Multiphysics simulation of magnetoelectric micro core-shells for wireless cellular stimulation therapy via magnetic temporal interference

  • Ram Prasadh Narayanan ,

    Roles Formal analysis, Investigation, Methodology, Validation, Writing – original draft, Writing – review & editing

    ram.p.narayanan@ntnu.no

    Affiliation Institute of Electronic Systems, Norwegian University of Science and Technology, Trondheim, Norway

  • Ali Khaleghi,

    Roles Conceptualization, Methodology, Supervision, Writing – review & editing

    Affiliations Institute of Electronic Systems, Norwegian University of Science and Technology, Trondheim, Norway, Intervention Center, Oslo University Hospital, Oslo, Norway

  • Mladen Veletić,

    Roles Investigation, Validation, Writing – review & editing

    Affiliations Institute of Electronic Systems, Norwegian University of Science and Technology, Trondheim, Norway, Intervention Center, Oslo University Hospital, Oslo, Norway

  • Ilangko Balasingham

    Roles Funding acquisition, Project administration, Resources, Supervision, Writing – review & editing

    Affiliations Institute of Electronic Systems, Norwegian University of Science and Technology, Trondheim, Norway, Intervention Center, Oslo University Hospital, Oslo, Norway

Abstract

This paper presents an innovative approach to wireless cellular stimulation therapy through the design of a magnetoelectric (ME) microdevice. Traditional electrophysiological stimulation techniques for neural and deep brain stimulation face limitations due to their reliance on electronics, electrode arrays, or the complexity of magnetic induction. In contrast, the proposed ME microdevice offers a self-contained, controllable, battery-free, and electronics-free alternative, holding promise for targeted precise stimulation of biological cells and tissues. The designed microdevice integrates core shell ME materials with remote coils which applies magnetic temporal interference (MTI) signals, leading to the generation of a bipolar local electric stimulation current operating at low frequencies which is suitable for precise stimulation. The nonlinear property of the magnetostrictive core enables the demodulation of remotely applied high-frequency electromagnetic fields, resulting in a localized, tunable, and manipulatable electric potential on the piezoelectric shell surface. This potential, triggers electrical spikes in neural cells, facilitating stimulation. Rigorous computational simulations support this concept, highlighting a significantly high ME coupling factor generation of 550 V/m·Oe. The high ME coupling is primarily attributed to the operation of the device in its mechanical resonance modes. This achievement is the result of a carefully designed core shell structure operating at the MTI resonance frequencies, coupled with an optimal magnetic bias, and predetermined piezo shell thickness. These findings underscore the potential of the engineered ME core shell as a candidate for wireless and minimally invasive cellular stimulation therapy, characterized by high resolution and precision. These results open new avenues for injectable material structures capable of delivering effective cellular stimulation therapy, carrying implications across neuroscience medical devices, and regenerative medicine.

Introduction

Stimulation therapy is the targeted application of controlled signals to biological tissues, predominantly focusing on responsive cells, such as neurons [1] to elicit specific physiological responses, offering therapeutic avenues for myriad medical conditions. By engaging specific neural pathways and circuits, stimulation therapy can modulate, rejuvenate, or regulate neural activity, potentially mitigating symptoms linked to neurological disorders [2, 3]. The recent advances with (non) invasive wearables and medical implants have boosted the research in wireless and battery-free solutions [4] for biological stimulation, sensing, communication and wireless power transfer [59]. Wireless stimulation therapy, either through micro-coils [10] or external wearable devices, which emit alternating magnetic or electrical fields for stimulation [1113]. In this, Transcranial Magnetic Stimulation (TMS) is an FDA approved non-invasive form of brain stimulation that uses alternating magnetic field, as either a single pulse mechanism or as repetitive pulse modulation. TMS plays an important role in elucidating the plasticity of neural networks in the brain for therapeutic applications including Parkinson’s Disease [14, 15], dystonia, stroke, deep brain stimulation (DBS) [16] and other psychiatric and neurological conditions [17]. However, it is not without concerns–issues related to penetration depth and potential side effects, like headaches and nausea, persists [17, 18]. On the other hand, transcranial Direct Current Stimulation (tDCS) involves the application of weak pulsating currents (1–2 mA) across the cortex using electrodes placed on the scalp [19]. Evidence suggests the efficacy of tDCS for specific neurological disorders, but different regions in the brain react differently to various neuromodulation signals, thus requiring better theoretical models and custom pulse modulations [20]. Moreover, tDCS also has evidence of side effects such as burning sensation, headache etc [21, 22]. These existing stimulation techniques remains unsuitable precise stimulation requirements (e.g. deep brain stimulation), due to the inability to penetrate the ganglion region of the brain, thus offering relief for conditions including paralysis rehabilitation [18, 19, 23], and connection to brain machine interface (BMI) systems [24, 25]. With the advances in micro-fabrication and the use of micro-electrode arrays for neuro-modulation, precise interaction with the physiological medium and subsequent electrical stimulation has improved the efficiency of neural sensing and the threshold of neural stimulation/modulation [26]. Electrophysiological stimulation utilizes microelectrode elements or arrays for precision interaction with biology, applicable across diverse scales—from organs to individual cells [27]. It is important on two fronts: it not only facilitates targeted interventions but also yields valuable physiological insights, acting as a conduit between electrical activity and biological responses. This method, supported by extensive research and practical implementations, holds promise for applications such as diagnosis and neuromodulation [28]. Powering these micro-electrode arrays and external stimulators could be through conventional wired bio-electrodes [29, 30] to contemporary wireless solutions [3134]. The wired paradigm employs bioelectrodes tethered to an external power source. Though effective, it can be invasive, penetrating through tissues and vascular systems, carry inherent risks—breakage, infections, tissue damage, and other long-term complications [35]. In contrast, wireless modalities, characterized by their non-invasive nature, dispatch electrical pulses through the skin to activate underlying nerves [36, 37], although these techniques also have challenges related to limited penetration depths and complications with the channel and external manipulation of the wireless signal [38].

Using of TMS or tDCS techniques, with or without micro-coils or micro-electrode arrays, a trade-off exists in these methods between the focal depth and the intensity of the stimulation [39, 40]. For target regions confined to a specific area, a temporal interference signalling method is required not only to modulate and increase the focality of the applied signal but also to reduce unnecessary stimulation of non-target areas [41]. Magnetic Temporal Interference (MTI) is a concept intertwined with TMS and tDCS to improve the focality and precision of stimulation through amplitude-modulated interference signalling [42, 43]. Recent studies have shown that with a four-coil MTI system, the maximum achieved depth, focality and the coil electric field intensity were 1.6 cm, 5.1 cm2, and, 523.93 V/m, at a sinusoidal oscillation frequency of 10 Hz (Table 1, in [41]). The intensity of the induced electrical field is proportional to the applied frequency and the current intensity therefore requiring higher current intensity for the coil operation (~1kA) [41]. An intuitive hypothesis to address the gaps and challenges in the existing techniques of neural stimulation, external modulation, and wireless applicators, is to bridge the existing wireless coil based external control, to semi-implantable micro-coils or particle clusters or arrays. The implant micro coils, though are effective and precise, have limitation on the power consumption [44].

Magnetoelectric (ME) composites have been exploited for their combined property of magnetostriction and piezoelectricity, exhibiting magnetoelectric Multiphysics behaviour. ME composites, including micro and nanoparticles, have been simulated [45, 46], synthesized [47] and reported for their high ME coupling [48], to be used in various applications including, wireless power transfer [7, 49, 50], radiating antennas [5155], bio-stimulation [37, 56], magnetic sensing [51, 57, 58] and targeted drug delivery [5961]. In this paper, we introduce the concept of ME microparticles mediated MTI for minimally invasive and precise stimulation. We hypothesize that our system allows the use of high-frequency magnetic fields to achieve a higher electric current density closer to the cell/tissue surface. The ME composites could be embedded into tissue layers or be present as an array in the inner layers below the scalp. The overall idea is to utilise high frequency non-ionising magnetic fields with less implementation complexity, and generate low frequency demodulated, we electric field pulses from the ME core-shells. In a recent work, a composite combination of Magnetoelectric layers with a rectifying electron transport layer made of ZnO, was shown [62]. The ME effect of the composite material with the rectifying layer allows for the self-rectification of the resultant electrical field intensity. The noise behaviour and the non-linear coefficient was studied for a wide frequency range (1–30 kHz) for the use of ME laminates as magnetic field sensors. The carrier frequency was optimized to improve the field sensitivity of the ME laminate [63].

In our hypothesis, we investigate the use of ME non-linear property, for the active demodulation of the temporal interference frequency component from the two applied high-frequency fields, as shown in Fig 1. We show that the material non-linearity could itself be used for frequency demodulation and to produce multiple eigenmodes at their resonant frequencies with varied deformation patterns on their surface, for precise stimulation. This results in the generation of specific electric potential patterns as hotspots of higher electric field intensity on their surface. Moreover, we show that by utilizing the nonlinear magnetostriction of the core, we can demodulate amplitude-modulated higher frequency components that can directly interact with biological cells, enabling localized, low-frequency stimulation at the single-cell level. Moreover, these hotspots can be adjusted in intensity, facilitating high-resolution cell or tissue stimulation.

thumbnail
Fig 1.

(A. Schematic of conventional MTI, where the power spectral density shows the presence of the two frequency components f1 and f2, and (B., ME-MTI, where the ME core-shells potentially function as a localized stimulant and a frequency demodulator due to their non-linear nature. This behavior of the (shown in the inset of B) demodulates the frequency component of the beating envelope, which could be utilized as a localized source for low frequency electrical stimulation. The power spectral density of the ME-MTI signal shows the demodulated MTI component (shown in the grey shaded area).

https://doi.org/10.1371/journal.pone.0297114.g001

In this study, we examine the detailed design elements of size, geometry, and material properties, as well as the impact of magnetic fields on the intensity of the demodulated waveform for electrical stimulation. Additionally, the study focuses on interactions with biological tissues, particularly neurons, aiming for targeted and controlled stimulation.

The paper is organized into four sections. The materials and methods section begins with the analysis, explaining the materials and methods, including the simulation setup and governing equations for Multiphysics coupling. The results section presents results for magnetostriction, electrical field distribution, and the use of the ME microdevice for neural bio-stimulation. In the discussion section, we discuss the ME microdevice in comparison with existing literature. The last section concludes the paper and outlines future work.

Materials and methods

Here, we discuss the materials and the composite equations used in COMSOL Multiphysics, to model the magnetostriction and the magnetoelectric behavior. This study primarily focuses on the simulation and modeling of an individual element of the ME composite. We have incorporated the non-linear behavior of the materials, especially when subjected to multi-frequency fields.

Magnetoelectric physics

The property of magnetostriction, is the strain observed in ferromagnetic materials due to applied magnetic fields. The strain is a result of a strong orientation and weak perturbation of the electron spin in the ferromagnetic material, resulting in the lattice deformation of the ferromagnetic structures [24]. In other words, magnetostriction is caused due to the interaction of magnetic and elastic forces, thereby considered as a useful property for energy conversion [51]. Coupling with a piezoelectric domain will enable the subsequent conversion of the generated elastic forces into equivalent electrical fields. This Multiphysics is defined by the Joule and Villari effect which is expressed as [64], (1) (2) (3) where, C is the elastic stiffness tensor, e and q are the piezoelectric and piezomagnetic constant tensors respectively, k and μ are the electric permittivity and the magnetic permeability tensors, α is the magnetoelectric coefficient tensor, ϵ and σ are the mechanical strain and stress, respectively; E and D, are the electrical field and displacement, respectively; and H and B are the magnetic field and flux density, respectively.

Multiphysics entities, piezoelectricity and magnetostriction are included with the coupling of solid mechanics, electrostatics, and magnetic fields modules. The magnetic field modeling was performed based on the constitutive B-H relation [64] wherein the boundary condition, Ampere’s Law, was set for all the domains, except the magnetostrictive material as, (4) (5)

The magnetostriction effect is defined with the boundary condition, Ampere’s Law, Magnetostrictive, as, (6) where B is the magnetic flux density, H is the magnetic field intensity, J is the volumetric current density, μ0 is the permeability of free space, μr is the magnetic relative permeability, M is the magnetization, Smech is the stress tensor, and Mr is the remanent magnetization, which is set as zero, since the material will not have residual magnetization after the external source is removed. The magnetostrictive stress is modeled as a non-linear isotropic entity given as, (7) where, λs and Ms are the saturation magnetostriction and magnetization, respectively. The magnetization M, is given as a function of effective magnetic field intensity Heff and Ms as, (8) (9) where, L is the Langevin function and S is the stress tensor. For the piezoelectric shell, The Charge Conservation boundary condition governs the operating piezo shell given as, (10) where, F is the electric flux density, and ρv is the volume charge density.

Simulation setup

To validate our initial concept of employing nonlinear magnetoelectric (ME) materials for biological stimulation, we have chosen a spherical core-shell geometry for our study. In this configuration, the spherical core is exposed to a magnetic field, resulting in deformations in all the directions that are fully coupled to the piezoelectric material comprising the shell. Using this geometry all the spherical vibration modes are transferred to the shell for maximum energy conversion. To validate, a single ME geometry is simulated to demonstrate the feasibility of generating a local voltage difference on the piezoelectric shell for potential cell and tissue stimulation.

Fig 2 depicts a schematic of the simulated COMSOL geometry, where two remote coils and the ME device are shown. The inset shows the magnetoelastic conversion of the induced magnetic field to mechanical stress and the subsequent piezoelectric coupling to produce an electric potential difference on the piezo shell. We used the commercially available 2628 MB MetGlas model as the magnetostrictive core due to its biocompatibility [45, 53] and high reported magneto mechanical coupling coefficient [65]. A piezoelectric coating of Aluminum Nitride (AlN) was added to the magnetostrictive core. The poling axis of the piezo shell was set to the same global axis coordinates. Table 1 provides the material characteristics of the selected magnetostrictive material (MetGlas) and piezoelectric material (ALN). Table 2 depicts the optimized dimensions of the microdevice. To model the piezoelectric material parameters, the material compliance (sE), coupling matrix (d), and the permittivity (ϵT) are required, and given as Table 3.

thumbnail
Fig 2. Simulation setup of the COMSOL geometry showing the coils and the ME device.

The distance between the coils were set to 800 μm in the simulation. Both the coils are at an equal distance from each other and the ME particle. The MTI frequency, which is the difference frequencies of f1 and f2, from coils 1 and 2 respectively, is used to induce magnetostriction in the core, that results in the surface voltage.

https://doi.org/10.1371/journal.pone.0297114.g002

thumbnail
Table 1. Material properties of MetGlas (magnetostrictive core) and AlN (piezo shell).

https://doi.org/10.1371/journal.pone.0297114.t001

thumbnail
Table 3. Piezoelectric ALN material parameters: Material compliance (sE), coupling matrix (d), and the permittivity (ϵT).

https://doi.org/10.1371/journal.pone.0297114.t003

The contribution of losses is considered by replacing ϵT with (1—jtanδ) ϵT and cE with (1 + s cE in which The loss values are given in Table 1.

We limited the dimensions of the simulated magnetostrictive core to 100 μm, making it suitable for use as an injectable device. In practical applications, this core could be incorporated into an array within a transplantable biological scaffold. The modeling methodology employed in our study is depicted in Fig 3. The core shell was placed between two symmetrically positioned coils driven by a constant AC current source of 100 mA. Two mechanical resonant frequencies (126 and 188 MHz) with a considerable frequency difference were selected as the AC excitation of the magnetic fields applied to each coil. The choice of the AC excitation frequencies is detailed in the results section. We apply a user-controlled mesh, where triangular mesh is applied for the coil and particle domains, additionally, boundary layer meshing was added to the coil and the particle boundaries to increase the mesh density distribution. We were able to achieve an average mesh element quality of 0.8496.

thumbnail
Fig 3. Modeling method to analyze the non-linear frequency demodulation and the magnetoelectric coupling.

Step 1 shows the non-linear variation of Magnetostriction for an applied DC bias sweep. The DC bias field where maximum slope of Magnetostriction is observed, is selected as the Optimum DC bias, HDC(opt). Step 2 shows the selection of AC excitation field through Eigen mode analysis. For the simulation, we selected 126 and 188 MHz (shown by the red and blue vertical lines). In step 3, we analyze the frequency demodulation at the HDC(opt), and the selected AC excitations.

https://doi.org/10.1371/journal.pone.0297114.g003

Regarding the model simulation, the time step for the time domain analysis was set at 0.1 × 0.

05/(f2), where f2 is 188 MHz. Consequently, each time step (sampling time) is approximately in the order of 100s of nanoseconds in duration. This resulted in a higher number of simulation steps, amounting to 200,000 entries, in the model. The number of degrees of freedom, with user-controlled meshing, was 7700 entries (refer to [66]), requiring around 24 GB (200000×7700×8B×2) of memory. Since this is a time-domain simulation, the factor of two is also multiplied because the time derivatives are also stored. Reduced sampling time would result in a decreased number of output times, which would subsequently reduce disk usage, albeit at the expense of diminished resolution of the time steps. The complexity of the ME Multiphysics and the requirement for time derivatives due to dual-frequency excitation are the primary reasons for the lengthy run times and high disk space consumption.

In contrast to the model presented in [45], which explores a magnetoelectric (ME) ’nanoparticle’ using a frequency domain solver in COMSOL at a low frequency of 50 Hz, our design approach diverges significantly in multiple key aspects. First, we utilize micro-sized particles instead of nanoparticles. Second, our model operates at a very high frequency (VHF) range, which results in multi-modal magnetostrictive deformations—essential for creating high-field intensity on the particle. Additionally, we account for the nonlinear behavior of the magnetostrictive core, particularly when two frequencies are applied simultaneously to demonstrate MTI effect.

Given this complexity, our numerical methodology relies on a finite element time domain solver, specifically employing the highly nonlinear PARADISO solver settings in COMSOL. To improve computational efficiency, and demonstrate the nonlinear simulation features we selected two AC frequencies with a significant difference of 62 MHz, thus reducing the simulation’s time steps and enabling a thorough demonstration of the ME core-shell’s nonlinear functionalities. It is important to note that, in practical applications, the difference frequency should fall within the range suitable for biological stimulation response, specifically units of 100s of Hz (Table 1 in [60]).

Results

We have organized our results into three distinct sections. First, we present findings related to the static magnetic field bias applied to our core-shell geometry. Second, we delve into the behavior of the magnetostrictive core when subjected to high-frequency AC field, focusing on frequency multiplication effects resulting from nonlinearity. Finally, we discuss the integration of the piezoelectric shell with the magnetostrictive core, which leads to electric field polarization and the conversion of mechanical deformations into electrical signals capable of interacting with biological systems.

Static magnetic bias

The magnetostrictive core’s elasticity-induced deformation, magnetostriction (λ) and subsequent electric polarization (on the piezo shell) are maximum when the ME device is operated at the optimum static magnetic bias HDC(opt). At this static magnetic bias, maximum magnetostrictivity is also observed. MetGlas has a magnetostriction of 12 ppm and a saturation magnetization of 0.88T (700282 A/m) [67], and the magnetostrictive strain is limited to {-0.88, 0.88} T. Fig 4 shows the non-linear variation of the magnetostrictive strain tensor for applied static magnetic field in the range {-1, 1} T, for the selected MetGlas material (Table 1) and dimensions (Table 2). The optimum bias field HDC(opt) was identified in the range 200–350 mT (maximum sloop of the curve). Through the subsequent simulations, the optimum magnetic DC bias for producing maximum coupling between the piezo shell and the core was observed at 293 mT.

thumbnail
Fig 4. Non-linear magnetization curve showing the rise and fall magnetostrictive strain at ≈ {-0.88, 0.88} T and the maximum slope of the magnetostriction in the range 200–350 mT.

https://doi.org/10.1371/journal.pone.0297114.g004

AC excitation

At the mechanical resonant frequency, the magnetostriction, magnetization, and deformation of the magnetostrictive core are maximum [68]. Thus, the non-linear characteristic of the magnetostrictive device operating in its Eigen mode can produce irregular but intense deformation (in the core) and thus couple electrical hotspots on the subsequent piezo shell. This theory is directly related to the mechanical resonant modes’ relationship with the elasticity caused by magnetostriction. Here, we analyzed the magnetostrictive sphere in its eigenmode mechanical resonances. In the simulation, the number of desired Eigen frequencies was set to 1000 and searched in the range of 30–240 MHz. The core shell geometry is biased at HDC(opt) of 293 mT. The damping ratio for the resulting Eigen frequencies was calculated as , as depicted in Fig 5.

thumbnail
Fig 5.

A.) Damping ratio for different Eigen frequencies (30–240 MHz) for the ME core shell, externally biased at 293 mT. B and C.) The damping ratio for the selected AC excitation frequencies of 126 MHz and 188 MHz is lower than 10−15.

https://doi.org/10.1371/journal.pone.0297114.g005

This analysis was conducted to identify two resonant mode frequencies at which the damping ratio (ζ) is extremely low (<10−10) (i.e., the device is nearly “critically damped”), thus providing maximum mechanical deformation. As displayed in Fig 5, multiple combinations of such frequencies exist. We opted for two resonant frequencies with a significant difference, specifically 126 MHz and 188 MHz, to enable faster simulations and reduce computation time. Selecting frequencies closer to each other would necessitate smaller timesteps and extended simulation time. Given that the focus of this work is to provide simulation proof for non-linear demodulation effects, we conducted the simulations at these higher frequencies. We then analyzed frequency demodulation at the difference frequency of 62 MHz.

The elastic deformation at sampled frequencies within the 30–240 MHz range is illustrated in Fig 6. As demonstrated, increased frequency leads to greater deformation and a higher number of peaks, which can result in a relative increase in polarization in the subsequent piezoelectric shell.

thumbnail
Fig 6. Deformation of the ME core shell for different Eigen frequencies.

Irregular deformation at certain frequencies could generate the hypothesized electrical hotspots. Displacement in certain spots is comparatively higher than other spots, for certain Eigen modes, confirming our hypothesis.

https://doi.org/10.1371/journal.pone.0297114.g006

Frequency demodulation

We conducted a time-domain simulation on the magnetostrictive core, incorporating both the two AC excitation frequencies and the DC bias. Fig 7 shows the magnetic field induced within the core at various time instances, as well as the resulting deformation of the core due to this induced magnetic field, as illustrated by the geometry of the sphere.

thumbnail
Fig 7.

A 2D axis-symmetric simulation showing the normalized magnetic field induced on the magnetostrictive core (shown with color bar) for fixed DC bias of 293 mT (A-F). The deformation of the core due to the induced magnetic field as the movement in the circle boundary. The observed average magnetic field is appx. 66.7 kA/m.

https://doi.org/10.1371/journal.pone.0297114.g007

Magnetic Temporal Interference (MTI) can be generated by overlapping two different magnetic field frequencies, resulting in an amplitude-modulated (beating) waveform at the difference frequency [69]. In our system, this MTI signal interacts with the magnetoelectric (ME) microdevice, producing a nonlinear multiplication of the applied frequencies. We demonstrate the impact of MTI through simulations in both an air medium and within the ME–MTI system. Fig 8 depicts the time-domain variations of the magnetic field within the magnetostrictive core (Fig 8A), as well as a spherical air medium in COMSOL (Fig 8C). The magnetostrictive core reveals waveform deformation, where the spectral density displays the core’s nonlinear behavior in the magnetic field, resulting in the generation of a difference frequency at 62 MHz (Fig 8B), whereas the amplitude of the magnetic field in the air medium follows a beating envelope, with the power spectral density incorporating both applied frequencies (Fig 8D). We have validated these findings using different frequency excitations, and the results consistently demonstrate the nonlinear demodulation effect.

thumbnail
Fig 8.

Normalized magnetic fields induced in A) the Magnetostrictive core and B) Power Spectral Density (FFT) on the core. Normalized magnetic fields induced in C) the Air and B) Power Spectral Density (FFT) on the air medium. The applied magnetic fields are at frequencies f1 (126 MHz) and f2 (188 MHz). The demodulated frequency component (Δf = f2 –f1) of 62 MHz is highlighted in B, demonstrating the ME-MTI system’s unique frequency characteristics.

https://doi.org/10.1371/journal.pone.0297114.g008

Electric field and ME coefficient

In this section, we explore the electric field distribution and the magnetoelectric coupling achieved through the addition of a piezoelectric shell. We examined the distribution of the induced electric field and polarization across this piezoelectric layer due to the applied magnetic field.

Electric field distribution

An aluminum nitride (AlN) piezoelectric coating was integrated with the magnetostrictive core to transduce its mechanical deformations into stress within the piezoelectric shell. Computations were performed using COMSOL, incorporating coupled Multiphysics domains of solid mechanics, electrostatics, and magnetic fields in single time domain simulations. The piezo shell’s thickness was optimized to 37.5 μm, guided by the coupling coefficient data (Fig 11). Fig 9 illustrates the resulting electric field distribution on the piezo shell, generated by the applied magnetic field to the medium. We observed dynamic patterns in the electric field across the piezo shell, corresponding to variations in the core magnetic field. Due to the multimodal nature of these magneto-mechanical deformations, we identified multiple intensified electric field distributions on the piezo shell, especially (see Fig 9). These irregular deformations, follows the same non-linear behavior which stem from the core material’s non-linear magnetostrictive properties, led to a localized, non-linear electrical "hotspots" on the piezoelectric shell. Meaning that the electric field observations indicate the same frequency conversion observed in the magnetic field of the core device.

thumbnail
Fig 9. Normalized magnetic field induced on the magnetostrictive core (top scale, in A/m) and the normalized electric field (bottom scale, in V/m) induced on the piezo shell (37.5 μm thickness) for an applied DC bias of 293 mT, at the MTI frequency of 62 MHz.

(A-F). The non-linear induced electric field is shown as the varying bright red spots on the piezoelectric shell, at varying time instances.

https://doi.org/10.1371/journal.pone.0297114.g009

Moreover, Fig 10 showcases the time-dependent average and peak electric fields, measured on the surface of the piezo shell. The maximum surface electric field recorded ranged between 20 and 70 mV/μm. Additionally, at specific time points and locations, the electric field exceeded 100 mV/μm.

thumbnail
Fig 10. Surface maximum and surface average electrical field values on the Piezo shell (37.5 μm thickness), for an applied DC bias of 293 mT.

https://doi.org/10.1371/journal.pone.0297114.g010

Magnetoelectric coupling coefficient

The magnetoelectric (ME) coupling factor (αME) serves as a crucial metric for understanding the non-linear interaction among the magnetic, elastic, and electric domains within the core-shell structure. This factor quantifies the transformation of induced magnetization into electrical polarization and is mathematically defined as: (11) where E and H represent the electrical and magnetic fields generated in the piezoelectric and magnetostrictive layers, respectively.

Based on our simulations, variations in the applied HDC and piezo shell thickness influenced the value of αME. Fig 11A depicts how αME varies for shell thicknesses ranging between 35 and 45 μm, at the difference frequency of 62 MHz, and for a DC bias range of 0–1000 mT. Fig 11B zooms in on this relationship by fine-tuning the piezo thickness between 37 and 37.7 μm, in increments of 0.1 μm, to pinpoint an optimal coupling coefficient. A peak coupling factor of approximately was observed at a shell thickness of approximately 37.5 μm, as shown in Fig 11B.

thumbnail
Fig 11.

A) ME coupling factor analysis for 35–45 μm (top) and B) step reduced analysis (bottom). The optimum Piezo shell thickness is determined to be 37.5 μm at which maximum αME is obtained. The shaded region shows the area within which αME is maximum.

https://doi.org/10.1371/journal.pone.0297114.g011

Interestingly, within the operational DC bias range of 200–350 mT (indicated by the shaded region), αME peaks, suggesting minimal secondary losses such as thermal power loss. This peak value indicates that maximum conversion efficiency has been attained. In summary, our work successfully achieved a maximal magnetoelectric coupling, obtained by optimizing both the DC bias and piezoelectric shell thickness.

ME core shell for bio-stimulation

In this section, we incorporate the core-shell Magneto-Electric (ME) microdevice into a biological environment populated with neurons to explore its capabilities for neural activation. We utilize the well-established Hodgkin–Huxley (HH) neuron model [70, 71] for our simulations, aiming to assess how the electrical potential difference generated on the surface of the piezoelectric shell, as detailed in previous sections, can trigger neural action potentials. It is crucial to highlight that we have adjusted the stimulation frequency to fall within a biologically relevant range of 35–380 Hz for inducing action potentials in neurons. This is different from the 62 MHz MTI frequency employed in our earlier COMSOL simulations, which was chosen for computational efficiency and memory-saving purposes. To resolve this, we introduce a corresponding difference frequency for the external signals that are applied to the coils.

Fig 12 provides a graphical representation that visualizes the interaction between two arbitrary points on the core-shell device and a neuron’s membrane surface. This figure also includes a generic circuit model to conceptualize the intricate bio-electro-mechanical interactions involved.

thumbnail
Fig 12.

(A) Schematic of the core shell interaction with a cell membrane and (B) the equivalent circuit representing the cell. V0 is the voltage difference between two random points on the core shell interacting with the cell. RIC is the intra-cellular resistance inside the cell.

https://doi.org/10.1371/journal.pone.0297114.g012

In HH model, the neuronal electrical activity is described through the membrane potential (vm), which is dependent on voltage-gated potassium (K+) channels, voltage-gated sodium (Na+) channels, a leak current, and an induced current (iind) [70]: (12) where Cm is specific membrane capacitance; VK, VNa, and VL are Nernst potentials for K+ ions; Na+ and other ions are combined as “leak” channels; and gk, gNa, and gL are the corresponding membrane conductances. Voltage-gated conductances change with time during an action potential. represent the opening probabilities for K+ and Na+ channels, respectively. The gating variables mK, mNa, and hNa and the relevant parameters are define here [70]. For the reproducibility of the results, we have also provided the Github link for the MATLAB file used to create the spike voltages [72].

In Fig 13, we display the electric potential difference between two random points on the piezoelectric shell surface, with peak-to-peak potentials ranging between approximately 250–400 mV. Given that the MTI frequency resides in the high-frequency (HF) spectrum (64 MHz), the pulse width for a single signal train is around 200 ns. This duration is insufficient for effective electrical neural stimulation, however we time scale signal to fall into the neural stimulation frequency range.

thumbnail
Fig 13. Time variation of electric potential difference between two random points on the piezo shell surface.

The shaded areas are expanded and shown related to regions A—D.

https://doi.org/10.1371/journal.pone.0297114.g013

To calculate the injected current at the biological environment, and use the HH model, we calculated the static resistance between two fixed points on the cell surface by using numerical full wave EM simulations using CST Microwave Studio ® Finite Element Method (FEM) tool. The ohmic impedance at the low frequency range for two spaced electrodes of area 400nm2 and the distance of 3 μm is in the range 0.45–1 MΩ. This information is crucial for optimizing the interaction between the core-shell ME microdevice and the biological medium for potential neural stimulation applications.

The current signal, iind, comprises various frequency components, which include the two initially applied RF frequencies, the temporal interference frequency, as well as the harmonics generated by the device’s non-linearity. While the original applied RF frequencies are not capable of cellular stimulation, the difference frequency serves as the signal responsible for stimulation. To isolate this component, a low-pass filter with a 100 MHz cut-off is applied to remove the RF signal from the stimulation pattern.

To analyze the impact of this filtered signal on neural spike generation, the time scale was adjusted by a factor of 106. This adjustment compensates for the high-frequency difference (62 MHz) selected in the simulation, effectively mapping the MTI frequency to 62 Hz for biological relevance. When this filtered and time-scaled current signal is input into a HH model, it becomes evident that the signal’s amplitude substantially exceeds the necessary stimulation threshold. To address this, the amplitude is scaled down within a range of 200–5000, and the resulting number of membrane potential spikes (shown in Fig 14) is then counted.

thumbnail
Fig 14. Induced current and the evoked plasma membrane potential using amplitude scaling factor.

For the applied induced current spike, a neural spike was generated, conforming to the voltage levels of the repolarization (-70 mV) and depolarization (30–40 mV).

https://doi.org/10.1371/journal.pone.0297114.g014

Fig 15 illustrates the number of generated spike potentials across various amplitude scaling factors. It’s worth noting that this amplitude scaling is functionally equivalent to either reducing the current in the external coil relative to the defined distance between the coils and the Magneto-Electric (ME) device or increasing the separation distance between the external applicator and the implanted ME device while maintaining the same current. As demonstrated in Fig 15, a scaling factor of 200 resulted in a high number of generated spikes, but also raised the risk of hyper-polarization or premature depolarization. As we increased the scaling factor, we observed a plateau phase featuring stable and temporally consistent spike generation, where the peak amplitude of iind reached 15 μA/cm2. Additionally, when iind dropped below 8 μA/cm2, spike generation became increasingly irregular, and no spikes were generated when iind was less than 3 μA/cm2. Consequently, we conclude that for effective localized cellular or tissue stimulation via the ME core-shell device, the external parameters should be tuned to produce an iind value falling within the shaded region shown in the figure. We also estimate that an exposure time of approximately 10 ms is sufficient to elicit a single spike.

thumbnail
Fig 15. This graph displays the number of generated spike potentials as a function of the scaled amplitude values of iind.

The shaded region highlights the plateau where spike generation stabilizes.

https://doi.org/10.1371/journal.pone.0297114.g015

Discussion

The Magnetoelectric (ME) effect, with its promising applications in medical sensing and stimulation, is gaining momentum in research. While most ME research has been devoted to characterizing and quantifying material properties, the potential of non-linear magnetostriction in ME cores for frequency demodulators remains relatively untapped. In this study, we introduce a novel battery-free ME core-shell structure. This design enhances bio-stimulation capabilities and eliminates the need for additional circuitry for frequency demodulation. We highlight that the non-linear traits of ME core shells can facilitate demodulation, allowing for the extraction of low-frequency components from AC excitations. Our results provide a foundation for understanding ME core shells in resonant operation and their potential applications in medical patch sensing and implantable devices.

To exploit this potential of ME core shells, multiple control parameters must be taken to consideration, which include, the core radius, magnetostriction of the core material, piezo shell thickness, placement of the external AC excitation coils and the DC bias value. Moreover, the operation of the ME in its Eigen frequencies, amplifies the magnetostriction induced deformation of the core and subsequently increases the electrical field distribution on the piezo layer.

In our fundamental simulation, two coils were placed symmetrical to the MetGlas core and excited with two different AC magnetic fields. The DC magnetic bias was fixed to a constant value, where maximum magnetostriction was observed. During the perturbation process, the core deforms as an intermediate result of the externally applied magnetic fields. The random compression and expansion in the axial and radial axes are a result of the placement of the AC excitation coils and the applied DC magnetic bias. Controlled change in the properties of the external DC and AC fields, could tune the deformation in the magnetostrictive core. In the subsequent simulations, we added the piezo shell layer and optimized its thickness for maximum coupling. The strain generated on the magnetostrictive core is transferred to the piezo shell as an electrical field distribution. Fig 11 shows the variation in the ME coupling coefficient for different piezo thickness. The Eigen frequency analysis was carried out to select the two AC excitation frequencies with minimum loss. At resonance, the deformation of the core is maximum, resulting in maximum electrical field on the piezo shell. To justify the selected AC excitation frequencies, we analyzed the damping factor for the available mechanical resonances in a range of selected frequencies (from 30–240 MHz). Two frequencies with damping ratios lower than 10−15, and with a substantial bandwidth difference, were selected. The mechanical resonances were non-uniform in many modes, which is also an advantage, so that, with precise placement of coils, selection of frequencies, the resultant electrical field distribution on the piezo shell could be controlled on specific areas on the shell. These predefined areas on the shell could be in contact with the cell or tissue layers (as depicted in Fig 12A). It should be noted that resonance frequencies vary for different dimensions of the core shell.

ME devices can be exploited for their multifunctional use through their non-linear eigenmodes. An interesting application could involve targeted drug delivery to neural tissue, where non-resonant frequency excitations could loosen the bonds between the drug and the ME particle [61]. Subsequent excitation of the ME device at its resonance could trigger electrical pulses on the piezo shell for bio-stimulation.

ME nanoparticles were experimentally tested in low-frequency non-resonant modes [56], where the use of such particles as a coagulated entity was experimentally tested. While the use of ME core shells in the nanoscale is required for cellular uptake and to cross the Blood-Brain Barrier, the controllability of single entities in a coagulation is complex. Our design suggests the use of micro core shells that adhere to the mechanical resonant modes, enabling future fabrication experiments to fabricate subsets of core shells with slightly different dimensions. This allows for intricate tuning of the external magnetic fields to control each of the subset which are identified through their unique resonant modes.

Additionally, the modeling of single ME core-shell nanoparticle was recently published by Fiocchi et al. in [45]. They achieved a coupling of 0.28 V/cm·Oe with their ME core-shell nanoparticles. In contrast, our design achieved a coupling of approximately 5.5 V/cm·Oe. This increase is primarily attributed to the larger dimensions of our core shell and the operation of the ME device in mechanical resonance modes. It is important to note that reducing the size of our core shells to the nanoscale increases mechanical resonance, indirectly increasing the complexity of the external coils and the required circuitry to process GHz/THz frequency inputs. We utilize the mechanical resonance modes of our designed core shells, which are in the order of the 100 MHz range. Therefore, we limit our size to the microscale to remain within the MHz frequency range. We fixed the MetGlas core to be 50μm in radius and then characterized the magnetostrictive core for its maximum magnetostriction.

To confirm the validity of our simulation models, we first simulated an existing experimental study of a ME structure. Joy et al. analyzed a miniature ME antenna as a planar structure [68], and their first step was to calculate the optimum DC field at which the structure needed to be biased. They conducted experimental verification and obtained a DC bias value. We replicated the same structure in COMSOL and observed that we also obtained the same DC bias field calculated through their experiments. This confirmed the validity and boundary conditions of our modeling for further simulations.

Conclusion

This study provides a proof-of-concept and analytical methodology, demonstrating the use of a Magnetic Temporal Interference for frequency demodulation through non-linear ME core shells to stimulate biological cells and tissues. Our central discovery is the ability of these remotely powered, resonant ME core shells to produce electric fields potent enough to stimulate cells and tissues effectively. Capitalizing on the irregular deformation and non-linearity of magnetostriction, it is possible to precisely localize the electric field on the piezoelectric shell. This precise localization is achieved through strategic adjustments in geometry, magnetic bias control, and frequency selection. By fine tuning external excitation frequencies, we can induce optimal temporal interference low-frequency magnetic fields on the magnetostrictive core, thereby generating electric fields on the piezo shell, tailored for cellular stimulation.

Our proposed method, termed ME-MTI, offers notable advantages over conventional Magnetic Temporal Interference. Notably, it provides enhanced spatial resolution and energy efficiency, leading to a reduction in the size and complexity of external units. The capability of ME core shell microdevices to produce low-frequency generation and demodulation responses means that cells and tissues can effectively filter out high-frequency fields, relying solely on MTI fields for stimulation. This research underscores the promise of ME core shells in remote neural stimulation and brain control applications, especially pertinent for deep brain stimulation or for inhibiting specific sensations in the brain cortex. However, further validation is essential. Future investigations should focus on analyzing the thermal effects of the implantable microdevice, evaluating the Specific Absorption Rate (SAR) of the external coils, and exploring the broader potential of ME-MTI core shells in the realm of biomedical applications.

Acknowledgments

All figures were either created or edited with BioRender.com.

References

  1. 1. Krauss JK, Lipsman N, Aziz T, Boutet A, Brown P, Chang JW, et al. Technology of deep brain stimulation: current status and future directions. Nat Rev Neurol. 2021;17: 75–87. pmid:33244188
  2. 2. Verdes M, Mace K, Margetts L, Cartmell S. Status and challenges of electrical stimulation use in chronic wound healing. Curr Opin Biotechnol. 2022;75: 102710. pmid:35398709
  3. 3. Soliman H, Gourcerol G. Gastric Electrical Stimulation: Role and Clinical Impact on Chronic Nausea and Vomiting. Front Neurosci. 2022;16: 909149. pmid:35620661
  4. 4. Nair V, Dalrymple AN, Yu Z, Balakrishnan G, Bettinger CJ, Weber DJ, et al. Miniature battery-free bioelectronics. Science (80-). 2023;382: eabn4732. pmid:37943926
  5. 5. Khaleghi A, Dumanli S, Balasingham I. An Overview of Medical Implant Antennas. 17th European Conference on Antennas and Propagation, EuCAP 2023. 2023. pp. 1–5.
  6. 6. Tai WW, Zulkefli MS, Soh PJ. A review of implantable and ingestible antenna for wireless capsule endoscopy system. 15Th Univ Malaysia Teren Annu Symp 2021 (Umtas 2021). 2023;2746: 020025.
  7. 7. Khan SR, Pavuluri SK, Cummins G, Desmulliez MPY. Wireless power transfer techniques for implantable medical devices: A review. Sensors (Switzerland). 2020;20: 1–58. pmid:32575663
  8. 8. Lu D, Yan Y, Avila R, Kandela I, Stepien I, Seo MH, et al. Bioresorbable, Wireless, Passive Sensors as Temporary Implants for Monitoring Regional Body Temperature. Adv Healthc Mater. 2020;9: 2000942. pmid:32597568
  9. 9. Tian X, Zeng Q, Kurt SA, Li RR, Nguyen DT, Xiong Z, et al. Implant-to-implant wireless networking with metamaterial textiles. Nat Commun. 2023;14: 4335. pmid:37468458
  10. 10. Lee SW, Fallegger F, Casse BDF, Fried SI. Implantable microcoils for intracortical magnetic stimulation. Sci Adv. 2016;2: e1600889. pmid:27957537
  11. 11. Koo J, MacEwan MR, Kang SK, Won SM, Stephen M, Gamble P, et al. Wireless bioresorbable electronic system enables sustained nonpharmacological neuroregenerative therapy. Nat Med. 2018;24: 1830–1836. pmid:30297910
  12. 12. Wirsing PG, Habrom AD, Zehnder TM, Friedli S, Blatti M. Wireless micro current stimulation—an innovative electrical stimulation method for the treatment of patients with leg and diabetic foot ulcers. Int Wound J. 2015;12: 693–698. pmid:24373098
  13. 13. Zhang C, Pan C, Chan KF, Gao J, Yang Z, Leung KKC, et al. Wirelessly powered deformable electronic stent for noninvasive electrical stimulation of lower esophageal sphincter. Sci Adv. 2023;9: eade8622. pmid:36888700
  14. 14. Boutet A, Madhavan R, Elias GJB, Joel SE, Gramer R, Ranjan M, et al. Predicting optimal deep brain stimulation parameters for Parkinson’s disease using functional MRI and machine learning. Nat Commun. 2021;12: 3043. pmid:34031407
  15. 15. Limousin P, Foltynie T. Long-term outcomes of deep brain stimulation in Parkinson disease. Nat Rev Neurol. 2019;15: 234–242. pmid:30778210
  16. 16. Mahoney JJ, Hanlon CA, Marshalek PJ, Rezai AR, Krinke L. Transcranial magnetic stimulation, deep brain stimulation, and other forms of neuromodulation for substance use disorders: Review of modalities and implications for treatment. J Neurol Sci. 2020;418: 117149. pmid:33002757
  17. 17. Hallett M. Transcranial Magnetic Stimulation: A Primer. Neuron. 2007;55: 187–199. pmid:17640522
  18. 18. Major BP, Rogers MA, Pearce AJ. Using transcranial magnetic stimulation to quantify electrophysiological changes following concussive brain injury: A systematic review. Clin Exp Pharmacol Physiol. 2015;42: 394–405. pmid:25603731
  19. 19. Chase HW, Boudewyn MA, Carter CS, Phillips ML. Transcranial direct current stimulation: a roadmap for research, from mechanism of action to clinical implementation. Mol Psychiatry. 2020;25: 397–407. pmid:31455860
  20. 20. Jacobson L, Koslowsky M, Lavidor M. TDCS polarity effects in motor and cognitive domains: A meta-analytical review. Exp Brain Res. 2012;216: 1–10. pmid:21989847
  21. 21. Fregni F, Nitsche MA, Loo CK, Brunoni AR, Marangolo P, Leite J, et al. Regulatory considerations for the clinical and research use of transcranial direct current stimulation (tDCS): Review and recommendations from an expert panel. Clin Res Regul Aff. 2015;32: 22–35. pmid:25983531
  22. 22. Matsumoto H, Ugawa Y. Adverse events of tDCS and tACS: A review. Clin Neurophysiol Pract. 2017;2: 19–25. pmid:30214966
  23. 23. Razza LB, Palumbo P, Moffa AH, Carvalho AF, Solmi M, Loo CK, et al. A systematic review and meta-analysis on the effects of transcranial direct current stimulation in depressive episodes. Depress Anxiety. 2020;37: 594–608. pmid:32101631
  24. 24. Slutzky MW. Brain-Machine Interfaces: Powerful Tools for Clinical Treatment and Neuroscientific Investigations. Neuroscientist. 2019;25: 139–154. pmid:29772957
  25. 25. Shanechi MM. Brain–machine interfaces from motor to mood. Nat Neurosci. 2019;22: 1554–1564. pmid:31551595
  26. 26. Pancrazio JJ, Deku F, Ghazavi A, Stiller AM, Rihani R, Frewin CL, et al. Thinking Small: Progress on Microscale Neurostimulation Technology. Neuromodulation. 2017;20: 745–752. pmid:29076214
  27. 27. Goyal A, Goetz S, Stanslaski S, Oh Y, Rusheen AE, Klassen B, et al. The development of an implantable deep brain stimulation device with simultaneous chronic electrophysiological recording and stimulation in humans. Biosens Bioelectron. 2021;176: 112888. pmid:33395569
  28. 28. Shan Y, Cui X, Chen X, Li Z. Recent progress of electroactive interface in neural engineering. Wiley Interdiscip Rev Nanomedicine Nanobiotechnology. 2023;15: e01827. pmid:35715994
  29. 29. Zhang Z, Zheng T, Zhu R. Microchip with Single-Cell Impedance Measurements for Monitoring Osteogenic Differentiation of Mesenchymal Stem Cells under Electrical Stimulation. Anal Chem. 2020;92: 12579–12587. pmid:32859132
  30. 30. Shadmani A, Viswam V, Chen Y, Bounik R, Dragas J, Radivojevic M, et al. Stimulation and Artifact-Suppression Techniques for In Vitro High-Density Microelectrode Array Systems. IEEE Trans Biomed Eng. 2019;66: 2481–2490. pmid:30605090
  31. 31. Frederick RA, Troyk PR, Cogan SF. Wireless microelectrode arrays for selective and chronically stable peripheral nerve stimulation for hindlimb movement. J Neural Eng. 2021;18: 56058. pmid:34592725
  32. 32. Lee J, Leung V, Lee AH, Huang J, Asbeck P, Mercier PP, et al. Neural recording and stimulation using wireless networks of microimplants. Nat Electron. 2021;4: 604–614.
  33. 33. Khalifa A, Liu Y, Karimi Y, Wang Q, Eisape A, Stanaćević M, et al. The Microbead: A 0.009 mm3 Implantable Wireless Neural Stimulator. IEEE Trans Biomed Circuits Syst. 2019;13: 971–985. pmid:31484132
  34. 34. Burton A, Won SM, Sohrabi AK, Stuart T, Amirhossein A, Kim JU, et al. Wireless, battery-free, and fully implantable electrical neurostimulation in freely moving rodents. Microsystems Nanoeng. 2021;7: 62. pmid:34567774
  35. 35. Co ML, Khouzam JP, Pour-Ghaz I, Minhas S, Basu-Ray I. Emerging Technologies in Cardiac Pacing From Leadless Pacers to Stem Cells. Curr Probl Cardiol. 2021;46: 100797. pmid:33561694
  36. 36. Bok I, Vareberg A, Gokhale Y, Bhatt S, Masterson E, Phillips J, et al. Wireless agents for brain recording and stimulation modalities. Bioelectron Med. 2023;9: 20. pmid:37726851
  37. 37. Zhang Y, Wu X, Ding J, Su B, Chen Z, Xiao Z, et al. Wireless-Powering Deep Brain Stimulation Platform Based on 1D-Structured Magnetoelectric Nanochains Applied in Antiepilepsy Treatment. ACS Nano. 2023;17: 15796–15809. pmid:37530448
  38. 38. Zheng Y, Xia P, Dong L, Tian L, Tian C. Development and review of a sub-millimeter-sized cell-scale micro-magnetic stimulation device. Biomed Phys Eng Express. 2021;7: 042001. pmid:34020437
  39. 39. Deng Z De Lisanby SH, Peterchev A V. Coil design considerations for deep transcranial magnetic stimulation. Clin Neurophysiol. 2014;125: 1202–1212. pmid:24411523
  40. 40. Esmaeilpour Z, Kronberg G, Reato D, Parra LC, Bikson M. Temporal interference stimulation targets deep brain regions by modulating neural oscillations. Brain Stimul. 2021;14: 55–65. pmid:33186778
  41. 41. Xin Z, Kuwahata A, Liu S, Sekino M. Magnetically Induced Temporal Interference for Focal and Deep-Brain Stimulation. Front Hum Neurosci. 2021;15: 693207. pmid:34646125
  42. 42. Khalifa A, Abrishami SM, Zaeimbashi M, Tang AD, Coughlin B, Rodger J, et al. Magnetic temporal interference for noninvasive and focal brain stimulation. J Neural Eng. 2023;20: 16002. pmid:36651596
  43. 43. Boldrey J, Higgs Z, Jiles DC. Temporal interference for dual site transcranial magnetic stimulation. AIP Adv. 2023;13.
  44. 44. Amar A Ben, Kouki AB, Cao H. Power approaches for implantable medical devices. Sensors (Switzerland). 2015;15: 28889–28914. pmid:26580626
  45. 45. Fiocchi S, Chiaramello E, Marrella A, Suarato G, Bonato M, Parazzini M, et al. Modeling of core-shell magneto-electric nanoparticles for biomedical applications: Effect of composition, dimension, and magnetic field features on magnetoelectric response. PLoS One. 2022;17. pmid:36149898
  46. 46. Chen JC, Kan P, Yu Z, Alrashdan F, Garcia R, Singer A, et al. A wireless millimetric magnetoelectric implant for the endovascular stimulation of peripheral nerves. Nat Biomed Eng. 2022;6: 706–716. pmid:35361934
  47. 47. Yan L, Wang Z, Xing Z, Li J, Viehland D. Magnetoelectric and multiferroic properties of variously oriented epitaxial BiFeO3—CoFe2 O4 nanostructured thin films. J Appl Phys. 2010;107.
  48. 48. Staaf H, Sawatdee A, Rusu C, Nilsson D, Schäffner P, Johansson C. High magnetoelectric coupling of Metglas and P(VDF-TrFE) laminates. Sci Rep. 2022;12: 5233. pmid:35347177
  49. 49. Truong BD, Andersen E, Casados C, Roundy S. Magnetoelectric wireless power transfer for biomedical implants: Effects of non-uniform magnetic field, alignment and orientation. Sensors Actuators, A Phys. 2020;316: 112269.
  50. 50. Saha O, Truong BD, Roundy S. A review of wireless power transfer using magnetoelectric structures. Smart Mater Struct. 2022;31: 113001.
  51. 51. Zaeimbashi M, Nasrollahpour M, Khalifa A, Romano A, Liang X, Chen H, et al. Ultra-compact dual-band smart NEMS magnetoelectric antennas for simultaneous wireless energy harvesting and magnetic field sensing. Nat Commun. 2021;12: 3141. pmid:34035237
  52. 52. Burnside EA, Tiwari S, Burnside SR, Candler RN, Henderson R, Grimm S, et al. An axial mode magnetoelectric antenna: Radiation predictions via multiphysics modeling with experimental validations. J Appl Phys. 2023;134.
  53. 53. Zaeimbashi M, Lin H, Dong C, Liang X, Nasrollahpour M, Chen H, et al. NanoNeuroRFID: A Wireless Implantable Device Based on Magnetoelectric Antennas. IEEE J Electromagn RF Microwaves Med Biol. 2019;3: 206–215.
  54. 54. Yu Z, Alrashdan FT, Wang W, Parker M, Chen X, Chen FY, et al. Magnetoelectric backscatter communication for millimeter-sized wireless biomedical implants. Proceedings of the Annual International Conference on Mobile Computing and Networking, MOBICOM. 2022. pp. 432–445.
  55. 55. Chen S, Li J, Gao Y, Li J, Dong H, Gu Z, et al. A micromechanical transmitter with only one BAW magneto-electric antenna. Micromachines. 2022;13: 272. pmid:35208399
  56. 56. Kozielski KL, Jahanshahi A, Gilbert HB, Yu Y, Erin , Francisco D, et al. Nonresonant powering of injectable nanoelectrodes enables wireless deep brain stimulation in freely moving mice. Sci Adv. 2021;7: 1742. pmid:33523872
  57. 57. Bichurin M, Petrov R, Sokolov O, Leontiev V, Kuts V, Kiselev D, et al. Magnetoelectric magnetic field sensors: A review. Sensors. 2021;21: 6232. pmid:34577439
  58. 58. Israel C, Mathur ND, Scott JF. A one-cent room-temperature magnetoelectric sensor. Nat Mater. 2008;7: 93–94. pmid:18219324
  59. 59. Stewart TS, Nagesetti A, Guduru R, Liang P, Stimphil E, Hadjikhani A, et al. Magnetoelectric nanoparticles for delivery of antitumor peptides into glioblastoma cells by magnetic fields. Nanomedicine. 2018;13: 423–438. pmid:29345190
  60. 60. Singer A, Dutta S, Lewis E, Chen Z, Chen JC, Verma N, et al. Magnetoelectric Materials for Miniature, Wireless Neural Stimulation at Therapeutic Frequencies. Neuron. 2020;107: 631–643.e5. pmid:32516574
  61. 61. Rodzinski A, Guduru R, Liang P, Hadjikhani A, Stewart T, Stimphil E, et al. Targeted and controlled anticancer drug delivery and release with magnetoelectric nanoparticles. Sci Rep. 2016;6: 20867. pmid:26875783
  62. 62. Chen JC, Bhave G, Alrashdan F, Dhuliyawalla A, Hogan KJ, Mikos AG, et al. Self-rectifying magnetoelectric metamaterials for remote neural stimulation and motor function restoration. Nat Mater. 2023; 1–8. pmid:37814117
  63. 63. Liu Y, Jiao J, Ma J, Ren B, Li L, Zhao X, et al. Frequency conversion in magnetoelectric composites for quasi-static magnetic field detection. Appl Phys Lett. 2013;103.
  64. 64. Wang Y, Su Y, Li J, Weng GJ. A theory of magnetoelectric coupling with interface effects and aspect-ratio dependence in piezoelectric-piezomagnetic composites. J Appl Phys. 2015;117.
  65. 65. Meeks SW, Hill JC. Piezomagnetic and elastic properties of metallic glass alloys Fe 67CO18B14Si1 and Fe81B13.5Si 3.5C2. J Appl Phys. 1983;54: 6584–6593.
  66. 66. Bloom N, Reenen J Van. 済無No Title No Title No Title. In: NBER Working Papers [Internet]. 2013 p. 89. Available: http://www.nber.org/papers/w16019
  67. 67. Bichurin MI, Petrov R V., Leontiev VS, Sokolov O V., Turutin A V., Kuts V V., et al. Self-biased bidomain linbo3 /ni/metglas magnetoelectric current sensor. Sensors (Switzerland). 2020;20: 1–19. pmid:33322153
  68. 68. Joy B, Cai Y, Bono DC, Sarkar D. Cell Rover—a miniaturized magnetostrictive antenna for wireless operation inside living cells. Nat Commun. 2022;13: 1–11. pmid:36138011
  69. 69. Mirzakhalili E, Barra B, Capogrosso M, Lempka SF. Biophysics of Temporal Interference Stimulation. Cell Syst. 2020;11: 557–572.e5. pmid:33157010
  70. 70. Hodgkin AL, Huxley AF. A quantitative description of membrane current and its application to conduction and excitation in nerve. J Physiol. 1952;117: 500–544. pmid:12991237
  71. 71. Ye H, Barrett L. Somatic inhibition by microscopic magnetic stimulation. Sci Rep. 2021;11: 1–18. pmid:34193906
  72. 72. Github. Supporting Data—NARAYANAN et. al.—PLOSONE. In: neuronStim [Internet]. 2023. Available: https://github.com/mladenveletic/neuronStim.git