Comparison of visual vs. automated detection of lipemic, icteric and hemolyzed specimens: can we rely on a human eye?

  • Ana-Maria Simundic , Nora Nikolac , Valentina Ivankovic , Dragica Ferenec-Ruzic , Bojana Magdic , Marina Kvaternik and Elizabeta Topic

Abstract

Background: Results from hemolyzed, icteric, and lipemic samples may be inaccurate and can lead to medical errors. These preanalytical interferences may be detected using visual or automated assessment. Visual inspection is time consuming, highly subjective and not standardized. Our aim was to assess the comparability of automated spectrophotometric detection and visual inspection of lipemic, icteric and hemolyzed samples.

Methods: This study was performed on 1727 routine biochemistry serum samples. Automated detection was performed using the Olympus AU2700 analyzer. We assessed: 1) comparability of visual and automated detection of lipemic, icteric and hemolyzed samples, 2) precision of automated detection, and 3) inter-observer variability for visual inspection.

Results: Weighted κ coefficients for comparability of visual and automated detection were: 0.555, 0.529 and 0.638, for lipemic, icteric and hemolyzed samples, respectively. The precision for automated detection was high for all interferences, with the exception of samples being only slightly lipemic. The best overall agreement between observers was present in assessing lipemia (mean weighted κ=0.698), whereas the lowest degree of agreement was observed in assessing icterus (mean weighted κ=0.476).

Conclusions: Visual inspection of lipemic, icteric and hemolyzed samples is highly unreliable and should be replaced by automated systems that report serum indices.

Clin Chem Lab Med 2009;47:1361–5.


Corresponding author: Assist. Prof. Ana-Maria Simundic, PhD, University Department of Chemistry, Department of Molecular Diagnostics, University Hospital “Sestre Milosrdnice”, Zagreb, Croatia

Received: 2009-6-30
Accepted: 2009-7-17
Published Online: 2009-09-25
Published in Print: 2009-11-01

©2009 by Walter de Gruyter Berlin New York

Downloaded on 29.3.2024 from https://www.degruyter.com/document/doi/10.1515/CCLM.2009.306/html