CT 画像二次元 NPS と仮想スリット法による一次元 NPS の中央断面定理に基づいた関係
成田 啓廣大久保 真樹深谷 貴広能登 義幸
著者情報
ジャーナル フリー

2021 年 77 巻 8 号 p. 828-832

詳細
抄録

Purpose: The method using a numerical slit (slit method) is used commonly to obtain the one-dimensional (1D) noise power spectrum (NPS) in computed tomography. However, the relationship between the 1D-NPS obtained by the slit method and the original two-dimensional (2D) NPS derived by the 2D Fourier transformation has not been elucidated clearly. The purpose of this study was to clarify their relationship based on the well-known central slice theorem (projection slice theorem) and validate it using computer simulation analysis. Methods: With the application of the central slice theorem, we described that the 1D-NPS obtained by the slit method was equal to the central slice (profile) in the 2D-NPS when we set the slit length to the maximum (i.e. the matrix size of the noise image). To verify this, we generated computer-simulated noise images with the known 2D-NPS (true 2D-NPS). From those images, we obtained the 1D-NPS that was obtained by the slit method and compared it with the central slice in the true 2D-NPS. Results: When we set the slit length to the maximum, the 1D-NPS obtained by the slit method showed good agreement with the central slice in the true 2D-NPS. Conclusion: We clarified the relationship between the 1D-NPS obtained by the slit method and the 2D-NPS using a theoretical approach and the computer simulation. We had to maximize the slit length to achieve the accurate measurement of the 1D-NPS using the slit method.