Original Research |
1 Department of Radiodiagnostic, Baskent University Faculty of Medicine, Fevzi
Çakmak caddesi 10.sokak, No: 45, 06490 Bahçelievler, Ankara,
Turkey.
2 Department of General Surgery, Baskent University Faculty of Medicine, Ankara,
Turkey.
3 Department of Public Health, Baskent University Faculty of Medicine, Ankara,
Turkey.
OBJECTIVE. The objective of our study was to evaluate the ability of a respiratory navigator-triggered T2-weighted turbo spin-echo (TSE) sequence with a prospective acquisition correction (PACE) technique for MR cholangiopancreatography (MRCP) to depict the biliary anatomy of living donor liver transplantation (LDLT) donors.
SUBJECTS AND METHODS. Forty potential LDLT donors who ranged in age from 19 to 54 years were prospectively evaluated with preoperative MRCP. MRCP was performed with a 1.5-T magnetic field using T2-weighted PACE TSE sequence. MRCP source data sets were processed with maximum-intensity-projection (MIP) and shaded surface display (SSD) algorithms. Findings were compared with intraoperative cholangiography. Biliary anatomy was classified according to the classification proposed by Huang and colleagues. The sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of MRCP for the detection of aberrant biliary anatomy were calculated.
RESULTS. Intraoperative cholangiography and biliary exploration revealed that 27 donor candidates (67.5%) had conventional and 13 (32.5%) had aberrant biliary anatomy. Two donors (5%) had type B biliary anatomy; eight donors (20%), type C; two donors (5%), type D; and one donor (2.5%), unclassified. The sensitivity of MRCP source data sets in differentiating aberrant biliary anatomies from nonaberrant ones was 100%, the specificity was 88.9%, and the accuracy was 92.5%. PPV and NPV were 81.3% and 100%, respectively. The sensitivity of MIP images in differentiating aberrant biliary anatomies was 100%, the specificity was 88.9%, and the accuracy was 92.5%. PPV and NPV were 81.3% and 100%, respectively. The sensitivity, specificity, accuracy, PPV, and NPV of the SSD images in detecting aberrant biliary anatomies were 100%, 77.8%, 85%, 68.4%, and 100%, respectively.
CONCLUSION. Preoperative MRCP using a respiratory navigator-triggered T2-weighted TSE sequence with a PACE technique accurately depicts the biliary anatomy in LDLT donors and may guide intraoperative management of the biliary tract.
Keywords: anatomy biliary system liver disease liver transplantation living donor liver transplantation MR cholangiopancreatography
![]()