J Adv Prosthodont. 2022 Dec;14(6):346-359. English.
Published online Dec 22, 2022.
© 2022 The Korean Academy of Prosthodontics
Original Article

Biomechanical investigation of maxillary implant-supported full-arch prostheses produced with different framework materials: a finite elements study

Mirac Berke Topcu Ersöz,1 and Emre Mumcu1,2,3
    • 1Department of Prosthodontics, Faculty of Dentistry, Eskisehir Osmangazi University, Eskisehir, Turkey.
    • 2Advanced Material Technologies Application and Research Center, Eskisehir Osmangazi University, Eskisehir, Turkey.
    • 3Translational Medicine Research and Clinical Center, Eskisehir Osmangazi University, Eskisehir, Turkey.
Received August 20, 2022; Revised October 18, 2022; Accepted October 19, 2022.

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

PURPOSE

Four and six implant-supported fixed full-arch prostheses with various framework materials were assessed under different loading conditions.

MATERIALS AND METHODS

In the edentulous maxilla, the implants were positioned in a configuration of four to six implant modalities. CoCr, Ti, ZrO2, and PEEK materials were used to produce the prosthetic structure. Using finite element stress analysis, the first molar was subjected to a 200 N axial and 45° oblique force. Stresses were measured on the bone, implants, abutment screw, abutment, and prosthetic screw. The Von Mises, maximum, and minimum principal stress values were calculated and compared.

RESULTS

The maximum and minimum principal stresses in bone were determined as CoCr < ZrO2 < Ti < PEEK. The Von Mises stresses on the implant, implant screw, abutment, and prosthetic screws were determined as CoCr < ZrO2 < Ti < PEEK. The highest Von Mises stress was 9584.4 Mpa in PEEK material on the prosthetic screw under 4 implant-oblique loading. The highest maximum principal stress value in bone was found to be 120.89 Mpa, for PEEK in 4 implant-oblique loading.

CONCLUSION

For four and six implant-supported structures, and depending on the loading condition, the system accumulated different stresses. The distribution of stress was reduced in materials with a high elastic modulus. When choosing materials for implant-supported fixed prostheses, it is essential to consider both the number of implants and the mechanical and physical attributes of the framework material.

Keywords
Biomechanics; Finite element analysis; Polyetheretherketone; Zirconia; Titanium

INTRODUCTION

Tooth loss is an irreversible condition that severely impacts an individual’s overall quality of life. Edentulism leads to the loss of function as well as physical, mental, and social impairments for the patient.1 Implant therapy has superceded traditional treatments as the treatment of choice for the replacement of single and multiple tooth loss. Full mouth implant treatments with fixed prostheses are especially popular among edentulous patients and have become common in contemporary dentistry.2, 3

A concept of four implant-supported fixed prostheses, often known as ‘All-on-4’, developed by Malo et al., is an alternative to Dr. Misch’s conventional fixed implant treatments.4, 5 By deliberately placing a limited number of implants and avoiding complicated surgical procedures, this method aimed to save recovery time and expenses (such as bone grafting and maxillary sinus augmentation). Two axial implants and two distally tilted (30° - 45°) implants are positioned anteriorly and posteriorly, respectively.5, 6, 7 In addition, a modest number of implants may be used to provide prosthetic rehabilitation for patients who are unable to get fixed prosthetic therapy. Despite research comparing the biomechanics of full-arch prosthesis supported by four implants, the lack of studies comparing six- and four-implant modalities over a wide range of materials is significant.8, 9, 10, 11, 12, 13 Previous research has shown that designs with 4-6 implants provide the desired functional, biological, and cosmetic outcomes.14, 15 However, they must be complemented with other materials.

For optimal clinical performance, dental materials must possess the necessary physical, chemical, and biological characteristics. The material for the permanent restoration must be chosen depending on the area being restored, aesthetic standards, the dentist’s suggestions, and the patient’s budget. For implant-supported fixed prostheses, there are a variety of framework material alternatives. Resins, polymeric materials, metal alloys, and ceramics are commonly used in therapeutic applications due to their aesthetic and mechanical properties. When choosing a material for implant-supported prostheses, it is essential to examine the biomechanical properties. Due to their enhanced physical and mechanical features, including durability, biocompatibility, corrosion resistance, and bond strength with ceramics, cobalt-chromium alloys have a wide variety of applications.16 Titanium alloys, which have a high melting temperature, resistance to deformation, and low modulus of elasticity, are preferred for implants, abutments, and prosthetic frameworks.17, 18 In comparison to other ceramics, zirconia provides improved mechanical and biological features, including a reduced retention of bacterial plaque and excellent shade compatibility.19 High-performance polymer polyetheretherketone (PEEK) is a non-toxic, biocompatible product. It has a high level of abrasion resistance, is chemically stable, and supports implants, interim abutments, gingiva formers, and frameworks for implant-supported prostheses in the field of dental implantology.20, 21 Regarding CoCr, ZrO2, Ti, and current PEEK material, whose in vitro testing for clinical outcomes are still being conducted, it is essential to assess alternative implant configurations in terms of clinical implications, in comparison to the literature. The use of distally tilted implants is significant when considering the clinical indications from a wider perspective and improving the quality of life with long-term usage of 4 and 6 implant modalities.9, 10, 22, 23

Various in vitro experiments are performed on dental materials to clarify their physical and mechanical behaviour. Finite element analysis (FEA) is a method for digitally simulating complex structures by using mathematical principles. FEA provides quantitative data for the thorough examination of complex structures, such as bones, implants, and prostheses.24

This study aimed to assess the biomechanical properties of the 4 and 6 implant configurations with distally tilted implants and different framework materials used in the fabrication of implant-supported full-arch fixed maxillary prostheses by using 3-dimensional finite element analysis. The null hypothesis was that implant configurations and framework materials affect stresses in the bone, implant, implant screw, framework and prosthetic screw.

MATERIALS AND METHODS

The upper jaw model with teeth (Frasaco, Tettnang, Germany) was scanned using a 3D desktop scanner (Dental Wings Inc., Montreal, Canada) and transmitted to 3D design software (Exocad, Darmstadt, Germany). In the software, tooth components on the model are eliminated and rendered implant-compatible (Fig. 1A). The edentulous models were saved in .stl format, and implant placement was accomplished using a 3D modelling tool (Ansys, Canonsburg, PA, USA). Implants with internal conical connection (Bioinfinity Dental Implant System, Istanbul, Turkey), a diameter of 4.2 mm and a length of 14 mm are appropriately positioned at bone level on 3D edentulous models (Fig. 1B). Mucosa was not included in the model since its effect on the distribution of stress in the peri-implant bone is minimal.

Fig. 1
(A) Raw image of edentulous model, (B) Six and four implant placement, (C) Implant, abutment (straight and 30° angled) and prosthetic screw, (D) Six and four implant supported framework design.

For the four implant-supported designs, implant placement is bilateral, axial to the lateral incisors, and 30° distal inclination in the second premolars (Fig. 1C). Implants were installed bilaterally, axially to the lateral incisors and first premolars, and 30° distal inclination to the first molars in six implant-supported configurations. As indicated in earlier research, the accumulation of stress on implants with more than 30° demonstrates a considerable increase, hence this angulation was selected. The trial version of the program Rhino 7.0 (Robert McNeel & Associates, Seattle, WA, USA) was used to insert straight and angled multi-unit abutments with a 2 mm gingival height on implants. Using the appropriate dental program Exocad (Darmstadt, Germany), the one-piece hybrid structure was designed. Model is created for four implant-supported designs with a 10 mm cantilever length and six implant-supported designs without cantilevers (Fig. 1D). The connector thickness for both designs was fixed at 5 mm fasciolingually and occlusoapically. By attaching prosthetic screws to the framework design, the main model was created.

Geometric models were constructed and solid meshes were used for 3D static analysis (Fig. 2A). Models were created using 10-node tetrahedral elements in order to achieve the highest-quality mesh structure with the maximum number of nodes possible. In locations near the center of the structures in the models, fewer nodal elements are employed when required to complete the structure. In order to assist the computation, the finest quality mesh structure with the greatest number of nodal elements has been created using this modeling approach. To improve the analysis’s validity, elements as many as feasible were created. In addition, the mesh structure was increased and compressed in areas where force would be applied (Fig. 2B). The mathematical model employs 492409 elements and 778014 nodes for the four-implant design and 484334 elements and 758370 nodes for the six-implant design. Convergence test is the process of approximating the real result to equations solved using the Newton-Raphson method in the background while Ansys is running. Ansys does numerical analysis instead of analytical analysis. The result is always shown in numerical analysis; that is, if the result converges, a solution is achieved. This signifies that the analysis has reached a point of convergence. The non-converging analysis indicates that the boundary conditions are not correctly constructed or that the model contains an inaccuracy. Since the results of our study were compatible with and comparable to those of the present literature, no further tests were conducted beyond the software’s convergence test. According to earlier research, the thickness of the cortical bone in the maxilla was 1 mm and the rest was cancellous bone to simulate Type 3 bone, to make the results comparable.8, 9, 25 All materials evaluated were considered homogeneous, isotropic, and linearly elastic. Table 1 contains the elastic modulus and Poisson’s ratio values, which define the physical properties of the structures necessary for model construction.9, 26, 27 On the model, implant and prosthetic structures are specified to match the real morphology.

Fig. 2
(A) Finite element model with mesh structure, (B) Element density in the force applied region, Definition of vertical (C) and angled (D) forces.

Table 1
Elasticity modulus and Poisson’s ratio of materials

The created model is fixed by holding it in all degrees of freedom. As stated in previous research, cortical and cancellous bone, bone implants, and all connected components such as abutments, frameworks, and screws were thought to be perfectly attached along their contact surfaces, with no relative movement throughout junctions. It was also assumed that osseointegration at the bone-implant contact was completed.9, 10 The following describes the applied forces into the system: 1. assuming centric occlusion, 200 N applied vertically on the palatal tuberosity of the first molar (Fig. 2C). 2. 200 N applied at a 45° angle to the palatal tuberosity of the first molar (Fig. 2D). As a result, a total of 16 finite element analyses were performed for 2 designs, 4 framework materials, and 2 loading conditions (4 implant-axial loading = 4A, 4 implant-oblique loading = 4O, 6 implant-axial loading = 6A, 6 implant-oblique loading = 6O).

Ansys was utilized for the FEA. For ductile materials, the Von Mises stress (σVM) was determined, whereas for non-ductile materials, the maximum principal stress (σMax) and minimum principal stress (σMin) were determined for each framework material. Additionally, the total deformation of the system was examined.

RESULTS

The stress values obtained in all groups are shown in Figure 3A-F. Regardless of the number of implants, the PEEK framework showed the highest values when the σMax was measured in bone. Similar results were found in CoCr, ZrO2, and Ti frameworks. In both loading circumstances, the bone area corresponding to the implant neck showed the highest values (Fig. 4). The highest σMax value for the bone was obtained at 120.89 MPa in the PEEK framework in 4O. The lowest σMin value was obtained as -121,7 MPa in 6O. In addition, the σMax values in the 6-implant design are lower than the 4-implant design.

Fig. 3
σmax, σmin values (MPa) in bone (A) and σVM values (MPa) occurring on the implant (B), abutment (C), prosthetic screw (D), implant screw (E), framework (F) and total deformation (G) under all boundary conditions on the models.

Fig. 4
σmax, σmin stress maps for bone. 1: CoCr, 2: Ti, 3: ZrO2, 4: PEEK for different loading conditions.

The PEEK framework provided the highest values for all scenarios when the σVM stresses on the implant were investigated. The highest σVM value was 2140.2 MPa on 4O. The lowest σVM value was 358.1 MPa in CoCr on 6A. The neck of the implant received higher stresses in all conditions. Comparing the 4 and 6-implant designs, the stresses were significantly reduced in the 6-implant design under both loading conditions (Fig. 5).

Fig. 5
σVM stress maps for implant. 1: CoCr, 2: Ti, 3: ZrO2, 4: PEEK for different loading conditions.

The highest σVM on the abutment in PEEK was 4900.5 MPa on 4O and the lowest value was 151.9 MPa in CoCr at 6A design. It was also found that the stress for PEEK decreases by 94.6% for axial loading and 74.11% for oblique loading. The stresses were concentrated on the implant-abutment connection side (Fig. 6).

Fig. 6
σVM stress maps for abutment. 1: CoCr, 2: Ti, 3: ZrO2, 4: PEEK for different loading conditions.

The highest σVM achieved on an implant screw was 462.1 MPa in PEEK under 4O conditions. The lowest value measured for CoCr under 6A conditions was 38.3 MPa (Fig. 7).

Fig. 7
σVM stress maps for implant screw. 1: CoCr, 2: Ti, 3: ZrO2, 4: PEEK for different loading conditions.

PEEK had the highest σVM stress value at 554.9 MPa on 4O. Ti had the lowest value at 143.5 MPa on 6A. The stress in 4O and 4A is localized in the second premolar region (Fig. 8), whereas the stress in 6A and 6O is concentrated in the molar and premolar regions, respectively. On 6O, only PEEK accumulates stresses in the molar area. Lower stress accumulation was observed for all materials in 6 implant-supported frameworks under both loading scenarios.

Fig. 8
σVM stress maps for framework and prosthetic screw. 1: CoCr, 2: Ti, 3: ZrO2, 4: PEEK for different loading conditions.

The highest σVM stresses on the prosthetic screw were found on 4O, 9584.4 MPa in PEEK (Fig. 8). CoCr showed the lowest values, 57.6 MPa on 6A. Under all conditions, the neck of the screw is the area where the force is concentrated. However, in 6 implant-supported structures, lower stress values were achieved.

The maximum deformation of the system is 2,514 mm at 4O in the PEEK framework. The lowest total deformation was 0.027 mm in the CoCr framework at 6A (Fig. 3G).

DISCUSSION

Compared to implant-supported overdentures, treatment with implant-supported fixed prostheses is preferred by clinicians and patients for the rehabilitation of edentulous patients because it improves masticatory function and patient satisfaction. However, in order to distribute forces and offer a biomechanical benefit, this kind of fixed prosthesis necessitates the insertion of multiple implants. One of the most significant objectives of implant dentistry practice is to provide successful treatment while employing a small number of implants and avoiding complex surgical methods. Planning is the key to successful prosthetic treatments, and it’s critical to anticipate the impacts of potential requirement on the system. Finite element analysis can be used to examine objects with complex geometries and a variety of materials.24 Since FEA is a simulation, the accuracy of the data is improved by increasing the mesh density in the regions to be evaluated on the model. Like previous finite element analysis-based implantology research, the current work makes the assumption that all modelled structures are in constant contact with each other.28, 29, 30 Clinically, however, complete bone-implant contact is not achievable. Therefore, when interpreting the results, this restriction should be considered. In these treatment modalities, it is advised to select implants with a minimum width of 4 mm and a length of 10 - 18 mm for both posterior and anterior position.5, 6 As indicated in earlier research, the accumulation of stress on implants with more than 30° demonstrates a dramatic increase, so the posterior implants were positioned at this angle.12, 31, 32 There is currently no standard method for delivering stress to fixed full-arch restorations, despite of previous research using a wide variety of loading factors and locations. However, in the present study, force was applied to molar area where mastication is mostly active in order to observe the effect of cantilever on the system.28, 33, 34 Therefore, the loading condition of 200 N on molar area applied unilaterally was designated. Based on the FEA results, the σVM, σMax and σMin were assessed. The structure’s tensile stresses were represented by the σMax, whereas the compressive stresses are σMin. Von Mises stresses are recommended for the interpretation of ductile materials and principal stresses for brittle materials.35

Many different materials such as CoCr, titanium, zirconia, PEEK, PEKK, reinforced plastic, and chromium-nickel can be used as framework materials in implant-supported fixed prosthesis.34, 36, 37, 38 These four framework materials, CoCr, titanium, zirconia, and PEEK, which are the most commonly used and developed, were used in our research.

σMax and σMin are compared when evaluating the forces exerted on the bone. When the compressive force exceeds 170-190 MPa and the tensile force exceeds 100-130 MPa in the cortical bone, it is considered that the risk of resorption owing to excessive stress build-up.39 In all analysis, the alveolar crest experienced greater stress. The σMax and σMin are higher in PEEK under all conditions, whereas the CoCr material transmits the least force to bone. The values obtained are lower than those reported for bone resorption, which is also consistent with earlier research.9, 10 The σVM was concentrated at the implant’s neck in each tested location, framework, and loading scenario.8, 40, 41, 42

σVM accumulated at the neck of the abutment in 4 implant-supported designs. However, in 6 implant-supported designs, they were concentrated on the implant-abutment junction for all materials in all loading conditions, except on the cervical area in PEEK under oblique loading. Under oblique loading, PEEK with the lowest modulus of elasticity exhibited the same behaviour. For both loading scenarios and all materials, σVM on the 4-implant supported design accumulated on the 2nd premolar and posterior implant, and these findings are consistent with previous research.8, 9 For oblique loading, only PEEK with the 6O condition showed concentration in the molar area, whereas all other materials deposited force in the premolar region. PEEK accumulated stress during oblique loading rather than distributing it, in contrast to the stiffer CoCr, ZrO2 and Ti. As a result, the stresses on the CoCr were the highest under oblique loading while those on the PEEK were the highest under vertical loading. Stiffer materials also contributed to a more uniform distribution of stresses throughout the abutment, implant, and framework. PEEK with a low elastic modulus transmits more stress to implant structures and bone. However, it absorbs stress in the presence of a cantilever. This indicates that it protects the prosthetic structure and is characterized by its polymeric structure. However, high stress levels may result in bone resorption and implant fracture with long-term usage.

In comparing the loading of CoCr and Ag-Pd frameworks, Rubo et al. found that CoCr transferred less stress to the implant and abutment.28 However, in our study, the PEEK experienced increased stress on the implant-abutment complex. Similarly, PEEK crowns caused greater stress accumulation on the abutment, according to the study by Manchikalapudi et al.43 Similarly, according to Kelkar et al.,22 PEEK full-arc prosthetic framework showed higher stress values than zirconia and titanium. In accordance with our findings, Yu et al. found that zirconia and metals caused lower stresses on bone and implant, and greater stresses on framework, as compared to polymeric frameworks.44, 45 According to the results of the study in which full-arch implant supported prosthesis were evaluated with different frameworks by strain gauge analysis, PEEK showed higher deformation values than CoCr and ZrO2 in the presence of cantilever.11

Under both loading parameters, the majority of the stress on the prosthetic screw was concentrated in the cervical region. The load on the screw was concentrated in the cervical region because it was unable to dissipate due to the force applied to the prosthesis’ terminal location. In the study by Dayan et al.,9 which examined the force distribution on the system employing Ti, ZrO2, PEEK, and PEKK frameworks in 4 implant-supported designs, the force on the prosthesis screw was larger with PEEK and PEKK materials, and the load was concentrated in the cervical area of the screw. Additionally, Bhering et al.8 examined the impact of material differences on the system in 4 and 6 implant supported structures, and they observed that the load on the prosthetic screw was higher in Ti and the force on the prosthetic screw was higher in 6-implant supported structures. In our study, the force on the prosthetic screw was determined in decreasing order as PEEK, Ti, ZrO2, and CoCr in all scenarios. However, six implant-supported designs had lower stresses on the prosthesis screws, and these findings varied. PEEK and Ni-Cr bars were used in the study of 4 implant-supported fixed prostheses by Jaros et al. They concluded that stresses are concentrated on PEEK and, similar to our study, the stress on bone, implant, and implant components is higher.46

In the present study, all of the scenarios with the presence of a cantilever on the system cause excessive stress.22, 33 Considering the influence of the number of implants on the structure, it is evident that six implants distribute stress more uniformly. In the study of Fazi et al.,33 which investigated stress distribution in 3-4-5 implant-supported prosthesis, it was also shown that load accumulation on the system decreased with 5 implant-supported prosthesis. Similar to the present study, Almeida et al.34 examined maxillary prostheses supported by 4 and 6 implants. They concluded that shortening the cantilever length provides a reduction in the total stress on the system.

It is also interesting that PEEK can respond differently under oblique and axial load among the results of a study. This is based on the material’s capability for shock absorption and low elastic modulus. In the study by Sirandoni et al.,10 in which they investigated the loading of Ti, CoCr, ZrO2, PEEK and PMMA materials in implant-supported prostheses, PEEK and PMMA frameworks with a low modulus of elasticity had the highest total deformation, while CoCr and ZrO2 had the lowest.

Current dental materials have a continually expanding range of applications. Numerous variables influence the long-term clinical success of a novel material. Due to the low elastic modulus in clinical applications within the distance between implants and biomechanical rules, these characteristics must be taken into account for the long-term clinical success of PEEK frameworks when evaluating the results of this study and applying PEEK, which has a wide range of applications. Higher confidence intervals exist for conventional materials such as CoCr and ZrO2. With the increase in lifetime and longevity of the produced system, 6 implant model is considered a more reliable system in terms of stress distribution and implant survival than 4 implant structures. Due to anatomical variances and constraints, implants placed up to the posterior area offer a secure foundation for therapeutically applicable frameworks.

The outputs of FEA studies rely on the initial data provided in the system, which are considered as limitations. All materials are introduced to the system as homogenous, isotropic, and linear elastic throughout the creation of the model. Despite the use of this method in FEA investigations, the real reaction of oral tissues cannot be adequately examined. Further in vitro testing of the mechanical behavior of the current framework materials and implant modalities under dynamic loads, followed by prospective clinical observational studies, is necessary to confirm the results.

CONCLUSION

The following conclusions were drawn from the results of the study: the distribution of stress on a structure is influenced by implant quantity and framework material. The existence of a cantilever increases the accumulation of stress on the bone, the implant, the abutment, the framework, and the prosthetic screw. As the material’s elastic modulus increases, the forces transferred to the bone, implant, abutment, and prosthetic screw decrease. The PEEK material that had the lowest elastic modulus displayed varied biomechanics depending on the loading conditions. For the long-term success of implant-supported fixed prostheses, it is important to have a good understanding of the mechanical and physical properties of the implant design, implant localization, and framework material, and needs to be supported by further more clinical and laboratory studies.

Notes

This research was supported by Eskisehir Osmangazi University Scientific Research Projects and assigned Project Code No: 202045A216.

ACKNOWLEDGEMENTS

The authors acknowledge Fatih Uysal for his contributions to the study.

References

    1. Emami E, de Souza RF, Kabawat M, Feine JS. The impact of edentulism on oral and general health. Int J Dent 2013;2013:498305
    1. Att W, Stappert C. Implant therapy to improve quality of life. Quintessence Int 2003;34:573–581.
    1. Heydecke G, Locker D, Awad MA, Lund JP, Feine JS. Oral and general health-related quality of life with conventional and implant dentures. Community Dent Oral Epidemiol 2003;31:161–168.
    1. Resnik R, Misch CE. Prosthetic options in implant dentistry. In: Resnik R, editor. Misch’s contemporary implant dentistry. 4th ed. Canada: Elsevier Health Sciences; 2020. pp. 436-449.
    1. Maló P, Rangert B, Nobre M. All-on-4 immediate-function concept with Brånemark System implants for completely edentulous maxillae: a 1-year retrospective clinical study. Clin Implant Dent Relat Res 2005;7 Suppl 1:S88–S94.
    1. Maló P, de Araújo Nobre M, Lopes A, Ferro A, Nunes M. The All-on-4 concept for full-arch rehabilitation of the edentulous maxillae: A longitudinal study with 5-13 years of follow-up. Clin Implant Dent Relat Res 2019;21:538–549.
    1. Maló P, de Araújo Nobre M, Lopes A, Francischone C, Rigolizzo M. “All-on-4” immediate-function concept for completely edentulous maxillae: a clinical report on the medium (3 years) and long-term (5 years) outcomes. Clin Implant Dent Relat Res 2012;14 Suppl 1:e139–e150.
    1. Bhering CL, Mesquita MF, Kemmoku DT, Noritomi PY, Consani RL, Barão VA. Comparison between all-on-four and all-on-six treatment concepts and framework material on stress distribution in atrophic maxilla: A prototyping guided 3D-FEA study. Mater Sci Eng C Mater Biol Appl 2016;69:715–725.
    1. Dayan SC, Geckili O. The influence of framework material on stress distribution in maxillary complete-arch fixed prostheses supported by four dental implants: a three-dimensional finite element analysis. Comput Methods Biomech Biomed Engin 2021;24:1606–1617.
    1. Sirandoni D, Leal E, Weber B, Noritomi PY, Fuentes R, Borie E. Effect of different framework materials in implant-supported fixed mandibular prostheses: a finite element analysis. Int J Oral Maxillofac Implants 2019;34:e107–e114.
    1. Shetty R, Singh I, Sumayli HA, Jafer MA, Abdul Feroz SM, Bhandi S, Raj AT, Patil S, Ferrari M. Effect of prosthetic framework material, cantilever length and opposing arch on peri-implant strain in an all-on-four implant prostheses. Niger J Clin Pract 2021;24:866–873.
    1. Tribst JPM, Campanelli de Morais D, Melo de Matos JD, Lopes GDRS, Dal Piva AMO, Souto Borges AL, Bottino MA, Lanzotti A, Martorelli M, Ausiello P. Influence of framework material and posterior implant angulation in full-arch all-on-4 implant-supported prosthesis stress concentration. Dent J (Basel) 2022;10:12
    1. Tiossi R, Gomes ÉA, Faria ACL, Rodrigues RCS, Ribeiro RF. Biomechanical behavior of titanium and zirconia frameworks for implant-supported full-arch fixed dental prosthesis. Clin Implant Dent Relat Res 2017;19:860–866.
    1. Maló P, Nobre Md, Lopes A. The rehabilitation of completely edentulous maxillae with different degrees of resorption with four or more immediately loaded implants: a 5-year retrospective study and a new classification. Eur J Oral Implantol 2011;4:227–243.
    1. Maló P, de Araújo Nobre M, Lopes A, Moss SM, Molina GJ. A longitudinal study of the survival of All-on-4 implants in the mandible with up to 10 years of follow-up. J Am Dent Assoc 2011;142:310–320.
    1. Al Jabbari YS, Koutsoukis T, Barmpagadaki X, Zinelis S. Metallurgical and interfacial characterization of PFM Co-Cr dental alloys fabricated via casting, milling or selective laser melting. Dent Mater 2014;30:e79–e88.
    1. Özcan M, Hämmerle C. Titanium as a reconstruction and implant material in dentistry: advantages and pitfalls. Materials 2012;5:1528–1545.
    1. Roach M. Base metal alloys used for dental restorations and implants. Dent Clin North Am 2007;51:603–627.
    1. Al-Amleh B, Lyons K, Swain M. Clinical trials in zirconia: a systematic review. J Oral Rehabil 2010;37:641–652.
    1. Najeeb S, Zafar MS, Khurshid Z, Siddiqui F. Applications of polyetheretherketone (PEEK) in oral implantology and prosthodontics. J Prosthodont Res 2016;60:12–19.
    1. Zoidis P, Papathanasiou I, Polyzois G. The use of a modified Poly-Ether-Ether-Ketone (PEEK) as an alternative framework material for removable dental prostheses. a clinical report. J Prosthodont 2016;25:580–584.
    1. Kelkar KC, Bhat V, Hegde C. Finite element analysis of the effect of framework materials at the bone–implant interface in the all-on-four implant system. Dent Res J (Isfahan) 2021;18:1
    1. Fabris D, Moura JPA, Fredel MC, Souza JCM, Silva FS, Henriques B. Biomechanical analyses of one-piece dental implants composed of titanium, zirconia, PEEK, CFR-PEEK, or GFR-PEEK: Stresses, strains, and bone remodeling prediction by the finite element method. J Biomed Mater Res B Appl Biomater 2022;110:79–88.
    1. Reddy MS, Sundram R, Eid Abdemagyd HA. Application of finite element model in implant dentistry: a systematic review. J Pharm Bioallied Sci 2019;11 Suppl 2:S85–S91.
    1. Lekholm U. Patient selection and preparation. In: Brånemark PI, Zarb GA, Albrektsson T, editors. Tissue-integrated prosthesis: osseointegration in clinical dentistry. Chicago: USA: Quintessence Publishing; 1985. pp. 199-209.
    1. Barbier L, Vander Sloten J, Krzesinski G, Schepers E, Van der Perre G. Finite element analysis of non-axial versus axial loading of oral implants in the mandible of the dog. J Oral Rehabil 1998;25:847–858.
    1. Ferreira MB, Barão VA, Faverani LP, Hipólito AC, Assunção WG. The role of superstructure material on the stress distribution in mandibular full-arch implant-supported fixed dentures. A CT-based 3D-FEA. Mater Sci Eng C Mater Biol Appl 2014;35:92–99.
    1. Rubo JH, Capello Souza EA. Finite-element analysis of stress on dental implant prosthesis. Clin Implant Dent Relat Res 2010;12:105–113.
    1. Silva GC, Mendonça JA, Lopes LR, Landre J Jr. Stress patterns on implants in prostheses supported by four or six implants: a three-dimensional finite element analysis. Int J Oral Maxillofac Implants 2010;25:239–246.
    1. Faverani LP, Barão VA, Ramalho-Ferreira G, Delben JA, Ferreira MB, Garcia Júnior IR, Assunção WG. The influence of bone quality on the biomechanical behavior of full-arch implant-supported fixed prostheses. Mater Sci Eng C Mater Biol Appl 2014;37:164–170.
    1. Sannino G. All-on-4 concept: a 3-dimensional finite element analysis. J Oral Implantol 2015;41:163–171.
    1. Begg T, Geerts GA, Gryzagoridis J. Stress patterns around distal angled implants in the all-on-four concept configuration. Int J Oral Maxillofac Implants 2009;24:663–671.
    1. Fazi G, Tellini S, Vangi D, Branchi R. Three-dimensional finite element analysis of different implant configurations for a mandibular fixed prosthesis. Int J Oral Maxillofac Implants 2011;26:752–759.
    1. Almeida EO, Rocha EP, Freitas Júnior AC, Anchieta RB, Poveda R, Gupta N, Coelho PG. Tilted and short implants supporting fixed prosthesis in an atrophic maxilla: a 3D-FEA biomechanical evaluation. Clin Implant Dent Relat Res 2015;17 Suppl 1:e332–e342.
    1. Gokhale NS, Deshpande SS, Bedekar SV, Thite AN. Basics of statics and strength of materials. In: Gokhale NS, editor. Practical finite element analysis. India: Finite to infinite; 2008. pp. 35-49.
    1. Heimer S, Schmidlin PR, Roos M, Stawarczyk B. Surface properties of polyetheretherketone after different laboratory and chairside polishing protocols. J Prosthet Dent 2017;117:419–425.
    1. Nistor L, Grădinaru M, Rîcă R, Mărășescu P, Stan M, Manolea H, Ionescu A, Moraru I. Zirconia use in dentistry - manufacturing and properties. Curr Health Sci J 2019;45:28–35.
    1. Revilla-León M, Sánchez-Rubio JL, Pérez-López J, Rubenstein J, Özcan M. Discrepancy at the implant abutment-prosthesis interface of complete-arch cobalt-chromium implant frameworks fabricated by additive and subtractive technologies before and after ceramic veneering. J Prosthet Dent 2021;125:795–803.
    1. Baggi L, Pastore S, Di Girolamo M, Vairo G. Implant-bone load transfer mechanisms in complete-arch prostheses supported by four implants: a three-dimensional finite element approach. J Prosthet Dent 2013;109:9–21.
    1. Bayrak A, Yaramanoğlu P, Kılıçarslan MA, Yaramanoğlu B, Akat B. Biomechanical comparison of a new triple cylindrical implant design and a conventional cylindrical implant design on the mandible by three-dimensional finite element analysis. Int J Oral Maxillofac Implants 2020;35:257–264.
    1. Ozan O, Kurtulmus-Yilmaz S. Biomechanical comparison of different implant inclinations and cantilever lengths in all-on-4 treatment concept by three-dimensional finite element analysis. Int J Oral Maxillofac Implants 2018;33:64–71.
    1. Özdemir Doğan D, Polat NT, Polat S, Şeker E, Gül EB. Evaluation of "all-on-four" concept and alternative designs with 3D finite element analysis method. Clin Implant Dent Relat Res 2014;16:501–510.
    1. Manchikalapudi G, Basapogu S. Finite element analysis of effect of cusp inclination and occlusal contacts in PFM and PEEK implant-supported crowns on resultant stresses. Med J Armed Forces India 2022;78:80–87.
    1. Yu W, Chen S, Ma L, Ma X, Xu X. Biomechanical analysis of different framework design, framework material and bone density in the edentulous mandible with fixed implant-supported prostheses: a three-dimensional finite element study. J Prosthodont. 2022 May 11; [doi: 10.1111/jopr.13532]
      Epub ahead of print.
    1. Yu W, Li X, Ma X, Xu X. Biomechanical analysis of inclined and cantilever design with different implant framework materials in mandibular complete-arch implant restorations. J Prosthet Dent 2022;127:783.e1–783.e10.
    1. Jaros OAL, De Carvalho GAP, Franco ABG, Kreve S, Lopes PAB, Dias SC. Biomechanical behavior of an implant system using polyether ether ketone bar: finite element analysis. J Int Soc Prev Community Dent 2018;8:446–450.