ノート
動体追跡システムを用いた 呼吸同期 VMAT-SBRT の線量検証
川村 哲朗藤井 雄介都倉 真典金田 和典藤田 雅大村上 直基栗山 由紀治宮脇 大輔土師 守
著者情報
ジャーナル フリー

2020 年 76 巻 7 号 p. 674-688

詳細
抄録

Recently, the introduction of various novel technologies in clinical settings has improved the accuracy of radiation therapy. Stereotactic body radiation therapy (SBRT) involves the delivery of an accurate radiation dose to the tumor with a minimal impact on normal tissues using various measures to address changes in the tumor position due to respiratory displacement. The SyncTraX FX4 real-time tumor tracking system (Shimadzu Corporation) introduced in our hospital tracks the actual tumor location by radioscopically monitoring a metallic marker that is placed in the vicinity of the tumor. However, there have been no reports yet on respiratory-gated volumetric modulated arc therapy (VMAT)-SBRT using a real-time tumor tracking system. This study aimed to develop an irradiation procedure for respiratory-gated VMAT-SBRT using a real-time tumor tracking system and to evaluate radiation doses therein. In this study, we found that absolute doses with respiratory gating did not deviate by more than ±1.0% from those without respiratory gating. In addition, the pass rate in gamma analysis using GAFCHROMIC EBT3 was ³95% with the pass criteria in dose difference, distance to agreement, and threshold being 2%, 2 mm, and 10%, respectively. Furthermore, a trajectory log file analysis did not reveal any significant error causes. Thus, these data indicate that respiratory-gated VMAT-SBRT can be applied clinically.