Immobilization of DNA on Quartz Crystal Microbalance Sensor Modified with Self-Assembled Monolayer of Thiol Derivative
Hirobumi ShibataJunpei NemotoShota ShibaYusuke YamaneGota KawaiKazuaki Hashimoto
Author information
JOURNAL FREE ACCESS

2020 Volume 69 Issue 3 Pages 271-276

Details
Abstract

In this study, we investigated the direct detection of DNA, without pretreatment, using a quartz crystal microbalance (QCM) sensor. This sensor is modified by a self-assembled monolayer of a thiol derivative that has an amino group as the terminal functional group. Contact angle values and the attenuated total reflectance Fourier transform infrared (ATR/FT-IR) spectra of the QCM sensors after immersion into an ethanol solution of thiol derivatives clearly showed that self-assembled monolayers of the derivatives were formed on the QCM sensors. Although QCM measurements using unmodified and carboxylic group-modified sensors could not detect DNA-Na salt, the sensor modified with amino groups could detect the DNA. This system can be used for the analysis of the interaction between DNA and DNAbinding proteins.

Content from these authors
© 2020 by Japan Oil Chemists' Society
Previous article