揮発性有機化合物のガスクロマトグラフィー/質量分析法におけるキャニスター法の評価
有害大気汚染物質の分析に関する共同実験(第2報)
今村 清江口 正治大平 修平田代 恭久立花 茂雄田中 稔平井 恭三藤方 豊藤瀬 剛一前田 泰昭矢坂 裕太藁科 宗博
著者情報
ジャーナル フリー

1999 年 48 巻 7 号 p. 717-723

詳細
抄録

Round-robin tests were conducted in order to investigate a canister method for determining standard gas mixtures of volatile organic hazardous air pollutants by 14 members of the Association for Environmental Analysis in Osaka. A first standard gas containing seven analytes (acrylonitrile, dichloromethane, chloroform, 1, 2-dichloroethane, benzene, trichloroethylene, and tetrachloroethylene at ca. 300 ppm each) was prepared in a glass vessel from their mixed solution according to our previous paper. An aliquot of this gas was transferred into a canister (61) using a syringe, and diluted (ca. 10 ppb each). The resulting standard gas mixture was evaluated by GC/MS after concentration. The obtained results prove this method to be both accurate and precise for the preparation of a standard gas mixture at low ppb levels. Another gas mixture containing nine analytes (vinyl chloride and 1, 3-butadiene in addition to the above-mentioned seven compounds at 3 ppb each) was also prepared in a canister, and similarly treated by GC/MS. It was found that the canister method is remarkably accurate and precise. However, the acrylonitrile values had in large standard deviations.

著者関連情報
© The Japan Society for Analytical Chemistry