Article

Replacement of the Mouse LD50 Assay for Determination of the Potency of AbobotulinumtoxinA with a Cell-Based Method in Both Powder and Liquid Formulations

1
Ipsen Bioinnovation, Abingdon OX14 4RY, UK
2
Ipsen Biopharm Ltd., Wrexham LL13 9UF, UK
*
Author to whom correspondence should be addressed.
Toxins 2023, 15(5), 314; https://doi.org/10.3390/toxins15050314
Submission received: 16 March 2023 / Revised: 18 April 2023 / Accepted: 21 April 2023 / Published: 29 April 2023
(This article belongs to the Special Issue Botulinum Neurotoxins: From Toxin to Medicine)

Abstract

:
Botulinum neurotoxins (BoNTs) are important therapeutic agents. The in vivo median lethal dose (LD50) assay has been commonly used to measure the potency of BoNT commercial preparations. As an alternative, we developed cell-based assays for abobotulinumtoxinA in both powder (Dysport®, Azzalure®) and liquid (Alluzience®) formulations using the in vitro BoCell® system. The assays demonstrated linearity over 50–130% of the expected relative potency, with a correlation coefficient of 0.98. Mean recoveries of 90–108% of the stated potency were observed over this range. The coefficients of variation for powder and liquid formulations, respectively, were 3.6% and 4.0% for repeatability and 8.3% and 5.0% for intermediate precision. A statistically powered comparability assessment of the BoCell® and LD50 assays was performed. Equivalence was demonstrated between the assays for the liquid formulation at release and end of shelf life using a paired equivalence test with predefined equivalence margins. For the powder formulation, the assays were also shown to be equivalent for release samples and when determining loss of potency following thermal degradation. The BoCell® assay was approved for establishing the potency of abobotulinumtoxinA for both powder and liquid formulations in Europe and for the powder formulation only in the USA.
Key Contribution: First example of a cell-based assay approved for use in Europe for potency release testing of a ready-to-use liquid botulinum neurotoxin A product.

1. Introduction

Botulinum neurotoxins (BoNTs) are important therapeutics used widely in many medical applications, such as the treatment of neuromuscular spasms (e.g., spasticity and cervical dystonia). The therapeutic utility of BoNTs is expanding, and novel indications continue to emerge [1,2]. BoNTs are also widely used in aesthetic medicine.
All BoNTs are modular proteins consisting of a heavy chain and a light chain linked by a disulfide bond [3,4]. The activity of BoNTs involves a multistep process of heavy-chain-mediated cell binding to neuronal cells, endocytosis, and translocation of the light chain into the cytosol, followed by cleavage by the light chain of soluble N-ethylmaleimide-sensitive factor attachment receptor (SNARE) proteins [3,4]. Cleavage of SNARE proteins by the zinc endopeptidase action of the light chain results in disrupted fusion of synaptic vesicles containing neurotransmitters with the plasma membrane; this blocks the release of neurotransmitters and paralyzes muscle activity [3,4]. BoNT serotypes differ in their target SNARE proteins and cleavage sites; for example, BoNT-A and -E cleave synaptosomal-associated proteins of 25 kDa (SNAP-25), whereas BoNT-B, -D, -F, and -G cleave vesicle-associated membrane proteins, and BoNT-C cleaves both SNAP-25 and syntaxin [3,4].
There are several commercially available BoNT-A-containing products that are not medically interchangeable owing to differences in manufacturing processes and their measurement of potency; as such, each BoNT product has its own indications and dosage [5,6,7]. Historically, commercial BoNT-A products have been formulated as a powder for reconstitution prior to injection. However, new products in a ready-to-use liquid formulation are coming to market, with products being approved in Korea [8] and in Europe (abobotulinumtoxinA (aboBoNT-A) in a novel liquid formulation (Alluzience®)). Potency testing is a requirement for the registration of pharmaceuticals for human use [9]. The mouse median lethal dose (LD50) assay has long been used to assess BoNT potency. However, several limitations are associated with the LD50 assay, including extensive animal use, requirements for highly skilled personnel and facilities, lack of automation, and laboratory-dependent outcomes [9,10]. These drawbacks have led to the search for alternative and reliable approaches to assess the potency of BoNTs, with the consideration that any replacement potency assay must reflect all steps in the cellular intoxication pathway of BoNTs [10,11,12,13]. Although several alternative methodologies have been described previously, these have often been limited to the detection of the BoNT protein or have only assessed a single element of the BoNT mechanism of action, such as endopeptidase activity. Frequently proposed alternatives to the LD50 have been published without sufficient consideration of the regulatory and commercial requirements to be able to adopt these methodologies for the testing of pharmaceutical products [14]. The progress in the development of in vitro cell-based methods for BoNT potency determination provides an approach that incorporates the key aspects of the BoNT mechanism of action.
Dong et al. developed fluorescence reporters that respond to BoNT activity, including activity in living cells [15]. The reporter technology was further developed by BioSentinel Inc. (Madison, WI, USA) and used to develop the novel BoCell® cell line and assay [16,17], which was subsequently licensed by BioSentinel Inc. to Ipsen Biopharm Ltd (Wrexham, UK). The BoCell® cell line is an engineered murine neuroblastoma cell line that stably expresses a reporter composed of SNAP-25 fused to cyan and yellow fluorescent proteins (CFP, YFP) on the N- and C-terminals of SNAP-25, respectively. Following exposure, BoNT-A binds to and is then internalized by the cells where the light chain cleaves the SNAP-25 moiety of the reporter. Following cleavage, the C-terminal fragment containing the YFP fluorophore is rapidly degraded. Loss of YFP fluorescence relative to the BoNT-insensitive CFP fluorescence provides the quantitative readout of the assay (Figure 1) [16,17].
The objective of the current study was to assess the suitability of a potency method that uses the BoCell® assay as a replacement for the LD50 assay to test aboBoNT-A in both powder (Dysport®, Azzalure®) and liquid (Alluzience®) formulations. Specifically, the goal was to validate the BoCell® assay to determine its capability as a release test for commercial batches and to conduct a statistically powered comparability assessment of data from the cell-based assay (CBA) and the LD50 assay using release and stability samples of aboBoNT-A in both powder and ready-to-use liquid formulations.

2. Results

2.1. Assay Specificity and Demonstration of Toxin Uptake

AboBoNT-A (Ipsen, Paris, France) at a vial strength of 500 LD50 units (U) was used, with further additional confirmatory analyses being conducted with 125 U/vial for some evaluations, as indicated.
When the BoCell® cell line was incubated with a range of dilutions of aboBoNT-A or BoNT-A, a concentration–response relationship was observed (Figure 2). A control sample that contained only excipients (i.e., without aboBoNT-A, placebo) showed no response with the BoCell® assay; thus, specificity (i.e., the sensitivity of the fluorescence reporter in the BoCell® cells to aboBoNT-A activity) was established. Assay responsiveness was reduced when BoCell® cells were incubated with 5 pM of a fragment of BoNT-A that contained the heavy-chain receptor-binding domain of BoNT-A (HC/A), together with increasing concentrations of aboBoNT-A (Figure 2A). The receptor-binding domain acts as a competitive antagonist of the full-length toxin; therefore, inhibition of assay responsiveness indicates that the effect of aboBoNT-A requires its binding domain (widely accepted as the first step in the cellular mode of action of BoNTs). In addition, the BoCell® assay was not responsive to a recombinant preparation that contained the light chain of BoNT-A (LC/A), which lacked receptor-binding activity but had the proteolytic activity of the light chain (Figure 2B). Again, this indicates that BoNT-A cellular receptor-binding activity is required to elicit a response in the BoCell® assay. These data showed that the developed cell line exhibited selective concentration responsiveness and that it reflected key steps of the cellular intoxication pathway.

2.2. Assay Linearity, Accuracy, Repeatability, and Intermediate Precision of the BoCell® Assay

Following the successful development of the assay for testing aboBoNT-A in both powder and liquid formulations, we progressed to validation of the assay. The potency method for aboBoNT-A was a relative potency method, with a test and reference standard on each plate. To determine the potency for the powder formulation, a nonlinear quantitative response analysis using a four-parameter logistic curve fit was utilized; for the liquid formulation, a linear model utilizing parallel line analysis was employed. The BoCell® assay was validated according to requirements from the International Conference of Harmonisation [19] and guidance from the US Food and Drug Administration (FDA) [20]. The predefined acceptance criteria were selected for this assay based on the requirements of the method’s intended purpose (i.e., for lot release and stability potency testing of aboBoNT-A) and were in line with widely used criteria in the field of CBA development for biologics. A summary of the predefined criteria is presented in Table 1. Using an aboBoNT-A powder formulation at 500 U/vial and a liquid formulation at 200 U/vial, the linearity of the analytical procedure was tested at 50%, 70%, 80%, 100%, 120%, and 130% (for powder) and 50%, 75%, 100%, 115%, and 130% (for liquid) of the expected relative potency to assess whether results obtained with the BoCell® assay were directly proportional to the concentration of aboBoNT-A in the sample. When estimated potencies were plotted against the expected results, linearity was observed, with correlation coefficients (r2) of 0.9753 and 0.9777 for the powder and liquid formulations, respectively (Figure 3, Table 1). When the sample concentration was tested at a nominal 130%, relative to the reference standard in five and six independent assays for the powder and liquid formulations, respectively, accuracy was 94%, 95%, 94%, 101%, and 90% for powder and 102%, 112%, 102%, 108%, 108%, and 112% for liquid. At the nominal 100% level, accuracy was 91%, 98%, 96%, 99%, 95%, and 91% for powder and 108%, 114%, 105%, 109%, 105%, and 105% for liquid and, at the nominal 50% level, accuracy was 93%, 103%, and 96% for powder and 102%, 110%, 111%, 104%, 112%, and 99% for liquid. All results met the predefined acceptance criteria. Overall, recovery of 90–103% for powder and 99–112% for liquid of the stated potency over the range of 50–130% was recorded, which is within the acceptance criteria for assay accuracy (Table 1).
The repeatability of the assay for the powder formulation was assessed using 500 U/vial and 125 U/vial samples of aboBoNT-A. When each sample was assayed six times, the coefficient of variation (%CV) was 3.6% for the 500 U/vial sample and 3.3% for the 125 U/vial sample (Table 1 and Table 2). To assess the repeatability of the assay with the liquid formulation, the %CV from results obtained for the same sample, assessed in three assays and performed by the same analyst at three test levels (total of nine results), was calculated, and a %CV of 4% was obtained. The results for both powder and liquid formulations were within the acceptance criterion of a %CV of 15% or below for repeatability.
When the powder formulation was tested by different analysts on different days to evaluate intermediate precision, a difference of 1.6% was observed between the means of the two analysts’ data for 500 U/vial samples, with an overall %CV of 8.3%, demonstrating acceptable intermediate precision (Table 1 and Table 3). For the liquid formulation, the %CV of the six reportable results at each test level was calculated. For intermediate precision to be considered acceptable, each level must have a %CV of 15% or below. The results are summarized in Table 4.
The performance of the BoCell® assay for aboBoNT-A was similar to the mouse LD50 assay when tested using 500 U/vial samples. The r2 correlation coefficient for the linearity of the mouse LD50 assay over a range of 50–130% was 0.99, and accuracy was 99–104%. For repeatability, triplicate determinations of potency were performed with a 500 U/vial sample diluted to 50%, 100%, and 130% of the target, and the %CV ranged from 4% to 10% over the three dilutional levels tested. For intermediate precision, six assays at the 100% level were performed, and the %CV of the LD50 assay was 2% with a 500 U/vial sample.

2.3. Comparability Assessment of Data from the LD50 Assay and BoCell® Assay Using Release and Stability Samples

Release samples from 30 batches of aboBoNT-A in the powder formulation and 11 batches of aboBoNT-A in the liquid formulation were used in the comparability studies. The mean (range) result from the BoCell® assay, as a percentage of the LD50 assay result, was 102% (89–112%) for powder and 101% (83–122%) for liquid.
Figure 4 shows a graphical representation of paired potency results from the LD50 assay and BoCell® assay for samples of 500 U/vial powder product, and Table 5 shows the summarized potency data for the liquid formulation batches. Results for the comparison of variance between the LD50 and CBA release comparability data sets indicated significantly higher variability in potency results from the BoCell® assay than from the LD50 assay (p = 0.02) in the powder formulation; however, a significant difference in variance between the LD50 and CBA data sets for the liquid formulation was not observed (p = 0.443). In the paired two one-sided t-test (TOST) for equivalence, the 90% confidence interval (CI) for the difference between the means for the BoCell® assay and LD50 assay was 0.9967–1.044 for powder and 0.9386–1.079 for liquid, which is within the equivalence margin of 0.9–1.1 in both cases. For the powder formulation, Pearson’s correlation coefficient was 0.93, demonstrating a highly positive linear correlation of the potency response over the two product strengths tested and within the acceptance criteria (ρ > 0.80).
In the stability analysis, loss in potency over time was determined for each powder formulation batch stored at 40 °C (Figure 5A). Graphical representation of paired stability slope results from the LD50 assay and BoCell® assay are shown in Figure 5B. Results from the F-test indicated that the LD50 assay slope variance was not significantly different from the variance in the BoCell® assay slope (p = 0.06). Potency data from CBA and LD50 from the 13 liquid formulation batches stored for 12 months at 2–8 °C are shown in Table 6. When the paired TOST for equivalence was performed, the 90% CI for the difference between the means for the BoCell® assay and LD50 assay was 0.8701–1.132 for powder and 0.9072–0.9976 for liquid, which was within the equivalence margin of 0.85–1.15 and 0.90–1.10, respectively. In the photostability study, which was not statistically powered, the BoCell® assay was capable of detecting photodegradation of aboBoNT-A and did so at a similar rate to the mouse LD50 assay (Figure 6).

3. Discussion

To ensure drug efficacy and patient safety, regulatory authorities worldwide request that manufacturers of BoNTs establish the potency of all batches that are intended for clinical or commercial use.
CBAs have the potential to replace mouse-based LD50 assays, leading to a drastic reduction in animal-based testing. However, any assay needs to be rigorously validated, and in cases in which an existing method is already in use, suitable bridging or comparability studies need to be performed before the new assay can be considered acceptable for use by a pharmaceutical company for regulatory batch-testing purposes [11,21,22]. Moreover, the replacement assay must be able to be used in a quality-controlled environment and at the high capacity needed to support commercial production. A cell-based potency assay was developed for onabotulinumtoxinA (Botox, Allergan Inc., Dublin, Ireland) that utilizes differentiated human neuroblastoma SiMa cells and a sandwich enzyme-linked immunosorbent assay readout, which measures BoNT-A-dependent intracellular increase of cleaved SNAP-25 [23]. A CBA has also been developed for incobotulinumtoxinA (Xeomin, Merz, Frankfurt, Germany [24]), although few details about the assay are publicly available.
The data from the BoCell® assay for aboBoNT-A reported here showed selective responsiveness and the ability to reflect key steps of the cellular intoxication process. The assay also showed specificity, linearity, accuracy, repeatability, and intermediate precision, all of which were within the acceptance criteria chosen for aboBoNT-A in this study. The assay was also shown to have suitable performance characteristics for use with both powder and liquid formulations. This is the first example of a CBA being used for potency release of a product containing BoNT in a liquid formulation. These acceptance criteria were established based on the product requirements and the authors’ experience in developing CBAs for biologics, as well as according to requirements from the International Conference of Harmonisation [15] and guidance from the FDA [16]. In this study, the BoCell® assay for aboBoNT-A was equivalent to the mouse-based LD50 assay for determining the potency of release samples for both powder and novel ready-to-use liquid formulations in statistically powered comparability assessments. Consistent with the release analysis, the stability analysis demonstrated the closeness of the LD50 assay and BoCell® assay data with respect to the mean stability slope observed in the powder formulation batches and the 12-month real-time stability results from the liquid formulation batches. Statistical analysis demonstrated equivalence and a positive linear correlation between the assays.
Based on these findings, the BoCell® assay has the potential to replace the mouse-based LD50 assay for potency determination of aboBoNT-A. The mouse LD50 assay, which has been the standard for BoNT potency and stability testing, has many serious disadvantages, including the use of many laboratory animals, expensive facilities, the need for highly trained operators, a lack of standardized methodology, and a lack of an international reference standard; all of these result in high variability in the assay and between laboratories [9,11]. The CBA reported here emulates all the essential steps in the BoNT intoxication process and has been shown to be selective, sensitive, specific, robust, and with acceptable levels of reproducibility and precision, which results in low variability in a quality-control setting of BoNT manufacture for patient use. Furthermore, the use of an established cell line facilitates automation of the assay and, in the future, the development and establishment of a common standardized methodology and reference standard for BoNT testing. It is acknowledged that a downside to the use of a genetically engineered immortalized cell line as the basis of an assay method is the potential for instability of the cell line over the lifetime of the assay. To mitigate this risk, the number of division cycles the cell line undergoes after its creation has been minimized by the use of a master cell bank from which working cell banks for the supply of cells to be cultured for the assay are generated. Cell bank stability protocols for both master and working cell banks have also been implemented.
To conclude, this study demonstrates that the BoCell® assay for aboBoNT-A achieves a level of precision and accuracy comparable to the mouse LD50 assay and is suitable to be used as a replacement for the LD50 assay to support product potency and quality testing of aboBoNT-A. Following the assay validation and comparability studies presented here, this in vitro assay has received approval from regulatory authorities in the European Union (both powder and liquid formulations) and in the USA (FDA; powder formulation) for establishing the potency, for batch release and stability purposes, of aboBoNT-A; submissions and approvals in other geographies are ongoing. In all countries in which Ipsen has received regulatory approval for the use of the CBA methodology presented here, Ipsen is actively using the CBA for aboBoNT-A powder and liquid product batch release and stability testing in place of the mouse LD50 method.

4. Materials and Methods

4.1. BoNT-A Preparations

Research-grade BoNT-A was purchased from Metabiologics Inc. (Madison, WI, USA). HC/A was made at BioSentinel Inc. (Madison, WI, USA), and LC/A was purchased from List Biological Laboratories Inc. (Campbell, CA, USA). AboBoNT-A powder formulation (Dysport®) was manufactured by Ipsen (Paris, France); 500 U/vial were used, with further confirmatory analyses being conducted with 125 U/vial for some tests. Vials were reconstituted with complete BoCell® assay media. After reconstitution, vials were pooled to provide a sufficient volume of BoNT-A to perform the assay.
AboBoNT-A liquid formulation (Alluzience®) was manufactured by Ipsen (Paris, France); 200 U/vial were used, and vials were pooled before being dialyzed against BoCell® assay media to provide a sufficient volume of aboBoNT-A to perform the assay.

4.2. BoCell® Assay Methodology

Cell culture and aboBoNT-A treatment were performed according to BoCell® technology (BioSentinel Inc). Briefly, 20,000 engineered BoCell® Neuro-2a cells/well were exposed to increasing concentrations of aboBoNT-A for a period of 48 h on 96-well microtiter plates. Mean relative fluorescence units (RFU) for YFP and CFP fluorescence output were recorded using a Tecan Infinite M1000Pro (Tecan Group, Männedorf, Switzerland), and mean background fluorescence was determined and subtracted from both emissions spectra. BoNT-A relative potency was calculated with PLA software (Stegmann Systems GmbH, Rodgau, Germany) using the half-maximal effective concentration values generated from fitting either a four-parameter logistic equation (for powder formulation) or a linear regression model (for liquid formulation) to the ratio of blank-subtracted YFP and CFP measurements for nine and seven (powder and liquid formulations, respectively) concentration points for both a reference and test sample. These reference and test curves were tested for parallelism before being fitted using a constrained model in which the values for the slope and the asymptotes were shared between the reference and test sample curves. Reference standards are produced internally by Ipsen Biopharm Ltd. using representative commercial batches of products that have undergone extended product characterization and potency testing.

4.3. Validation of the BoCell® Assay to Determine its Capability as a Release Test

The BoCell® assay was validated according to requirements from the International Conference of Harmonisation [19] and guidance from the FDA [20], and based on the authors’ experience in developing CBAs for biologics. A summary of the method validation parameters tested and the acceptance criteria is presented in Table 1.
Specificity was assessed using placebo vials that contained the excipients but no BoNT-A. The ability of the BoCell® assay to measure toxin potency through a mechanism thought to be the mode of action (i.e., cellular binding as its first step and SNARE cleavage activity as the final step) was also tested. BoCell® cells were incubated with either of the following, each in the range of 0.3–100 pM: (a) aboBoNT-A or aboBoNT-A together with HC/A; or (b) BoNT-A or LC/A, which lacks receptor-binding activity. Mean RFU for YFP and CFP fluorescence output were recorded as above.
The accuracy of the BoCell® assay was determined by comparing the potency obtained for each of the 500 U/vial samples at a nominal 50%, 100%, and 130% of the expected concentration. Acceptance criteria for the accuracy study were that recovery for the potency was within 84% and 116% of the target at each of the three levels tested (50%, 100%, and 130% of nominal concentration). Data from the accuracy study were supplemented with assays performed at 70%, 80%, and 120% potency levels to determine the linearity of the method (i.e., its ability to obtain results that are directly proportional to the concentration of analyte in the sample). Estimated potencies were plotted against the expected result for each of the concentrations with an acceptance criterion of a correlation coefficient (r2) of at least 0.90. For assay repeatability, samples of aboBoNT-A at strengths of 500 U/vial and 125 U/vial were assayed six times, and the precision of the assay was determined by evaluating the %CV obtained for each sample. An acceptance criterion for %CV of 15% or below was used. The intermediate precision of the assay was evaluated using the means and %CVs obtained from 12 assays of 500 U/vial samples, performed by two different operators on different days. Acceptance criteria used were a difference of no more than 10% between operator means and a %CV of 15% or below for different analysts on different days.

4.4. Comparability Assessment of Data from the LD50 Assay and the BoCell® Assay Using Release and Stability Samples

Statistically powered assessments comparing data from the LD50 assay and BoCell® assays were performed for both release and stability samples of aboBoNT-A in the powder formulation at 500 U/vial and in the liquid formulation at 200 U/mL.
In the mouse LD50 assay, results were expressed as LD50 U/vial, calculated according to the Spearman–Kärber method [25]. Animal experiments were performed by a third-party provider in accordance with the UK Animal (Scientific Procedures) Act 1986, with ethical approval from the UK Home Office granted under license number P8350E94C; 15 April 2019.
In order to demonstrate statistical comparability, an equivalence-based approach was preferred because this allows for an appropriate null hypothesis to be defined and avoids the pitfall of the inappropriate application of a difference test, where statistical equivalence is falsely declared when the test is unable to show a statistical difference between data sets. To analyze the comparability of release samples, a sample size of 30 batches for the powder formulation and at least 10 batches for the liquid formulation was selected. Sample sizes were chosen based on a prospective power calculation with the following assumptions: the ability to identify a difference of 10% between the LD50 methods; and BoCell® assay at α of 0.05, with a power of > 90% for powder and > 80% for liquid formulation. Data were analyzed using a paired TOST equivalence test performed in Minitab 17 statistical software (Minitab Ltd., Coventry, UK). In total, 30 powder formulation and 11 liquid formulation batches were tested using the BoCell® assay, and the mean results were expressed as a percentage of data obtained using the LD50 assay. A test for the equality of the variances between the LD50 and CBA data sets was performed using Bonnett’s method for the powder formulation and Levene’s test for the liquid formulation (the choice of the Levene’s test over the Bonnett’s method was due to the smaller size of the data set for the liquid formulation). A predefined equivalence margin of ±10% was set for the ratio of CBA:LD50 comparability data sets. The TOST for equivalence assessed if the difference between the potency determined by both assays was within the equivalence margin by assessing if the 90% CI of the mean of the CBA:LD50 ratio data set crossed either the upper or lower bound of the 10% equivalence margin. For the powder data set, a regression analysis was performed by pooling release data from 30 batches of 500 U/vial and seven batches of 300 U/vial data to ensure the regression was performed over a wide enough range of potencies. Pearson’s correlation coefficient was calculated for the relationship between the BoCell® and LD50 methods, with an acceptance criterion of ρ > 0.8.
For analyzing the comparability of stability in thermally stressed samples, seven powder formulation batches were placed at 40 ± 2 °C/75 ± 5% relative humidity to obtain sufficient and substantial degradation in a reasonable time frame. Loss of potency over time (slope), according to linear regression, was determined for each batch and data set from each assay. Thirteen liquid formulation batches were placed on stability at 2–8 °C and, after 12 months of storage, were tested for potency in both LD50 and CBA methods, with the results being expressed as a percentage of CBA:LD50 potency. The test for the equality of the variances and paired equivalence test were performed as above, with predetermined equivalence margins of ±15% for stability slopes for the powder formulation data and ±10% for the difference in CBA:LD50 potency at 12 months (2–8 °C) for liquid formulation samples. Non-statistically powered stability data were also generated to determine if there was a difference in the ability of the LD50 and BoCell® assays to detect reduced potency due to photostress. Samples from a single batch of aboBoNT-A powder 500 U/vial were exposed to 0, 300, 600, 900, or 1200 kLux.

Author Contributions

Conceptualization, all authors; methodology, E.F., E.M., L.-M.F., R.S. and S.C.; formal analysis, E.F., E.M., L.-M.F., R.S. and S.C.; investigation, E.F., E.M., L.-M.F. and R.S.; data curation, E.F., L.-M.F. and R.S.; writing—original draft preparation, E.M. and S.C.; writing—review and editing, all authors; visualization, E.F., E.M., L.-M.F., R.S. and S.C.; supervision, E.F., D.H., S.C. and A.K.; project administration, E.F., L.-M.F., R.S. and S.C. All authors have read and agreed to the published version of the manuscript.

Funding

The study and the article processing charge was funded by Ipsen Biopharm Ltd.

Data Availability Statement

Data underlying this manuscript may be available upon request.

Acknowledgments

The authors thank all team members who contributed to this study. Medical writing/editorial support, provided by Emma Marshman of Zenith Healthcare Communications Ltd, and Emma Westcott of Oxford PharmaGenesis Ltd, was funded by Ipsen.

Conflicts of Interest

Elizabeth Marks, Rebecca Schofield, Sam Coward, Daniel Higazi, and Alistair Kippen are employees of Ipsen. Elena Fonfria and Lisa-Marie Foulkes were employees of Ipsen at the time the work was completed.

References

  1. Fonfria, E.; Maignel, J.; Lezmi, S.; Martin, V.; Splevins, A.; Shubber, S.; Kalinichev, M.; Foster, K.; Picaut, P.; Krupp, J. The expanding therapeutic utility of botulinum neurotoxins. Toxins 2018, 10, 208. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Cocco, A.; Albanese, A. Recent developments in clinical trials of botulinum neurotoxins. Toxicon 2018, 147, 77–83. [Google Scholar] [CrossRef] [PubMed]
  3. Montal, M. Botulinum neurotoxin: A marvel of protein design. Annu. Rev. Biochem. 2010, 79, 591–617. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  4. Pirazzini, M.; Rossetto, O.; Eleopra, R.; Montecucco, C. Botulinum neurotoxins: Biology, pharmacology, and toxicology. Pharmacol. Rev. 2017, 69, 200–235. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  5. Samizadeh, S.; De Boulle, K. Botulinum neurotoxin formulations: Overcoming the confusion. Clin. Cosmet. Investig. Dermatol. 2018, 11, 273–287. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  6. Field, M.; Splevins, A.; Picaut, P.; van der Schans, M.; Langenberg, J.; Noort, D.; Snyder, D.; Foster, K. AbobotulinumtoxinA (Dysport®), OnabotulinumtoxinA (Botox®), and IncobotulinumtoxinA (Xeomin®) neurotoxin content and potential implications for duration of response in patients. Toxins 2018, 10, 535. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  7. Ferrari, A.; Manca, M.; Tugnoli, V.; Alberto, L. Pharmacological differences and clinical implications of various botulinum toxin preparations: A critical appraisal. Funct. Neurol. 2018, 33, 7–18. [Google Scholar] [CrossRef] [PubMed]
  8. Kim, J.E.; Song, E.J.; Choi, G.S.; Lew, B.L.; Sim, W.Y.; Kang, H. The efficacy and safety of liquid-type botulinum toxin type A for the management of moderate to severe glabellar frown lines. Plast. Reconstr. Surg. 2015, 135, 732–741. [Google Scholar] [CrossRef] [PubMed]
  9. International Council for Harmonisation Harmonised Tripartite Guideline Specifications. Test Procedures and Acceptance Criteria for Biotechnological/Biological Products Q6b. Available online: https://database.ich.org/sites/default/files/Q6B%20Guideline.pdf (accessed on 23 November 2022).
  10. Pellett, S. Progress in cell based assays for botulinum neurotoxin detection. Curr. Top. Microbiol. Immunol. 2013, 364, 257–285. [Google Scholar] [PubMed] [Green Version]
  11. Adler, S.; Bicker, G.; Bigalke, H.; Bishop, C.; Blümel, J.; Dressler, D.; Fitzgerald, J.; Gessler, F.; Heuschen, H.; Kegel, B.; et al. The current scientific and legal status of alternative methods to the LD50 test for botulinum neurotoxin potency testing. The report and recommendations of a ZEBET Expert Meeting. Altern. Lab. Anim. 2010, 38, 315–330. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  12. von Berg, L.; Stern, D.; Pauly, D.; Mahrhold, S.; Weisemann, J.; Jentsch, L.; Hansbauer, E.-M.; Müller, C.; Avondet, M.A.; Rummel, A.; et al. Functional detection of botulinum neurotoxin serotypes A to F by monoclonal neoepitope-specific antibodies and suspension array technology. Sci. Rep. 2019, 9, 5531. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  13. Behrensdorf-Nicol, H.A.; Wild, E.; Bonifas, U.; Klimek, J.; Hanschmann, K.M.; Krämer, B.; Kegel, B. In vitro potency determination of botulinum neurotoxin serotype A based on its receptor-binding and proteolytic characteristics. Toxicol. In Vitro 2018, 53, 80–88. [Google Scholar] [CrossRef] [PubMed]
  14. Pickett, A. Commentary: A cell line for detection of botulinum neurotoxin type B. Front. Pharmacol. 2018, 9, 1056. [Google Scholar] [CrossRef] [PubMed]
  15. Dong, M.; Tepp, W.H.; Johnson, E.A.; Chapman, E.R. Using fluorescent sensors to detect botulinum neurotoxin activity in vitro and in living cells. Proc. Natl. Acad. Sci. USA 2004, 101, 14701–14706. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  16. Atapattu, D.; Olivares, H.A.; Piazza, T.M.; Crawford, D.; Zeytin, N.; Tucker, W. Performance and mouse bioassay comparability of the BoCell™ A cell-based assay. Toxins 2013, 68, 105. [Google Scholar] [CrossRef]
  17. Piazza, T.M.; Geurs, T.L.; Hendricksen, S.F.; Zeytin, F.N.; Tucker, W.C. Development of a cell-based relative potency assay for botulinum neurotoxin-based drug products. In Proceedings of the 2nd Annual US BioPharmaceutical Emerging Best Practices Association (BEPBA) Bioassay Conference, San Diego, CA, USA, 7–9 March 2018. [Google Scholar]
  18. BioSentinel. Assays for Botulinum Neurotoxin Detection and Quantification. Available online: https://biosentinelpharma.com/files/BioSentinel_Brochure_2017-09.pdf (accessed on 23 November 2022).
  19. International Council for Harmonisation Harmonised Tripartite Guideline Specifications. Validation of Analytical Procedures: Text and Methodology Q2(R1). Available online: https://database.ich.org/sites/default/files/Q2%28R1%29%20Guideline.pdf (accessed on 23 November 2022).
  20. US Food and Drug Administration. Analytical Procedures and Methods Validation for Drugs and Biologics [Guidance Document]. Available online: https://www.fda.gov/files/drugs/published/Analytical-Procedures-and-Methods-Validation-for-Drugs-and-Biologics.pdf (accessed on 23 November 2022).
  21. Singh, A.K.; Stanker, L.H.; Sharma, S.K. Botulinum neurotoxin: Where are we with detection technologies? Crit. Rev. Microbiol. 2013, 39, 43–56. [Google Scholar] [CrossRef]
  22. National Institutes of Health. Report on the ICCVAM-NICEATM/ECVAM Scientific Workshop on Alternative Methods to Refine, Reduce or Replace the Mouse LD50 Assay for Botulinum Toxin Testing. Available online: https://ntp.niehs.nih.gov/iccvam/docs/biologics-docs/bontwkshprept.pdf (accessed on 23 November 2022).
  23. Fernández-Salas, E.; Wang, J.; Molina, Y.; Nelson, J.B.; Jacky, B.P.S.; Aoki, K.R. Botulinum neurotoxin serotype A specific cell-based potency assay to replace the mouse bioassay. PLoS ONE 2012, 7, e49516. [Google Scholar] [CrossRef] [Green Version]
  24. Merz Pharma GmbH & Co. KGaA. Alternative Test Method for Botulinum Neurotoxin Now Approved in Europe [Press Release]. Available online: https://www.merz.com//wp-content/uploads/2015/12/20151211PM-BfArM-Zulassung-alternative-Testmethode_EN_v.pdf (accessed on 23 November 2022).
  25. Morgan, B.J.T. Analysis of Quantal Response Data; Chapman and Hall: London, UK, 1992; Chapter 7. [Google Scholar]
Figure 1. Cartoon representation of the BoCell® cell-based assay (adapted from [18]). BoNT, botulinum neurotoxin; CFP, cyan fluorescent protein; YFP, yellow fluorescent protein.
Figure 1. Cartoon representation of the BoCell® cell-based assay (adapted from [18]). BoNT, botulinum neurotoxin; CFP, cyan fluorescent protein; YFP, yellow fluorescent protein.
Toxins 15 00314 g001
Figure 2. Concentration–response curves for demonstration of toxin-binding requirement: (A) aboBoNT-A, or aboBoNT-A together with 5 pM HC/A; and (B) BoNT-A or LC/A, lacking receptor-binding activity. AboBoNT-A, abobotulinumtoxinA; BoNT-A, botulinum toxin A; HC/A, heavy-chain receptor-binding domain of BoNT-A; LC/A, light chain of BoNT-A; U, units.
Figure 2. Concentration–response curves for demonstration of toxin-binding requirement: (A) aboBoNT-A, or aboBoNT-A together with 5 pM HC/A; and (B) BoNT-A or LC/A, lacking receptor-binding activity. AboBoNT-A, abobotulinumtoxinA; BoNT-A, botulinum toxin A; HC/A, heavy-chain receptor-binding domain of BoNT-A; LC/A, light chain of BoNT-A; U, units.
Toxins 15 00314 g002
Figure 3. Linearity of the BoCell® assay for detecting the potency of aboBoNT-A in: (A) powder formulation at 500 U/vial; and (B) liquid formulation at 200 U/vial. AboBoNT-A, abobotulinumtoxinA; U, units.
Figure 3. Linearity of the BoCell® assay for detecting the potency of aboBoNT-A in: (A) powder formulation at 500 U/vial; and (B) liquid formulation at 200 U/vial. AboBoNT-A, abobotulinumtoxinA; U, units.
Toxins 15 00314 g003
Figure 4. Graphical representation of paired potency results from the mouse LD50 assay and BoCell® assay using release samples of aboBoNT-A at 500 U/vial. AboBoNT-A, abobotulinumtoxinA; LD50, median lethal dose; U, units.
Figure 4. Graphical representation of paired potency results from the mouse LD50 assay and BoCell® assay using release samples of aboBoNT-A at 500 U/vial. AboBoNT-A, abobotulinumtoxinA; LD50, median lethal dose; U, units.
Toxins 15 00314 g004
Figure 5. Stability of thermally degraded samples (stored at 40 °C) of aboBoNT-A at 500 U/vial, as detected by the mouse LD50 assay and the BoCell® assay. (A) Stability profiles of each sample. (B) Graphical representation of paired stability slopes. AboBoNT-A, abobotulinumtoxinA; LD50, median lethal dose; U, units.
Figure 5. Stability of thermally degraded samples (stored at 40 °C) of aboBoNT-A at 500 U/vial, as detected by the mouse LD50 assay and the BoCell® assay. (A) Stability profiles of each sample. (B) Graphical representation of paired stability slopes. AboBoNT-A, abobotulinumtoxinA; LD50, median lethal dose; U, units.
Toxins 15 00314 g005
Figure 6. Stability profiles for photodegraded samples (exposed to up to 1200 kLux) of aboBoNT-A at 500 U/vial, as detected by the mouse LD50 assay and the BoCell® assay. AboBoNT-A, abobotulinumtoxinA; LD50, median lethal dose; U, units.
Figure 6. Stability profiles for photodegraded samples (exposed to up to 1200 kLux) of aboBoNT-A at 500 U/vial, as detected by the mouse LD50 assay and the BoCell® assay. AboBoNT-A, abobotulinumtoxinA; LD50, median lethal dose; U, units.
Toxins 15 00314 g006
Table 1. Acceptance criteria and results for method validation of the BoCell® assay.
Table 1. Acceptance criteria and results for method validation of the BoCell® assay.
ParameterAcceptance CriteriaResults for the BoCell® Assay
Powder FormulationLiquid FormulationPowder FormulationLiquid Formulation
SpecificityNo response when placebo tested vs. reference standardNo response when placebo tested vs. reference standardNo response seen with placeboNo response seen with placebo
Linearityr2 ≥ 0.90r2 ≥ 0.97
CI of the slope for linear regression must include 1
CI for the Y-intercept must include 0
r2 = 0.98r2 = 0.98
CI for slope = 0.99 to 1.13
CI for Y-intercept = −7.51 to 6.45
RangeAcceptable accuracy, precision, and linearity over the relative potency range of 50–130%Acceptable accuracy, precision, and linearity over the relative potency range of 50–130%Accurate, precise, and linear response observed in the range of 50–130%Accurate, precise, and linear response observed in the range of 50–130%
AccuracyRecovery for estimated potency of 84–116% of target potency over the interval of 50–130% relative potencyRecovery for estimated potency of 85–115% of target potency over the interval of 50–130% relative potencyRecovery of 90–103%Recovery of 106–108%
Precision–repeatability%CV ≤ 15%%CV ≤ 15%%CV = 3.6% with 500 U/vial
%CV = 3.3% with 125 U/vial
%CV = 4%
Intermediate precision≤10% difference between operator meansN/A1.6% difference with 500 U/vialN/A
%CV ≤ 15%At each of three potency levels (50%, 100%, and 130%), %CV ≤ 15%%CV = 8.3% with 500 U/vialPotency level%CV
50%4%
100%3%
130%5%
CI, confidence interval; %CV, coefficient of variation; N/A, not applicable; U, units.
Table 2. Acceptance criteria and results for method validation of the BoCell® assay.
Table 2. Acceptance criteria and results for method validation of the BoCell® assay.
Sample IDAssay Plate No.Sample Potency Results Ratio vs. Reference StandardUnweighted Mean (n = 3)Individual Assay %CVRepeatability %CV
500 U/vial10.8950.9137.73.6
20.853
30.991
10.9930.9834.4
21.021
30.936
10.8500.96010.3
21.042
30.987
11.0110.9948.0
20.906
31.064
11.0800.94911.9
20.879
30.889
10.9750.9136.2
20.865
30.898
125 U/vial10.9900.9815.63.3
20.923
31.031
10.9020.9727.1
21.040
30.975
10.9490.9212.7
20.900
30.914
11.0590.90319.6 1
20.711
30.940
10.9890.9576.1
20.890
30.993
10.9680.9703.0
20.941
31.000
1 The overall repeatability acceptance criteria was a %CV of 15% or below. The assay precision was slightly higher for this replicate but met the requirements of a %CV of 20% or below. %CV, coefficient of variation; ID, identification; U, units.
Table 3. Intermediate precision results with 500 U/vial powder formulation samples.
Table 3. Intermediate precision results with 500 U/vial powder formulation samples.
Sample IDAssay Plate No.Sample Potency Results Ratio vs. Reference StandardUnweighted Mean (n = 3)Individual Assay %CVRepeatability %CV
Analyst 110.8950.9137.710.4
20.853
30.991
10.9930.9834.4
21.021
30.936
10.8500.96010.3
21.042
30.987
11.0110.9948.0
20.906
31.064
11.0800.94911.9
20.879
30.889
10.9750.9136.2
20.865
30.898
10.7170.7144.1
20.683
30.741
Analyst 210.8580.8707.15.0
20.815
30.937
10.9590.9411.7
20.932
30.932
10.8940.8439.8
20.748
30.888
10.9390.9165.7
20.953
30.856
10.9530.9483.2
20.976
30.916
Analyst 1 mean0.9180
Analyst 2 mean0.9036
Average0.912
SD0.076
%CV8.3
%CV, coefficient of variation; ID, identification; SD, standard deviation; U, units.
Table 4. Summary of data for accuracy and intermediate precision assessment from BoCell® assay validation for liquid formulation.
Table 4. Summary of data for accuracy and intermediate precision assessment from BoCell® assay validation for liquid formulation.
Test LevelRun No.Weighted Combination
Reportable Result
(% Rel. Pot.)
Result Expressed as % RecoveryOverall Mean % Recovery (Accuracy)%CV
(Intermediate Precision)
50%4511021074
556112
651102
1354108
1454108
1556112
100%11081083
2114
3105
10109
11105
12105
130%71321021065
8143110
9144111
16135104
17145112
1812999
%CV, coefficient of variation. The %CV at all three levels was below 15%. The intermediate precision of the method was considered acceptable at all levels.
Table 5. Summary of comparability data for release samples in the BoCell® assay for liquid formulation.
Table 5. Summary of comparability data for release samples in the BoCell® assay for liquid formulation.
BatchPurpose of BatchCNT52120 BAS BatchCBA, U/mLLD50, U/mLRatio, %
L17564PPQ and stabilityL04793202201100
L18325PPQ and stabilityK02858229219105
L19836PPQ and stabilityK0285820122191
L20499PPQ and stabilityL0479317720985
N14521Potency comparabilityL1061318722483
N15416Potency comparabilityL10613260232112
N15791Potency comparabilityL10613245214114
N16407Potency comparabilityL0479318619098
N16906Potency comparabilityL0479319321689
N17823Potency comparabilityL04793189172110
N17580Potency comparabilityL04793217178122
Mean208207101
BAS, bulk active substance; CBA, cell-based assay; LD50, median lethal dose; PPQ, process performance qualification; U, units.
Table 6. Summary of comparability data for stability samples (12-month storage at 2–8 °C) in the BoCell® assay for liquid formulation.
Table 6. Summary of comparability data for stability samples (12-month storage at 2–8 °C) in the BoCell® assay for liquid formulation.
BatchPurpose of BatchCNT52120 BAS BatchCBA,
U/mL
LD50,
U/mL
Ratio,
%
L17564 INVPPQ and stabilityL04793189189100
L18325 INVPPQ and stabilityK0285819220494
L18325 UPR17818795
L19836 INVPPQ and stabilityK0285817419788
L19836 UPR14417781
L20499 INVPPQ and stabilityL0479318720094
L20499 UPR16117194
L13489StabilityK02858198194102
N15416Potency comparabilityL10613218206106
N15791Potency comparabilityL10613172170101
N16407Potency comparabilityL0479314517981
N16906Potency comparabilityL0479315918486
N17823Potency comparabilityL04793186160116
N17580Potency comparabilityL0479318820094
Mean17818795
BAS, bulk active substance; CBA, cell-based assay; INV, inverted; LD50, median lethal dose; PPQ, process performance qualification; U, units; UPR, upright.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Fonfria, E.; Marks, E.; Foulkes, L.-M.; Schofield, R.; Higazi, D.; Coward, S.; Kippen, A. Replacement of the Mouse LD50 Assay for Determination of the Potency of AbobotulinumtoxinA with a Cell-Based Method in Both Powder and Liquid Formulations. Toxins 2023, 15, 314. https://doi.org/10.3390/toxins15050314

AMA Style

Fonfria E, Marks E, Foulkes L-M, Schofield R, Higazi D, Coward S, Kippen A. Replacement of the Mouse LD50 Assay for Determination of the Potency of AbobotulinumtoxinA with a Cell-Based Method in Both Powder and Liquid Formulations. Toxins. 2023; 15(5):314. https://doi.org/10.3390/toxins15050314

Chicago/Turabian Style

Fonfria, Elena, Elizabeth Marks, Lisa-Marie Foulkes, Rebecca Schofield, Daniel Higazi, Sam Coward, and Alistair Kippen. 2023. "Replacement of the Mouse LD50 Assay for Determination of the Potency of AbobotulinumtoxinA with a Cell-Based Method in Both Powder and Liquid Formulations" Toxins 15, no. 5: 314. https://doi.org/10.3390/toxins15050314

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop