Abstract
Loose-fitting powered air-purifying respirators (PAPRs) are used in healthcare settings to reduce exposure to high-risk respiratory pathogens. Innovative computational fluid dynamics (CFD) models were developed for evaluating loose-fitting PAPR performance. However, the computational results of the CFD models have not been validated using actual experimental data.
Experimental testing to evaluate particle facepiece leakage was performed in a test laboratory using two models of loose-fitting PAPRs. Each model was mounted on a static (non-moving) advanced headform placed in a sodium chloride (NaCl) aerosol test chamber. The headform performed cyclic breathing via connection to a breathing machine. High-efficiency particulate air (HEPA)-filtered air was supplied directly to the PAPR facepiece using laboratory compressed supplied-air regulated with a mass-flow controller. One model was evaluated with six supplied-air flowrates from 50–215 L/min (Lpm) and the other model with six flowrates from 50–205 Lpm. Three different workrates (minute volumes) were evaluated: low (25 Lpm), moderate 46 (Lpm), and high 88 (Lpm). Manikin penetration factor (mPF) was calculated as the ratio of chamber particle concentration to the in-facepiece concentration.
Overall, data analyses indicated that the mPF results from the simulations were well correlated with the experimental laboratory data for all data combined (r = 0.88). For data at the three different workrates (high, moderate, low) for both models combined, the r-values were 0.96, 0.97, and 0.77, respectively. The CFD models of the two PAPR models were validated and may be utilized for further research.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Similar content being viewed by others
References
Wizner K, Stradtman L, Novak D, Shaffer R (2016) Prevalence of respiratory protective devices in US health care facilities: implications for emergency preparedness. Workplace Health Saf 64:359–368
Federal Register (1995) National Institute for Occupational Safety and Health (NIOSH). Respiratory protective devices. US Government Printing Office, Office of the Federal Register, Washington, DC. 42 CFR Part 84
Martin S, Moyer E, Jensen P (2006) Integrated unit performance testing of powered, air-purifying particulate respirators using a DOP challenge aerosol. J Occup Environ Hyg 3:631–641
Anderson NJ, Casidy PE, Janssen LL, Dengel DR (2006) Peak inspiratory flows of adults exercising at low, moderate and heavy work loads. J Int Soc Respir Prot 23:53–63
Gao S, McKay RT, Yermakov M, Kim J, Reponen T, He X, Kimura K, Grinshpun SA (2016) Performance of an improperly sized and stretched-out loose-fitting powered air-purifying respirator: manikin-based study. J Occup Environ Hyg 13:169–176
Lei Z, Yang J, Zhuang Z, Roberge R (2012) Simulation and evaluation of respirator faceseal leaks using computational fluid dynamics and infrared imaging. Ann Occup Hyg 57:493–506
Lei Z, Yang J (2014) Computing carbon dioxide and humidity in filtering facepiece respirator cavity during breathing cycles. In: ASME 2014 international design engineering technical conferences and computers and information in engineering conference. American Society of Mechanical Engineers, p V01AT02A077
Lei Z, Zhuang Z, Bergman M (2017) Application of digital human modeling for evaluating loose-fitting powered air-purifying respirators. In: Proceedings of the 5th international digital human modeling symposium
Federal Register (2006) Occupational Safety and Health Administration (OSHA). Assigned protection factors: Final rule. 29 CFR Parts 1910, 1915, 1926 71: 50121–50192
Bergman MS, Basu R, Lei Z, Niezgoda G, Zhuang Z (2017) Development of a manikin-based performance evaluation method for loose-fitting powered air-purifying respirators. J Int Soc Respir Prot 34(1):40–57
Bergman MS, Zhuang Z, Hanson D, Heimbuch BK, McDonald MJ, Palmiero AJ, Shaffer RE, Harnish D, Husband M, Wander JD (2014) Development of an advanced respirator fit-test headform. J Occup Environ Hyg 11(2):117–125
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Ethics declarations
The findings and conclusions in this report are those of the author(s) and do not necessarily represent the official position of the National Institute for Occupational Safety and Health, Centers for Disease Control and Prevention. Mention of a company or product name does not constitute endorsement by NIOSH. Mention of a product or use of a photo does not constitute NIOSH endorsement.
Rights and permissions
Copyright information
© 2019 Springer Nature Switzerland AG
About this paper
Cite this paper
Bergman, M., Lei, Z., Xu, S., Strickland, K., Zhuang, Z. (2019). Validation of Computational Fluid Dynamics Models for Evaluating Loose-Fitting Powered Air-Purifying Respirators. In: Bagnara, S., Tartaglia, R., Albolino, S., Alexander, T., Fujita, Y. (eds) Proceedings of the 20th Congress of the International Ergonomics Association (IEA 2018). IEA 2018. Advances in Intelligent Systems and Computing, vol 819. Springer, Cham. https://doi.org/10.1007/978-3-319-96089-0_20
Download citation
DOI: https://doi.org/10.1007/978-3-319-96089-0_20
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-96088-3
Online ISBN: 978-3-319-96089-0
eBook Packages: Intelligent Technologies and RoboticsIntelligent Technologies and Robotics (R0)