X線CTにおけるtime sensitivity profile(TSP)の考案と測定法の開発 : non helical scan, single slice helical scan, multi slice helical scanの比較
辻岡 勝美井田 義宏大坪 寛知高橋 康方丹羽 正厳
著者情報
ジャーナル フリー

2000 年 56 巻 12 号 p. 1461-1469

詳細
抄録

We focused on the time element contained in a single CT image, and devised the concept of a time-sensitivity profile (TSP) describing how the time element is translated into an image. We calculated the data collection time range when the helical pitch is changed in non helical scans, single slice helical scans, and multi slice helical scans. We then calculated the time sensitivity profile (TSP) from the weighting applied when the data collection time range is translated into an image. TSP was also measured for each scanning method using our self-made moving phantom. TSPs obtained from the calculation and the experiments were very close. TSP showed interesting characteristics with each scanning method, especially in the case of multi slice helical scanning, in which TSP became shorter as helical pitch increased. We referred to the TSP's FWHM as the effective scanning time. When we conducted multi slice helical scanning at helical pitch 3, the effective scanning time increased to about 24% longer than that of a non helical scan. When we conducted multi slice helical scanning at helical pitch 5 or 6, the effective scanning time was about half that of a non helical scan. The time sensitivity profile (TSP) is a totally new concept that we consider an important element in discussing the time resolution of a CT scanner. The results of this review will provide significant data in determining the scanning parameters when scanning a moving object.

著者関連情報
© 2000 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top