Visual Abstract. Effect of Wearing a Novel Electronic Wearable Device on Hand Hygiene Compliance Among Health Care Workers
Effect of Wearing a Novel Electronic Wearable Device on Hand Hygiene Compliance Among Health Care Workers
Figure 1.  Flowchart of the Study
Flowchart of the Study

ABHR indicates alcohol-based handrub; HCWs, health care workers; ITT, intention-to-treat; PP, per protocol.

Figure 2.  HH Compliance by Periods, Months, and Group
HH Compliance by Periods, Months, and Group
Table 1.  Characteristics of Wards and Health Care Workers (HCWs) by Group at Baseline
Characteristics of Wards and Health Care Workers (HCWs) by Group at Baseline
Table 2.  Length of the Study, Number of HH Opportunities, Actions Observed, and HH Compliance in Total, by Group and Period
Length of the Study, Number of HH Opportunities, Actions Observed, and HH Compliance in Total, by Group and Period
Table 3.  Effect of Real-Time Feedback Provided by SmartRub on Compliance With HH Across Calendar Time and Exposure to the Interventiona
Effect of Real-Time Feedback Provided by SmartRub on Compliance With HH Across Calendar Time and Exposure to the Interventiona
1.
Cassini  A, Plachouras  D, Eckmanns  T,  et al.  Burden of six healthcare-associated infections on European population health: estimating incidence-based disability-adjusted life years through a population prevalence-based modelling study.   PLoS Med. 2016;13(10):e1002150. doi:10.1371/journal.pmed.1002150 PubMedGoogle Scholar
2.
Suetens  C, Latour  K, Kärki  T,  et al; The Healthcare-Associated Infections Prevalence Study Group.  Prevalence of healthcare-associated infections, estimated incidence and composite antimicrobial resistance index in acute care hospitals and long-term care facilities: results from two European point prevalence surveys, 2016 to 2017.   Euro Surveill. 2018;23(46). doi:10.2807/1560-7917.ES.2018.23.46.1800516 PubMedGoogle Scholar
3.
Zingg  W, Holmes  A, Dettenkofer  M,  et al; Systematic Review and Evidence-Based Guidance on Organization of Hospital Infection Control Programmes (SIGHT) study group.  Hospital organisation, management, and structure for prevention of health care–associated infection: a systematic review and expert consensus.   Lancet Infect Dis. 2015;15(2):212-224. doi:10.1016/S1473-3099(14)70854-0 PubMedGoogle ScholarCrossref
4.
World Health Organization.  Guidelines on core components of infection prevention and control programmes at the national and acute care facility level.  Published November 2016. Accessed December 11, 2020. https://www.who.int/gpsc/ipc-components-guidelines/en/
5.
World Health Organization.  WHO guidelines on hand hygiene in health care: first global patient safety challenge: clean care is safer care.  Published 2009. Accessed December 11, 2020. https://www.who.int/gpsc/5may/tools/who_guidelines-handhygiene_summary.pdf
6.
World Health Organization.  Guide to implementation: a guide to the implementation of the WHO multimodal hand hygiene improvement strategy.  Published 2009. Accessed December 11, 2020. https://www.who.int/gpsc/5may/Guide_to_Implementation.pdf
7.
Widmer  AF, Conzelmann  M, Tomic  M, Frei  R, Stranden  AM.  Introducing alcohol-based hand rub for hand hygiene: the critical need for training.   Infect Control Hosp Epidemiol. 2007;28(1):50-54. doi:10.1086/510788 PubMedGoogle ScholarCrossref
8.
Laustsen  S, Lund  E, Bibby  BM, Kristensen  B, Thulstrup  AM, Kjølseth Møller  J.  Effect of correctly using alcohol-based hand rub in a clinical setting.   Infect Control Hosp Epidemiol. 2008;29(10):954-956. doi:10.1086/590393PubMedGoogle ScholarCrossref
9.
Bellissimo-Rodrigues  F, Pires  D, Soule  H, Gayet-Ageron  A, Pittet  D.  Assessing the likelihood of hand-to-hand cross-transmission of bacteria: an experimental study.   Infect Control Hosp Epidemiol. 2017;38(5):553-558. doi:10.1017/ice.2017.9 PubMedGoogle ScholarCrossref
10.
Erasmus  V, Daha  TJ, Brug  H,  et al.  Systematic review of studies on compliance with hand hygiene guidelines in hospital care.   Infect Control Hosp Epidemiol. 2010;31(3):283-294. doi:10.1086/650451 PubMedGoogle ScholarCrossref
11.
Pires  D, Soule  H, Bellissimo-Rodrigues  F, Gayet-Ageron  A, Pittet  D.  Hand hygiene with alcohol-based hand rub: how long is long enough?   Infect Control Hosp Epidemiol. 2017;38(5):547-552. doi:10.1017/ice.2017.25 PubMedGoogle ScholarCrossref
12.
Pires  D, Soule  H, Bellissimo-Rodrigues  F, de Kraker  MEA, Pittet  D.  Antibacterial efficacy of handrubbing for 15 versus 30 seconds: EN 1500-based randomized experimental study with different loads of Staphylococcus aureus and Escherichia coli.   Clin Microbiol Infect. 2019;25(7):851-856. doi:10.1016/j.cmi.2018.10.012 PubMedGoogle ScholarCrossref
13.
Kramer  A, Pittet  D, Klasinc  R,  et al.  Shortening the application time of alcohol-based hand rubs to 15 seconds may improve the frequency of hand antisepsis actions in a neonatal intensive care unit.   Infect Control Hosp Epidemiol. 2017;38(12):1430-1434. doi:10.1017/ice.2017.217 PubMedGoogle ScholarCrossref
14.
Harnoss  JC, Dancer  SJ, Kaden  CF,  et al.  Hand antisepsis without decreasing efficacy by shortening the rub-in time of alcohol-based handrubs to 15 seconds.   J Hosp Infect. 2020;104(4):419-424. doi:10.1016/j.jhin.2019.09.004 PubMedGoogle ScholarCrossref
15.
Tschudin-Sutter  S, Rotter  ML, Frei  R,  et al.  Simplifying the WHO ‘how to hand rub’ technique: three steps are as effective as six-results from an experimental randomized crossover trial.   Clin Microbiol Infect. 2017;23(6):409.e1-409.e4. doi:10.1016/j.cmi.2016.12.030 PubMedGoogle ScholarCrossref
16.
Tschudin-Sutter  S, Sepulcri  D, Dangel  M, Ulrich  A, Frei  R, Widmer  AF.  Simplifying the World Health Organization protocol: 3 steps versus 6 steps for performance of hand hygiene in a cluster-randomized trial.   Clin Infect Dis. 2019;69(4):614-620. doi:10.1093/cid/ciy948 PubMedGoogle ScholarCrossref
17.
Pires  D, Bellissimo-Rodrigues  F, Soule  H, Gayet-Ageron  A, Pittet  D.  Revisiting the WHO “how to handrub” hand hygiene technique: fingertips first?   Infect Control Hosp Epidemiol. 2017;38(2):230-233. doi:10.1017/ice.2016.241 PubMedGoogle ScholarCrossref
18.
Bellissimo-Rodrigues  F, Soule  H, Gayet-Ageron  A, Martin  Y, Pittet  D.  Should alcohol-based handrub use be customized to healthcare workers’ hand size?   Infect Control Hosp Epidemiol. 2016;37(2):219-221. doi:10.1017/ice.2015.271 PubMedGoogle ScholarCrossref
19.
Zingg  W, Haidegger  T, Pittet  D.  Hand coverage by alcohol-based handrub varies: volume and hand size matter.   Am J Infect Control. 2016;44(12):1689-1691. doi:10.1016/j.ajic.2016.07.006 PubMedGoogle ScholarCrossref
20.
Jeanes  A, Coen  PG, Gould  DJ, Drey  NS.  Validity of hand hygiene compliance measurement by observation: a systematic review.   Am J Infect Control. 2019;47(3):313-322. doi:10.1016/j.ajic.2018.08.004 PubMedGoogle ScholarCrossref
21.
Boyce  JM.  Electronic monitoring in combination with direct observation as a means to significantly improve hand hygiene compliance.   Am J Infect Control. 2017;45(5):528-535. doi:10.1016/j.ajic.2016.11.029 PubMedGoogle ScholarCrossref
22.
Srigley  JA, Gardam  M, Fernie  G, Lightfoot  D, Lebovic  G, Muller  MP.  Hand hygiene monitoring technology: a systematic review of efficacy.   J Hosp Infect. 2015;89(1):51-60. doi:10.1016/j.jhin.2014.10.005 PubMedGoogle ScholarCrossref
23.
Hsu  Y-W, Yu  C-Y.  Hand surface area estimation formula using 3D anthropometry.   J Occup Environ Hyg. 2010;7(11):633-639. doi:10.1080/15459624.2010.514259 PubMedGoogle ScholarCrossref
24.
International Consortium for Prevention & Infection Control.  Abstracts from the 5th International Conference on Prevention & Infection Control (ICPIC 2019).   Antimicrob Resist Infect Control. 2019;8(Suppl 1):175, poster 431. doi:10.1186/s13756-019-0567-6Google Scholar
25.
Pittet  D, Hugonnet  S, Harbarth  S,  et al.  Effectiveness of a hospital-wide programme to improve compliance with hand hygiene—infection control programme.   Lancet. 2000;356(9238):1307-1312. doi:10.1016/S0140-6736(00)02814-2PubMedGoogle ScholarCrossref
26.
Allegranzi  B, Sax  H, Bengaly  L,  et al; World Health Organization “Point G” Project Management Committee.  Successful implementation of the World Health Organization hand hygiene improvement strategy in a referral hospital in Mali, Africa.   Infect Control Hosp Epidemiol. 2010;31(2):133-141. doi:10.1086/649796 PubMedGoogle ScholarCrossref
27.
World Health Organization.  Hand hygiene technical reference manual: to be used by health-care workers, trainers and observers of hand hygiene practices.  Published 2009. Accessed December 11, 2020. https://apps.who.int/iris/handle/10665/44196
28.
Hemming  K, Taljaard  M.  Sample size calculations for stepped wedge and cluster randomised trials: a unified approach.   J Clin Epidemiol. 2016;69:137-146. doi:10.1016/j.jclinepi.2015.08.015 PubMedGoogle ScholarCrossref
29.
Baio  G, Copas  A, Ambler  G, Hargreaves  J, Beard  E, Omar  RZ.  Sample size calculation for a stepped wedge trial.   Trials. 2015;16:354. doi:10.1186/s13063-015-0840-9 PubMedGoogle ScholarCrossref
30.
 Abstracts from the 5th International Conference on Prevention & Infection Control (ICPIC 2019): Geneva, Switzerland. 10-13 September 2019.   Antimicrob Resist Infect Control 2019;8(Suppl 1):176, poster 434. doi:10.1186/s13756-019-0567-6Google ScholarCrossref
31.
Chen  LF, Vander Weg  MW, Hofmann  DA, Reisinger  HS.  The Hawthorne effect in infection prevention and epidemiology.   Infect Control Hosp Epidemiol. 2015;36(12):1444-1450. doi:10.1017/ice.2015.216 PubMedGoogle ScholarCrossref
32.
Meng  M, Sorber  M, Herzog  A, Igel  C, Kugler  C.  Technological innovations in infection control: a rapid review of the acceptance of behavior monitoring systems and their contribution to the improvement of hand hygiene.   Am J Infect Control. 2019;47(4):439-447. doi:10.1016/j.ajic.2018.10.012 PubMedGoogle ScholarCrossref
33.
Conway  LJ.  Challenges in implementing electronic hand hygiene monitoring systems.   Am J Infect Control. 2016;44(5)(suppl):e7-e12. doi:10.1016/j.ajic.2015.11.031 PubMedGoogle ScholarCrossref
34.
Ward  MA, Schweizer  ML, Polgreen  PM, Gupta  K, Reisinger  HS, Perencevich  EN.  Automated and electronically assisted hand hygiene monitoring systems: a systematic review.   Am J Infect Control. 2014;42(5):472-478. doi:10.1016/j.ajic.2014.01.002 PubMedGoogle ScholarCrossref
Original Investigation
Infectious Diseases
February 8, 2021

Effect of Wearing a Novel Electronic Wearable Device on Hand Hygiene Compliance Among Health Care Workers: A Stepped-Wedge Cluster Randomized Clinical Trial

Author Affiliations
  • 1Infection Control Programme and World Health Organization Collaborating Centre on Patient Safety—Infection Control & Improving Practices, University of Geneva Hospitals and Faculty of Medicine, Geneva, Switzerland
  • 2Faculty of Health Sciences, University of Malta, Msida, Malta
  • 3iQati, Sion, Switzerland
  • 4Haute école du paysage, d’ingénierie et d’architecture de Genève (HEPIA), Geneva, Switzerland
  • 5Social Medicine Department, Ribeirão Preto Medical School, University of São Paulo, Ribeirão Preto, São Paulo, Brazil
JAMA Netw Open. 2021;4(2):e2035331. doi:10.1001/jamanetworkopen.2020.35331
Key Points

Question  Does providing real-time feedback on a simplified hand hygiene (HH) action with a novel electronic wearable device improve the compliance with the “5 Moments” and the quality of the HH action according to World Health Organization (WHO) guidelines?

Findings  In this stepped-wedge, cluster randomized clinical trial involving 97 health care workers at the University of Geneva Hospitals, the device did not improve compliance with the “5 Moments” of HH. The volume of alcohol-based handrub and the duration of hand rubbing per HH increased after the activation of the feedback.

Meaning  The use of the device did not change HH compliance, but improved quality of the HH action.

Abstract

Importance  Hand hygiene (HH) is essential to prevent hospital-acquired infections.

Objective  To determine whether providing real-time feedback on a simplified HH action improves compliance with the World Health Organization’s “5 Moments” and the quality of the HH action.

Design, Setting, and Participants  This open-label, cluster randomized, stepped-wedge clinical trial was conducted between June 1, 2017, and January 6, 2018 (with a follow-up in March 2018), in a geriatric hospital of the University of Geneva Hospitals, Switzerland. All 12 wards and 97 of 306 eligible health care workers (HCWs) volunteered to wear a novel electronic wearable device that delivered real-time feedback on duration of hand rubbing and application of a hand-sized customized volume of alcohol-based handrub (ABHR).

Interventions  This study had 3 sequential periods: baseline (no device), transition (device monitoring without feedback), and intervention (device monitoring and feedback). The start of the transition period was randomly allocated based on a computer-generated block randomization.

Main Outcomes and Measures  The primary outcome was HH compliance, according to the direct observation method during intervention as compared with baseline. Secondary outcomes included the volume of ABHR and duration of hand rubbing measured by the device during intervention as compared with transition.

Results  All wards and respective HCWs were evenly assigned to group 1 (26 participants), 2 (22 participants), 3 (25 participants), or 4 (24 participants). Twelve HCWs did not fully complete the intervention but were included in the analysis. During 759 observation sessions, 6878 HH opportunities were observed. HH compliance at intervention (62.9%; 95% CI, 61.1%-64.7%) was lower than at baseline (66.6%; 95% CI, 64.8%-68.4%). After adjusting for covariates, HH compliance was not different between periods (odds ratio, 1.03; 95% CI, 0.75-1.42; P = .85). Days since study onset (OR, 0.997; 95% CI, 0.994-0.998; P < .001), older age (OR, 0.97; 95% CI, 0.95-0.99; P = .015), and workload (OR, 0.29; 95% CI, 0.20-0.41; P < .001) were independently associated with reduced HH compliance. The median (interquartile range) volume of ABHR and duration of hand rubbing in transition and intervention increased from 1.12 (0.76-1.68) mL to 1.71 (1.01-2.76) mL and from 6.5 (4.5-10.5) seconds to 8 (4.5-15.5) seconds, respectively. There were no serious adverse events.

Conclusions and Relevance  The use of this device did not change HH compliance, but increased the duration of hand rubbing and volume of ABHR used by HCWs.

Trial Registration  isrctn.org Identifier: ISRCTN25430066

Introduction

Health care–associated infections (HAI) and the spread of antimicrobial resistance are major public health concerns1,2 that are largely avoidable in health care by effective implementation of infection prevention and control best practices.3,4 The World Health Organization (WHO) recommends that performing hand hygiene (HH) correctly (“How to Handrub”) at the correct time (“5 Moments”) is the most effective measure to prevent HAI.5,6 The appropriate HH action is crucial to assure proper antimicrobial efficacy.7-9 Unfortunately, HH compliance remains suboptimal,10 and new strategies are needed to improve its implementation.

A number of studies suggest that the HH action, as endorsed by the WHO, could be simplified without compromising efficacy. Proposed changes include shortening duration of hand rubbing to 15 seconds (instead of 30 seconds)11-14 and changing the number or order of steps (“3 steps” or “fingertips-first” instead of the “6 step” technique).15-17 Additionally, standardization of the hand-size (“palm full”) concept recommended by the WHO has been proposed.18,19 It is hypothesized that these changes could lead to an increase in the quality of the HH action performed by health care workers (HCWs).13,16 In addition, some small-sized studies have suggested that a simplified HH action could also improve compliance with the “5 Moments.”13,16

Monitoring and feedback are essential parts of the WHO multimodal strategy to improve HH.6 However, direct observation and timely feedback are time-consuming, costly, and prone to bias.20,21 There has been a growing interest in the use of electronic monitoring, which can increase the number of monitored actions and remove the observation bias.21

We developed a novel wearable electronic device that monitors and provides real-time, personalized feedback to HCWs based on a simplified and customized HH action. This device is the result of an investigator-initiated partnership between 3 Swiss institutions: the University of Geneva Hospitals (HUG) and Faculty of Medicine, the Haute école du paysage, d’ingénierie et d’architecture de Genève (HEPIA), and iQati, a start-up medical device company.

We aimed to test the impact of the electronic device on compliance with the WHO “5 Moments” for HH and on the quality of hand rubbing in daily patient-care activities. We hypothesized that the use of the device would create a permanent sense of being observed on the HCWs, the so-called positive Hawthorne effect,22 that is known to influence compliance with HH. We also hypothesized that the use of the device would directly increase the quality of HH action reflected by the volume of ABHR used and the duration of HH.

Methods
Study Design

This study was a Swiss National Research Foundation–funded, investigator-initiated, single-center (ie, HUG) project. We conducted a stepped-wedge, cluster randomized, and controlled open-label trial. This study design was preferred to a classic parallel cluster randomized trial because the device was assumed to be a quality improvement tool to which all participating HCWs should have access.

The trial was approved by the Regional Research Ethics Committee, and all participating HCWs provided written informed consent. This study followed Consolidated Standards of Reporting Trials (CONSORT) reporting guideline. A full trial protocol is available in Supplement 1.

Participants

All wards of the geriatric hospital at HUG, including outpatient clinics and the emergency department, were eligible to participate if at least 5 permanent HCWs volunteered. HCWs were not eligible if, during the study period, they: (1) planned to leave the ward; (2) worked in more than 1 ward (risk of contamination bias); (3) had more than 3 consecutive weeks of vacations scheduled for the study period; or (4) used an ABHR agent other than the standard at HUG due to skin allergy. HCWs were recruited over the course of 26 10-minute information sessions at clinical nursing and medical staff meetings, as well as by posters and leaflets (eFigure 1 in Supplement 2).

Settings

HUG is a 1900-bed, tertiary care university hospital with approximately 50 000 admissions per year. The geriatric hospital is a 300-bed, freestanding building, located at a separate site in Geneva, with approximately 10 000 admissions per year.

Intervention

The intervention consisted of providing real-time individual feedback to HCWs after 15 seconds of hand rubbing and the application of a hand-sized volume of ABHR. Feedback was provided by the novel electronic wearable device (SmartRub), which consists of 2 elements—a bottle and a wristband (eFigure 2 in Supplement 2). Each individual ABHR bottle that is widely used at HUG was equipped, during the transition and intervention periods, with a volumetric flow meter. The flow meter measured the volume of ABHR poured onto hands and during the intervention period it also provided feedback by vibrating as soon as the predefined volume had been applied. The volume was determined for each HCW by taking into account the surface area of each individual’s hands,23 as described in previous studies.9,12,18 The wristband, made from silicone, was worn during the transition and intervention periods. It measured the duration of each HH action, and during the intervention period it also vibrated after 15 seconds independently of the hand rubbing duration performed by the HCW.11,12

Instructions regarding the duration of hand rubbing and the volume of ABHR were solely delivered by the vibration of the electronic device during the intervention period. No additional educational sessions were organized during the study.

HCWs were asked to place the device into a charging station at the end of each shift and to collect it upon starting the subsequent shift. The device recorded the date and time of use, volume of applied ABHR, duration of hand rubbing, and whether feedback was provided to the HCW or not for all of the HH actions. Overall sensitivity and specificity of this novel electronic wearable device (to correctly identify a HH action) were 94.1% (95% CI, 91.4%-96.2%) and 99.0% (95% CI, 97.5%-99.7%), respectively.24

Study Periods and Randomization

Study duration, including baseline, transition, and intervention periods was approximately 6 months, followed by a 2-month washout and a 1-month follow-up period. HH observations were performed at least once a month throughout the baseline, transition, intervention, and follow-up periods.

At baseline period, HCWs did not wear the devices. During the transition period, the novel electronic wearable device was worn but the feedback mode (vibration) was not activated, although the device actively monitored the volume of ABHR and duration of hand friction for each HH action. During the intervention period, the feedback mode of the device was activated and the monitoring of practices by the device continued. At follow-up, HCWs did not use the device.

After an initial common 1-month baseline period, 3 wards per month were randomly assigned to start with the 1-month transition period, followed by the intervention period. The length of the baseline and intervention periods were thus inversely related and varied from 1 to 4 months according to the group of randomization (eFigure 3 in Supplement 2).

In total, 12 wards were randomized following a computer-generated block randomization (1:1:1:1) performed by an independent statistician. Numbered opaque envelopes allowed allocation concealment.

Wards were informed the day before shifting from baseline to transition period. Because of the nature of the study, it was not possible to mask study participants or observers after the baseline period.

Primary Outcome

The primary outcome was overall HH compliance, as monitored by direct observation of HH and applying the WHO guidelines.5 HH opportunities were identified, and rubbing or washing actions were recorded.25 HH compliance was calculated as a proportion of HH actions divided by the identified HH opportunities. Three validated observers performed HH observations during audit sessions on weekdays and day shifts. Each HCW was observed at least once per month, with each observation period consisting of a minimum of 6 HH opportunities. HH compliance at the individual HCW level (closed cohort) between intervention and baseline was then compared.

Secondary Outcomes

Secondary outcomes were: (1) volume of ABHR and duration of hand rubbing for each HH action in transition and intervention periods, as automatically recorded by the device; (2) adherence to device use (ie, hours of device use per day) and frequency of HH (HH events per hour), as automatically recorded by the device; (3) HH compliance of participants during follow-up by direct HH observation; (4) ABHR consumption per ward and study period, as per pharmacy dispensing and device data; (5) adverse events spontaneously reported to the study team by HCWs; and (6) satisfaction and perception of device usefulness assessed by mean of a poststudy questionnaire and focus group discussions.

Statistical Analysis

Sample size was estimated by hypothesizing that wearing the device would improve HH compliance by a relative 20%, going from 69% (the 2015 HH compliance at HUG) in baseline to 83% in the intervention period. This corresponds to a 0.35 standardized difference in proportions.26 We used an intraclass correlation coefficient of 0.015 (based on the 2013-2014 HH compliance data at HUG), leading to a design effect of 1.06.27 Considering a study power of 80%, and an α error fixed at 5% (2-sided), we estimated that 12 wards (3 per group, 4 groups) with at least 5 HCWs per ward would be needed to test our study hypothesis.28,29 Continuous variables were reported by means (with SDs or 95% CIs) or medians (interquartile ranges [IQR]) where appropriate. Categorical variables were reported by numbers with relative frequency.

All analyses were conducted on an intention-to-treat (ITT) and a per-protocol (PP) basis, the latter excluding HCWs that did not have at least 1 HH observation in each study period or did not complete the intervention period. Pooled HH compliance between the intervention and the baseline periods was modeled using mixed-effects logistic regression. The effect of length of active device use on HH compliance was assessed by testing for interaction between the randomization group and the study period (baseline vs intervention). The clustering of data was taken into account by including nested random effects for HH opportunities within each audit session, HCWs within each ward, and wards within each group (ie, the hierarchical levels of clustering within the trial were in order of groups, wards, HCWs, sessions, and opportunities).

All analyses were adjusted for covariates comprising HCWs’ demographic characteristics (ie, age, sex, profession, years of work experience, years since last HH training, full-/part-time work), ward specialty (internal medicine, geriatrics, emergency department, ambulatory care), workload (mean HH opportunities per minute) and the type of ABHR used (gel, rinse). The models also included time of exposure to the study (number of days between the start of the study and the individual HH session).

We compared separately the volume of applied ABHR and the duration of hand rubbing between the transition and intervention periods by performing a mixed linear regression model (with each outcome used as dependent variables). We assessed the effect of length of active device use on volume and duration of hand rubbing by testing for interaction between the randomization group and the study period (transition vs intervention). Analysis used HCW-level data clustered within the ward level and group level (ie, hierarchical levels of nested random effects, in order: groups, wards, and HCWs). All analyses were adjusted for the same covariates as for the primary analysis. In addition, we calculated the proportion of correct HH actions, which were defined as a sufficient volume ABHR applied (as per hand size) and hand rubbing for at least 15 seconds. Finally, we calculated the HH compliance of all HCWs and per group for the follow-up period.

Two-sided P values < .05 were considered statically significant. All analyses were performed using Stata IC version 16.0 (StataCorp).

Results

All 12 wards of the geriatric hospital at HUG were eligible and were recruited from June 1 to 30, 2017. There were 4 medical wards, 6 geriatric wards, 1 ambulatory unit, and the emergency department included. Of the 306 eligible HCWs, 97 volunteered and were included in the study (Figure 1). There were 80 women participants, and the median (IQR) age was 42 years (33-53); 63 were nurses, 32 were assistant nurses, and 2 were physiotherapists. Baseline characteristics of wards and HCWs per group are presented in Table 1. See eFigure 3 in Supplement 2 for a summary of the dates of the study periods.

Overall, 750 HH observation sessions were conducted, resulting in 6878 HH opportunities. The median (IQR) duration of each HH session was 18 (15-23) minutes. We observed a median of 72 (61-84) HH opportunities per HCW during a median of 8 (7-9) observation sessions.

HH compliance rates during baseline, transition, and intervention periods were 66.6% (95% CI, 64.8%-68.4%), 67.9% (95% CI, 65.4%-70.3%), and 62.9% (95% CI, 61.1%-64.7%), respectively. Compliance by group and study period and month are summarized in Table 2 and Figure 2.

Crude analysis reveals that, overall, HH compliance at the intervention period was lower than at baseline (OR, 0.62; 95% CI, 0.53-0.72; P < .001). After adjusting for covariates, HH compliance was not different between baseline and intervention periods (OR, 1.03; 95% CI, 0.75-1.42; P = .85) in the ITT analysis. The randomization group, which took the different lengths of intervention into account, was not associated with a change in HH compliance. Thus, the interaction term was not included in the final model. However, HH compliance in all periods combined was significantly higher in groups 1, 2, and 3 than in group 4 (eg, group 1 vs group 4: 68.6%; 95% CI, 66.5%-70.6% vs 57.3%; 95% CI, 55.1%-59.5%; P = .02). HH compliance was inversely correlated with study duration (ie, study days before individual HH observation) (OR, 0.997; 95% CI, 0.994-0.998; P < .001), older age (OR, 0.97; 95% CI, 0.95-0.99; P = .015), and workload (OR, 0.29; 95% CI, 0.20-0.41; P < .001) (Table 3).

In the per-protocol analysis, 81 (83.5%) HCWs were analyzed: 22 (91.7%) in group 1, 17 (77.3%) in group 2, 23 (92%) in group 3, and 19 (79.2%) in group 4. The fitted model confirmed that overall HH compliance was not significantly different between intervention and baseline periods (OR, 0.99; 95% CI, 0.71-1.40; P = .99).

Ninety-three of the included 97 HCWs (95.9%) wore the novel electronic wearable device. The bottles and bracelets recorded 65 818 and 45 813 actions in total (eTable in Supplement 2).

The individually applied ABHR volume increased during the intervention period (feedback on) (median [IQR], 1.71 [1.01-2.76] mL) as compared with the transition period (feedback off) (1.12 [0.76-1.68] mL). Correct HH actions, as per correct application of volume of ABHR only, increased from 10.2% (95% CI, 9.8%-10.6%) during transition to 30.5% (95% CI, 30.0%-30.9%) during intervention (eTable in Supplement 2). After adjusting for covariates, the effect of the intervention on higher applied ABHR volume was significant in every group.

Notably, the duration of hand rubbing also increased during intervention (median [IQR], 8 [4.5-15.5] seconds as compared with the transition period, 6.5 [4.5-10.5] seconds). Correct HH actions, as per correct duration only, increased from 11.4% (95% CI, 10.6%-11.9%) in transition to 26.9% (95% CI, 26.4%-27.4%) (eTable in Supplement 2) during intervention. After adjusting for covariates, the effect of the device on the duration of hand rubbing was significant in every group.

Follow-up HH observations at 3 months after the end of the intervention (March 2018) were performed on 78 (80.4%) HCWs. A total of 790 opportunities were observed. The overall HH compliance was 63.4% (95% CI, 60.1%-66.8%). Compliance at follow-up was not different from baseline (66.6%; 95% CI, 64.8%-68.4%; P = .09) or intervention periods (62.9%; 95% CI, 61.1%-64.7%; P = .79). Compliance at follow-up per group was: 63.6% (95% CI, 57.2%-69.9%) in group 1, 63.2% (95% CI, 55.0%-71.4%) in group 2, 63.9% (95% CI, 57.6%-70.2%) in group 3, and 62.9% (95% CI, 56.2%-69.6%) in group 4 (P = .99).

The total participants experiencing any adverse event was 19. All events were skin related: 6 reports of dry skin due to changing from gel to rinse (which subsided after changing back to gel), 3 reports of itching from the bracelet (which subsided after temporarily stopping bracelet use), and 9 reports of dry skin due to using more ABHR than previously. These adverse events directly motivated 2 dropouts (1 related to bracelet itching and 1 to dry skin). As judged by investigators, no serious adverse event was reported.

Device satisfaction and perception were evaluated by questionnaires and focus group discussions. Results are reported elsewhere.30

Discussion

This stepped-wedge cluster randomized clinical trial tested the effect of a novel electronic wearable device that provides feedback on appropriately applied ABHR volume and sufficient duration of hand rubbing, on both compliance and quality of HH practices. Our study showed that wearing the novel electronic wearable device did not improve HH compliance. However, wearing the device did improve the quality of the HH action, with a significant increase in both ABHR volume applied and in the duration of hand rubbing.

Intriguingly, we observed a gradual decline in HH compliance throughout the study, from 73.5% (95% CI, 70.5%-76.5%) during the first month to 56.6% (95% CI, 53.2%-60.1%) during the last month. This time trend was unexpected and, likely, hampered the assessment of the effect of the device on HH compliance. We hypothesize there was a significant Hawthorne effect when HCWs were observed for the first time during the baseline period, which resulted in overestimated HH compliance. The same auditors performed repeated observations of the same HCWs, and HCWs may have become accustomed to them over time. Thus, the presence of the auditors may not have sparked immediate behavior change in subsequent observations. This would result in regression of an initial Hawthorne effect, resulting in HH compliance falling back to routine behavior. This phenomenon of habituation has been previously described in Chen et al.31

In contrast, monitoring of HCWs’ behavior with a wearable device might be less affected by those variables. This advantage probably allowed detecting the effect of the device feedback on HH quality. Notably, we observed an absolute increase of around 20% in the correct compliance with both the volume of applied ABHR and on the 15-second duration of hand rubbing. It has been widely shown that these parameters are major determinants of the microbial efficacy of the HH action.11-13,18,19 However, the clinical significance of these results remains unknown, as our study was not designed to assess the impact of improving the quality of the HH action on the reduction of HAI.

The implementation of electronic monitoring systems raises issues related with transparency and confidentiality,32 accuracy, and adaptation to workflow.33 We observed good buy-in from HCWs, as the proportion of dropouts related to the use of the device was very small. In a questionnaire performed after the study, the majority (55/70 participants) of HCWs providing feedback agreed that the device was a helpful reminder for correct HH. However, 27% (19/70 participants) were concerned about confidentiality, and 23% (16/70 participants) reported that using the wristband interfered with their clinical activities.30 Importantly, the device might not be applicable in settings where a “bare below elbows” strict policy applies. However, we believe that its material (medical silicone), its ergonomics, and the ease of disinfection using any detergent disinfectant might allow for an implementation of the device in these settings for occasional educational interventions. Finally, we have previously tested the device’s sensitivity and specificity in a range of laboratory and clinical conditions.24

The most recent systematic review on the effect of monitoring technologies on HH adherence suggested that electronic devices have the potential to change HH behavior.32 However, there is a lack of clinical trials with solid designs including system-independent, relevant outcomes.22,32,34 We believe that the design of this stepped-wedge, cluster randomized clinical trial contributed to raise the standards of methodological quality of studies on this topic.

Our study is a proof of concept of the benefits of a wearable device on improving the quality of HH in clinical practice and it opens perspectives for new strategies on HH improvement. The great majority of HH monitoring devices focus on dispensing events or proxies for HH indications.21 This electronic wearable device is unique in its potential to improve HH by interacting directly with the HCW during HH, in real time and during daily routine.

Limitations

Our study had several limitations. This was a novel device and, not surprisingly, we faced a series of technical challenges. Problems included random inactivation of the devices (more frequent with the wristband) and loss of data because of errors in data transfer. These prevented us from calculating predetermined secondary outcomes, such as the frequency of HH events, adherence to device use, and ward-level ABHR consumption. The lack of such data did not allow us to analyze if trends of observed HH compliance were different from trends of performed HH events. In addition, the implementation of the study revealed some difficulties. For example, we faced problems related to the production and delivery of the devices that induced delays in the pre-established dates of rollout from baseline to transition in some groups. However, we did manage to respect the stepped-wedge design of the study.

Conclusions

Wearing a novel electronic wearable device did not improve HH compliance but did lead to an improvement of the quality of HH action as measured by an increase in the volume of ABHR per hand action and in the duration of hand rubbing. The gradual decline in HH compliance may have been unrelated to the use of the device and associated with a high-magnitude Hawthorne effect at baseline. In order to reach its full potential, this innovative device merits further development and adequate testing in future studies.

Back to top
Article Information

Accepted for Publication: December 3, 2020.

Published: February 8, 2021. doi:10.1001/jamanetworkopen.2020.35331

Open Access: This is an open access article distributed under the terms of the CC-BY License. © 2021 Pires D et al. JAMA Network Open.

Corresponding Author: Didier Pittet, MD, MS, Infection Control Programme and World Health Organization Collaborating Centre on Patient Safety, University of Geneva Hospitals and Faculty of Medicine, 4 Rue Gabrielle-Perret-Gentil, 1211 Geneva, Switzerland (didier.pittet@hcuge.ch).

Author Contributions: Drs Pire, Gayet-Ageron, and Pittet had full access to all of the data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis.

Concept and design: Pires, Gayet-Ageron, Soule, Bellissimo-Rodrigues, Martin, Zingg, Pittet.

Acquisition, analysis, or interpretation of data: Pires, Gayet-Ageron, Guitart, Robert, Fankhauser, Tartari, Peters, Tymurkaynak, Fourquier, Beuchat, Bellissimo-Rodrigues, Martin, Zingg, Pittet.

Drafting of the manuscript: Pires, Gayet-Ageron, Tymurkaynak, Fourquier, Zingg.

Critical revision of the manuscript for important intellectual content: Pires, Gayet-Ageron, Guitart, Robert, Fankhauser, Tartari, Peters, Soule, Beuchat, Bellissimo-Rodrigues, Martin, Zingg, Pittet.

Statistical analysis: Pires, Gayet-Ageron.

Obtained funding: Gayet-Ageron, Bellissimo-Rodrigues, Martin, Pittet.

Administrative, technical, or material support: Pires, Guitart, Robert, Fankhauser, Tartari, Peters, Tymurkaynak, Fourquier, Soule, Beuchat, Bellissimo-Rodrigues, Martin, Zingg, Pittet.

Supervision: Pires, Guitart, Fankhauser, Bellissimo-Rodrigues, Zingg, Pittet.

Conflict of Interest Disclosures: Dr Guitart reported receiving grants from Fond National Suisse during the conduct of the study. Dr Robert reported a patent pending (PCT/CH2018/000027) with iQati. Dr Timurkaynak reported receiving grants from Swiss National Science Foundation (No. 32003B_163262) during the conduct of the study. Dr Martin reported receiving grants from Swiss National Science Foundation during the conduct of the study; in addition, Dr Martin reported a patent for a system and method to improve hand hygiene quality pending; he is a cofounder of iQati. Dr Pittet reported working with the World Health Organization (WHO) in the context of the WHO initiative Private Organizations for Patient Safety—Hand Hygiene, which aims to harness industry strengths to align and improve implementation of WHO recommendations for hand hygiene in health care in different parts of the world. No other disclosures were reported.

Funding/Support: This study was supported by the Swiss National Research Foundation (grant No. 2003B_163262). This work was also supported by the Infection Control Program (SPCI) and Faculty of Medicine, University of Geneva Hospitals, Geneva, Switzerland.

Role of the Funder/Sponsor: The funders had no role in the design and conduct of the study; collection, management, analysis, and interpretation of the data; preparation, review, or approval of the manuscript; and decision to submit the manuscript for publication.

Meeting Presentation: Part of the data was presented at the International Conference of Prevention and Infection Control (ICPIC); September 12, 2019; Geneva, Switzerland.

Disclaimer: The World Health Organization takes no responsibility for the information provided or the views expressed in this paper. The authors alone are responsible for the views expressed in this article and they do not necessarily represent the views, decisions, or policies of the institutions with which they are affiliated.

Additional Contributions: We would like to thank all the health care workers that participated in the study and the important help of all head nurses and of the nurses in the infection prevention and control unit at the University of Geneva Hospitals, Valerie Sauvan, BS, and Josiane Sztajzel-Boissard, BS, in the implementation of this study. Contributors received no additional compensation.

Data Sharing Statement: See Supplement 3.

References
1.
Cassini  A, Plachouras  D, Eckmanns  T,  et al.  Burden of six healthcare-associated infections on European population health: estimating incidence-based disability-adjusted life years through a population prevalence-based modelling study.   PLoS Med. 2016;13(10):e1002150. doi:10.1371/journal.pmed.1002150 PubMedGoogle Scholar
2.
Suetens  C, Latour  K, Kärki  T,  et al; The Healthcare-Associated Infections Prevalence Study Group.  Prevalence of healthcare-associated infections, estimated incidence and composite antimicrobial resistance index in acute care hospitals and long-term care facilities: results from two European point prevalence surveys, 2016 to 2017.   Euro Surveill. 2018;23(46). doi:10.2807/1560-7917.ES.2018.23.46.1800516 PubMedGoogle Scholar
3.
Zingg  W, Holmes  A, Dettenkofer  M,  et al; Systematic Review and Evidence-Based Guidance on Organization of Hospital Infection Control Programmes (SIGHT) study group.  Hospital organisation, management, and structure for prevention of health care–associated infection: a systematic review and expert consensus.   Lancet Infect Dis. 2015;15(2):212-224. doi:10.1016/S1473-3099(14)70854-0 PubMedGoogle ScholarCrossref
4.
World Health Organization.  Guidelines on core components of infection prevention and control programmes at the national and acute care facility level.  Published November 2016. Accessed December 11, 2020. https://www.who.int/gpsc/ipc-components-guidelines/en/
5.
World Health Organization.  WHO guidelines on hand hygiene in health care: first global patient safety challenge: clean care is safer care.  Published 2009. Accessed December 11, 2020. https://www.who.int/gpsc/5may/tools/who_guidelines-handhygiene_summary.pdf
6.
World Health Organization.  Guide to implementation: a guide to the implementation of the WHO multimodal hand hygiene improvement strategy.  Published 2009. Accessed December 11, 2020. https://www.who.int/gpsc/5may/Guide_to_Implementation.pdf
7.
Widmer  AF, Conzelmann  M, Tomic  M, Frei  R, Stranden  AM.  Introducing alcohol-based hand rub for hand hygiene: the critical need for training.   Infect Control Hosp Epidemiol. 2007;28(1):50-54. doi:10.1086/510788 PubMedGoogle ScholarCrossref
8.
Laustsen  S, Lund  E, Bibby  BM, Kristensen  B, Thulstrup  AM, Kjølseth Møller  J.  Effect of correctly using alcohol-based hand rub in a clinical setting.   Infect Control Hosp Epidemiol. 2008;29(10):954-956. doi:10.1086/590393PubMedGoogle ScholarCrossref
9.
Bellissimo-Rodrigues  F, Pires  D, Soule  H, Gayet-Ageron  A, Pittet  D.  Assessing the likelihood of hand-to-hand cross-transmission of bacteria: an experimental study.   Infect Control Hosp Epidemiol. 2017;38(5):553-558. doi:10.1017/ice.2017.9 PubMedGoogle ScholarCrossref
10.
Erasmus  V, Daha  TJ, Brug  H,  et al.  Systematic review of studies on compliance with hand hygiene guidelines in hospital care.   Infect Control Hosp Epidemiol. 2010;31(3):283-294. doi:10.1086/650451 PubMedGoogle ScholarCrossref
11.
Pires  D, Soule  H, Bellissimo-Rodrigues  F, Gayet-Ageron  A, Pittet  D.  Hand hygiene with alcohol-based hand rub: how long is long enough?   Infect Control Hosp Epidemiol. 2017;38(5):547-552. doi:10.1017/ice.2017.25 PubMedGoogle ScholarCrossref
12.
Pires  D, Soule  H, Bellissimo-Rodrigues  F, de Kraker  MEA, Pittet  D.  Antibacterial efficacy of handrubbing for 15 versus 30 seconds: EN 1500-based randomized experimental study with different loads of Staphylococcus aureus and Escherichia coli.   Clin Microbiol Infect. 2019;25(7):851-856. doi:10.1016/j.cmi.2018.10.012 PubMedGoogle ScholarCrossref
13.
Kramer  A, Pittet  D, Klasinc  R,  et al.  Shortening the application time of alcohol-based hand rubs to 15 seconds may improve the frequency of hand antisepsis actions in a neonatal intensive care unit.   Infect Control Hosp Epidemiol. 2017;38(12):1430-1434. doi:10.1017/ice.2017.217 PubMedGoogle ScholarCrossref
14.
Harnoss  JC, Dancer  SJ, Kaden  CF,  et al.  Hand antisepsis without decreasing efficacy by shortening the rub-in time of alcohol-based handrubs to 15 seconds.   J Hosp Infect. 2020;104(4):419-424. doi:10.1016/j.jhin.2019.09.004 PubMedGoogle ScholarCrossref
15.
Tschudin-Sutter  S, Rotter  ML, Frei  R,  et al.  Simplifying the WHO ‘how to hand rub’ technique: three steps are as effective as six-results from an experimental randomized crossover trial.   Clin Microbiol Infect. 2017;23(6):409.e1-409.e4. doi:10.1016/j.cmi.2016.12.030 PubMedGoogle ScholarCrossref
16.
Tschudin-Sutter  S, Sepulcri  D, Dangel  M, Ulrich  A, Frei  R, Widmer  AF.  Simplifying the World Health Organization protocol: 3 steps versus 6 steps for performance of hand hygiene in a cluster-randomized trial.   Clin Infect Dis. 2019;69(4):614-620. doi:10.1093/cid/ciy948 PubMedGoogle ScholarCrossref
17.
Pires  D, Bellissimo-Rodrigues  F, Soule  H, Gayet-Ageron  A, Pittet  D.  Revisiting the WHO “how to handrub” hand hygiene technique: fingertips first?   Infect Control Hosp Epidemiol. 2017;38(2):230-233. doi:10.1017/ice.2016.241 PubMedGoogle ScholarCrossref
18.
Bellissimo-Rodrigues  F, Soule  H, Gayet-Ageron  A, Martin  Y, Pittet  D.  Should alcohol-based handrub use be customized to healthcare workers’ hand size?   Infect Control Hosp Epidemiol. 2016;37(2):219-221. doi:10.1017/ice.2015.271 PubMedGoogle ScholarCrossref
19.
Zingg  W, Haidegger  T, Pittet  D.  Hand coverage by alcohol-based handrub varies: volume and hand size matter.   Am J Infect Control. 2016;44(12):1689-1691. doi:10.1016/j.ajic.2016.07.006 PubMedGoogle ScholarCrossref
20.
Jeanes  A, Coen  PG, Gould  DJ, Drey  NS.  Validity of hand hygiene compliance measurement by observation: a systematic review.   Am J Infect Control. 2019;47(3):313-322. doi:10.1016/j.ajic.2018.08.004 PubMedGoogle ScholarCrossref
21.
Boyce  JM.  Electronic monitoring in combination with direct observation as a means to significantly improve hand hygiene compliance.   Am J Infect Control. 2017;45(5):528-535. doi:10.1016/j.ajic.2016.11.029 PubMedGoogle ScholarCrossref
22.
Srigley  JA, Gardam  M, Fernie  G, Lightfoot  D, Lebovic  G, Muller  MP.  Hand hygiene monitoring technology: a systematic review of efficacy.   J Hosp Infect. 2015;89(1):51-60. doi:10.1016/j.jhin.2014.10.005 PubMedGoogle ScholarCrossref
23.
Hsu  Y-W, Yu  C-Y.  Hand surface area estimation formula using 3D anthropometry.   J Occup Environ Hyg. 2010;7(11):633-639. doi:10.1080/15459624.2010.514259 PubMedGoogle ScholarCrossref
24.
International Consortium for Prevention & Infection Control.  Abstracts from the 5th International Conference on Prevention & Infection Control (ICPIC 2019).   Antimicrob Resist Infect Control. 2019;8(Suppl 1):175, poster 431. doi:10.1186/s13756-019-0567-6Google Scholar
25.
Pittet  D, Hugonnet  S, Harbarth  S,  et al.  Effectiveness of a hospital-wide programme to improve compliance with hand hygiene—infection control programme.   Lancet. 2000;356(9238):1307-1312. doi:10.1016/S0140-6736(00)02814-2PubMedGoogle ScholarCrossref
26.
Allegranzi  B, Sax  H, Bengaly  L,  et al; World Health Organization “Point G” Project Management Committee.  Successful implementation of the World Health Organization hand hygiene improvement strategy in a referral hospital in Mali, Africa.   Infect Control Hosp Epidemiol. 2010;31(2):133-141. doi:10.1086/649796 PubMedGoogle ScholarCrossref
27.
World Health Organization.  Hand hygiene technical reference manual: to be used by health-care workers, trainers and observers of hand hygiene practices.  Published 2009. Accessed December 11, 2020. https://apps.who.int/iris/handle/10665/44196
28.
Hemming  K, Taljaard  M.  Sample size calculations for stepped wedge and cluster randomised trials: a unified approach.   J Clin Epidemiol. 2016;69:137-146. doi:10.1016/j.jclinepi.2015.08.015 PubMedGoogle ScholarCrossref
29.
Baio  G, Copas  A, Ambler  G, Hargreaves  J, Beard  E, Omar  RZ.  Sample size calculation for a stepped wedge trial.   Trials. 2015;16:354. doi:10.1186/s13063-015-0840-9 PubMedGoogle ScholarCrossref
30.
 Abstracts from the 5th International Conference on Prevention & Infection Control (ICPIC 2019): Geneva, Switzerland. 10-13 September 2019.   Antimicrob Resist Infect Control 2019;8(Suppl 1):176, poster 434. doi:10.1186/s13756-019-0567-6Google ScholarCrossref
31.
Chen  LF, Vander Weg  MW, Hofmann  DA, Reisinger  HS.  The Hawthorne effect in infection prevention and epidemiology.   Infect Control Hosp Epidemiol. 2015;36(12):1444-1450. doi:10.1017/ice.2015.216 PubMedGoogle ScholarCrossref
32.
Meng  M, Sorber  M, Herzog  A, Igel  C, Kugler  C.  Technological innovations in infection control: a rapid review of the acceptance of behavior monitoring systems and their contribution to the improvement of hand hygiene.   Am J Infect Control. 2019;47(4):439-447. doi:10.1016/j.ajic.2018.10.012 PubMedGoogle ScholarCrossref
33.
Conway  LJ.  Challenges in implementing electronic hand hygiene monitoring systems.   Am J Infect Control. 2016;44(5)(suppl):e7-e12. doi:10.1016/j.ajic.2015.11.031 PubMedGoogle ScholarCrossref
34.
Ward  MA, Schweizer  ML, Polgreen  PM, Gupta  K, Reisinger  HS, Perencevich  EN.  Automated and electronically assisted hand hygiene monitoring systems: a systematic review.   Am J Infect Control. 2014;42(5):472-478. doi:10.1016/j.ajic.2014.01.002 PubMedGoogle ScholarCrossref
×