Unveiling Morphine: A Rapid and Selective Fluorescence Sensor for Forensic and Medical Analysis

Faculty of Science and Technology, Bournemouth University, Talbot Campus, Fern Barrow, Poole BH12 5BB, UK
*
Author to whom correspondence should be addressed.
Sensors 2024, 24(6), 1722; https://doi.org/10.3390/s24061722
Submission received: 5 February 2024 / Revised: 29 February 2024 / Accepted: 4 March 2024 / Published: 7 March 2024
(This article belongs to the Special Issue Fluorescence Sensors for Biological and Medical Applications)

Abstract

:
Opioid use, particularly morphine, is linked to CNS-related disorders, comorbidities, and premature death. Morphine, a widely abused opioid, poses a significant global health threat and serves as a key metabolite in various opioids. Here, we present a turn-off fluorescent sensor capable of detecting morphine with exceptional sensitivity and speed in various samples. The fluorescent sensor was developed through the dimerization process of 7-methoxy-1-tetralone and subsequent demethylation to produce the final product. Despite morphine possessing inherent fluorophoric properties and emitting light in an approximately similar wavelength as the sensor’s fluorescent blue light, the introduction of the target molecule (morphine) in the presence of the sensor caused a reduction in the sensor’s fluorescence intensity, which is attributable to the formation of the sensor–morphine complex. By utilizing this fluorescence quenching sensor, the chemo-selective detection of morphine becomes highly feasible, encompassing a linear range from 0.008 to 40 ppm with an impressive limit of detection of 8 ppb. Consequently, this molecular probe demonstrates a successful application in determining trace amounts of morphine within urine, yielding satisfactory analytical results. The study also explores the effect of several variables on the sensor’s response and optimizes the detection of morphine in urine using a response surface methodology with a central composite design.

1. Introduction

The illegal usage of controlled substances and illicit drugs presents a significant threat to human well-being, family cohesion, and societal equilibrium [1,2,3]. Often, there are indications of the use or abuse of such drugs found at the crime scene [4,5,6,7]. This has made the rapid and accurate detection of such drugs a compelling subject of research. Moreover, many of these chemical compounds hold immense value in the medical field due to their significance in patient recovery and overall health [8,9,10]. Consequently, pharmaceutical industries stand to gain considerable benefits from the availability of sensitive and selective sensors for quality control, ensuring the production of pure products and the monitoring of safe dosages in modern drug-delivery setups [11,12,13,14,15]. In medicine, an early diagnosis of patients’ conditions (particularly in cases of overdose resulting from drug abuse, misuse, or drug–drug interactions) can be life-saving [16,17,18,19,20,21]. Consequently, scientists are actively working on the development of novel sensors that offer better sensitivity, selectivity, ease of use, fast response, and affordability for the on-site and in-lab detection of illicit drugs [22,23,24]. Such advancements also hold tremendous importance in the prevention and reduction of drug-related crimes.
Morphine, one of the most powerful analgesics in use worldwide [25,26], is primarily extracted from the opium poppy plants [27]. It is a commonly used narcotic drug with extensive applications in clinical medicine [28,29,30]. One of the most significant drawbacks of morphine is its high potential for addiction, making it prone to abuse [31,32,33]. Morphine induces a rewarding effect and hyperlocomotion, which lead to addiction through a dopamine-reward pathway [34,35]. Morphine targets and binds to the μ-opioid receptor in the brain, and this binding leads to an inhibition of the release of a range of neurotransmitters, such as GABA [36]. This inhibition leads to an increase in dopamine release in the brain’s reward system, particularly in the nucleus accumbens [37,38]. The surge of dopamine creates a feeling of pleasure and euphoria, reinforcing the rewarding effects of morphine and contributing to its potential for abuse [39,40]. Additionally, this interaction with the dopamine-reward pathway is responsible for the pain-relieving properties of morphine, as it alters the perception of pain signals in the brain [41,42]. Intentionally administered morphine or morphine generated through the metabolism of other drugs, such as heroin, may potentially cause respiratory depression and even lead to fatal outcomes when taken in high doses [43,44].
A variety of methods have been developed for the detection of controlled and illicit drugs; they mainly include high-performance liquid chromatography (HPLC) [45,46,47], gas chromatography mass spectrometry (GCMS) [48,49,50], electrochemical sensors [7,51,52], electrochemical luminescence [53,54,55], fluorescence spectroscopy [56,57], and surface-enhanced Raman spectroscopy (SERS) [58,59,60]. However, these methods have certain shortcomings, such as a complexity in analysis, the need for highly trained operators, long pre-treatment periods, and high costs [59,61,62]. Fluorescence spectroscopy, with the aid of chemical fluorophores or biochemical aptamers, stands out as a promising detection technique that can overcome the limitations of the mentioned methods. The key strengths of this technique, other than low cost, simple preparation, and ease of use, lie in its high sensitivity and rapid response, and its ability to eliminate a wide range of interferences without requiring any pretreatment, as many compounds do not generate fluorescence emissions [1,63,64,65]. Moreover, fluorescence sensors have the potential to be fine-tuned by anchoring selective functional groups to enhance their selectivity for compounds like morphine and other opioids in various solutions [56,57].
Other than the importance of rapid detection of overdose and poisoning cases in medicine, the early hours of a crime scene investigation are extremely important for forensic analysts, as they play a critical role in obtaining vital information and insights, as well as acquiring data [66,67,68]. Portable and handheld spectrofluorometers, integrated with selective and highly sensitive chemical sensors, can be employed to facilitate quantitative and highly accurate presumptive tests [69,70,71]. Such selective chemical sensors also hold the promise of being utilized as confirmatory detection methodologies using cutting-edge stationary spectrofluorometers within laboratory environments as well. This paper introduces a novel selective fluorophore enabling the rapid, selective, and sensitive detection of morphine in various samples, with a particular focus on urine, a critical pharmaceutical and forensic biological fluid. The study employs Central Composite Design-Response Surface Methodology (CCD-RSM) as a systematic approach to analyze the influence of various variables on the sensor’s response to morphine in urine samples while optimizing the sensor’s performance to its fullest potential.

2. Experimental Section

2.1. Materials

All chemical reagents and solvents were purchased from Fisher Scientific (Loughborough, UK) and were of analytical reagent grade. Morphine, benzylpiperazine, nicotine, midazolam, and ketamine were purchased from either Sigma–Aldrich (Gillingham, UK) or LGC Standards (Middlesex, UK). Urine samples were voluntarily provided by the authors of this study (Ethics ID: 55013).

2.2. Software

The experimental design analysis and subsequent regression analysis of the urine analysis data were conducted using Design-Expert, statistical software version 11.0 (Stat-Ease, Minneapolis, MN, USA). The creation of figures and mathematical correction of baselines (when necessary and as indicated in the main text) were conducted using OriginPro 2022 (OriginLab, Northampton, MA, USA). Fluoracle (Edinburgh Instruments, Livingston, UK) was used as the operating software for the spectrofluorometer.

2.3. Instrumentation

Photoluminescence (PL) spectra were acquired using an FS5 Spectrofluorometer (Edinburgh Instruments, UK) equipped with a PMT-EXT detector, providing spectral coverage up to 980 nm. A quartz (170–2700 nm) micro-fluorescence cuvette with the designated volume of 0.7 mL and minimal volume of 0.1 mL was used for PL studies. Fourier-transform infrared (FTIR) spectra were recorded using an ATR-FTIR spectrophotometer (Agilent, Santa Clara, CA, USA). Raman spectra were collected with an XploRA™ PLUS Raman spectrometer (Horiba, Northampton, UK), equipped with a 532 nm laser and 50× lens.

2.4. Synthesizing the Sensor

Previously, 7,7′-Dimethoxy-1,1′-binaphthalene and its modified derivative, 7′-Methoxy-[1,1′-binaphthalen]-7-ol, were synthesized as part of a published research endeavor [72]. We have now embarked on a fresh scientific exploration, one that harnesses the potential of these synthesized materials in an entirely novel context. The optimized synthesis will briefly follow as outlined below:
The process of synthesis of 7,7′-Dimethoxy-1,1′-binaphthalene began with the mixing of 10 g of 7-methoxy-1-tetralone and 0.1 g of HgCl2 in a solution of 50 mL of toluene and 30 mL of ethanol. The addition of 1.6 g of aluminum foil resulted in a transformation of the solution to a grayish color. After 6 h of reflux heating, 50 mL of 10% diluted HCl were introduced in two portions, causing the solution to change to white and then yellow. Following the separation and drying of the organic phase, an orange solution was obtained. When subjected to cooling and agitation, yellow crystals were formed, which were subsequently filtered and washed with acetic anhydride and acetic acid.
The resulting compound was demethylated by dissolving 4 g of it in 15 mL of dichloromethane to produce 7′-Methoxy-[1,1′-binaphthalen]-7-ol. At −78 °C, 2 mL of BBr3 were added drop by drop, and the mixture was stirred for 3 h. Subsequently, 2 g of ice were introduced, leading to the conversion of one of the methoxy groups of the molecule into an alcohol group (Figure 1). After an additional hour, the organic phase was carefully separated, washed with water, and dried under a fume hood.

2.5. Standard Solutions

Owing to the inherent degradation of morphine within aqueous matrices, resulting primarily in the generation of pseudomorphine, alongside the less frequent emergence of morphine-N-oxide and potentially apomorphine [73], a deliberate choice has been made to formulate standard solutions within a methanolic medium. This selection is underpinned by the advantageous miscibility properties exhibited by water and methanol, ensuring the preservation of morphine’s integrity and stability, as well as facilitating its blending with water-based biological fluids, such as urine. Simultaneously, the developed sensing material demonstrates significantly enhanced solubility in methanol compared to water, owing to its distinct polarity characteristics.
The developed sensor, 7′-Methoxy-[1,1′-binaphthalen]-7-ol, was dissolved in methanol within a separate container and then introduced into the standard solutions. This infusion was carefully calibrated to achieve a final concentration of 100 ppm for the sensing material within the testing standards. Standard solutions of morphine were produced using pure methanol, encompassing a concentration gradient spanning from 1 ppb to 100 ppm. This comprehensive range was established to investigate the sensor’s linear response capabilities.

2.6. Biological Sample Preparation

Each urine sample was prepared by initially transferring 2 mL of untreated urine into a container. Subsequently, the requisite volumes of both the morphine solution (in methanol) and the sensor solution (in methanol) were added to the mixture (as illustrated in Table 1 and Table 2). To achieve a final sample volume of 5 mL, methanol was added into the solution. Adjustments to the concentrations of the sensor and morphine were made to align with the targeted concentrations dictated by the experimental design within the CCD-RSM framework.

2.7. Measurements of Photophysical Properties

Excitation and emission mapping were employed to determine the optimal excitation wavelength for the developed sensor solution (400 ppm in methanol). This mapping procedure involved measuring the fluorescence emission spectra within the range of 320 to 800 nm while systematically varying the excitation wavelength from 225 to 450 nm. The excitation wavelength associated with the maximal intensity of the fluorescence peak was ascertained as the optimal excitation wavelength. This wavelength was subsequently utilized to measure the fluorescence emission of the sensor in the presence of various drugs and in urine samples. Additionally, the same excitation wavelength was used to analyze the emission spectra of morphine in the absence of the sensor for comparison.

2.8. FTIR and Raman Spectroscopy

Regarding the FTIR and Raman spectroscopy, sample preparation necessitated the utilization of approximately 0.05 mg of dry, powdered samples. These samples were then transformed into tablets directly on the diamond surface of the ATR-FTIR instrument through the utilization of a specialized sample holder integrated with the ATR system. Subsequently, the compressed powder was delicately transferred onto a pristine glass sample holder for subsequent Raman spectroscopic analysis, subsequent to FTIR assessment.
For FTIR analysis, signal enhancement was accomplished through the accumulation of data derived from 24 scans of the identical sample, thereby augmenting the signal-to-noise ratio. Spectral data acquisition encompassed a range spanning from 650 to 3650 cm−1.
Concerning Raman spectroscopy, a laser beam having a wavelength of λex = 532 nm was employed, accompanied by a grating featuring 600 grooves per millimeter, a 50% filter, a 100 μm slit width, and a 200 μm aperture. Data were accumulated over a period of 10 s, with a total accumulation count of 8, and a Read-to-Determine (RTD) time interval of 10 s. The acquisition span for Raman spectra encompassed the range from 0 to 3500 cm−1. This rigorous approach to sample preparation and data collection serves to ensure the veracity and precision of the spectroscopic outcomes.

2.9. CCD-RSM Design

The primary objective of this research is to create a chemo-selective probe capable of analyzing samples on site with minimal or no prior sample pretreatment. To address the challenge posed by the high miscibility of methanol and water, we utilized a solution containing the dissolved sensor (in methanol) and untreated urine to assess the sensor’s response. On-site urine drug-testing products provide a valuable opportunity for conducting immediate drug screenings directly at the point of care [74]. Due to the significance of on-site drug testing in the context of crime scene investigations, emergency rooms, and dependency clinics [75,76,77,78], it is crucial to understand how certain variables influence the sensor’s response. This understanding is essential for ensuring the sensor’s consistency to generate reliable results, underscoring the importance of comprehensive performance evaluations. These evaluations serve to highlight any limitations associated with the developed testing products.
The current investigation employed a simultaneous optimization technique known as Central Composite Design (CCD) within the framework of Response Surface Methodology (RSM). This integrated approach combines aspects of experimental design, mathematical modeling, and optimization, providing a systematic means to explore intricate response functions with a streamlined set of variable combinations [79,80,81,82]. In this research, we utilized a 24 full-factorial design in the CCD format to create mathematical models that would enable a quantitative assessment of sensor behavior under the influence of various parameters. We systematically explored the influence of alterations in morphine and sensor concentrations, temperature variations, and the duration since the mixing of morphine and the sensor on the fluorescence emission intensity response of the sensor. The experimental design employed a five-level CCD, resulting in a total of 30 experimental runs to facilitate the optimization process. Table 1 and Table 2 provide an overview of the five-level CCD used to explore the effects of the aforementioned variables on electrical current, highlighting its utility in optimizing sensor performance.

3. Results and Discussion

3.1. Fourier-Transform Infrared Spectroscopy

Figure 2 depicts the spectra of 7-methoxy-1-tetralone, its dimerized form 7,7′-Dimethoxy-1,1′-binaphthalene, and its derivative 7′-Methoxy-[1,1′-binaphthalen]-7-ol. In the spectrum of 7-methoxy-1-tetralone, a prominent C=O stretching bond is observed at approximately 1674 cm−1 [83]. This peak vanishes in both dimerized structures, confirming that dimerization occurred at the carbonyl site. A comparison of the spectra of 7,7′-Dimethoxy-1,1′-binaphthalene and 7′-Methoxy-[1,1′-binaphthalen]-7-ol reveals two new peaks: C−O stretching at 1106 cm−1 [84] and O−H stretching at 3349 cm−1 [85]. The presence of O-H peaks confirms the conversion of the functional group on the diene from methoxy to an alcoholic group.
It has been established that demethylation by a strong reagent such as BBr3 can either convert all [86] or just half [72,87] of the methoxy groups to alcoholic groups. The detection of peaks between 2800 and 3000 cm−1, corresponding to the methoxy group (–OCH3) [88], along with the appearance of O-H peaks in 7′-Methoxy-[1,1′-binaphthalen]-7-ol, suggests that only one of the methyl groups in the molecule was replaced by hydrogen, forming –OH, while the other remained unchanged [72].

3.2. Raman Spectroscopy

Raman spectroscopy was employed as another characterisation technique in the comparative analysis of the molecular structures of 7-methoxy-1-tetralone and the developed fluorescence sensor, 7′-Methoxy-[1,1′-binaphthalen]-7-ol (Figure 3). While the spectrum for the initial reagent, 7-methoxy-1-tetralone, was clear with minimum noise, it is imperative to acknowledge the inherent inefficiency of Raman scattering, particularly in the context of fluorescent compounds such as the developed 7′-Methoxy-[1,1′-binaphthalen]-7-ol. In most instances, the generation of a single Raman photon necessitates the incidence of approximately 106–108 laser photons upon the sample [89,90]. Consequently, even minute concentrations of fluorescent species (either in pure compound or as interferences) within the sample can obscure or complicate the detection and interpretation of the weak Raman-scattered photon signals [91,92]. If the fluorescence baseline is elevated, the shot noise emanating from this signal may equate to or even surpass the intensity of the Raman signal itself, effectively veiling the Raman photon signals [91,93].
In an endeavor to rectify this issue, other than minimizing the volume of the sample and changing the scan duration and intensity of the excitation beam [93], we employed the base-line correction of Origin software (v9), employing mathematical techniques aimed at refining the baseline correction [94]. However, it is crucial to underscore that despite these efforts, Raman peaks, specifically the ones associated with 7′-Methoxy-[1,1′-binaphthalen]-7-ol, may still persistently elude detection amidst the backdrop of noise and fluorescence interference.
Looking at Figure 3, the C-O stretching peak for the methoxy group can be seen at approximately 1141 cm−1 [95]. The peak at 1239 cm−1 could represent C-O symmetric stretching [95]. However, since this frequency is too high, it might correspond to a coupled C-O + O-CH3 deformation. The O-CH3 bending peaks typically appear between 1450–1475 cm−1, but in this case, they are observed at 1432 cm−1 for 7-methoxy-1-tetralone and 1487 cm−1 for the dimerization products [96]. This shift is likely due to the specific molecular structure of the analysed samples. The CH3 symmetric stretching peak is located at 2885 cm−1, while the CH3 asymmetric stretching peak for the methoxy group is at 2950 cm−1 [97,98,99]. The dangling O-H bond of the sensor is centered at 3460 cm−1 [100]. The C-O-C symmetric stretching peak is observed at 878 cm−1 [95]. The aromatic C-H asymmetric stretching peaks are located at 3079 cm−1 for 7-methoxy-1-tetralone and 3042 cm−1 for the dimers 7,7′-dimethoxy-1,1′-binaphthalene and 7′-methoxy-[1,1′-binaphthalen]-7-ol [95,96].

3.3. Fluorescence Spectroscopy

Owing to the conjugated double-bond system and the high mobility of their π-electrons, the developed sensor (7′-Methoxy-[1,1′-binaphthalen]-7-ol) was able to exhibit a strong fluorescence emission [64]. The three-dimensional fluorescence spectrum is visually represented in Figure 4a,b, unveiling a prominent fluorescence peak at an emission wavelength of 369 nanometers, which is elicited by an excitation wavelength of 325 nanometers. Subsequently, in Figure 4c, the emission and excitation scans of the sensor at the optimized excitation and emission wavelengths are displayed.
In Figure 4d, the emission peak color resulting from excitation with a 325 nm wavelength is characterized, within the International Commission on Illumination (CIE 1976) color space, as blue. Within this color space, the u’ component corresponds to the u-prime axis, representing hues ranging from green to yellow, while the v’ component signifies the v-prime axis, encompassing hues from blue to red. Collectively, the u’ and v’ components jointly determine the chromaticity coordinates within a cylindrical color space.
Synchronous fluorescence spectroscopy (SFS) serves as a powerful tool for the detection of impurities or contaminants within chemical samples, even when present at trace levels [101,102,103]. Often, these impurities exhibit distinct fluorescence properties in comparison to the primary product being analyzed. In our study, we utilized SFS to scan the synthesized product, 7′-Methoxy-[1,1′-binaphthalen]-7-ol. Scans encompassed the spectral range from 250 nm to 550 nm, employing 5 nm incremental adjustments within the 40 nm-to-60 nm offset range. The objective was to systematically eliminate the presence of unexpected fluorescence signals, which could serve as indicators of impurities [65,104] (Figure 5a), which suggests the relatively pure product. Simultaneously, to ensure the stability of the developed compound in its solution form over time, and the absence of self-quenching effects [105,106], especially at high concentrations of 7′-Methoxy-[1,1′-binaphthalen]-7-ol, an examination was conducted on the effect of time on the intensity of the emission peak. This involved generating emission spectra spanning the short range (from 350 nm to 420 nm) for a solution containing 500 ppm of the sensor in methanol, recorded over a duration of 10 min (Figure 5b). The resulting data convincingly demonstrate the stability of the sensor’s response, reinforcing its reliability for practical applications.
The morphine molecule features five rings within its three-dimensional structure, with three of these rings lying approximately in the same plane, bonded to hydroxyl groups. Meanwhile, the remaining two rings, including one with oxygen and the other with nitrogen, are oriented at an angle to the others. This configuration, and a nitrogen atom possessing a lone electron pair, provides morphine with the capability to produce fluorescence emissions [64]. To further investigate this property, emission spectra were recorded for standard solutions of pure morphine in methanol. Additionally, the response of the sensor to changes in morphine concentration was observed over a range spanning from 1 ppb to 50 ppm. These observations were subsequently compared with the emission response of the sensor.
As seen in Figure 6a, pure-morphine standard solutions (in methanol) produced an emission peak of 366 nm (λex = 325 nm), which closely matches the fluorescence peak wavelength of 7′-Methoxy-[1,1′-binaphthalen]-7-ol at 369 nm. However, the intensity of the fluorescence emission of the sensor is significantly higher than that of pure morphine. As expected, with an increase in the concentration of morphine in the solution, the intensity of its emission peak shows a linear increase (R2 = 0.999; Figure 6c). However, for concentrations lower than 2.5 ppm of the morphine solution (in the absence of sensor), the emission-peak intensity is found to be very low, nearly indistinguishable from noise and the background spectra of pure methanol. At this range, differentiating between noise, the background spectra of pure methanol, and the morphine emission peak becomes challenging.
While both morphine and the developed sensor emit fluorescence peaks at nearly identical wavelengths (λmorphine = 366 nm, λsensor = 369 nm), the introduction of morphine to the sensor results in quenching, leading to a decrease in the intensity of the emission peak of 7′-Methoxy-[1,1′-binaphthalen]-7-ol. The higher the concentration of morphine, the lower the intensity of the sensor’s emission peak. This intriguing phenomenon has been harnessed to establish a linear relationship between the concentration of morphine and the intensity of the sensor’s fluorescence peak. Notably, the linear range of the turn-off sensor extends from 8 ppb to 40 ppm (R2 = 0.987), boasting a remarkable detection limit of 8 ppb. In addition to the observation that 8 ppb was the lowest-tested concentration that caused a noticeable quenching in the sensor response, the detection limit was determined using the formula LoD = 3.3Sb/m. Here, “Sb” signifies the standard deviation obtained from three repeated runs (Sb = 540.33247), and “m” represents the slope derived from the sensor’s linear graph (m = 213544.57). Consequently, the calculated LoD for the sensor is calculated to be approximately 0.00835 ppm. It is worth highlighting that for concentrations exceeding 40 ppm, the fluorescence peak of pure morphine itself serves as a robust indicator of the presence of a high concentration of morphine in the solution (Figure 6c).
To assess the chemo-selectivity of the developed sensor before examining its behavior in complex matrices such as urine, a series of tests were conducted to observe variations in the sensor’s output signal when exposed to standard solutions containing various drugs. Morphine, ketamine, midazolam, nicotine, and benzyl-piperazine (BZP) share several commonalities despite their distinct effects, uses, mechanisms of action, and therapeutic or recreational purposes [107,108,109,110,111]. Notably, all of them feature structures composed of multiple aromatic rings and harbor active functional groups, allowing these substances to primarily exert their influence on the central nervous system. They cause the modulation of brain function and neurotransmitter activity, resulting in alterations in perception, mood, consciousness, or pain perception, and all of them have the potential for abuse, which can lead to addiction or dependence [7].
Standard solutions with a concentration of 20 ppm for each of these drugs were individually prepared in methanol. The sensor’s response in the absence and presence of these drugs was utilized to demonstrate the selectivity of the sensor, as depicted in Figure 7a. Interestingly, only morphine induced quenching effects, while the other tested drugs either did not alter the fluorescence-emission intensity of the sensor or caused an increase in emission intensity. The observed increase in intensity, as seen with molecules such as ketamine and nicotine, could be attributed to changes in the local environment surrounding the sensor molecule or the stabilization resulting from the formation of more complex structures. In another experiment aimed at assessing the sensor’s affinity for morphine, standard solutions were prepared by combining previously tested drugs (each at 20 ppm) with morphine (20 ppm). Subsequently, the sensor’s response to morphine within this mixture was evaluated. Figure 7b presents the results, indicating a notable affinity of the sensor for morphine compared to all other tested drugs. While fluorescence quenching of the sensor is observed in all mixed samples upon the presence of morphine (F/F′), it is also noticeable by the comparison between Figure 7a,b. The fluorescence intensity in the nicotine mixture sample, despite showing quenching, is comparatively higher than in other mixed samples. This observation may suggest a potential competition between nicotine and morphine molecules for binding sites on the developed sensor.
When two different fluorescent molecules, both serving as sensor and analyte, are brought into close proximity, they can exhibit a phenomenon in which one molecule quenches the fluorescence of the other. This quenching can occur through various mechanisms, primarily involving energy-transfer processes. One of the most prevalent mechanisms responsible for this phenomenon are Förster Resonance Energy Transfer (FRET) [64]. Since the emission wavelength of morphine (366 nm) is significantly distant (more than 10 nm apart) from the excitation wavelength of the sensor (325 nm) and does not exhibit spectral overlap with it, it precludes the possibility of an energy transfer from one fluorescent molecule (the donor) to another (the acceptor) [112]. This energy-transfer mechanism, relies on spectral overlap for effective quenching of the donor molecule’s fluorescence and is thus ruled out in this context.
Dynamic quenching, collisional quenching, and photo-induced electron transfers (PETs) are temperature-dependent processes [64]. Typically, PET rates increase with higher temperatures due to enhanced molecular motion and collision rates. However, it is important to note that no significant changes in the emission-peak intensity of the sensor, while maintaining a constant concentration of morphine, are observed in response to variations in temperature (Figure 7d). This indicates that the quenching has occurred due to a strong binding interaction between the sensor and quencher (morphine). It can result from the formation of a non-fluorescent complex through non-covalent interactions, such as van der Waals forces, hydrogen bonding, or electrostatic interactions, between 7′-Methoxy-[1,1′-binaphthalen]-7-ol and morphine molecules (Figure 7c) [64,113].
Morphine (Figure 7a,c) exhibits the potential to engage in various non-covalent interactions, contributing to its ability to induce static quenching with the sensor. These interactions include the formation of van der Waals bonds facilitated by its aromatic carbon rings (rings comprising carbons 1, 2, 7, and 8), the establishment of hydrogen bonds using its alcohol groups (located on carbons 3 and 6), and the possibility of electro-static interactions, primarily through its tertiary nitrogen atom, which is connected between Carbons 9 and 16. This multifaceted nature of morphine’s molecular structure, which makes it interact with μ opioid receptors in the body [114], makes it a particularly suitable candidate for causing static quenching phenomena in this context.
In addition, 7′-Methoxy-[1,1′-binaphthalen]-7-ol (Figure 1) comprises four aromatic rings in its structure, endowing it with the capability to engage in van der Waals interactions. Furthermore, its planar configuration enables the sensor to establish stacking interactions with other planar or nearly planar molecules. Additionally, the presence of oxygen atoms within the molecule, found in both the alcohol and methoxy moieties, facilitates the formation of potent non-covalent bonds, namely hydrogen bonds, with other molecules. The sensor’s remarkable selectivity for morphine, as demonstrated in Figure 7a, when compared to four other tested drugs (benzylpiperazine, ketamine, midazolam, and nicotine), strongly implies that multiple types of non-covalent bonds may be contributing to the static quenching of the sensor.

3.4. Comprehensive Urine Analysis

The current standard procedure for opiate urine testing involves the collection of urine samples, their transfer to laboratories, and a subsequent analysis using liquid chromatography with tandem mass spectrometry (LC-MS/MS) [115,116,117]. Detecting the presence of morphine, whether as the primary drug of use or as a metabolite of other substances like heroin, in urine becomes exceedingly challenging after 48 h due to its limited half-life of 2–4 h. This underscores the critical importance of analyzing collected samples and obtaining results as swiftly as possible. Therefore, any techniques capable of reducing the sample transportation time from the collection site to the laboratory or delivering faster results compared to conventional methods like LC-MS/MS would be highly advantageous.
As demonstrated earlier, our objective was to develop a selective molecular fluorescence sensor suitable for use with both portable spectrofluorometers, ensuring accurate on-site testing, and stationary spectrofluorometers, providing faster responses than techniques such as LC-MS/MS. To verify that the sensor can still produce a distinguishable emission spectrum, unobstructed by the emissions of other molecules within the complex urine matrix, a series of tests were conducted, as depicted in Figure 8.
The results indicate a significant difference in peak wavelength between pure urine (428 nm) and the emission from the sensor (382 nm) within the urine matrix. The small shift in the fluorescence emission of the sensor (13 nm blue shift) can be attributed to alterations in the solvent’s polarity, viscosity, and hydrogen-bonding characteristics [64,118]. Notably, methanol dilution does not diminish the fluorescence intensity of the urine. However, the introduction of a morphine standard solution (in methanol) to the urine sample leads to an increase in the fluorescence intensity of the broad urine-emission peak.
While the increase in the concentration of morphine in urine samples still leads to a linear reduction in the sensor’s fluorescence emission (Figure 9a), the chemometric technique of central composite design-response surface methodology was utilized to examine the influence of various variables on the sensor’s response within the complex urine sample. The aim was to identify and optimize key operational parameters to enhance sensor performance. It is worth mentioning that due to the small peak shift resulting from changes in the solvent, as discussed earlier, the alterations in the intensity of the emission peak at 382 nm were used for CCD-RSM analysis.
In Table 2, the design matrix and outcomes were presented, stemming from the central composite full-factorial experimental design. The primary focus of this investigation was centered on the assessment of reproducibility and errors, specifically at the “center points” of the design. The “factorial points”, constituting 2k data points (with k representing the involvement of four factors), were strategically employed to facilitate the estimation of first-order effects and two-factor interactions. Additionally, the “axial points”, comprising 2k data points, were utilized to gauge the exclusive quadratic effects within the experimental framework. Leverage quantifies the potential for a data point to exert a strong influence on the regression analysis, and it depends on the predictor values; high leverage points (close to 1.0) are considered outliers with respect to the independent variables. The average leverage of the 30 runs was 5.0 × 10−1, and neither leverage was 1.0 or higher.
The optimization of the sensor efficiency was achieved through the application of response-surface methodology. Within this framework, we employed an analysis of variance (ANOVA) in the context of a standard response surface design to investigate the key parameters affecting the performance of the electrochemical sensor. In this study, we opted for a quadratic model due to its capacity to accommodate higher-order polynomial terms that carry significant implications, while the cubic model was rendered inconclusive. The results of the ANOVA, presented in Table 3, offer valuable insights into the significance of each variable. The Model F-value, which stands at 44.86, underscores the overall significance of our model. The probability of encountering such a substantial F-value purely by chance is a mere 0.01%. p-values below 0.0500 indicate that the model terms hold substantial importance. In our investigation, the terms A, C, D, BC, A², B², and D² were all found to be statistically significant. Conversely, values exceeding 0.1000 signify that the model terms are not statistically significant. In cases where numerous terms fall into this category, excluding those essential for maintaining the hierarchy, may enhance model performance [119,120,121]. To assess the suitability of the model, we conducted a Lack of Fit analysis, which should ideally yield an insignificant result for the model to align effectively with the experimental design. In our study, the Lack of Fit F-value, which is 3.65, does not reach statistical significance, affirming the adequacy of our model.
Upon analyzing the data, it becomes evident that the Predicted R², standing at 0.8778, aligns reasonably well with the Adjusted R², which reaches 0.9549. This implies that the difference between the two values is less than 0.2, signifying a close match. Additionally, the Adequacy Precision analysis, which gauges the signal-to-noise ratio, reveals that our model boasts a ratio of 28.462, which is well above the desirable threshold of 4. This strong signal-to-noise ratio reaffirms the model’s suitability for navigating the design space.
Furthermore, the equation expressed in terms of coded factors serves as a valuable tool for predicting responses based on various factor levels. This coded equation allows for a comparative assessment of the relative impact of individual factors through the examination of their coefficients. The equation is as follows:
E m i s s i o n = 789,026.99 61,501.38 × A 15,784.00 & × B + 169,072.98 × C 35,336.13 × D   + 7404.21 × A B + 14,880.30 × A C 15,238.75 × A D 41,704.39 × & B C   11,917.54 × & B D + 9447.28 × & C D + 99,043.50 × A 2 + 43,927.54 & × B 2   2254.98 & × C 2 + ( 29,954.79 × D 2 )
In essence, these findings underscore the robustness of our model, making it a reliable tool for navigating the intricacies of the design space and providing valuable insights into the impact of individual factors on emissions.
The normal plot of residuals serves as a valuable tool for assessing the assumption of normality in error-term distribution [122,123]. This plot employs a diagonal line, aligning with the lower and upper quartiles of the theoretical distribution, to visually gauge the linearity of the relationship between theoretical and sample percentiles. When this relationship appears approximately linear, as suggested by the outcome plot of this study (Figure 9b), it indicates that the error terms follow a normal distribution. Consequently, we can confidently proceed with the assumption that the error terms exhibit normal distribution properties [120,121,124]. Another essential visualization tool is the actual-versus-predicted emission graph (Figure 9c), which offers a visual assessment of the model’s fit by depicting variations attributed to random effects [125,126]. This graph plots observed emission values against the predicted emission values. In an ideal scenario, data points would be evenly distributed along the 45-degree line. A well-fitted model would exhibit data points closely aligned with the fitted line. It is worth noting that the residual, defined as the difference between actual and predicted (or fitted) response variable values, should not be confused with the normal residual plot in this context. Both plots, as depicted in Figure 9b,c, confirm that there are no anomalies in the experimental data, and they align well with the predicted data within the quadratic model. This robust alignment proves the model’s effectiveness in accurately predicting fluorescence quenching.
To gain a deeper understanding of the variables with the most significant influence on quenching, ultimately resulting in minimized emissions, and to explore the interactions between these variables, we turn our attention to Figure 10a–f. These graphical representations provide a valuable tool for predicting optimal conditions, and the presence of curvatures in these response surfaces hints at intricate interactions among the variables.
As expected, Figure 10a vividly illustrates a trade-off relationship between the concentration of morphine and the sensor’s concentration. As the morphine concentration increases, there is a notable reduction in emissions, while elevating the sensor’s concentration amplifies fluorescence emission at the specified peak wavelength. This behavior aligns seamlessly with our initial expectations, and a similar trend can be observed in Figure 10b,c. Remarkably, the enhancement of the sensor’s concentration emerges as a pivotal factor influencing fluorescence intensity, as anticipated. However, it is worth noting that changes in morphine concentration have a significant impact even within ppb ranges, whereas sensor-concentration changes operate at ppm levels. Conversely, two variables, mixing time and temperature (Figure 10d), exhibit a relatively weak mutual influence, yielding a nearly flat response with only a slight increase in quenching observed between temperatures of 20–25 °C, coupled with a modest extension in mixing time. However, when examining the interplay between mixing time and morphine concentration on emissions (Figure 10e), a somewhat distinct relationship emerges. For higher morphine concentrations, a slightly prolonged mixing time before initiating fluorescence studies appears to enhance quenching effects. Additionally, Figure 10f demonstrates that the sensor exhibits substantial quenching as the morphine concentration increases, particularly within the temperature range of approximately 15–30 °C. These findings provide valuable insights into the intricate interactions of these variables and their impact on the fluorescence emissions of the sensor.
Understanding the intricate interplay among these variables and their impacts on the sensor’s emissions is invaluable. Consequently, we can deduce that the primary influencers in the realm of fluorescence quenching and emission in urine are the concentrations of morphine and the sensor, with nuanced interaction effects discernible between them. Mixing time, in this context, assumes a subsidiary role, while temperature exerts the least amount of influence on these processes. Armed with this knowledge, we can tailor the optimization of sensors for urine matrices, meticulously considering all variables and the magnitude of their effects.
Building on this comprehension, we delineate our desired conditions with precision. Our primary objective is to achieve the lowest possible emission, indicative of the highest quenching effect, while concurrently minimizing the concentration of morphine. In this pursuit, we must also consider the economic aspects, encompassing the affordability and production costs of the sensor. Consequently, one of our paramount desirability settings involves determining the minimum requisite amount of the sensor. Our desirability scale ranges from 0.0 (indicating undesirability) to 1.0 (representing high desirability), and this configuration allows us to reach an optimal condition with a commendable desirability rate of 0.825.
As depicted in Figure 11a–f, it becomes evident that the lowest concentration of morphine capable of eliciting maximum quenching in a urine matrix is precisely 21.81 ppb. Moreover, the optimal working conditions include a temperature of 18.48 degrees Celsius and a waiting time (after mixing) of 7.45 min, both of which yield the smallest quenching effect.
To validate the outcomes derived from the modeled response and to ensure their reliability, a confirmation sample was employed. A confirmation sample comprises a series of supplementary trials conducted at a specific combination of factor settings, situated within the studied range and aligned with the optimized conditions. The average response obtained from this confirmation sample is then compared to the prediction interval (PI). When the average observation from the confirmation experiment falls within the prediction interval, it affirms the accuracy and dependability of the model.
For this purpose, three separate samples were prepared and analyzed using a spectrofluorometer. The selected parameter settings were as follows: a morphine concentration of 21.8 ppb, a sensor concentration of 75 ppm, a temperature of 18.5 °C, and an additional mixing time of 7.5 min. The resulting data are summarized in Table 4, where a 95% confidence level was applied. It is noteworthy that all measured responses, as well as the mean of repeated tests, fall comfortably within the prediction intervals of the quadratic model. This comprehensive alignment serves to confirm the precision of the quadratic model and the replicability of the sensor’s results.

3.5. Evaluating Morphine Sensors in Comparison

As previously discussed, morphine holds a critical role as a potent analgesic medication, particularly in medical contexts like surgical procedures [127]. For effective pain relief, it is recommended that the optimal dose should not exceed 15 mg per 70 kg of body weight when administered through an extended-release liposome injection or an intrathecal injection [128,129,130,131,132]. In the case of oral administration, the maximum daily dose can reach as high as 30 mg, considering the effects of first-pass metabolism [133]. It is worth noting that there is not a well-defined toxic dose or specific plasma/blood concentration for morphine [44]. However, its use carries a substantial risk of addiction and misuse, contributing to a range of serious health complications [134]. Importantly, the administration of higher doses of morphine can lead to respiratory depression and potentially life-threatening health issues [135]. The use of morphine for analgesia has been associated with a significant number of drug-related deaths, with estimated incidence rates ranging from 0.3% to 4% [44,136]. Additionally, morphine can be detected in trace amounts not only in biological fluids but also in aqueous environments [137]. Consequently, considerable research efforts have been dedicated to the development of innovative detection techniques and sensors to rapidly and accurately quantify morphine in various sample types. In this context, we seek to compare the sensor developed in our research with 15 other different sensing systems, as presented in Table 5. This comparison aims to highlight the distinct advantages of our turn-off sensor in terms of its performance and capabilities.
Upon comparing the sensors detailed in Table 5, it becomes evident that our developed sensor boasts several distinct advantages. With a detection limit as low as 8 ppb, our fluorescence sensor not only rivals but also surpasses established methods such as GC-MS, which has a detection limit of 3 ppb [138]. This is particularly noteworthy due to the inherent ease of use and affordability associated with fluorescence spectroscopes. In addition to its impressive sensitivity, our sensor offers the distinct advantage of providing rapid results in under 10 s. Furthermore, the biological (urine) sample-preparation process is remarkably straightforward and swift, involving the addition of the sensor solution to urine and a brief 7.45-min mixing period before measuring emissions (as depicted in Figure 11c,e,f) to achieve highly optimized outcomes. This results in a total analysis time of less than 7.5 min. Moreover, our sensor exhibits a significantly wider linear range, surpassing the capabilities of all other sensors in the comparison.
The developed sensor boasts an impressive linear range that spans from 0.008 to 40 ppm, a range that is more extensive than nearly all other methods considered, including GC-MS (0.0025–2 ppm) [138], surface plasmon resonance imaging (1–50 ppm) [142], colorimetric assay (19.97–856 ppb) [143], piezoelectric sensor (0.25–2500 ppb) [140], and various other sensor types [62,139,141,146]. This practical broader linear range enhances the versatility and adaptability of our sensor for use with a wide range of concentration levels.
Notably, the detection limit, linear range, and response time of our fluorescence sensor are on par with electrochemical morphine sensors [26,144,145], even though they employ a different detection method. It is worth mentioning that the significantly wider linear range offered by our sensor in this study surpasses that of all other morphine fluorescence sensors [56,147,148], including the fluorescein-gold nanoparticle-based fluorescence sensor, which, despite its competitive detection limits (0.015 ppb), suffers from a limited linear range (0.001–13.9 ppb) [149].

4. Conclusions

This study embarked on the comprehensive journey of synthesizing and characterizing 7′-Methoxy-[1,1′-binaphthalen]-7-ol from scratch, subjecting it to rigorous characterisation and purity tests employing spectroscopy techniques. During this investigation, we unveiled its remarkable photoluminescence properties, notably the ability to emit intense, blue-colored fluorescence. This emission peak became the focal point of our research, as we explored its potential as a fluorescence sensor for morphine. Our findings revealed that it exhibited quenching exclusively in the presence of morphine without responding to other tested drugs or a range of diverse molecules typically found in complex samples like urine.
After confirming these initial results, we harnessed CCD-RSM to not only delve into the intricate interplay of various variables and their influence on each other but also to optimize the testing conditions for the sensor. Our objective was to achieve maximum quenching with the lowest possible concentration of morphine. Subsequently, we rigorously assessed the reliability and reproducibility of the developed quadratic model.
In the subsequent comparative analysis with a range of other morphine sensors, our sensor was found to excel in terms of response time, ease of use, linear range, and detection limit. This promising performance positions it as a compelling choice for morphine quantification in various sample types, potentially surpassing more complex and costly analytical methods. Due to these achievements, our developed sensor holds significant promise as an affordable, user-friendly, and reliable detection technique for morphine concentration measurements in biological fluids, particularly urine. Its potential applications span forensic and medical studies where it can serve as a valuable tool for precise morphine quantification.

Author Contributions

Conceptualization, R.B.; Methodology, R.B. and A.B.; Software, R.B.; Validation, A.B., R.P. and S.M.; Formal analysis, R.B.; Investigation, R.B.; Resources, A.B., R.P. and S.M.; Data curation, R.B.; Writing—original draft, R.B.; Writing—review & editing, A.B., R.P. and S.M.; Visualization, R.B.; Project administration, R.B. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Informed Consent Statement

Patient consent was waived due to self-experimentation of authors. Ethics ID: 55013.

Data Availability Statement

Data are contained within the article.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Elmizadeh, H.; Bardajee, G.R.; Moaddeli, A. Ultrasensitive and rapid detection of methamphetamine in forensic biological fluids using fluorescent apta-nanobiosensors based on CdTe quantum dots. Microchem. J. 2023, 189, 108519. [Google Scholar] [CrossRef]
  2. Armenian, P.; Vo, K.T.; Barr-Walker, J.; Lynch, K.L. Fentanyl, fentanyl analogs and novel synthetic opioids: A comprehensive review. Neuropharmacology 2018, 134, 121–132. [Google Scholar] [CrossRef]
  3. Zanfrognini, B.; Pigani, L.; Zanardi, C. Recent advances in the direct electrochemical detection of drugs of abuse. J. Solid State Electrochem. 2020, 24, 2603–2616. [Google Scholar] [CrossRef]
  4. Bruijns, B.; Tiggelaar, R.; Knotter, J.; van Dam, A. Use of Lateral Flow Assays in Forensics. Sensors 2023, 23, 6201. [Google Scholar] [CrossRef]
  5. Park, J.; Hong, S.; Min, S.; Lee, N.Y.; Chung, H.; Han, E. Awareness survey on drug crime scene investigation and drug detection kits among drug-related police officers. J. Forensic Leg. Med. 2023, 94, 102470. [Google Scholar] [CrossRef]
  6. de Araujo, W.R.; Cardoso, T.M.G.; da Rocha, R.G.; Santana, M.H.P.; Muñoz, R.A.A.; Richter, E.M.; Paixão, T.R.L.C.; Coltro, W.K.T. Portable analytical platforms for forensic chemistry: A review. Anal. Chim. Acta 2018, 1034, 1–21. [Google Scholar] [CrossRef]
  7. Boroujerdi, R.; Paul, R. Graphene-Based Electrochemical Sensors for Psychoactive Drugs. Nanomaterials 2022, 12, 2250. [Google Scholar] [CrossRef]
  8. Volkow, N.D.; Jones, E.B.; Einstein, E.B.; Wargo, E.M. Prevention and Treatment of Opioid Misuse and Addiction: A Review. JAMA Psychiatry 2019, 76, 208–216. [Google Scholar] [CrossRef]
  9. Anzar, N.; Suleman, S.; Parvez, S.; Narang, J. A review on Illicit drugs and biosensing advances for its rapid detection. Process Biochem. 2022, 113, 113–124. [Google Scholar] [CrossRef]
  10. dos Santos, B.P.; Eller, S.; Borges, G.R.; de Gouveia, G.C.; Sebben, V.C.; Arbo, M.D.; de Oliveira, T.F. A multi-analyte LC-MS/MS method for the determination of 57 pharmaceuticals and illicit drugs in plasma, and its application to poisoning cases. J. Pharm. Biomed. Anal. 2023, 222, 115082. [Google Scholar] [CrossRef] [PubMed]
  11. Zafar Razzacki, S.; Thwar, P.K.; Yang, M.; Ugaz, V.M.; Burns, M.A. Integrated microsystems for controlled drug delivery. Adv. Drug Deliv. Rev. 2004, 56, 185–198. [Google Scholar] [CrossRef]
  12. Kumar, C.S.S.R.; Mohammad, F. Magnetic nanomaterials for hyperthermia-based therapy and controlled drug delivery. Adv. Drug Deliv. Rev. 2011, 63, 789–808. [Google Scholar] [CrossRef]
  13. Peltonen, L. Practical guidelines for the characterization and quality control of pure drug nanoparticles and nano-cocrystals in the pharmaceutical industry. Adv. Drug Deliv. Rev. 2018, 131, 101–115. [Google Scholar] [CrossRef]
  14. Busacca, C.A.; Fandrick, D.R.; Song, J.J.; Senanayake, C.H. The Growing Impact of Catalysis in the Pharmaceutical Industry. Adv. Synth. Catal. 2011, 353, 1825–1864. [Google Scholar] [CrossRef]
  15. Boroujerdi, R.; Paul, R.; Abdelkader, A. Rapid Detection of Amitriptyline in Dried Blood and Dried Saliva Samples with Surface-Enhanced Raman Spectroscopy. Sensors 2022, 22, 8257. [Google Scholar] [CrossRef]
  16. Ashbourne, J.F.; Olson, K.R.; Khayam-Bashi, H. Value of rapid screening for acetaminophen in all patients with intentional drug overdose. Ann. Emerg. Med. 1989, 18, 1035–1038. [Google Scholar] [CrossRef]
  17. Boroujerdi, R.; Abdelkader, A.; Paul, R. Highly selective detection of ethanol in biological fluids and alcoholic drinks using indium ethylenediamine functionalized graphene. Sens. Diagn. 2022, 1, 566–578. [Google Scholar] [CrossRef]
  18. Pomes, L.M.; Guglielmetti, M.; Bertamino, E.; Simmaco, M.; Borro, M.; Martelletti, P. Optimising migraine treatment: From drug-drug interactions to personalized medicine. J. Headache Pain 2019, 20, 56. [Google Scholar] [CrossRef]
  19. Seymour, R.M.; Routledge, P.A. Important Drug-Drug Interactions in the Elderly. Drugs Aging 1998, 12, 485–494. [Google Scholar] [CrossRef]
  20. Kuhlman, J.J., Jr.; McCaulley, R.; Valouch, T.J.; Behonick, G.S. Fentanyl Use, Misuse, and Abuse: A Summary of 23 Postmortem Cases. J. Anal. Toxicol. 2003, 27, 499–504. [Google Scholar] [CrossRef] [PubMed]
  21. Lombardi, A.R.; Arya, R.; Rosen, J.G.; Thompson, E.; Welwean, R.; Tardif, J.; Rich, J.D.; Park, J.N. Overdose Detection Technologies to Reduce Solitary Overdose Deaths: A Literature Review. Int. J. Environ. Res. Public Health 2023, 20, 1230. [Google Scholar] [CrossRef]
  22. Ahmed, S.R.; Chand, R.; Kumar, S.; Mittal, N.; Srinivasan, S.; Rajabzadeh, A.R. Recent biosensing advances in the rapid detection of illicit drugs. TrAC Trends Anal. Chem. 2020, 131, 116006. [Google Scholar] [CrossRef]
  23. Mahajan, G. Role of urine drug testing in the current opioid epidemic. Analgesia 2017, 125, 2094–2104. [Google Scholar] [CrossRef]
  24. Florea, A.; de Jong, M.; De Wael, K. Electrochemical strategies for the detection of forensic drugs. Curr. Opin. Electrochem. 2018, 11, 34–40. [Google Scholar] [CrossRef]
  25. Dehdashtian, S.; Gholivand, M.B.; Shamsipur, M.; Kariminia, S. Construction of a sensitive and selective sensor for morphine using chitosan coated Fe3O4 magnetic nanoparticle as a modifier. Mater. Sci. Eng. C 2016, 58, 53–59. [Google Scholar] [CrossRef] [PubMed]
  26. Pu, S.; Zhou, M.; Tang, T.; Cheng, H.; Yan, X.; Hu, G. Boron-cluster-based porous BCN material modified electrode for electrochemical determination of morphine in serum. Microchim. Acta 2023, 190, 307. [Google Scholar] [CrossRef]
  27. Hagel, J.M.; Facchini, P.J. Dioxygenases catalyze the O-demethylation steps of morphine biosynthesis in opium poppy. Nat. Chem. Biol. 2010, 6, 273–275. [Google Scholar] [CrossRef] [PubMed]
  28. Klee, W.A.; Streaty, R.A. Narcotic Receptor Sites in Morphine-dependent Rats. Nature 1974, 248, 61–63. [Google Scholar] [CrossRef]
  29. Holbrook, T.L.; Galarneau, M.R.; Dye, J.L.; Quinn, K.; Dougherty, A.L. Morphine Use after Combat Injury in Iraq and Post-Traumatic Stress Disorder. N. Engl. J. Med. 2010, 362, 110–117. [Google Scholar] [CrossRef] [PubMed]
  30. Ekström, M.; Ferreira, D.; Chang, S.; Louw, S.; Johnson, M.J.; Eckert, D.J.; Fazekas, B.; Clark, K.J.; Agar, M.R.; Currow, D.C.; et al. Effect of Regular, Low-Dose, Extended-release Morphine on Chronic Breathlessness in Chronic Obstructive Pulmonary Disease: The BEAMS Randomized Clinical Trial. JAMA 2022, 328, 2022–2032. [Google Scholar] [CrossRef]
  31. Gilson, A.M.; Ryan, K.M.; Joranson, D.E.; Dahl, J.L. A reassessment of trends in the medical use and abuse of opioid analgesics and implications for diversion control: 1997–2002. J. Pain Symptom Manag. 2004, 28, 176–188. [Google Scholar] [CrossRef] [PubMed]
  32. Nicolakis, J.; Gmeiner, G.; Reiter, C.; Seltenhammer, M.H. Aspiration in lethal drug abuse—A consequence of opioid intoxication. Int. J. Leg. Med. 2020, 134, 2121–2132. [Google Scholar] [CrossRef] [PubMed]
  33. Smith, M.T.; Mun, C.J.; Remeniuk, B.; Finan, P.H.; Campbell, C.M.; Buenaver, L.F.; Robinson, M.; Fulton, B.; Tompkins, D.A.; Tremblay, J.-M.; et al. Experimental sleep disruption attenuates morphine analgesia: Findings from a randomized trial and implications for the opioid abuse epidemic. Sci. Rep. 2020, 10, 20121. [Google Scholar] [CrossRef]
  34. Narita, M.; Mizuo, K.; Mizoguchi, H.; Sakata, M.; Narita, M.; Tseng, L.F.; Suzuki, T. Molecular evidence for the functional role of dopamine D3 receptor in the morphine-induced rewarding effect and hyperlocomotion. J. Neurosci. 2003, 23, 1006–1012. [Google Scholar] [CrossRef]
  35. Wise, R.A. Drug-activation of brain reward pathways. Drug Alcohol Depend. 1998, 51, 13–22. [Google Scholar] [CrossRef]
  36. Chieng, B.; Williams, J.T. Increased opioid inhibition of GABA release in nucleus accumbens during morphine withdrawal. J. Neurosci. 1998, 18, 7033–7039. [Google Scholar] [CrossRef]
  37. Olds, M.E. Reinforcing effects of morphine in the nucleus accumbens. Brain Res. 1982, 237, 429–440. [Google Scholar] [CrossRef]
  38. Mayberry, H.L.; Bavley, C.C.; Karbalaei, R.; Peterson, D.R.; Bongiovanni, A.R.; Ellis, A.S.; Downey, S.H.; Toussaint, A.B.; Wimmer, M.E. Transcriptomics in the nucleus accumbens shell reveal sex- and reinforcer-specific signatures associated with morphine and sucrose craving. Neuropsychopharmacology 2022, 47, 1764–1775. [Google Scholar] [CrossRef]
  39. Zarrabian, S.; Riahi, E.; Karimi, S.; Razavi, Y.; Haghparast, A. The potential role of the orexin reward system in future treatments for opioid drug abuse. Brain Res. 2020, 1731, 146028. [Google Scholar] [CrossRef]
  40. Di Chiara, G.; Bassareo, V. Reward system and addiction: What dopamine does and doesn’t do. Curr. Opin. Pharmacol. 2007, 7, 69–76. [Google Scholar] [CrossRef]
  41. Reisi, Z.; Bani-Ardalan, M.; Zarepour, L.; Haghparast, A. Involvement of D1/D2 dopamine receptors within the nucleus accumbens and ventral tegmental area in the development of sensitization to antinociceptive effect of morphine. Pharmacol. Biochem. Behav. 2014, 118, 16–21. [Google Scholar] [CrossRef]
  42. Sardari, M.; Seddighfar, M.; Sardari, S. Dopamine receptors in the anterior cingulate cortex implicate in nicotine enhanced morphine analgesia. Psychopharmacology 2021, 238, 3311–3323. [Google Scholar] [CrossRef]
  43. Strandberg, J.J.; Kugelberg, F.C.; Alkass, K.; Gustavsson, A.; Zahlsen, K.; Spigset, O.; Druid, H. Toxicological analysis in rats subjected to heroin and morphine overdose. Toxicol. Lett. 2006, 166, 11–18. [Google Scholar] [CrossRef]
  44. Noufal, Y.; Kringel, D.; Toennes, S.W.; Dudziak, R.; Lötsch, J. Pharmacological data science perspective on fatal incidents of morphine treatment. Pharmacol. Ther. 2023, 241, 108312. [Google Scholar] [CrossRef]
  45. Hupka, Y.; Beike, J.; Roegener, J.; Brinkmann, B.; Blaschke, G.; Köhler, H. HPLC with laser-induced native fluorescence detection for morphine and morphine glucuronides from blood after immunoaffinity extraction. Int. J. Leg. Med. 2005, 119, 121–128. [Google Scholar] [CrossRef]
  46. Tagliaro, F.; Franchi, D.; Dorizzi, R.; Marigo, M. High-performance liquid chromatographic determination of morphine in biological samples: An overview of separation methods and detection techniques. J. Chromatogr. B Biomed. Sci. Appl. 1989, 488, 215–228. [Google Scholar] [CrossRef]
  47. Buratti, E.; Cippitelli, M.; Mietti, G.; Scendoni, R.; Froldi, R.; Cerioni, A.; Cingolani, M. Validation of an HPLC–HR-MS Method for the Determination and Quantification of Six Drugs (Morphine, Codeine, Methadone, Alprazolam, Clonazepam and Quetiapine) in Nails. J. Anal. Toxicol. 2023, 47, 488–493. [Google Scholar] [CrossRef]
  48. Sachs, H.; Arnold, W. Results of Comparative Determination of Morphine in Human Hair Using RIA and GC/MS. Clin. Chem. Lab. Med. 1989, 27, 873–878. [Google Scholar] [CrossRef] [PubMed]
  49. Meatherall, R. GC-MS Confirmation of Codeine, Morphine, 6-Acetylmorphine, Hydrocodone, Hydromorphone, Oxycodone, and Oxymorphone in Urine. J. Anal. Toxicol. 1999, 23, 177–186. [Google Scholar] [CrossRef] [PubMed]
  50. Ghazali, A.R.; Soon, C.K.; Othman, N.A.B. Stability of Extracted Morphine from Human Urine Samples at 4 °C for 7 Days. Bul. Sains Kesihat. 2023, 7, 20–23. [Google Scholar]
  51. Bahrami, G.; Ehzari, H.; Mirzabeigy, S.; Mohammadi, B.; Arkan, E. Fabrication of a sensitive electrochemical sensor based on electrospun magnetic nanofibers for morphine analysis in biological samples. Mater. Sci. Eng. C 2020, 106, 110183. [Google Scholar] [CrossRef]
  52. Beitollahi, H.; Garkani Nejad, F.; Tajik, S.; Di Bartolomeo, A. Screen-Printed Graphite Electrode Modified with Graphene-Co3O4 Nanocomposite: Voltammetric Assay of Morphine in the Presence of Diclofenac in Pharmaceutical and Biological Samples. Nanomaterials 2022, 12, 3454. [Google Scholar] [CrossRef] [PubMed]
  53. Qiu, B.; Chen, X.; Chen, H.-L.; Chen, G.-N. Electrochemiluminescence determination of codeine or morphine with an organically modified silicate film immobilizing Ru(bpy)32+. Luminescence 2007, 22, 189–194. [Google Scholar] [CrossRef] [PubMed]
  54. Ya, Y.; Xiaoshu, W.; Qing, D.; Lin, J.; Yifeng, T. Label-free immunosensor for morphine based on the electrochemiluminescence of luminol on indium–tin oxide coated glass functionalized with gold nanoparticles. Anal. Methods 2015, 7, 4502–4507. [Google Scholar] [CrossRef]
  55. Fei, W.; Chen, F.; Sun, L.; Li, Q.; Yang, J.; Wu, Y. Ultrasensitive electrochemiluminescent immunoassay for morphine using a gold electrode modified with CdS quantum dots, polyamidoamine, and gold nanoparticles. Microchim. Acta 2014, 181, 419–425. [Google Scholar] [CrossRef]
  56. Yin, Q.; Wang, Y.; Li, X.; Yang, D.; Yang, Y.; Yang, C.; Zhu, Y. Dual-Emission Carbon-Dot Ratiometric Fluorescence Sensor for Morphine Recognition in Biological Samples. Biosensors 2023, 13, 143. [Google Scholar] [CrossRef]
  57. Shcherbakova, E.G.; Zhang, B.; Gozem, S.; Minami, T.; Zavalij, P.Y.; Pushina, M.; Isaacs, L.D.; Anzenbacher, P., Jr. Supramolecular Sensors for Opiates and Their Metabolites. J. Am. Chem. Soc. 2017, 139, 14954–14960. [Google Scholar] [CrossRef]
  58. Li, W.; Li, X.; Yang, T.; Guo, X.; Song, Y. Detection of saliva morphine using surface-enhanced Raman spectroscopy combined with immunochromatographic assay. J. Raman Spectrosc. 2020, 51, 642–648. [Google Scholar] [CrossRef]
  59. Zhao, L.; Wei, Y.; Fu, H.; Yang, R.; Zhao, Q.; Zhang, H.; Cai, W. Solid chip-based detection of trace morphine in solutions via portable surface-enhanced Raman spectroscopy. Spectrochim. Acta Part A Mol. Biomol. Spectrosc. 2023, 302, 122977. [Google Scholar] [CrossRef]
  60. Feng, S.; Chen, W.; Huang, W.; Cheng, M.; Lin, J.; Li, Y.; Chen, R. Surface-enhanced Raman spectroscopy of morphine in silver colloid. Chin. Opt. Lett. 2009, 7, 1055–1057. [Google Scholar] [CrossRef]
  61. Cortade, D.L.; Wang, S.X. Quantitative and rapid detection of morphine and hydromorphone at the point of care by an automated giant magnetoresistive nanosensor platform. Anal. Bioanal. Chem. 2022, 414, 7211–7221. [Google Scholar] [CrossRef] [PubMed]
  62. Rohani Bastami, T.; Bayat, M.; Paolesse, R. Naked-Eye Detection of Morphine by Au@Ag Nanoparticles-Based Colorimetric Chemosensors. Sensors 2022, 22, 2072. [Google Scholar] [CrossRef] [PubMed]
  63. Shahzad, A.; Köhler, G.; Knapp, M.; Gaubitzer, E.; Puchinger, M.; Edetsberger, M. Emerging applications of fluorescence spectroscopy in medical microbiology field. J. Transl. Med. 2009, 7, 99. [Google Scholar] [CrossRef] [PubMed]
  64. Lakowicz, J.R. Principles of Fluorescence Spectroscopy; Springer: Berlin/Heidelberg, Germany, 2006. [Google Scholar]
  65. Pan, H.; Yu, H.; Wang, Y.; Liu, R.; Lei, H. Investigating variations of fluorescent dissolved organic matter in wastewater treatment using synchronous fluorescence spectroscopy combined with principal component analysis and two-dimensional correlation. Environ. Technol. 2018, 39, 2495–2502. [Google Scholar] [CrossRef]
  66. Kloosterman, A.; Mapes, A.; Geradts, Z.; van Eijk, E.; Koper, C.; van den Berg, J.; Verheij, S.; van der Steen, M.; van Asten, A. The interface between forensic science and technology: How technology could cause a paradigm shift in the role of forensic institutes in the criminal justice system. Philos. Trans. R. Soc. B Biol. Sci. 2015, 370, 20140264. [Google Scholar] [CrossRef]
  67. Yu, H.; Lee, H.; Cheong, J.; Woo, S.W.; Oh, J.; Oh, H.-K.; Lee, J.-H.; Zheng, H.; Castro, C.M.; Yoo, Y.-E.; et al. A rapid assay provides on-site quantification of tetrahydrocannabinol in oral fluid. Sci. Transl. Med. 2021, 13, eabe2352. [Google Scholar] [CrossRef]
  68. Hammett-Stabler, C.A.; Pesce, A.J.; Cannon, D.J. Urine drug screening in the medical setting. Clin. Chim. Acta 2002, 315, 125–135. [Google Scholar] [CrossRef]
  69. Griss, R.; Schena, A.; Reymond, L.; Patiny, L.; Werner, D.; Tinberg, C.E.; Baker, D.; Johnsson, K. Bioluminescent sensor proteins for point-of-care therapeutic drug monitoring. Nat. Chem. Biol. 2014, 10, 598–603. [Google Scholar] [CrossRef]
  70. Nelson, N.; Sander, D.; Dandin, M.; Prakash, S.B.; Sarje, A.; Abshire, P. Handheld Fluorometers for Lab-on-a-Chip Applications. IEEE Trans. Biomed. Circuits Syst. 2009, 3, 97–107. [Google Scholar] [CrossRef]
  71. Thomson-Laing, G.; Puddick, J.; Wood, S.A. Predicting cyanobacterial biovolumes from phycocyanin fluorescence using a handheld fluorometer in the field. Harmful Algae 2020, 97, 101869. [Google Scholar] [CrossRef]
  72. Boroujerdi, R.; Abdelkader, A.; Paul, R. Highly sensitive and selective detection of the antidepressant amitriptyline using functionalised graphene-based sensor. ChemNanoMat 2022, 8, e202200209. [Google Scholar] [CrossRef]
  73. Vermeire, A.; Remon, J.P. Stability and compatibility of morphine. Int. J. Pharm. 1999, 187, 17–51. [Google Scholar] [CrossRef] [PubMed]
  74. Beck, O.; Carlsson, S.; Tusic, M.; Olsson, R.; Franzen, L.; Hulten, P. Laboratory and clinical evaluation of on-site urine drug testing. Scand. J. Clin. Lab. Investig. 2014, 74, 681–686. [Google Scholar] [CrossRef] [PubMed]
  75. Andersson, M.; Stephanson, N.; Öhman, I.; Terzuoli, T.; Lindh, J.D.; Beck, O. Direct and efficient liquid chromatographic-tandem mass spectrometric method for opiates in urine drug testing—Importance of 6-acetylmorphine and reduction of analytes. Drug Test. Anal. 2014, 6, 317–324. [Google Scholar] [CrossRef]
  76. Kranenburg, R.F.; Ou, F.; Sevo, P.; Petruzzella, M.; de Ridder, R.; van Klinken, A.; Hakkel, K.D.; van Elst, D.M.J.; van Veldhoven, R.; Pagliano, F.; et al. On-site illicit-drug detection with an integrated near-infrared spectral sensor: A proof of concept. Talanta 2022, 245, 123441. [Google Scholar] [CrossRef] [PubMed]
  77. Dagar, M.; Yadav, S.; Sai, V.V.R.; Satija, J.; Bhatia, H. Emerging trends in point-of-care sensors for illicit drugs analysis. Talanta 2022, 238, 123048. [Google Scholar] [CrossRef]
  78. Vinks, A.A.; Punt, N.C.; Menke, F.; Kirkendall, E.; Butler, D.; Duggan, T.J.; Cortezzo, D.E.; Kiger, S.; Dietrich, T.; Spencer, P.; et al. Electronic Health Record–Embedded Decision Support Platform for Morphine Precision Dosing in Neonates. Clin. Pharmacol. Ther. 2020, 107, 186–194. [Google Scholar] [CrossRef] [PubMed]
  79. Khani, R.; Ghiamati, E.; Boroujerdi, R.; Rezaeifard, A.; Zaryabi, M.H. A new and highly selective turn-on fluorescent sensor with fast response time for the monitoring of cadmium ions in cosmetic, and health product samples. Spectrochim. Acta Part A Mol. Biomol. Spectrosc. 2016, 163, 120–126. [Google Scholar] [CrossRef]
  80. Asfaram, A.; Ghaedi, M.; Goudarzi, A.; Rajabi, M. Response surface methodology approach for optimization of simultaneous dye and metal ion ultrasound-assisted adsorption onto Mn doped Fe3O4-NPs loaded on AC: Kinetic and isothermal studies. Dalton Trans. 2015, 44, 14707–14723. [Google Scholar] [CrossRef]
  81. Kicsi, A.; Cojocaru, C.; Macoveanu, M.; Bilba, D. Response surface methodology applied for zinc removal from aqueous solutions using sphagnum peat moss as sorbent. J. Environ. Prot. Ecol. 2010, 11, 614–622. [Google Scholar]
  82. Parsaee, Z.; Karachi, N.; Abrishamifar, S.M.; Kahkha, M.R.R.; Razavi, R. Silver-choline chloride modified graphene oxide: Novel nano-bioelectrochemical sensor for celecoxib detection and CCD-RSM model. Ultrason. Sonochemistry 2018, 45, 106–115. [Google Scholar] [CrossRef]
  83. Yilgor, I.; Yilgor, E.; Guler, I.G.; Ward, T.C.; Wilkes, G.L. FTIR investigation of the influence of diisocyanate symmetry on the morphology development in model segmented polyurethanes. Polymer 2006, 47, 4105–4114. [Google Scholar] [CrossRef]
  84. Sharma, S.K.; Sharma, G.; Sharma, A.; Bhardwaj, K.; Preeti, K.; Singh, K.; Kumar, A.; Pal, V.K.; Choi, E.H.; Singh, S.P.; et al. Synthesis of silica and carbon-based nanomaterials from rice husk ash by ambient fiery and furnace sweltering using a chemical method. Appl. Surf. Sci. Adv. 2022, 8, 100225. [Google Scholar] [CrossRef]
  85. Oh, S.Y.; Yoo, D.I.; Shin, Y.; Kim, H.C.; Kim, H.Y.; Chung, Y.S.; Park, W.H.; Youk, J.H. Crystalline structure analysis of cellulose treated with sodium hydroxide and carbon dioxide by means of X-ray diffraction and FTIR spectroscopy. Carbohydr. Res. 2005, 340, 2376–2391. [Google Scholar] [CrossRef]
  86. Li, J.; Li, M.; Zhou, L.-L.; Lang, S.-Y.; Lu, H.-Y.; Wang, D.; Chen, C.-F.; Wan, L.-J. Click and patterned functionalization of graphene by Diels–Alder reaction. J. Am. Chem. Soc. 2016, 138, 7448–7451. [Google Scholar] [CrossRef]
  87. Sousa, C.; Silva, P.J. BBr3-Assisted Cleavage of Most Ethers Does Not Follow the Commonly Assumed Mechanism. Eur. J. Org. Chem. 2013, 2013, 5195–5199. [Google Scholar] [CrossRef]
  88. Huberty, J.S.; Madix, R.J. An FTIR study of the bonding of methoxy on Ni(100): Effects of coadsorbed sulfur, carbon monoxide and hydrogen. Surf. Sci. 1996, 360, 144–156. [Google Scholar] [CrossRef]
  89. Ellis, D.I.; Goodacre, R. Metabolic fingerprinting in disease diagnosis: Biomedical applications of infrared and Raman spectroscopy. Analyst 2006, 131, 875–885. [Google Scholar] [CrossRef] [PubMed]
  90. Feng, S.; Chen, R.; Lin, J.; Pan, J.; Wu, Y.; Li, Y.; Chen, J.; Zeng, H. Gastric cancer detection based on blood plasma surface-enhanced Raman spectroscopy excited by polarized laser light. Biosens. Bioelectron. 2011, 26, 3167–3174. [Google Scholar] [CrossRef] [PubMed]
  91. Rojalin, T.; Kurki, L.; Laaksonen, T.; Viitala, T.; Kostamovaara, J.; Gordon, K.C.; Galvis, L.; Wachsmann-Hogiu, S.; Strachan, C.J.; Yliperttula, M. Fluorescence-suppressed time-resolved Raman spectroscopy of pharmaceuticals using complementary metal-oxide semiconductor (CMOS) single-photon avalanche diode (SPAD) detector. Anal. Bioanal. Chem. 2016, 408, 761–774. [Google Scholar] [CrossRef] [PubMed]
  92. Hu, Y.; Liu, C.; Wang, X.; Zhao, D. Adaptive handling of Rayleigh and Raman scatter of fluorescence data based on evaluation of the degree of spectral overlap. Spectrochim. Acta Part A Mol. Biomol. Spectrosc. 2018, 199, 170–178. [Google Scholar] [CrossRef] [PubMed]
  93. Wei, D.; Chen, S.; Liu, Q. Review of Fluorescence Suppression Techniques in Raman Spectroscopy. Appl. Spectrosc. Rev. 2015, 50, 387–406. [Google Scholar] [CrossRef]
  94. Osticioli, I.; Zoppi, A.; Castellucci, E.M. Fluorescence and Raman spectra on painting materials: Reconstruction of spectra with mathematical methods. J. Raman Spectrosc. 2006, 37, 974–980. [Google Scholar] [CrossRef]
  95. Chen, J.; Li, J.; Xu, L.; Hong, W.; Yang, Y.; Chen, X. The Glass-Transition Temperature of Supported PMMA Thin Films with Hydrogen Bond/Plasmonic Interface. Polymers 2019, 11, 601. [Google Scholar] [CrossRef]
  96. Roth, C.B.; Dutcher, J.R. Glass transition temperature of freely-standing films of atactic poly(methyl methacrylate). Eur. Phys. J. E 2003, 12, 103–107. [Google Scholar] [CrossRef]
  97. Monti, P.; Freddi, G.; Bertoluzza, A.; Kasai, N.; Tsukada, M. Raman spectroscopic studies of silk fibroin from Bombyx mori. J. Raman Spectrosc. 1998, 29, 297–304. [Google Scholar] [CrossRef]
  98. Takeuchi, H.; Harada, I. Ultraviolet resonance Raman spectroscopy of X—Proline bonds: A new marker band of hydrogen bonding at the imide C=O site. J. Raman Spectrosc. 1990, 21, 509–515. [Google Scholar] [CrossRef]
  99. Vidya, S.; Ravikumar, C.; Hubert Joe, I.; Kumaradhas, P.; Devipriya, B.; Raju, K. Vibrational spectra and structural studies of nonlinear optical crystal ammonium D, L-tartrate: A density functional theoretical approach. J. Raman Spectrosc. 2011, 42, 676–684. [Google Scholar] [CrossRef]
  100. Perera, P.N.; Fega, K.R.; Lawrence, C.; Sundstrom, E.J.; Tomlinson-Phillips, J.; Ben-Amotz, D. Observation of water dangling OH bonds around dissolved nonpolar groups. Proc. Natl. Acad. Sci. USA 2009, 106, 12230–12234. [Google Scholar] [CrossRef]
  101. Mukherjee, S.; Dhar, S. Synchronous fluorescence techniques for the detection and monitoring of selected fluorescent dyes in binary and multifluorophoric mixtures. Results Chem. 2023, 5, 100867. [Google Scholar] [CrossRef]
  102. Assi, S.; Abbas, I.; Arafat, B.; Evans, K.; Al-Jumeily, D. Authentication of COVID-19 Vaccines Using Synchronous Fluorescence Spectroscopy. J. Fluoresc. 2023, 33, 1165–1174. [Google Scholar] [CrossRef] [PubMed]
  103. El-Aziz, H.A.; Fathy, M.E.; El-Enany, N.; Aly, F.A.; Tolba, M.M. Concurrent estimation of some co-administered antimicrobial drugs applying conventional and first derivative synchronous fluorescence spectroscopy techniques. Spectrochim. Acta Part A Mol. Biomol. Spectrosc. 2022, 264, 120255. [Google Scholar] [CrossRef]
  104. Wakeham, S.G. Synchronous fluorescence spectroscopy and its application to indigenous and petroleum-derived hydrocarbons in lacustrine sediments. Environ. Sci. Technol. 1977, 11, 272–276. [Google Scholar] [CrossRef]
  105. Ghosh, M.; Nath, S.; Hajra, A.; Sinha, S. Fluorescence self-quenching of tetraphenylporphyrin in liquid medium. J. Lumin. 2013, 141, 87–92. [Google Scholar] [CrossRef]
  106. Chen, W.; Young, L.J.; Lu, M.; Zaccone, A.; Ströhl, F.; Yu, N.; Kaminski Schierle, G.S.; Kaminski, C.F. Fluorescence Self-Quenching from Reporter Dyes Informs on the Structural Properties of Amyloid Clusters Formed In Vitro and in Cells. Nano Lett. 2017, 17, 143–149. [Google Scholar] [CrossRef]
  107. Becerra, L.; Harter, K.; Gonzalez, R.G.; Borsook, D. Functional Magnetic Resonance Imaging Measures of the Effects of Morphine on Central Nervous System Circuitry in Opioid-Naive Healthy Volunteers. Anesth. Analg. 2006, 103, 208–216. [Google Scholar] [CrossRef]
  108. Wong, D.H.W.; Jenkins, L.C. An experimental study of the mechanism of action of ketamine on the central nervous system. Can. Anaesth. Soc. J. 1974, 21, 57–67. [Google Scholar] [CrossRef]
  109. Mandona, J.W.; Tuk, B.; van Steveninck, A.L.; Breimer, D.D.; Cohen, A.F.; Danhof, M. Pharmacokinetic-pharmacodynamic modeling of the central nervous system effects of midazolam and its main metabolite α-hydroxymidazolam in healthy volunteers. Clin. Pharmacol. Ther. 1992, 51, 715–728. [Google Scholar] [CrossRef] [PubMed]
  110. Silvette, H.; Hoff, E.C.; Larson, P.S.; Haag, H.B. The Actions of Nicotine on Central Nervous System Functions. Pharmacol. Rev. 1962, 14, 137. [Google Scholar] [PubMed]
  111. Hashimoto, K.; Ide, S.; Arata, M.; Nakata, A.; Ito, A.; Ito, T.K.; Kudo, N.; Lin, B.; Nunomura, K.; Tsuganezawa, K.; et al. Discovery of Benzylpiperazine Derivatives as CNS-Penetrant and Selective Histone Deacetylase 6 Inhibitors. ACS Med. Chem. Lett. 2022, 13, 1077–1082. [Google Scholar] [CrossRef] [PubMed]
  112. Broussard, J.A.; Rappaz, B.; Webb, D.J.; Brown, C.M. Fluorescence resonance energy transfer microscopy as demonstrated by measuring the activation of the serine/threonine kinase Akt. Nat. Protoc. 2013, 8, 265–281. [Google Scholar] [CrossRef] [PubMed]
  113. Noun, F.; Jury, E.A.; Naccache, R. Elucidating the Quenching Mechanism in Carbon Dot-Metal Interactions–Designing Sensitive and Selective Optical Probes. Sensors 2021, 21, 1391. [Google Scholar] [CrossRef] [PubMed]
  114. Kaserer, T.; Lantero, A.; Schmidhammer, H.; Spetea, M.; Schuster, D. μ Opioid receptor: Novel antagonists and structural modeling. Sci. Rep. 2016, 6, 21548. [Google Scholar] [CrossRef] [PubMed]
  115. French, D. The challenges of LC–MS/MS analysis of opiates and opioids in urine. Bioanalysis 2013, 5, 2803–2820. [Google Scholar] [CrossRef]
  116. Jakobsson, G.; Truver, M.T.; Wrobel, S.A.; Gréen, H.; Kronstrand, R. Heroin-Related Compounds and Metabolic Ratios in Postmortem Samples Using LC–MS-MS. J. Anal. Toxicol. 2021, 45, 215–225. [Google Scholar] [CrossRef]
  117. Yen, Y.-T.; Chang, Y.-J.; Lai, P.-J.; Chang, C.-L.; Chen, T.-Y.; Chyueh, S.-C. A Study of Opiate, Opiate Metabolites and Antihistamines in Urine after Consumption of Cold Syrups by LC-MS/MS. Molecules 2020, 25, 972. [Google Scholar] [CrossRef]
  118. Haidekker, M.A.; Brady, T.P.; Lichlyter, D.; Theodorakis, E.A. Effects of solvent polarity and solvent viscosity on the fluorescent properties of molecular rotors and related probes. Bioorganic Chem. 2005, 33, 415–425. [Google Scholar] [CrossRef]
  119. Yetilmezsoy, K.; Saral, A. Stochastic modeling approaches based on neural network and linear–nonlinear regression techniques for the determination of single droplet collection efficiency of countercurrent spray towers. Environ. Model. Assess. 2007, 12, 13–26. [Google Scholar] [CrossRef]
  120. Yetilmezsoy, K.; Demirel, S.; Vanderbei, R.J. Response surface modeling of Pb(II) removal from aqueous solution by Pistacia vera L.: Box–Behnken experimental design. J. Hazard. Mater. 2009, 171, 551–562. [Google Scholar] [CrossRef] [PubMed]
  121. Boroujerdi, R.; Paul, R. Introducing Graphene–Indium Oxide Electrochemical Sensor for Detecting Ethanol in Aqueous Samples with CCD-RSM Optimization. Chemosensors 2022, 10, 42. [Google Scholar] [CrossRef]
  122. Musabeygi, T.; Goudarzi, N.; Arab-Chamjangali, M.; Mirzaee, M. Fabrication of a magnetic composite by CoFe2O4 and an inorganic polymer for simultaneous photo-degradation of organic pollutants under visible LED light: Bandgap engineering, CCD-RSM modeling, and resolving spectral overlap of analytes. J. Mol. Liq. 2022, 362, 119692. [Google Scholar] [CrossRef]
  123. Khurshid, S.; Iqbal, S.; Arif, S.; Akbar, Q.-U.-A.; Iqbal, H.M.; Yousaf, S.; Uddin, R.; Ullah, R.; Fatima, N. Chemometric assisted RSM–CCD strategy for optimization of RP-HPLC conditions and its application for simultaneous determination of multiclass pesticides in plant protection products. J. Iran. Chem. Soc. 2023, 20, 701–711. [Google Scholar] [CrossRef]
  124. Liu, H.-L.; Lan, Y.-W.; Cheng, Y.-C. Optimal production of sulphuric acid by Thiobacillus thiooxidans using response surface methodology. Process Biochem. 2004, 39, 1953–1961. [Google Scholar] [CrossRef]
  125. Arafat Hossain, M.; Ganesan, P.; Jewaratnam, J.; Chinna, K. Optimization of process parameters for microwave pyrolysis of oil palm fiber (OPF) for hydrogen and biochar production. Energy Convers. Manag. 2017, 133, 349–362. [Google Scholar] [CrossRef]
  126. Ghaleb, A.A.; Kutty, S.R.; Ho, Y.-C.; Jagaba, A.H.; Noor, A.; Al-Sabaeei, A.M.; Almahbashi, N.M. Response Surface Methodology to Optimize Methane Production from Mesophilic Anaerobic Co-Digestion of Oily-Biological Sludge and Sugarcane Bagasse. Sustainability 2020, 12, 2116. [Google Scholar] [CrossRef]
  127. Afsharimani, B.; Cabot, P.; Parat, M.-O. Morphine Use in Cancer Surgery. Front. Pharmacol. 2011, 2, 46. [Google Scholar] [CrossRef]
  128. Denton, J.E.; Beecher, H.K. New analgesics. 2. A clinical appraisal of the narcotic power of methadone and its isomers. JAMA J. Am. Med. Assoc. 1949, 141, 1146–1153. [Google Scholar] [CrossRef]
  129. Lasagna, L.; Beecher, H.K. The Optimal Dose of Morphine. J. Am. Med. Assoc. 1954, 156, 230–234. [Google Scholar] [CrossRef] [PubMed]
  130. Wang, J.; Sun, H.; Sun, W.-T.; Sun, H.-P.; Tian, T.; Sun, J. Efficacy and safety of intrathecal morphine for pain control after spinal surgery: A systematic review and meta-analysis. Eur. Rev. Med. Pharmacol. Sci. 2021, 25, 2674–2684. [Google Scholar] [PubMed]
  131. Alam, M.; Hartrick, C.T. Extended-Release Epidural Morphine (DepoDur™): An Old Drug with a New Profile. Pain Pract. 2005, 5, 349–353. [Google Scholar] [CrossRef]
  132. Baraka, A.; Noueihid, R.; Hajj, S. Intrathecal injection of morphine for obstetric analgesia. Anesthesiology 1981, 54, 136–140. [Google Scholar] [CrossRef]
  133. Chatham, M.S.; Dodds Ashley, E.S.; Svengsouk, J.S.; Juba, K.M. Dose Ratios between High Dose Oral Morphine or Equivalents and Oral Methadone. J. Palliat. Med. 2013, 16, 947–950. [Google Scholar] [CrossRef]
  134. Meldon, K.; Anita, S.; Lynn, W.; Douglas, G.; Deana, M. Misuse of and dependence on opioids: Study of chronic pain patients. Can. Fam. Physician 2006, 52, 1081. [Google Scholar]
  135. Wang, C.; Luo, J.; Dou, H.; Raise, A.; Ali, M.S.; Fan, W.; Li, Q. Optimization and analytical behavior of a morphine electrochemical sensor in environmental and biological samples based on graphite rod electrode using graphene/Co3O4 nanocomposite. Chemosphere 2023, 326, 138451. [Google Scholar] [CrossRef] [PubMed]
  136. Krebs, E.E.; Becker, W.C.; Zerzan, J.; Bair, M.J.; McCoy, K.; Hui, S. Comparative mortality among Department of Veterans Affairs patients prescribed methadone or long-acting morphine for chronic pain. Pain 2011, 152, 1789–1795. [Google Scholar] [CrossRef] [PubMed]
  137. Campos-Mañas, M.C.; Ferrer, I.; Thurman, E.M.; Agüera, A. Opioid occurrence in environmental water samples—A review. Trends Environ. Anal. Chem. 2018, 20, e00059. [Google Scholar] [CrossRef]
  138. Özbunar, E.; Aydoğdu, M.; Döğer, R.; Bostancı, H.İ.; Koruyucu, M.; Akgür, S.A. Morphine Concentrations in Human Urine Following Poppy Seed Paste Consumption. Forensic Sci. Int. 2019, 295, 121–127. [Google Scholar] [CrossRef] [PubMed]
  139. Ghorani-Azam, A.; Balali-Mood, M.; Khatami, S.M.; Asoodeh, A.; Es’haghi, Z.; Riahi-Zanjani, B. Plant extract and herbal products as potential source of sorbent for analytical purpose: An experimental study of morphine and codeine determination using HPLC and LC–MSMS. J. Chromatogr. Sci. 2021, 59, 482–489. [Google Scholar] [CrossRef] [PubMed]
  140. Soltanabadi, Z.; Esmaeili, A.; Bambai, B. Fabrication of morphine detector based on quartz@Au-layer biosensor. Microchem. J. 2022, 175, 107127. [Google Scholar] [CrossRef]
  141. Chu, Z.; Fu, M.; Guo, J.; Wang, W.; Zhou, J.; Ma, X.; Guo, J. Magnetic Resistance Sensory System for the Quantitative Measurement of Morphine. IEEE Trans. Biomed. Circuits Syst. 2021, 15, 171–176. [Google Scholar] [CrossRef]
  142. Ke, H.; Du, X.; Wang, L.; Wang, X.; Zhu, J.; Gao, Y.; Peng, B.; Hao, H.; Cai, N. Detection of morphine in urine based on a surface plasmon resonance imaging immunoassay. Anal. Methods 2020, 12, 3038–3044. [Google Scholar] [CrossRef] [PubMed]
  143. Mohseni, N.; Bahram, M. Mean centering of ratio spectra for colorimetric determination of morphine and codeine in pharmaceuticals and biological samples using melamine modified gold nanoparticles. Anal. Methods 2016, 8, 6739–6747. [Google Scholar] [CrossRef]
  144. Habibi, M.M.; Ghasemi, J.B.; Badiei, A.; Norouzi, P. Simultaneous electrochemical determination of morphine and methadone by using CMK-5 mesoporous carbon and multivariate calibration. Sci. Rep. 2022, 12, 8270. [Google Scholar] [CrossRef] [PubMed]
  145. Imanzadeh, H.; Khataee, A.; Hazraty, L.; Amiri, M. Broken hollow carbon spheres decorated by gold nanodendrites as the advanced electrochemical sensing platform for sensitive tracing of morphine in human serum and saliva. Sens. Actuators B Chem. 2024, 398, 134738. [Google Scholar] [CrossRef]
  146. Liu, H.; Motlak, M.; Feng, Z.; Kaffash, A. Three-dimensional hierarchical Co3O4/carbon composite: Hydrothermal synthesis and morphine electrochemical sensing application. Inorg. Chem. Commun. 2023, 151, 110629. [Google Scholar] [CrossRef]
  147. Zhao, X.; Fu, Y.; Ren, C.; Guo, J.; Kang, Y. Quantitative detection of morphine based on an up-conversion luminescent system. Analyst 2021, 146, 989–996. [Google Scholar] [CrossRef]
  148. Zarad, W.; Shawky, A.; Ali, A.; Aboulella, Y.; Kamal, M.; Masujima, T.; Emara, S.; El-Gendy, H. Field amplified sample stacking and in-capillary derivatization for forensic analysis of morphine and morphine-6-glucuronide in human urine by capillary electrophoresis. Talanta Open 2021, 3, 100041. [Google Scholar] [CrossRef]
  149. Nebu, J.; Anjali Devi, J.S.; Aparna, R.S.; Aswathy, B.; Aswathy, A.O.; Sony, G. Fluorometric determination of morphine via its effect on the quenching of fluorescein by gold nanoparticles through a surface energy transfer process. Microchim. Acta 2018, 185, 532. [Google Scholar] [CrossRef]
Figure 1. Synthesis process of 7,7′-Dimethoxy-1,1′-binaphthalene and 7′-Methoxy-[1,1′-binaphthalen]-7-ol.
Figure 1. Synthesis process of 7,7′-Dimethoxy-1,1′-binaphthalene and 7′-Methoxy-[1,1′-binaphthalen]-7-ol.
Sensors 24 01722 g001
Figure 2. Comparison between the FTIR spectra of 7-methoxy-1-tetralone, 7,7′-Dimethoxy-1,1′-binaphthalene, and 7′-Methoxy-[1,1′-binaphthalen]-7-ol.
Figure 2. Comparison between the FTIR spectra of 7-methoxy-1-tetralone, 7,7′-Dimethoxy-1,1′-binaphthalene, and 7′-Methoxy-[1,1′-binaphthalen]-7-ol.
Sensors 24 01722 g002
Figure 3. Comparison between the Raman spectra of 7-methoxy-1-tetralone, 7,7′-Dimethoxy-1,1′-binaphthalene, and 7′-Methoxy-[1,1′-binaphthalen]-7-ol.
Figure 3. Comparison between the Raman spectra of 7-methoxy-1-tetralone, 7,7′-Dimethoxy-1,1′-binaphthalene, and 7′-Methoxy-[1,1′-binaphthalen]-7-ol.
Sensors 24 01722 g003
Figure 4. (a) The three-dimensional mapping graph illustrates how altering the excitation wavelength affects the intensity of the sensor’s emission peak. (b) The contour map displays the fluorescence emission and excitation spectra of the sensor, with the intersection denoting the position of the fluorescence peak (λex = 325 nm, λem = 369 nm). (c) A comparison of the emission (λex = 325 nm) and excitation (λem = 369 nm) spectra of 400 ppm 7′-Methoxy-[1,1′-binaphthalen]-7-ol dissolved in methanol. (d) The CIE chromaticity plot reveals the color coordinates of the emission beam, demonstrating that the molecule emits blue light upon excitation (λex = 325 nm).
Figure 4. (a) The three-dimensional mapping graph illustrates how altering the excitation wavelength affects the intensity of the sensor’s emission peak. (b) The contour map displays the fluorescence emission and excitation spectra of the sensor, with the intersection denoting the position of the fluorescence peak (λex = 325 nm, λem = 369 nm). (c) A comparison of the emission (λex = 325 nm) and excitation (λem = 369 nm) spectra of 400 ppm 7′-Methoxy-[1,1′-binaphthalen]-7-ol dissolved in methanol. (d) The CIE chromaticity plot reveals the color coordinates of the emission beam, demonstrating that the molecule emits blue light upon excitation (λex = 325 nm).
Sensors 24 01722 g004
Figure 5. (a) The three-dimensional synchronous fluorescence spectroscopy mapping. (b) The influence of time on the fluorescence-emission intensity of the developed sensor.
Figure 5. (a) The three-dimensional synchronous fluorescence spectroscopy mapping. (b) The influence of time on the fluorescence-emission intensity of the developed sensor.
Sensors 24 01722 g005
Figure 6. (a) Compares the variations in the intensity of the fluorescence-emission peak due to alterations in the concentration of morphine, both in the presence and absence of the developed sensor. (b) Illustrates the sensor’s linear range, in response to morphine, displaying an R² value of 0.98 (λex = 325 nm). (c) Demonstrates the linear range of morphine’s fluorescence peak in the absence of the sensor, accompanied by an R² value of 0.99 (λex = 325 nm).
Figure 6. (a) Compares the variations in the intensity of the fluorescence-emission peak due to alterations in the concentration of morphine, both in the presence and absence of the developed sensor. (b) Illustrates the sensor’s linear range, in response to morphine, displaying an R² value of 0.98 (λex = 325 nm). (c) Demonstrates the linear range of morphine’s fluorescence peak in the absence of the sensor, accompanied by an R² value of 0.99 (λex = 325 nm).
Sensors 24 01722 g006
Figure 7. (a) Bar chart illustrating variations in sensor fluorescence when exposed to various drugs (20 ppm each) individually. F represents the fluorescence response of the sensor in the presence of the drug, while F’ signifies the sensor’s fluorescence response in the absence of any drugs. (b) The competition experiment on the effect of other drugs on morphine detection by developing a mixture of morphine (20 ppm) with other drugs (20 ppm each) separately. (c) The possible reaction sites for the formation of a non-fluorescent complex between morphine and the sensor. (d) Effect of temperature on fluorescence-emission intensity of the sensor in the presence of a constant amount of morphine (20 ppm in methanol).
Figure 7. (a) Bar chart illustrating variations in sensor fluorescence when exposed to various drugs (20 ppm each) individually. F represents the fluorescence response of the sensor in the presence of the drug, while F’ signifies the sensor’s fluorescence response in the absence of any drugs. (b) The competition experiment on the effect of other drugs on morphine detection by developing a mixture of morphine (20 ppm) with other drugs (20 ppm each) separately. (c) The possible reaction sites for the formation of a non-fluorescent complex between morphine and the sensor. (d) Effect of temperature on fluorescence-emission intensity of the sensor in the presence of a constant amount of morphine (20 ppm in methanol).
Sensors 24 01722 g007
Figure 8. Comparison of fluorescence-emission spectra. The graph illustrates the fluorescence-emission spectra of the sensor (100 ppm in methanol) in a 1:1 mixture with urine, morphine (40 ppm in methanol) in a 1:1 mixture with urine, pure methanol in a 1:1 mixture with urine, pure urine without any additives, and pure analytical-grade methanol.
Figure 8. Comparison of fluorescence-emission spectra. The graph illustrates the fluorescence-emission spectra of the sensor (100 ppm in methanol) in a 1:1 mixture with urine, morphine (40 ppm in methanol) in a 1:1 mixture with urine, pure methanol in a 1:1 mixture with urine, pure urine without any additives, and pure analytical-grade methanol.
Sensors 24 01722 g008
Figure 9. (a) Sensor’s response to morphine concentration: This panel illustrates the dynamic response of the sensor as it detects variations in morphine concentration within urine samples. (b) Normal plot of residuals: In this plot, the residuals are visually examined for adherence to normal distribution assumptions, providing insights into the model’s performance. (c) Predicted-versus-actual graph: This graph showcases the predictive accuracy of the model by comparing the predicted sensor’s fluorescence emission with the actual values.
Figure 9. (a) Sensor’s response to morphine concentration: This panel illustrates the dynamic response of the sensor as it detects variations in morphine concentration within urine samples. (b) Normal plot of residuals: In this plot, the residuals are visually examined for adherence to normal distribution assumptions, providing insights into the model’s performance. (c) Predicted-versus-actual graph: This graph showcases the predictive accuracy of the model by comparing the predicted sensor’s fluorescence emission with the actual values.
Sensors 24 01722 g009
Figure 10. Three-dimensional plots of sensor’s fluorescence-emission response in urine samples: (a) interactive effect of morphine’s concentration and temperature; (b) interactive time and morphine’s concentration; (c) interactive effect of time and temperature; (d) interactive effect of sensor’s concentration and morphine’s concentration; (e) interactive effect of sensor’s concentration and temperature; (f) interactive effect of time and sensor’s concentration.
Figure 10. Three-dimensional plots of sensor’s fluorescence-emission response in urine samples: (a) interactive effect of morphine’s concentration and temperature; (b) interactive time and morphine’s concentration; (c) interactive effect of time and temperature; (d) interactive effect of sensor’s concentration and morphine’s concentration; (e) interactive effect of sensor’s concentration and temperature; (f) interactive effect of time and sensor’s concentration.
Sensors 24 01722 g010
Figure 11. Two-dimensional desirability graphs versus actual points: (a) Actual factors include temperature at 18.53 °C and addition time at 7.47 min; (b) Actual factors include morphine’s concentration at 21.83 ppb and addition time at 7.47; (c) Actual factors include morphine’s concentration at 21.83 ppb and temperature at 18.53 °C (d) Actual factors include sensor’s concentration at 75 ppm and addition time at 7.47 min; (e) Actual factors include temperature at 18.53 °C and sensor’s concentration at 75 ppm; (f) Actual factors include morphine’s concentration at 21.83 ppb and sensor’s concentration at 75 ppm.
Figure 11. Two-dimensional desirability graphs versus actual points: (a) Actual factors include temperature at 18.53 °C and addition time at 7.47 min; (b) Actual factors include morphine’s concentration at 21.83 ppb and addition time at 7.47; (c) Actual factors include morphine’s concentration at 21.83 ppb and temperature at 18.53 °C (d) Actual factors include sensor’s concentration at 75 ppm and addition time at 7.47 min; (e) Actual factors include temperature at 18.53 °C and sensor’s concentration at 75 ppm; (f) Actual factors include morphine’s concentration at 21.83 ppb and sensor’s concentration at 75 ppm.
Sensors 24 01722 g011
Table 1. Experimental factors and levels in the central composition design.
Table 1. Experimental factors and levels in the central composition design.
NotationFactorUnitRange and Levels
−α−10+1
AConcentration of morphine in the solutionppb020406080
BTemperature of the solution°C010203040
CConcentration of the sensor in the solutionppm5075100125150
DWaiting time after addition of morphinemin036912
Table 2. Design matrix and the results of the central composite full factorial design.
Table 2. Design matrix and the results of the central composite full factorial design.
RunABCDLeverageResponse (Emission)Space Type
1402015065.833 × 10−11.14 × 106Factorial
2404010065.833 × 10−11.01 × 106Factorial
3402010061.667 × 10−18.19 × 105Factorial
4402010061.667 × 10−11.667 × 10−1Factorial
520107595.833 × 10−18.40 × 105Factorial
6203012595.833 × 10−11.26 × 106Factorial
7402010061.667 × 10−17.83 × 105Factorial
8203012535.833 × 10−11.27 × 106Center
960107595.833 × 10−16.59 × 105Factorial
10603012595.833 × 10−11.10 × 106Factorial
1160307535.833 × 10−17.29 × 105Factorial
124020100125.833 × 10−17.91 × 105Factorial
13402010061.667 × 10−18.01 × 105Center
14201012595.833 × 10−11.10 × 106Center
15402010061.667 × 10−17.83 × 105Center
16402010061.667 × 10−18.04 × 105Factorial
1740010065.833 × 10−11.02 × 106Factorial
1860307595.833 × 10−17.95 × 105Factorial
19601012535.833 × 10−11.08 × 106Factorial
2020307595.833 × 10−18.57 × 105Factorial
21603012535.833 × 10−19.86 × 105Axial
22402010005.833 × 10−19.14 × 105Center
2320107535.833 × 10−19.05 × 105Axial
2460107535.833 × 10−16.61 × 105Axial
25802010065.833 × 10−11.11 × 106Axial
2602010065.833 × 10−11.25 × 106Axial
27601012595.833 × 10−11.09 × 106Axial
2840205065.833 × 10−14.07 × 105Center
29201012535.833 × 10−11.10 × 106Axial
3020307535.833 × 10−19.58 × 105Axial
Table 3. The ANOVA results of CCD for the developed morphine sensor in urine samples.
Table 3. The ANOVA results of CCD for the developed morphine sensor in urine samples.
SourceSum of SquaresDegree of FreedomMean SquareF-Valuep-Value
Model1.16 × 1012148.31 × 101044.91.16 × 10−9significant
A-Morphine’s Concentration9.08 × 101019.08 × 1010494.29 × 10−6
B-Temperature5.98 × 10915.98 × 1093.239.26 × 10−2
C-Sensor’s Concentration6.86 × 101116.86 × 10113705.52 × 10−12
D-Addition time2.99 × 101012.99 × 101016.21.11 × 10−3
AB8.77 × 10818.77 × 1080.4735.02 × 10−1
AC3.54 × 10913.54 × 1091.911.87 × 10−1
AD3.72 × 10913.72 × 1092.011.77 × 10−1
BC2.78 × 101012.78 × 1010151.50 × 10−3
BD2.27 × 10912.27 × 1091.232.86 × 10−1
CD1.43 × 10911.43 × 1090.7713.94 × 10−1
2.69 × 101112.69 × 10111454.08 × 10−9
5.29 × 101015.29 × 101028.68.19 × 10−5
1.39 × 10811.39 × 1080.07537.88 × 10−1
2.46 × 101012.46 × 101013.32.40 × 10−3
Residual2.78 × 1010151.85 × 109
Lack of Fit2.44 × 1010102.44 × 1093.658.2753 × 10−2not significant
Pure Error3.35 × 10956.69 × 108
Cor Total1.19 × 101229
Table 4. Quadratic model’s confirmation data.
Table 4. Quadratic model’s confirmation data.
Response TypePredicted MeanPredicted MedianNumber of RunsSE Pred a95% PI LowData Mean95% PI High
Emission7.59987 × 1057.59987 × 10533.25448 × 1046.90620 × 1058.11434 × 1058.29355 × 105
a Standard deviation associated with the prediction of observations.
Table 5. Comparison table of different morphine-sensing technologies.
Table 5. Comparison table of different morphine-sensing technologies.
Detection MethodSensing MaterialDetection LimitLinear RangeAnalysis Duration a Complex SampleRef.
Gas chromatography–mass spectrometryNot available3 ppb0.0025–2 ppm<70 minUrine[138]
SPME b RP-HPLC c and LC-MS/MSCarboxylated carbon nanotubes1 ppb1–10 ppb and 0.001–1 ppm d30 minFerula gummosa e[139]
Piezoelectric biosensor Anti-Morphine antibody on the gold coated quartz0.25 ppb0.25–2500 ppb8 minUrine[140]
Magnetic resistance sensorySuperparamagnetic nanoparticles0.1 ppb0.5–1.5 ppb10 sNone[141]
Surface plasmon resonance imagingActivated carboxyl groups on the chips9.59 ppb1–50 ppm20 minUrine[142]
Colorimetric determinationMelamine modified gold nanoparticles4.85 ppb19.97–856 ppb10 minUrine and serum[143]
Colorimetric determinationAu@Ag core–shell nanoparticles55 ppb0.055–30 ppm5 minUrine[62]
Square wave voltammetryMesoporous carbon nanostructures7.7 ppb0.0285–114.136 ppm>2 minUrine[144]
Differential pulse voltammetryGold nanodendrites—broken hollow carbon spheres2.37 ppb0.0029–85.6 ppm<10 sHuman serum and saliva[145]
Differential pulse voltammetryHierarchical CoO4-carbon composite25.68 ppb0.228–21.4 ppm6 minUrine and serum[146]
Linear sweep voltammetryHighly boron-doped BCN (p-BCN)5.08 ppb0.014–57.07 ppm<10 s fHuman serum[26]
Up-conversion luminescent systemNitrocellulose membrane0.1 ppb0.1–10 ppb>30 sHuman hair[147]
Capillary zone electrophoresis with fluorescence detectionDisodium tetraborate decahydrate (BGE solution)0.5 ppb0.002–2 ppm>32 minUrine[148]
Ratiometric fluorescence sensorNitrogen-doped carbon dot71.8 ppb0.25–25 ppm<31 minHuman plasma[56]
Turn-on fluorescence detectionFluorescein—Gold nanoparticles 0.015 ppb0.0013–13.942 ppb6 minUrine and serum[149]
Fluorescence quenching (turn-off) system7′-Methoxy-[1,1′-binaphthalen]-7-ol8 ppb0.008–40 ppm<10 s gUrineThis work
a Including pretreatment time for complex biological samples. b Solid-phase microextraction. c Reversed-phase high-performance liquid chromatography. d Two different linear ranges. e Dried herb. f The electrolyte solution had its dissolved oxygen removed by introducing nitrogen gas into the buffer solution for a duration of 20 min prior to each experiment. g Sensor can generate an acceptable and measurable response in less than 10 s, but allowing for 7.4 min mixing time with urine will offer the most optimized response.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Boroujerdi, R.; Butt, A.; Paul, R.; Majumder, S. Unveiling Morphine: A Rapid and Selective Fluorescence Sensor for Forensic and Medical Analysis. Sensors 2024, 24, 1722. https://doi.org/10.3390/s24061722

AMA Style

Boroujerdi R, Butt A, Paul R, Majumder S. Unveiling Morphine: A Rapid and Selective Fluorescence Sensor for Forensic and Medical Analysis. Sensors. 2024; 24(6):1722. https://doi.org/10.3390/s24061722

Chicago/Turabian Style

Boroujerdi, Ramin, Andrew Butt, Richard Paul, and Santanu Majumder. 2024. "Unveiling Morphine: A Rapid and Selective Fluorescence Sensor for Forensic and Medical Analysis" Sensors 24, no. 6: 1722. https://doi.org/10.3390/s24061722

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop