Regular Article
Analysis of Transport Experiments Using Pseudo-Absorbance Data

https://doi.org/10.1006/abio.2000.4748Get rights and content

Abstract

The measurement of the concentration distribution of a macromolecule across a solution column by absorption optics usually requires optical transmission profiles of both the sample solution and the buffer, measured under identical conditions, to calculate the absorbance as the logarithm of the ratio of reference to sample intensity. For transport experiments, however, where the changes in the local macromolecule concentration with time are measured, a reference buffer intensity is not necessarily required. We demonstrate that the logarithm of the light transmitted through the sample solution, referred to as pseudo-absorbance, can suffice to determine macromolecular transport parameters of interest, with little loss of precision. Local changes in illumination of the sample column or in the detection efficiency of the transmitted light, as well as temporal fluctuations of the light source intensity can be well-described by consideration of time-invariant and radial-invariant signal components in the pseudo-absorbance data, using the systematic noise decomposition techniques developed recently (Schuck, P., and Demeler, B. (1999) Biophys. J. 76, 2288–2296). The practical use of the method is demonstrated with double-sector and single-sector sedimentation velocity experiments, and with analytical electrophoresis experiments. It is shown that pseudo-absorbance analysis can increase the capacity of a sedimentation velocity experiment in ultracentrifugation, and, in general, can considerably simplify the requirements of optical design.

References (25)

  • N. Muramatsu et al.

    Anal. Biochem.

    (1988)
  • H.K. Schachman et al.

    Arch. Biochem. Biophys.

    (1962)
  • S. Hanlon et al.

    Arch. Biochem. Biophys.

    (1962)
  • A.T. Ansevin et al.

    Anal. Biochem.

    (1970)
  • W.F. Stafford

    Anal. Biochem.

    (1992)
  • P. Schuck et al.

    Biophys. J.

    (1999)
  • P. Schuck

    Anal. Biochem.

    (1999)
  • P. Schuck et al.

    Biophys. J.

    (1998)
  • P. Schuck

    Biophys. J.

    (1998)
  • P. Schuck

    Biophys. J.

    (2000)
  • T. Svedberg et al.

    The Ultracentrifuge

    (1940)
  • Longsworth, L. G. 1959, in, Electrophoresis, (, Bier, M, Ed.), Academic Press, New...
  • Cited by (43)

    • Structure and adsorption behavior of high hydrostatic pressure-treated β-lactoglobulin

      2021, Journal of Colloid and Interface Science
      Citation Excerpt :

      For all SV-AUC experiments, centerpieces with an optical path length of 12 mm were used. When converting intensity data to absorbance data, pseudo-absorption of each sample was calculated and analyzed as described by Kar et al. [33]. SV-AUC data was analyzed with the continuous c(s) model, which is implemented in the SEDFIT program (Version 16-1c) [34].

    • Characterizing biopharmaceuticals using analytical ultracentrifugation

      2019, Biophysical Characterization of Proteins in Developing Biopharmaceuticals
    • Tubulin dimer reversible dissociation: Affinity, kinetics, and demonstration of a stable monomer

      2016, Journal of Biological Chemistry
      Citation Excerpt :

      SV experiments were carried out at 50,000 rpm in an 8-hole An-50 Ti rotor at 20 °C, using 400 μl of samples in 12-mm double-sector charcoal-filled Epon centerpieces, using standard procedures as described in more detail elsewhere (40). To accommodate a large number of samples in absorbance experiments, data were acquired in the intensity mode (41). Tubulin mixtures were prepared by diluting the stock solution to reach the desired protein concentration, allowing for ∼2 h of temperature and chemical equilibration prior to sedimentation.

    View all citing articles on Scopus
    1

    To whom correspondence should be addressed at Bldg. 13, Room 3N17, National Institutes of Health, 13 South Drive, Bethesda, MD 20892-5766. Fax: (301) 496-6608. E-mail: [email protected].

    View full text