下腿DSA撮影における体動防止固定具の開発
安田 光慶加藤 京一先山 耕史内山 裕史浅沼 眞一藤村 一正鈴木 洋中澤 靖夫
著者情報
ジャーナル フリー

2010 年 66 巻 1 号 p. 49-56

詳細
抄録

In the peripheral angiography to evaluate blood flow below the knee levels in patients with arteriosclerosis obliterans, a motion artifact occurs as a result of body movements of a patient in normal fixation. This sometimes makes a correct evaluation difficult. Therefore, we developed a fixed appliance which can restrain body movement without affecting image quality and blood flow in digital subtraction angiography of a leg. The appliance is filled with the styrofoam of 1 mm diameter in a sealed bag, after air is aspirated from inside the bag. The appliance is stiffened to fit the shape of the crus. We measured signal to noise ratio / contrast to noise ratio / a resolution limit by visual evaluation to examine the influence of the image before and after usage of this appliance. In addition, the blood velocity of the dorsalis artery in ultrasound was measured to examine the effect on the blood flow. As a result, the fixed appliance did not affect blood flow in peripheral angiography to evaluate the clinical significance, the usual 5-point evaluate scale was used. The scale was significantly improved (p <0.01) after usage of this appliance. The newly developed fixed appliance for digital subtraction angiography of a leg is useful to avoid motion artifacts in clinical settings.