IEEJ Transactions on Sensors and Micromachines
Online ISSN : 1347-5525
Print ISSN : 1341-8939
ISSN-L : 1341-8939
Special Issue Letter
Detection of Heavy-metal Ions Based on Evaporative Concentration Using a Super-hydrophobic Surface
Isao YanagimachiNorihiro NashidaKoichiro IwasaHiroaki Suzuki
Author information
JOURNAL FREE ACCESS

2006 Volume 126 Issue 7 Pages 356-357

Details
Abstract

A concentrator chip which could detect a variety of heavy-metal ions was fabricated. To improve the detection sensitivity, a droplet of a sample solution was concentrated evaporatively using a super-hydrophobic surface formed with polytetrafluoroethylene (PTFE) beads. The system consists of a working electrode at the center, surrounded by an Ag/AgCl reference electrode. Square-wave anodic stripping voltammetry was conducted using concentrator chips with different working electrode materials. A significant increase in peak height was observed as the sensitive area decreased and the volume of the droplet increased. When a 5-μl droplet was used, the detection limit for lead, cadmium, and arsenic ions was 1 ppb.