Determination of the limit of quantification of the calorimeter using a mixture of benzoic acid and silicon dioxide
Krstić Vesna R.
Blagojević Branko D.
Gomidželović Lidija D.
Požega Emina D.
Petrović Jelena V.
Stanković Suzana D.
In recent years quality control has received a great attention in laboratory
work. Implementation of the international standard ISO/IEC 17025 is necessary
for any laboratory that wishes to establish quality control in its work. One
of the important factors for meeting the requirements of this standard is the
usage of the certified reference materials (CRM) in laboratory work. In order
to determine the performance of the calorimeter, benzoic acid as CRM, from
AlliedSignal Riedelda Haen, Ref.: 33045 and SiO2, Pro analyze, in various
mass ratios was used. The results showed that benzoic acid can be
successfully utilized for the control of the entire technical and
instrumental measuring range and resolve the problem of determination of the
limit of detection and quantification of the calorimeter.
Keywords: CRM, Benzoic acid, Limit of detection, Limit of quantification, C5000 Calorimeter System