Repeatability of Pentacam-derived intraocular lens decentration measurements and the level of agreement with OPD-Scan III: A prospective observational case series

  • Xiaobao Liu ,

    Contributed equally to this work with: Xiaobao Liu, Wenjie Wu

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Project administration, Validation, Visualization, Writing – original draft

    Affiliation Shengli Clinical Medical College of Fujian Medical University; Fujian Medical University, Fuzhou, China

  • Wenjie Wu ,

    Contributed equally to this work with: Xiaobao Liu, Wenjie Wu

    Roles Conceptualization, Data curation, Funding acquisition, Investigation, Methodology, Project administration, Supervision, Validation, Writing – review & editing

    Affiliations Shengli Clinical Medical College of Fujian Medical University; Fujian Medical University, Fuzhou, China, Department of Ophthalmology, Fujian Provincial Hospital, Fuzhou, China

  • Yulong Huang,

    Roles Data curation, Formal analysis, Investigation, Methodology, Software

    Affiliation Shengli Clinical Medical College of Fujian Medical University; Fujian Medical University, Fuzhou, China

  • Yabo Fu,

    Roles Data curation, Formal analysis, Methodology, Software

    Affiliations Shengli Clinical Medical College of Fujian Medical University; Fujian Medical University, Fuzhou, China, Department of Ophthalmology, Longyan People Hospital of Fujian, Longyan, China

  • Yue Huang,

    Roles Data curation, Formal analysis, Software, Validation

    Affiliation Shengli Clinical Medical College of Fujian Medical University; Fujian Medical University, Fuzhou, China

  • Qiong Li

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Validation, Visualization, Writing – review & editing

    liqiong2000@hotmail.com

    Affiliations Shengli Clinical Medical College of Fujian Medical University; Fujian Medical University, Fuzhou, China, Department of Ophthalmology, Fujian Provincial Hospital, Fuzhou, China

Abstract

Purpose

This study aimed to assess the repeatability of intraocular lens (IOL) decentration measurements obtained through Pentacam, based on corneal topographic axis (CTA) and pupillary axis (PA), and to evaluate the level of agreement between Pentacam and OPD-Scan III devices in measuring IOL decentration.

Methods

In this prospective observational case series, three measurements were performed with Pentacam to evaluate the repeatability of the measurements. The analysis included the calculation of the mean and standard deviations (SD), conducting a repeated measures analysis of variance (rANOVA), and determining an intraclass correlation coefficient (ICC) to assess the repeatability of the measurements. Moreover, Bland-Altman analysis was employed to assess the agreement between Pentacam and OPD-Scan III devices in measuring IOL decentration. IOL decentration measurements were obtained with respect to both CTA and PA.

Results

A total of 40 eyes from 40 patients were analyzed. The rANOVA revealed no significant difference among three consecutive measurements of IOL decentration obtained with Pentacam. The mean SD of all parameters ranged from 0.04 mm to 0.07 mm. With CTA as the reference axis, the ICC values for Pentacam measurements of IOL decentration were 0.82 mm for the X-axis, 0.76 mm for the Y-axis, and 0.82 mm for spatial distance. When using PA as the reference axis, the corresponding ICC values were 0.87, 0.89, and 0.77, respectively. The 95% limits of agreement for all IOL decentration measurements were wide when comparing Pentacam and OPD-Scan III.

Conclusions

Pentacam demonstrated high repeatability in measuring IOL decentration with respect to both CTA and PA. However, due to poor agreement between Pentacam and OPD-Scan III measurements, caution should be exercised when using data interchangeably between the two devices.

Introduction

The advancements in intraocular lens (IOL) design and prolonged lifespan of pseudophakic patients have led to greater expectations for positive visual and refractive outcomes following cataract surgery [13]. However, the decentration of IOLs can considerably impair postoperative optical quality and overall quality of life, particularly in cases involving intricate IOLs implantation, such as multifocal IOLs and toric IOLs. [46]. Therefore, precise measurement of IOL decentration is essential for estimation of postoperative visual performance after refractive cataract surgery.

Several techniques are currently employed for the measurement of IOL decentration [4, 5, 79]. One of these is the Scheimpflug method, which has garnered considerable prominence, featuring in nearly 30% of published studies focused on IOL decentration measurements [4, 1012]. The Pentacam, a non-invasive device based on the Scheimpflug principle, can produce precise, high-resolution images that enable an accurate assessment of the IOL’s position. A newer device that has gained widespread use in recent years is the OPD-Scan III, which can detect IOL decentration and provide supplementary optical quality data in a single scan. This feature streamlines the measurement process and improves the reliability and accessibility of subsequent optical quality analysis [5, 7, 1315]. Accordingly, the OPD-Scan III has become increasingly popular in the realm of IOL decentration measurements and optical quality analysis.

Historically, studies investigating the measurement of IOL decentration have predominantly relied on the pupillary axis (PA) as the reference axis, primarily due to its clinical ease of visualization [10, 1618]. Regrettably, the inherent variability of pupil size and location undermines the precision and consistency attainable with this reference axis [19, 20]. Consequently, the corneal topographic axis (CTA), positioned close to the subject-fixated coaxially sighted corneal light reflex axis, has emerged as a proposed universal standard axis for decentration measurement [17, 21]. However, to date, no studies have elucidated the repeatability of IOL decentration measurements specifically focused on the CTA utilizing Pentacam. Furthermore, the interchangeability of Pentacam-based IOL measurements referenced to both CTA and the PA with those obtained from OPD-Scan III remains unknown.

Therefore, the primary objective of the present study was to assess the repeatability of IOL decentration measurements derived from the Pentacam, focusing on both the PA and the CTA. Additionally, a secondary aim involved evaluating the level of agreement among the Pentacam and the OPD-Scan III with regards to IOL decentration measurements.

Materials and methods

A STROBE flowchart is shown in Fig 1.

Participants

This was a prospective observational case series, in which 43 eyes of 43 patients with age-related cataract were recruited from the Department of Ophthalmology, Fujian Provincial Hospital (Fujian, China), between October 1st, 2021 and January 1st, 2022. Patients with a history of trauma, uveitis, glaucoma, pseudoexfoliation syndrome, strabismus, dilated pupil diameter of less than 5.0 mm, axial length of less than 21 mm or more than 25 mm, keratoconus or other corneal disease, or previous ocular surgery or laser treatment were excluded. The study protocol was approved by the Institutional Review Board of Fujian Provincial Hospital on April 29th, 2020 (K2020-04-048), and all procedures were conducted in accordance with the principles outlined in the Declaration of Helsinki. Written informed consent was obtained from each patient prior to his or her participation.

All subjects underwent routine phacoemulsification and IOL implantation, with surgical parameters detailed in Table 1. At the 3-month follow-up visit, a comprehensive ophthalmic examination was conducted for each patient, including assessments of uncorrected visual acuity (UCVA), best corrected visual acuity (BCVA), slit-lamp microscopy, noncontact tonometry, fundus examination, and measurement of IOL decentration using both Pentacam AXL (Oculus, Wetzlar, Germany) and OPD-Scan III (Nidek Co., Ltd., Gamagori, Japan) devices.

Surgical technique

All cataract surgeries were conducted by a single surgeon (WJW) according to a standard procedure. A 2.4-mm temporal clear corneal incision and a side port incision were created. Subsequently, nuclear removal and cortical aspiration were performed within a precisely centered 5.5-mm anterior continuous curvilinear capsulorrhexis. Following the cataract removal and capsular polishing, IOL implantation was carried out. The IOL was carefully rotated to achieve a vertical position. Lastly, any remaining ophthalmic viscosurgical device was aspirated, and the incisions were hydrated.

IOL decentration measurement

The measurement of IOL decentration was conducted using the Pentacam and OPD-Scan III devices following complete mydriasis achieved with a mixture of 1% tropicamide and 2.5% phenylephrine hydrochloride. Before the study, both devices were calibrated by their respective technical support services. The order of the measurements for the patients was determined using a computerized table of random numbers. All measurements were performed in a standard dim-light room. Two optometrists, who were blinded to the study protocol, were involved in the measurements: one for Pentacam imaging and the other for OPD-Scan III acquisitions.

To measure the IOL decentration using Pentacam, horizontal (0–180°) and vertical (90–270°) cross-sections of Scheimpflug images were captured three times consecutively. Before each measurement, careful attention was paid to head positioning and eye alignment. Patients were instructed to blink immediately before the measurements and were asked to briefly leave the headrest and chinrest between two consecutive measurements. An image was acquired only when the data quality statement indicated "OK" and the edge of the IOL was clearly visible. Subsequently, Image J software (Image J version 1.52, NIH, USA) was utilized for image processing.

The identification of the CTA and PA followed the method described by previous studies [10, 16]. Two arcs were adjusted to fit the anterior and posterior surfaces of the IOL, and the IOL center was determined as the midpoint of the IOL plane passing through the two intersections of the arcs. The position of the CTA was obtained by connecting the fixation point to the corneal vertex. The PA was defined as the line perpendicular to the cornea and passing through the midpoint of the two visible segments of the pupil.

Given the high agreement between Pentacam and OPD-Scan III in measuring corneal curvature, the corneal vertex was chosen as the origin of the coordinate system to determine the positions of the IOL center and pupil center in both devices [22]. Using a scale of 0.02 mm/pixel, the locations of the PC and IOL center were represented as (Xp, Yp) and (Xi, Yi), respectively (Fig 2). The parameters of IOL decentration with reference to CTA and PA were calculated as shown in Table 2. Positive horizontal coordinates represented nasal decentrations in the right eye and temporal decentrations in the left eye, while positive vertical coordinates indicated superior decentrations and negative values indicated inferior decentrations.

thumbnail
Fig 2. Definition for all parameters in Sheimpflug image using Pentacam.

CTA, corneal topographic axis; IOL, intraocular lens; PC, pupil center. The blue line represents CTA. The orange line represents pupillary plane. The red line represents IOL plane.

https://doi.org/10.1371/journal.pone.0299064.g002

thumbnail
Table 2. Parameters definition and equation of IOL decentration measurement with Pentacam.

https://doi.org/10.1371/journal.pone.0299064.t002

For OPD-Scan III, the locations of the IOL center and pupil center relative to CTA were directly obtained from the instrument as distances (di’ and dp’, respectively) and orientations (αi and αp, respectively) (S1 Fig). The parameters were calculated according to Table 3.

thumbnail
Table 3. Parameters definition and equation of IOL decentration measurement with OPD-Scan III.

https://doi.org/10.1371/journal.pone.0299064.t003

Sample size

The determination of the sample size was carried out with a significance level (α) of 0.01 and a power of 90%. Drawing from prior studies and our own research data, the average IOL decentration level observed was approximately 0.3 mm, which was within the acceptable range [4, 5, 7]. Based on previous research, clinically significant IOL decentration has been defined as decentration exceeding 0.4 mm [23, 24]. As a result, a clinically relevant mean difference of 0.1 mm was considered for the calculation. Based on the results obtained from a pilot study, which indicated a mean measurement error of 0.15 mm for IOL decentration measurement, the minimum required sample size was determined to be 34.

Statistical analysis

Statistical analyses were conducted using SPSS for Windows Version 15.0 software (SPSS Inc., Chicago, IL, USA). All measurements were reported as mean ± standard deviation (SD). The statistical significance of differences between the three consecutive measurements obtained with the Pentacam was assessed using repeated-measures analysis of variance (rANOVA). A p-value less than 0.05 was considered statistically significant. Furthermore, the mean SD and intraclass correlation coefficients (ICCs) were calculated to evaluate the repeatability of the Pentacam measurements. A higher ICC value indicates better measurement consistency.

To assess the agreement between the measurements obtained with the Pentacam and OPD-Scan III, Bland-Altman plots were employed. The 95% limits of agreement (LoA) were determined as the mean ± 1.95 times the SD of the measurement difference between the two devices. In the Bland-Altman plots, the y-axis represented the difference between each pair of measurements, while the x-axis depicted the mean values. The central line on the plot represented the mean of the measurement differences between the two devices, and the broken lines indicated the 95% LoA.

Results

In this study, three eyes were excluded due to a failure to achieve pupil dilation in one patient and poor image quality in two patients. Therefore, a total of 40 eyes from 40 patients were included in the analysis. The demographic data and surgical parameters of the patients are presented in Table 1.

At the 3-month postoperative follow-up, the mean IOL decentration measurements, with reference to the CTA using the Pentacam, were as follows: -0.17 ± 0.12 mm in the x-axis, 0.07 ± 0.11 mm in the y-axis, and 0.24 ± 0.10 mm in spatial distance. When using the PC as the reference, the mean IOL decentration measurements were -0.20 ± 0.14 mm, 0.09 ± 0.18 mm, and 0.30 ± 0.13 mm, respectively. A rANOVA indicated no statistically significant differences among the three consecutive measurements for all parameters obtained with the Pentacam. The measurements of DCX, DCY, DC, DPX, DPY, and DP showed high repeatability, as indicated by ICCs ranging from 0.76 to 0.89, and mean SD ranging from 0.04 to 0.06 mm. For the OPD-Scan III, the IOL decentration measurements, with reference to the CTA, were -0.20 ± 0.22 mm in the x-axis, 0.28 ± 0.17 mm in the y-axis, and 0.25 ± 0.14 mm in spatial distance. When using the PC as the reference, the measurements were -0.32 ± 0.40 mm, 0.20 ± 0.35 mm, and 0.45 ± 0.27 mm, respectively (Table 4).

thumbnail
Table 4. Repeatability of IOL decentration measured by Pentacam and IOL decentration outcomes of OPD-Scan III.

https://doi.org/10.1371/journal.pone.0299064.t004

The agreement of IOL decentration measurements between the Pentacam and OPD-Scan III was analyzed using Bland-Altman analysis (Fig 3). The 95% LoAs for all parameters were wide when comparing the three measurements, indicating poor agreement of IOL decentration between the two devices. Specifically, in the measurements of DCX, 20.0% of the measurement differences between the two devices exceeded 0.3 mm; in the measurements of DCY, 32.5% exceeded 0.3 mm; in the measurements of DPX, 40% exceeded 0.3 mm; in the measurements of DPY, 50% exceeded 0.3 mm; in the measurements of DC, 5% exceeded 0.3 mm; and in the measurements of DP, 7.5% exceeded 0.3 mm. It should be noted that the measurement differences of IOL decentration based on CTA were smaller than those based on the PA.

thumbnail
Fig 3. Bland-Altman plots for intraocular lens (IOL) decentration measurements for Pentacam and OPD-Scan III.

The solid line indicates the mean difference (bias). The upper and lower lines represent the 95% limits of agreement. DCX, DCY and DC represent IOL decentration measured by Pentacam based on corneal topographic axis (CTA) in the x-axis, y-axis and spatial distance respectively. DPX, DPY and DP represent IOL decentration measured by Pentacam based on pupillary axis (PA) in the x-axis, y-axis and spatial distance respectively. DCX’, DCY’ and DC’ represent IOL decentration measured by OPD-Scan III based on CTA in the x-axis, y-axis and spatial distance respectively. DPX’, DPY’ and DP’ represent IOL decentration measured by OPD-Scan III based on PA in the x-axis, y-axis and spatial distance respectively.

https://doi.org/10.1371/journal.pone.0299064.g003

Discussion

The evolution of cataract surgery into a refractive procedure has underscored the growing significance of accurate IOL decentration measurements for assessing optical outcomes [1, 4]. Decentration of the IOL can impair optical outcomes, such as higher-order aberrations (mainly coma) and visual quality metrics (e.g., Modulation Transfer Function and Point Spread Function) [47]. Therefore, accurate measurement is essential to avoid overestimating or underestimating the impact of IOL decentration on optical outcomes, especially for sophisticated optical IOLs [46]. Previous investigations have explored various axes, including the visual axis (VA), CTA and PA, as potential reference centers for such measurements [10, 17]. However, it is important to note that the visual axis represents a theoretical mathematical construct that lacks corresponding anatomical landmarks, thereby posing challenges in achieving clinical correlation with real eyes [25, 26]. Conversely, the CTA and PA axes, which possess well-defined anatomical characteristics and can be detected by commonly employed devices, including the Pentacam and the OPD-Scan III, have emerged as more frequently utilized parameters in clinical measurement practice [7, 10, 12, 13, 18]. Notably, to the best of our knowledge, no prior studies have specifically examined the repeatability of Pentacam-based IOL decentration measurements using CTA as a reference, nor the agreement between the Pentacam and the OPD-Scan III regarding IOL decentration measurements. Consequently, the primary objective of this study was to evaluate the repeatability of IOL decentration measurements obtained with the Pentacam. Additionally, we aimed to ascertain the level of agreement between the Pentacam and the OPD-Scan III devices in measuring IOL decentration, utilizing both CTA and PA as reference axes.

Our findings revealed that IOL decentration measurements centered on the CTA exhibited commendable repeatability, a notion supported by a previous study conducted by Fan Zhang et al., who proposed that the CTA may serve as the most visually pertinent reference axis [10]. Notably, when employing the Pentacam, a strong correlation was observed between visual acuity and CTA-based IOL decentration, contrasting with the use of PA as a reference [10]. Additionally, Arbelaez’s study reported improved postoperative visual acuity outcomes when aligning the ablation center with the CTA, further emphasizing the significance of clinical performance of CTA [27]. Coupled with the robust repeatability of IOL decentration measurements observed in our study, the CTA axis emerges as an optimal choice for IOL decentration assessment.

In parallel, favorable repeatability was also achieved when utilizing PA as a reference axis, a finding consistent with observations made by Castro et al. [16]. However, their study was limited in reliability as it relied solely on the pupil axis to evaluate IOL decentration, without considering the CTA. The primary drawback of employing PA as a reference lies in the variability of pupil size and center, influenced by factors such as light intensity, accommodation, pharmacological interventions, trauma, and surgical interventions, including mydriasis [17, 19, 20]. Notably, the strong repeatability observed in our study may largely be attributed to the standardized light levels and consistent mydriasis conditions maintained throughout the trial. Moreover, the measured PA center under experimental conditions deviates from its position in daily life, suggesting that IOL decentration centered on the PA may exhibit limited relevance to postoperative optical quality in patients’ everyday activities. Thus, PA may be considered as an adjunctive reference axis for the measurement of IOL decentration, particularly in situations characterized by aberrant corneal morphology, such as corneal scarring, ectasia, or previous corneal surgeries.

In our study, we observed a relatively wide 95% limit of agreement (LoA) between the Pentacam and OPD-Scan III. Notably, measurements based on the PA exhibited lower concordance than those based on the CTA for both devices. This discrepancy may be attributed to differences in pupil conditions resulting from variations in lighting levels between the two devices. Furthermore, the Pentacam displayed a tendency towards superotemporal IOL decentration and greater centralization. It is important to acknowledge that while the OPD-Scan III utilizes a 2-dimensional view for its measurements, the Pentacam employs a 3-dimensional approach, incorporating corrections for optical and geometric distortions during data processing [10]. We hypothesize that the corneal magnification evident in retroillumination images captured by the OPD-Scan III may significantly contribute to the observed differences. However, further research is necessary to validate this hypothesis.

In recent years, novel devices have been developed to assess postoperative IOL decentration [8, 2830]. For instance, Ding et al. evaluated the repeatability of swept-source optical coherence tomography (SS-OCT) for IOL decentration and reported high ICCs when employing PA as the reference axis [28]. However, the limited availability of SS-OCT, particularly in regions such as China, hampers its widespread popularity as a tool for IOL decentration assessment.

Several limitations warrant mention. First, this study only compared two devices, and the feasibility of using alternative instruments interchangeably remains uncertain. Second, the absence of a gold standard comparison impedes a precise assessment of the accuracy of the two methods employed. Validation using a model eye might be advantageous in this regard. Thirdly, our study involved various types of IOLs. However, the four types of IOLs that we used were all 1-piece acrylic 360-degree square edge IOLs, which may have a negligible influence on IOL decentration measurements. Lastly, as the definition of IOL tilt differed between the two devices (physical tilt in the Pentacam and aberration parameters in the OPD-Scan III), the assessment of IOL tilt was not included in this study [15].

In conclusion, our study demonstrated the high repeatability of IOL decentration measurements obtained using the Pentacam, with both the CTA and PA serving as reliable reference axes. However, the poor agreement observed between the Pentacam and OPD-Scan III suggests that the interchangeable use of these two devices for IOL decentration measurement is not recommended.

Supporting information

S1 Fig. Definition for all parameters in retroillumination image using OPD-Scan III.

The red box represents the distance and the orientation of intraocular lens center with respect to corneal topographic axis (CTA); The orange box represents pupil center with respect to CTA.

https://doi.org/10.1371/journal.pone.0299064.s001

(DOCX)

References

  1. 1. Maurino V, Allan BD, Rubin GS, Bunce C, Xing W, Findl O. Quality of vision after bilateral multifocal intraocular lens implantation: a randomized trial—AT LISA 809M versus AcrySof ReSTOR SN6AD1. Ophthalmology. 2015;122(4):700–710. pmid:25537197
  2. 2. Kretz FTA, Bastelica A, Carreras H, et al. Clinical outcomes and surgeon assessment after implantation of a new diffractive multifocal toric intraocular lens. The British Journal of Ophthalmology. 2015;99(3):405–411. pmid:25249611
  3. 3. de Vries NE, Nuijts RMMA. Multifocal intraocular lenses in cataract surgery: literature review of benefits and side effects. Journal of Cataract and Refractive Surgery. 2013;39(2):268–278. pmid:23332253
  4. 4. Lawu T, Mukai K, Matsushima H, Senoo T. Effects of decentration and tilt on the optical performance of 6 aspheric intraocular lens designs in a model eye. J Cataract Refract Surg. 2019;45(5):662–668. pmid:30876781
  5. 5. Xu J, Zheng T, Lu Y. Effect of Decentration on the Optical Quality of Monofocal, Extended Depth of Focus, and Bifocal Intraocular Lenses. J Refract Surg. 2019;35(8):484–492. pmid:31393986
  6. 6. Eppig T, Scholz K, Löffler A, Messner A, Langenbucher A. Effect of decentration and tilt on the image quality of aspheric intraocular lens designs in a model eye. Journal of Cataract and Refractive Surgery. 2009;35(6):1091–1100. pmid:19465297
  7. 7. Liu X, Yu M, Huang Y, Li Q, Wu W. Intraocular lens tilt and decentration after cataract surgery with and without primary posterior continuous curvilinear capsulorhexis. Journal of Cataract and Refractive Surgery. 2023;49(5):492–498. pmid:36745843
  8. 8. Ding X, Wang Q, Xiang L, Chang P, Huang S, Zhao YE. Three-Dimensional Assessments of Intraocular Lens Stability With High-Speed Swept-Source Optical Coherence Tomography. J Refract Surg. 2020;36(6):388–394. pmid:32521026
  9. 9. Nishi Y, Hirnschall N, Crnej A, et al. Reproducibility of intraocular lens decentration and tilt measurement using a clinical Purkinje meter. Journal of Cataract and Refractive Surgery. 2010;36(9):1529–1535. pmid:20692566
  10. 10. Zhang F, Zhang J, Li W, et al. Correlative Comparison of Three Ocular Axes to Tilt and Decentration of Intraocular Lens and Their Effects on Visual Acuity. Ophthalmic Research. 2020;63(2):165–173. pmid:31846979
  11. 11. Wang D, Yu X, Li Z, et al. The Effect of Anterior Capsule Polishing on Capsular Contraction and Lens Stability in Cataract Patients with High Myopia. J Ophthalmol. 2018;2018:8676451. pmid:30425856
  12. 12. Kim KH, Kim WS. Intraocular lens stability and refractive outcomes after cataract surgery using primary posterior continuous curvilinear capsulorrhexis. Ophthalmology. 2010;117(12):2278–2286. pmid:20561685
  13. 13. Meng J, He W, Rong X, Miao A, Lu Y, Zhu X. Decentration and tilt of plate-haptic multifocal intraocular lenses in myopic eyes. Eye and Vision (London, England). 2020;7:17. pmid:32280721
  14. 14. Zhu X, He W, Zhang Y, Chen M, Du Y, Lu Y. Inferior Decentration of Multifocal Intraocular Lenses in Myopic Eyes. American Journal of Ophthalmology. 2018;188:1–8. pmid:29355482
  15. 15. He W, Qiu X, Zhang S, et al. Comparison of long-term decentration and tilt in two types of multifocal intraocular lenses with OPD-Scan III aberrometer. Eye (London, England). 2018;32(7):1237–1243. pmid:29520045
  16. 16. de Castro A, Rosales P, Marcos S. Tilt and decentration of intraocular lenses in vivo from Purkinje and Scheimpflug imaging. Validation study. J Cataract Refract Surg. 2007;33(3):418–429. pmid:17321392
  17. 17. Chang DH, Waring GO. The subject-fixated coaxially sighted corneal light reflex: a clinical marker for centration of refractive treatments and devices. American Journal of Ophthalmology. 2014;158(5):863–874. pmid:25127696
  18. 18. Kimura S, Morizane Y, Shiode Y, et al. Assessment of tilt and decentration of crystalline lens and intraocular lens relative to the corneal topographic axis using anterior segment optical coherence tomography. PloS One. 2017;12(9):e0184066. pmid:28863141
  19. 19. Porter J, Yoon G, Lozano D, et al. Aberrations induced in wavefront-guided laser refractive surgery due to shifts between natural and dilated pupil center locations. Journal of Cataract and Refractive Surgery. 2006;32(1):21–32. pmid:16516775
  20. 20. De Ortueta D, Schreyger F. Measurement of the spatial shift of the pupil center. Journal of Cataract and Refractive Surgery. 2006;32(6):906–907. pmid:16814035
  21. 21. Holladay JT, Calogero D, Hilmantel G, et al. Special Report: American Academy of Ophthalmology Task Force Summary Statement for Measurement of Tilt, Decentration, and Chord Length. Ophthalmology. 2017;124(1):144–146. pmid:27742457
  22. 22. Asgari S, Hashemi H, Jafarzadehpur E, Mohamadi A, Rezvan F, Fotouhi A. OPD-Scan III: a repeatability and inter-device agreement study of a multifunctional device in emmetropia, ametropia, and keratoconus. Int Ophthalmol. 2016;36(5):697–705. pmid:26854319
  23. 23. McKelvie J, McArdle B, McGhee C. The influence of tilt, decentration, and pupil size on the higher-order aberration profile of aspheric intraocular lenses. Ophthalmology. 2011, 118(9): 1724–1731. pmid:21665282
  24. 24. Holladay JT, Piers PA, Koranyi G, van der Mooren M, Norrby NES. A new intraocular lens design to reduce spherical aberration of pseudophakic eyes. Journal of Refractive Surgery (Thorofare, NJ: 1995). 2002, 18(6): 683–691. pmid:12458861
  25. 25. Harris WF. Nodes and nodal points and lines in eyes and other optical systems. Ophthalmic Physiol Opt. 2010;30(1):24–42. pmid:20444108
  26. 26. Harris WF. Chief nodal axes of a heterocentric astigmatic eye and the Thibos-Bradley achromatic axis. Vision Res. 2012;73:40–45. pmid:23041561
  27. 27. Arbelaez MC, Vidal C, Arba-Mosquera S. Clinical outcomes of corneal vertex versus central pupil references with aberration-free ablation strategies and LASIK. Investigative Ophthalmology & Visual Science. 2008;49(12):5287–5294. pmid:18658090
  28. 28. Ding X, Wang Q, Chang P, et al. The Repeatability Assessment of Three-Dimensional Capsule-Intraocular Lens Complex Measurements by Means of High-Speed Swept-Source Optical Coherence Tomography. PloS One. 2015;10(11):e0142556. pmid:26600254
  29. 29. Wang X, Dong J, Wang X, Wu Q. IOL tilt and decentration estimation from 3 dimensional reconstruction of OCT image. PloS One. 2013;8(3):e59109. pmid:23554982
  30. 30. Gu X, Chen X, Yang G, et al. Determinants of intraocular lens tilt and decentration after cataract surgery. Annals of Translational Medicine. 2020;8(15):921. pmid:32953721