dental Materials. Prep Manual For Undergraduates: First Edition. New Delhi. Elsevier. 2007. 49.

    Google Scholar 

  16. Costalos PA, Sarraf K, Cangialosi TJ, Efstratiadis S. Evaluation of the accuracy of digital model analysis for the American Board of Orthodontics objective grading system for dental casts. Am J Orthod Dentofacial Orthop. 2005; 128: 624–629

    Article  Google Scholar 

  17. Santoro M, Galkin S, Teredesai M, Nicolay OF, Cangialosi TJ (2003). Comparison of measurements made on digital and plaster models Am J Orthod Dentofacial Orthop. 124:101–105

    Article  Google Scholar 

  18. Rheude B. Sadowsky PL, Ferriera A., Jacobson A (2005). An evaluation of the use of digital study models in orthodontic diagnosis and treatment planning, Angle Orthod, 75: 300–304

    Google Scholar 

  19. Stevens DR., Flores-Mir C, Nebbe B, Raboud DW, Heo G, Major PW. Validity, reliability, and reproducibility of plaster vs digital study models: Comparison of peer assessment rating and Bolton analysis and their constituent measurements. Am J Orthod Dentofacial Orthop. 2006;129: 794–803

    Article  Google Scholar 

  20. Lundner AS. Comparison of comprehensive cast analysis on plaster and digital study models. Am J Orthod Dentofacial Orthop. 2007; 131: 812–813

    Google Scholar 

  21. Mullen SR, Martin CA, Ngan P, Gladwin M. Accuracy of space analysis with emodels and plaster models. Am J Orthod Dentofacial Orthop. 2007; 132: 346–352

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Zamri Radzi .

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2008 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Radzi, Z. et al. (2008). Validation of Cone Beam CT Scan for Measurement of Palatal Depth in Study Casts. In: Abu Osman, N.A., Ibrahim, F., Wan Abas, W.A.B., Abdul Rahman, H.S., Ting, HN. (eds) 4th Kuala Lumpur International Conference on Biomedical Engineering 2008. IFMBE Proceedings, vol 21. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-540-69139-6_188

Download citation

  • DOI: https://doi.org/10.1007/978-3-540-69139-6_188

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-540-69138-9

  • Online ISBN: 978-3-540-69139-6

  • eBook Packages: EngineeringEngineering (R0)

Publish with us

Policies and ethics