分析化学
Print ISSN : 0525-1931
ノート
減圧ヘリウム誘導結合プラズマ質量分析法によるハロゲンの定量
林 英男原 泰久田中 智一平出 正孝
著者情報
ジャーナル フリー

2003 年 52 巻 4 号 p. 275-278

詳細
抄録

A low-pressure ICP formed with helium has been utilized as an ion source for MS, which is capable of providing information about traces of iron, arsenic and selenium without any interference from ArO+, ArCl+ and Ar2+, respectively. The present paper describes the application of a novel ICP for the determination of fluorine, chlorine, bromine and iodine, because the ionization of these halogens is highly suppressed in the case of conventional argon ICP (e.g., 9×10−4% for fluorine). A 5 μl aliquot of an aqueous sample was placed on a looped tungsten filament and heated electrothermally at ca. 2500°C . The resulting sample vapor was transported to the ICP with a stream of helium. The analytes (19F+, 35Cl+, 79Br+ and 127I+) were then determined by a quadrupole mass analyzer. The detection limits (ng ml−1, based on 3σ) were 23 (F), 2.4 (Cl), 0.13 (Br) and 0.05 (I), which were much better than those of argon ICP. The absolute detection limits using 5 μl of a sample reached to the sub-picogram range. Especially, the detectability for fluorine was significantly improved, which allowed a direct determination of fluorine at the ng ml−1 level in water.

著者関連情報
© The Japan Society for Analytical Chemistry 2003