Performance of different abutment/implant joints as a result of a sealing agent
Camila Berbel SelotoHenrico Badaoui Strazzi-sahyonPaulo Henrique Dos SantosWirley Gonçalves Assunção
Author information
JOURNAL OPEN ACCESS

2021 Volume 65 Issue 4 Pages 489-494

Details
Abstract

Purpose This in vitro study aimed to evaluate the effectiveness of a sealing agent in sealing the abutment/implant interface and the preload maintenance of retaining screws after mechanical cycling.

Methods Six groups (n = 12) were evaluated according to the abutment/implant system (external-hexagon implant and UCLA abutments, EHU; Morse taper implant and UCLA abutments, MTU; and Morse taper implant and flexcone abutments, MTF) and the presence of an anaerobic gel sealing agent (control group, no sealing agent; experimental group, sealing agent). Toluidine blue (0.7 μL) was inserted into each implant and the abutments were attached to the implants using a digital torque wrench to evaluate the sealing of the abutment/implant interface. The specimens were tested through mechanical cycling (1 × 106 cycles, 2 Hz, and 130 N). Dye release from the abutment/implant interface was analyzed using a spectrophotometer, and the reverse torque values were obtained using a digital wrench. Reverse torque and dye release data were measured after mechanical cycling and analyzed using ANOVA and Tukey's test (α =.05).

Results All experimental groups showed higher reverse torque values than the control groups (P <.05). In general, the MTU and MTF experimental groups, as well as the MTF control group, showed no significant dye release at different periods (P >.05).

Conclusions The use of a sealing agent improved the preload maintenance of screw-retained implant-supported prostheses. The sealing agent was effective in sealing the Morse taper connection.

Content from these authors
© 2021 Japan Prosthodontic Society

This is an open-access article distributed under the terms of Creative Commons Attribution License 4.0 (CCBY 4.0), which allows users to distribute and copy the material in any format so long as attribution is given to the Japan Prosthodontic Society.
https://creativecommons.org/licenses/by/4.0/