Development of a fluorimetric sequential injection analysis (SIA) methodology for determination of quinine

Abstracts

A sequential injection analysis (SIA) methodology was developed for fluorimetric determination of quinine in soft drinks. Minimum dispersion and efficient mixing were achieved by aspirating 200 mL of sample between two 100 mL reagent zones (0.10 mol L-1 H2SO4) at flow rate of 250 mL s-1 and using a reaction coil length of 50 cm (0.8 mm internal diameter). Linear response for quinine concentrations (C) between 0.050 and 100.0 mg L-1 was described by: I = (532 ± 40) + (2.36 ± 0.04) C, with r = 0.999, where I is the relative fluorescence intensity. The coefficient of variation and limits of detection and quantification were 1.9% (0.50 mg L-1 n = 10), 2.3 and 4.5 mg L-1, respectively. The sampling throughput was 60 analyses per hour, consuming 2.2 mL of concentrated H2SO4 and producing 4 mL of wastes per analysis. No statistically significant differences were observed between the results obtained by the proposed SIA methodology and the ones obtained by the batch procedure.

sequential injection analysis; quinine; fluorimetry; dispersion; optimization


Desenvolveu-se uma metodologia de análise por injeção sequencial (SIA) para a determinação fluorimétrica de quinino em refrigerantes. Dispersão mínima e mistura eficiente foram obtidas aspirando-se 200 mL de amostra entre duas zonas de 100 mL de reagente (0,10 mol L-1 H2SO4) a uma vazão de 250 mL s-1, usando uma bobina de reação de 50 cm de comprimento (0,8 mm de diâmetro interno). Resposta linear para concentrações (C) entre 0,050 e 100,0 mg L-1 foi descrita por: I = (532 ± 40) + (2,36 ± 0,04) C, com r = 0.999, onde I é intensidade relativa de fluorescência. Coeficiente de variação e limites de detecção e quantificação foram 1,9% (0,50 mg L-1, n = 10), 2,3 e 4,5 mg L-1, respectivamente. A frequência de amostragem foi 60 amostras por hora, consumindo 2,2 mL de H2SO4 concentrado e produzindo 4 mL de resíduos por análise. Não foram observadas diferenças significativas entre os resultados obtidos pela metodologia SIA proposta e aqueles obtidos pelo procedimento em batelada.


ARTICLE

Development of a fluorimetric sequential injection analysis (SIA) methodology for determination of quinine

Carlos M. C. Infante; Jorge C. Masini* * e-mail: jcmasini@iq.usp.br

Instituto de Química, Universidade de São Paulo, Av. Prof. Lineu Prestes 748, 05508-000 São Paulo-SP, Brazil

ABSTRACT

A sequential injection analysis (SIA) methodology was developed for fluorimetric determination of quinine in soft drinks. Minimum dispersion and efficient mixing were achieved by aspirating 200 mL of sample between two 100 mL reagent zones (0.10 mol L-1 H2SO4) at flow rate of 250 mL s-1 and using a reaction coil length of 50 cm (0.8 mm internal diameter). Linear response for quinine concentrations (C) between 0.050 and 100.0 mg L-1 was described by: I = (532 ± 40) + (2.36 ± 0.04) C, with r = 0.999, where I is the relative fluorescence intensity. The coefficient of variation and limits of detection and quantification were 1.9% (0.50 mg L-1 n = 10), 2.3 and 4.5 mg L-1, respectively. The sampling throughput was 60 analyses per hour, consuming 2.2 mL of concentrated H2SO4 and producing 4 mL of wastes per analysis. No statistically significant differences were observed between the results obtained by the proposed SIA methodology and the ones obtained by the batch procedure.

Keywords: sequential injection analysis, quinine, fluorimetry, dispersion, optimization

RESUMO

Desenvolveu-se uma metodologia de análise por injeção sequencial (SIA) para a determinação fluorimétrica de quinino em refrigerantes. Dispersão mínima e mistura eficiente foram obtidas aspirando-se 200 mL de amostra entre duas zonas de 100 mL de reagente (0,10 mol L-1 H2SO4) a uma vazão de 250 mL s-1, usando uma bobina de reação de 50 cm de comprimento (0,8 mm de diâmetro interno). Resposta linear para concentrações (C) entre 0,050 e 100,0 mg L-1 foi descrita por: I = (532 ± 40) + (2,36 ± 0,04) C, com r = 0.999, onde I é intensidade relativa de fluorescência. Coeficiente de variação e limites de detecção e quantificação foram 1,9% (0,50 mg L-1, n = 10), 2,3 e 4,5 mg L-1, respectivamente. A frequência de amostragem foi 60 amostras por hora, consumindo 2,2 mL de H2SO4 concentrado e produzindo 4 mL de resíduos por análise. Não foram observadas diferenças significativas entre os resultados obtidos pela metodologia SIA proposta e aqueles obtidos pelo procedimento em batelada.

Introduction

Sequential injection analysis (SIA) is a robust single channel flow technique widely used to automate analytical determinations and process control. It is based on the measurement of a transient signal resulting from sequential injection of defined sample and reagent volumes in a carrier solution under controlled dispersion.1 The sample processing conditions can be changed by software control as required by the analytical procedure. The possibility of exploiting feedback mechanisms to change sample processing conditions using a unique physical configuration make these approaches attractive to design flow procedures for process monitoring.2,3 The characteristics of SIA enable automation of diverse operations such as sampling, chemical derivatization, dilution, standard addition and others, improving the reproducibility. Low sample and reagent consumption and small waste generation4,5 are remarkable features of this technique, following the principles of green chemistry6 applied to development of clean analytical methods. A typical system is configured in single line by a syringe pump, coupled to a two-way valve that can communicate with the carrier or with a multiport valve through a holding coil where the reaction zone is formed prior to injection toward the detector, placed in another port of the multiport valve, immediately after the reaction coil.7 The optimization process can be laborious and take a long time, depending on the chosen route and variables considered in the process. The whole process may be simplified by an initial evaluation of sample dispersion and mixing efficiency using a dye solution in place of either sample or reagent.7,8 In SIA systems this approach is useful because the physical configuration does not change, and all the parameters can be used for development of other methods if the tubing dimensions are not altered. As a second step the chemical parameters are optimized for the specific determination, usually considering: concentration of reagents, flow rate, stopped-flow time and others.

The alkaloid quinine ((R)-(6-methoxyquinolin-4-yl)((2S,4S,8R)-5-vinylquinuclidin-2-yl)methanol trihydrate), originally extracted from the Peruvian quina tree (Cinchona), was firstly used as antimalarial.9 Diverse uses have been reported: in medicine as antiparasitic, antiprotozoal, anti-arrhythmic, antispasmodic; in cosmetics to decrease hair loss; in drinks as flavor drinks and bitter digestive aid.10 Alkaloids are toxic if consumed in large doses; consequently the concentration control is mandatory. The Food and Drug Administration of the United States (FDA)11 has established a limit content for quinine in soft drink of 83 mg L-1 and in Brazil the National Health Surveillance Agency (ANVISA)12 established the concentration of 30 to 50 mg L-1 of quinine. Diverse procedures are employed for quinine determination,13 but the intense fluorescence in acidic medium is a quinine property that is widely used for its quantification.14,15 Fluorimetry provides low detection limits and wide linear dynamic range, but its application is restricted because the analyte must be fluorescent or a fluorescent derivate must be produced.16 Complex matrixes can be analyzed by high performance liquid chromatography (HPLC)17 in normal phase18 or based on ion-pair formation,19 considering the ionic nature of quinine. But this makes difficult the determination because the equilibrium between mobile and stationary phases is slow and the separation is susceptible to variations of temperature, pH and concentration of organic compounds in the mobile-phase. For quality control, a quick method that saves sample and reagent, and produces low quantities of waste is highly desirable.

The aim of this work was the development of a fluorimetric sequential injection analysis (SIA) methodology for quinine determination, suggesting a general protocol to start the optimization of SIA systems and applying the procedure for determination of quinine in soft drinks of several trademarks.

Experimental

Apparatus

A FIAlab 3500 (FIAlab Instruments, Bellevue, WA) instrument was used in all experiments in the sequential injection mode according to Figure 1. Solutions were driven by a 5.00 mL syringe pump and an eight port rotary valve (RV, Valco Instrument Co., Houston, TX). The holding coil, HC, was made of 3 m of 0.8 mm i.d. polytetrafluoroethylene (PTFE, Teflon®, DuPont) tubing. The reaction coil (RC) was 50 cm long, made of 0.8 mm i.d. PTFE tubing. All other tubing connections were made of 0.5 mm i.d. PTFE tubing and PTFE nuts and ferrules (Upchurch, Oak Harbor, WA). The control of the pump and valve was made with the FIAlab 5.0 software. Fluorescence detection was performed with a PMT-FL (FIAlab Instruments) fluorimeter with excitation at 340 nm (the source of light was an inexpensive LED), measuring the emission at 450 nm (selected with an interference filter) using a photomultiplier as detector, coupled to a 3.1 mL illuminated flow cell. Acquisition data was performed with the FIAlab 5.0 software.


The dispersion was studied by molecular absorbance spectrophotometry at 620 nm with a dye solution composed of 0.010 mol L-1 bromothymol blue (BTB) in 0.010 mol L-1 borate buffer (pH 9.1), using an USB 4000 spectrometer (Ocean Optics, Dunedin, FL, USA) for measurements. A Tungsten-Halogen lamp LS-1 (Ocean Optics, Dunedin, FL, USA) was used as light source with a Hellma U-flow cell with 10 mm light pathlength and 80 mL internal volume. Two optical fibers (600 µm diameter and 20 inches long) were used to transmit radiation from the source to the flow cell and to the spectrometer.

Reagents and solutions

All solutions were prepared with analytical-grade chemicals and freshly distilled-deionized water. The reagent (R) was prepared by dissolution of concentrated sulfuric acid (ca. 18 mol L-1), making the volume up to 1.0 L with water. A 1000 mg L-1 quinine stock solution was prepared by dissolution of the sulfate mono-hydrated (Riedel-de-Haën, Seelze-Germany) in water. Working solutions of concentrations between 0.050 and 100.0 mg L-1 were prepared by dilution of the stock solution in water.

Soft drink samples were purchased from a local supermarket and directly analyzed by SIA without any preparation. For the fluorimetric batch methodology the samples were dissolved in 0.050 mol L-1 H2SO4 and properly diluted.

Flow diagram and procedure

The flow system (Figure 1) is initially filled with carrier solution. To start the automated analyses, the syringe valve is set at position IN and the syringe pump aspirates 4500 mL of carrier solution inside the syringe at a flow rate of 300 mL s-1. Next, the syringe valve switches to position OUT and, with the rotary valve at position 4, the syringe pump aspirates 500 mL of sample; the rotary valve switches to port 2 and the syringe pump dispenses 1000 mL, cleaning the sample line, preparing the system for the determination. In the next step, the syringe pump stacks sequentially 100 mL of the reagent R (port 3), 200 mL of standard or sample S (port 4), and 100 mL of reagent R (port 3) inside the holding coil at a flow rate of 250 mL s-1. In the final step, the rotary valve switches to port 1 and the syringe pump dispenses 2500 mL of carrier at 250 mL s-1, injecting the mixture zone through the mixing coil and the flow cell of the detector. Simultaneously, the acquisition data system is activated and the transient fluorescence analytical signal is obtained. Next, the flow rate is increased to 450 mL s-1 and the syringe is emptied, flushing the system for the next analysis. All measurements were carried out in triplicate.

Results and Discussion

Sample to reagent volume ratios and dispersion

Initially the influence of the sample and reagent volumes on the dispersion coefficient (D) was studied with the dye solution using a spectrophotometer. The sequence of aspiration to compose the sample zone was established in the "sandwich" format, with the sample between aliquots of reagent. Diverse volumetric proportions of sample and reagent were tested to define the best compromise between minimal dispersion and efficient mixture. The increase of sample volume (Figure 2A) reduces dramatically the dispersion, but over 200 mL no significant variations are observed, since the dispersion coefficient approaches the unity. The increase of reagent volume (Figure 2B) has significant influence on sample dispersion only for sample volumes < 100 mL.



Reagent-based chemistries automated by sequential injection analysis require the identification of the time interval in which sample and reagent zones are mutually interdispersed, a condition that is achieved for D > 2 if, at the same time, sufficient excess of reagent is present.7 Increase in sample volume enhances the sensitivity of the measurements only if excess of reagent is fed to the sample zone. Thus, for sample volumes as large as the ones leading to 1 < D < 2, sandwiching the sample between reagent zones is recommended. Additionally, under these conditions it is recommended to use reagents with concentration larger than that ones used in flow injection systems configured with confluence points. Figure 3 shows the gradient profile of reagent and sample zones for a method configured by sandwiching different sample volumes between 100 mL reagent zones. Complete interdispersion is achieved for sample volume of 25 mL (D > 6) as shown in Figure 3A, but at the cost of low sensitivity of the measurement as a consequence of the large degree of sample dilution. With a sample volume of 200 mL (Figure 3B), it is still possible to feed reagent to the sample zone using the sandwich approach, but one must be careful preparing a sufficiently concentrated reagent to warrant its excess in the central portions of the reaction zone. Increasing the sample volume further to 400 mL (Figure 3C) is no longer recommended because the reaction will occur only in the edges of the sample zone, which will cause peak distortions leading to inaccuracy in the results, besides to increased sample consumption and decreased the sampling throughput.




Influence of the chemical and hydrodynamic parameters

The formation of fluorescent species was carried out in acid media, a condition in which the quinine molecule is rigid and the not radioactive deactivation is less likely. The spectrum of quinine in this condition indicates the presence of two intense maximum of excitation at 280 and 350 nm, the latter wavelength being chosen because it is easily implemented with a LED. The maximal emission is observed in 450 nm, and was monitored with the aid of an interference filter. Formation of the fluorescent species with sulfuric acid is slow, but the product is highly stable and the quenching effect of chloride ions is avoided.

The evaluation of the experimental conditions was carried out by the univariate method. As expected, increase in the concentration of sulfuric acid (Figure 4A) enhanced the signal, without causing Schlieren effect,20 which is likely to occur in case of spectrophotometric detection. A sulfuric acid concentration of 100 mmol L-1 was elected, considering the little increase in the signal observed for higher concentrations. The influence of the sample volume (sandwiched between two 100 mL zones of 100 mmol L-1 H2SO4) is shown in Figure 4B. The analytical signal increased with the sample volume, but the volume of 200 mL was selected to avoid the problems discussed in the previous section, related to lack of sulfuric acid in the central portion of the sample volume. Besides, sample volumes larger than 200 mL imply in high sample consumption and significant decrease in the sampling throughput. The influence of the residence time on the analytical signal was studied by varying the flow rate in the range from 50 to 400 mL s-1 and the reaction coil length (30, 50 and 100 cm). These parameters did not have significant influence on the analytical signal, indicating the reaction is fast and is completed in the time scale of the experiments. Flow rate of 250 mL s-1 and reaction coil of 50 cm were selected to obtain a high sampling throughput, providing quick clean-up of the system after the determination.



Figures of merit

Linear response was observed for quinine concentrations between 0.05 and 100.0 mg L-1 as quinine sulfate, described by the equation: I = (532 ± 40) + (2.36 ± 0.04) C with r = 0.999. The detection limit was estimated at 2.3 µg L-1 at the 99.7% confidence level. Sampling rate and coefficient of variation (n = 10) were estimated at 60 measurements per hour and 1.9%, respectively. The proposed procedure consumed the equivalent to 2.2 mL of concentrated sulfuric acid, which was 322-fold lower than the consumption in the batch procedure.15 The effluent volume was 4.0 mL per determination, which was 64-fold lower than the consumption in the batch procedure.15 Comparison of the reagent consumption with other procedures is difficult in view of the use of other reagents13 and solvents in chromatographic separation.19 However the proposed procedure can be considered a green analytical method in view of the consumption of low amounts of low toxicity reagents.

Selectivity

The main interference in the fluorimetric determination of quinine is related with vitamin B18 and other alkaloids of similar structure.19 In the case of soft drinks, it is not expected the presence of any other alkaloid and the quinine concentration is high, so the possible interference of vitamin B is negligible. Other interesting aspect in the flow-based procedure in view of the low residence time (ca. 30 s) is the minimization in the extension of side reactions, improving the selectivity.21

Analysis of samples

The proposed procedure was applied for quinine determination in soft drinks of all the commercially available trademarks in Brazil. Quinine concentrations found in all samples were smaller than the limit value established by FDA11 (108 mg L-1 as quinine sulfate), but in two cases these concentrations exceeded the limit defined by ANVISA12 (65 mg L-1 as quinine sulfate). Table 1 shows that there is no evidence of statistically significant differences between the results obtained by the SIA procedure and those ones obtained by the fluorimetric batch reference procedure15 at the 95% confidence level, by considering a paired student's t-test (n = 8). Other profitable characteristics are the possibility of direct determination of quinine, without prior dilution, and reduction of the analyses time (ca. 1 min), in comparison with about 5 min required for batch procedures.

Conclusions

The sequential injection analysis methodology shows profitable characteristics for the determination of quinine in soft drinks. Optimization is simple and is started by searching the sample to reagent volumetric ratio that provides efficient mixture and low dispersion. Sensitivity was improved by using a fluorimetric determination yielding a linear response range compatible with diverse requirements (e.g. the established by the US-FDA11 or ANVISA).12 Sampling throughput was improved (60 samples per hour) and reagent consumption was at least 322-fold reduced in relation to previous works15 as a consequence of the use of intermittent flow. The proposed procedure can be considered as a clean alternative for fast quinine determination in soft drinks.

Acknowledgements

The authors acknowledge the fellowships and financial support from the Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq) and from the Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP, grant 2008/09284-3).

Submitted: March 1, 2011

Published online: July 26, 2011

FAPESP has sponsored the publication of this article.

  • 1. Ruzicka, J.; Marshall, G. D.; Anal. Chim. Acta 1990,237,329.
  • 2. Mesquita, R. B. R.; Rangel A. O. S. S.; Anal. Chim. Acta 2009,648,7.
  • 3. Mervartová, K.; Polásek, M.; Martiánez Calatayud, J.; J. Pharm. Biomed. Anal. 2007,45,367.
  • 4. Economou, A.; TrAC, Trends Anal. Chem. 2005,24,416.
  • 5. Cerdà, V.; Cerdà, A.; Cladera, A.; Oms, M. T.; Mas, F.; Goámez, E.; Bauzaá, F.; Miroá, M.; Forteza, R.; Estela, J. M.; TrAC, Trends Anal. Chem. 2001,20,407.
  • 6. Anastas, P. T.; Kirchoff, M. M.; Acc. Chem. Res. 2002,35,686.
  • 7. Gubeli, T.; Christian, G. D.; Ruzicka, J.; Anal. Chem. 1991,63,2407.
  • 8. Masini, J. C.; Quim. Nova 2008,31,704.
  • 9. Samuelsson, G.; Drugs of Natural Origin,5th ed.; Swedish Pharmaceutical Press: Stockholm, 1992.
  • 10. Papariello, G. J.; Tishler, F. In Encyclopedia of Industrial Chemical Analysis; Snell, F. D.; Hilton, C. L., eds.; Interscience Publishers: New York, 1967, chapter 4.
  • 11
    http://www.fda.gov/Food/FoodIngredientsPackaging/FoodAdditives/ucm191033.htm accessed in January 2011.
    » link
  • 12
    http://www.anvisa.gov.br/legis/decretos/2314_97.htm accessed in January 2011.
    » link
  • 13. Ortega-Algar, S.; Ramos-Martos, N.; Molina-Díaz, A.; Microchim. Acta 2004,147,211.
  • 14. Guilbault, G. G.; Practical Fluorescence: Theory, Methods and Techniques,1st ed.; Marcel Decker: New York, 1973.
  • 15. O'Reilly, J. E.; J. Chem. Educ. 1975,52,610.
  • 16. Ruzicka, J.; Hansen, E. H.; Flow Injection Analysis, 2nd ed.; Wiley Interscience: New York, 1988.
  • 17. Gatti, R.; Gioia, M. G.; Cavrini, V.; Anal. Chim. Acta 2004,512,85.
  • 18. McCalley, D. V.; J. Chromatogr., A 1983,260,184.
  • 19. Chen, Q. C.; Wang, J.; J. Liq. Chromatogr. Relat. Technol. 2001,24,1341.
  • 20. Zagatto, E. A. G.; Arruda, M. A. Z.; Jacintho, A. O.; Mattos, I. L.; Anal. Chim. Acta 1990,234,153.
  • 21. Infante, C. M. C.; Soares, V. R. B.; Korn, M.; Rocha, F. R. P.; Microchim. Acta 2008,161,279.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      25 Oct 2011
    • Date of issue
      Oct 2011

    History

    • Received
      01 Mar 2011
    • Accepted
      26 July 2011
    Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
    E-mail: office@jbcs.sbq.org.br