Multiplexed micronutrient, inflammation, and malarial antigenemia assessment using a plasma fractionation device

  • Eleanor Brindle,

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Supervision, Writing – original draft, Writing – review & editing

    Affiliations Center for Studies in Demography and Ecology, University of Washington, Seattle, Washington, United States of America, PATH, Seattle, Washington, United States of America

  • Lorraine Lillis,

    Roles Investigation, Methodology, Writing – review & editing

    Affiliation PATH, Seattle, Washington, United States of America

  • Rebecca Barney,

    Roles Investigation, Methodology, Writing – review & editing

    Affiliation PATH, Seattle, Washington, United States of America

  • Pooja Bansil,

    Roles Data curation, Formal analysis, Investigation, Writing – review & editing

    Affiliation PATH, Seattle, Washington, United States of America

  • Francisco Arredondo,

    Roles Investigation, Methodology, Writing – review & editing

    Affiliation Dept of Medicine, Duke University, Durham, North Carolina, United States of America

  • Neal E. Craft,

    Roles Conceptualization, Formal analysis, Funding acquisition, Methodology, Supervision, Writing – review & editing

    Affiliation Craft Nutrition Consulting, Elm City, North Carolina, United States of America

  • Eileen Murphy,

    Roles Project administration, Supervision, Writing – review & editing

    Affiliation PATH, Seattle, Washington, United States of America

  • David S. Boyle

    Roles Conceptualization, Funding acquisition, Methodology, Supervision, Writing – original draft, Writing – review & editing

    dboyle@path.org

    Affiliation PATH, Seattle, Washington, United States of America

Abstract

Processing and storing blood samples for future analysis of biomarkers can be challenging in resource limited environments. The preparation of dried blood spots (DBS) from finger-stick collection of whole blood is a widely used and established method as DBS are biosafe, and allow simpler field processing, storage, and transport protocols than serum or plasma. Therefore, DBS are commonly used in population surveys to assess infectious disease and/or micronutrient status. Recently, we reported that DBS can be used with the Q-plex Human Micronutrient 7-plex Array (MN 7-plex), a multiplexed immunoassay. This tool can simultaneously quantify seven protein biomarkers related to micronutrient deficiencies (iodine, iron and vitamin A), inflammation, and malarial antigenemia using plasma or serum. Serum ferritin, an iron biomarker, cannot be measured from DBS due to red blood cell (RBC) ferritin content confounding the results. In this study, we assess a simple blood fractionation tool that passively separates plasma from other blood components via diffusion through a membrane into a plasma collection disc (PCD). We evaluated the concordance of MN 7-plex analyte concentrations from matched panels of eighty-eight samples of PCD, DBS, and wet plasma prepared from anticoagulated venous whole blood. The results showed good correlations of >0.93 between the eluates from PCD and DBS for each analyte except ferritin; while correlations seen for plasma/PCD were weaker. However, the recovery rate of the analytes from the PCD were better than those from DBS. The serum ferritin measures from the PCD were highly correlated to wet plasma samples (0.85). This suggests that surveillance for iron status in low resource settings can be improved over the current methods restricted to only measuring sTfR in DBS. When used in combination with the MN 7-plex, all seven biomarkers can be simultaneously measured using eluates from the PCDs.

Introduction

Micronutrient deficiency (MND) has been described as hidden hunger and an estimated 2 billion people are affected globally [1] with iodine, iron, and vitamin A representing three of the four primary MNDs of global significance [24]. Sufficient concentrations of key micronutrients (MN) are required for normal development of the fetus, infants, and young children. MND can impair cognitive, immune, and ocular function and physical health and development. Fetuses, young children, and women of reproductive age are particularly at risk from MND and the burden is greatest in low- and middle-income countries (LMICs) that are the least equipped to routinely assess MN status [5].

There are low cost and effective solutions to mitigate the effects of MND including dietary initiatives, fortification of common foodstuffs (e.g. salt or flour) or directly via supplementation [68]. The challenge to nutrition programs and researchers in LMICs is to effectively perform population surveillance to identify groups at most risk of MND before its effects manifest. Baseline data can identify the scale of MND and groups at greatest risk so that effective interventions can be introduced. Interventions can then be monitored via ongoing surveillance at scheduled intervals to establish their impact.

There is general consensus on a panel of blood-based biomarkers that are both informative on iodine, iron and vitamin A status [913], and practical for use in population-level surveillance. These act as surrogates to more direct MN measurement methods which involve challenging specimen collection (e.g. bone marrow aspirates for iron measurement [14]) or require highly skilled users and complex equipment (e.g. high pressure liquid chromatography for serum retinol analysis [15]). The blood specimens for these tests can be collected by venipuncture or less invasive methods, such as finger- or heel-stick capillary blood collection. Venous blood collection offers the advantage of much larger sample volumes and reduced chance of error resulting from the capillary blood collection technique (e.g. milking fingers), but venous samples are more difficult to collect, particularly from children, and more challenging to transport, process, and store [16]. Serum or plasma from whole blood must be rapidly processed to avoid hemolysis and must be kept under cold chain to prevent decay of the biomarkers. A perception of greater pain associated with venipuncture versus finger stick collection may also negatively influence participation in studies [17]. Blood and its derivatives pose biohazard risks, and handling should only be performed by staff with certified training for blood borne pathogens [18]. Therefore, the scaled collection of venous blood adds significant logistical, financial and technical challenges particularly in low resource settings where the need for effective MND surveillance is greatest.

The use of DBS as an alternative sample type meets many of the challenges presented with venous whole blood. Blood collection of up to 250 μL from a heel or finger prick via sterile lancet can be performed by less skilled users [19]. The blood is spotted on a paper card and left to dry before being stored in a sealed pouch with a desiccant to eliminate moisture. The preparation of a sample takes less time, needs no ancillary equipment (e.g. centrifuge or refrigeration) and uses fewer materials and consumables [16, 20, 21]. Once dried, the sample has greatly reduced biohazard risk during sample handling and processing. DBS cards have the further advantages of being relatively small, requiring significantly less space in freezers during long-term storage. As many biomarkers are stable in DBS at a broader range of temperatures, they can be shipped using gel packs instead of dry ice, or even out of cold chain [19, 20, 2224].

A primary challenge in using DBS is that the sample quantity is limited, with a typical collection yielding five spots of whole blood with approximately 50 μL each. Eluting DBS from filter paper results in dilution of the specimen, so assays must be capable of quantifying low concentrations. We recently highlighted that low concentration biomarkers such as thyroglobulin (Tg, μg/L) could be quantified from DBS eluates using the Q-Plex Human Micronutrient Array (Quansys Biosciences, Logan, Utah, USA; hereafter the MN 7-plex) [25, 26]. The MN 7-plex is a multiplexed immunoarray that enables the simultaneous quantification of biomarkers to deficiencies in iodine, iron, and vitamin A, in addition to inflammatory biomarkers and Plasmodium falciparum, from a single small volume of liquid serum or plasma [25].

While this evaluation found acceptable agreement between DBS and plasma for most analytes, using DBS for iron deficiency assessment is challenging because ferritin levels are grossly elevated by the co-elution of serum and red blood cell ferritin from lysed RBCs. Therefore, sTfR is the only iron biomarker that can be measured from DBS [26, 27]. While sTfR is one of the recommended biomarkers for iron status assessment, we and other groups have reported challenges in harmonizing absolute sTfR values derived from different immunoassay methods [2, 2529]. The ratio of sTfR to serum ferritin has been found to be a better indicator of iron deficiency than either analyte alone [2]. As such, we hypothesize that a parallel measurement of ferritin, inflammatory biomarkers and sTfR may give greater confidence in establishing iron status. To support this, we propose that devices developed to passively fractionate plasma from whole blood in austere settings, primarily for viral load testing of people living with HIV [19, 22], could be repurposed to include serum ferritin measurement and each of the other assays included on the MN 7-plex immunoarray.

In this manuscript we describe an evaluation of use of the MN 7-plex with matched sets of plasma, eluates of DBS, and eluates from a prototype of the plasma collection disc (PCD), a passive blood fractionation device [24]. In particular, we evaluate whether the PCD allows reliable measurement of ferritin using a method with the same logistical advantages as DBS.

Materials and methods

Collection of whole blood specimens

Specimens of ~8 mLs of human whole blood were procured from a commercial vendor, BioIVT (Westbury, New York). The WIRB protocol number for the collection of 80 blood samples was 20161665 and all specimens were obtained only after signed consent by each participant. The PATH Office of Research Affairs also deemed the use of these samples to be non-human subjects research.

Preparation of DBS, PCD and plasma panels

A panel of 80 heparinized anticoagulated whole blood samples (40 adult male, 40 adult female) was used for this study. The whole blood was delivered to the PATH laboratory the next day, and processed upon receipt to prepare three sample types (DBS, PCD, and plasma) within 24 hours of blood collection. Upon arrival in the laboratory, blood was continuously rotated at 4 °C before processing to ensure it remained thoroughly mixed. Matched sets of plasma, DBS and PCD were prepared in triplicate from the eighty blood specimens and the complete panels were measured in three different laboratories for α-acid glycoprotein (AGP), C-reactive protein (CRP), ferritin, histidine rich protein 2 (HRP2), retinol binding protein 4 (RBP4), soluble transferrin receptor (sTfR), and thyroglobulin (Tg) using the MN 7-plex. Because the blood panel was derived from US donors, it was not expected to include malarial parasitemia or a high prevalence of inflammation or iron deficiency [26]. Thus, twenty whole blood samples were randomly chosen from the set and a portion of the whole blood was spiked with calibrator solutions provided by Quansys to simulate samples positive for HRP2 and with elevated AGP, CRP, or sTfR levels. Each contrived sample was prepared by transferring 2 mL of whole blood to a new tube, adding spike solutions, and gently mixing by rotating tubes at 4 °C for at least 5 minutes (see Table 1). The resulting samples (Sample numbers 81–100) were then processed for DBS and PCD as described below (Fig 1).

thumbnail
Table 1. Contrived blood samples spiked with a range of elevated levels of AGP, CRP, HRP2 and sTfR.

https://doi.org/10.1371/journal.pone.0277835.t001

DBS were prepared from the 80 whole blood specimens and 20 spiked whole blood specimens by spotting 70 μL of whole blood onto Whatman 903 cards (n = 100). The cards were stored overnight to dry at room temperature and then individually stored at -80°C in sealed plastic pouches with desiccant packets.

Prototype PCDs were assembled from parts supplied by the manufacturer for use in these experiments. To produce the PCD, each absorbent plasma collection disc was loaded onto a custom-made manifold and covered by a separation membrane (Fig 2). A volume of 35 μL of whole blood was added to each PCD membrane using a wide-bore pipet tip. After approximately 90 minutes, the PCD membranes containing the RBCs and other cellular components were removed and discarded. The PCDs containing the adsorbed plasma were then removed from the manifold and stored in zipped plastic bags with desiccant at -80 °C. Samples 63–80 were omitted from this portion of the testing for lack of PCD units; the resulting final panel included PCDs from 62 unadulterated donor specimens and 20 PCDs from spiked specimens (total n = 82). Significant hemolysis was noted on one PCD (ID 12), and corresponding DBS spots appeared very pale. Some slight hemolysis above background was observed on PCDs for 13 samples (Samples 4, 5, 32, 37, 38, 50, 52, 83, 85, 86, 94, 96, and 100). Remaining whole blood was then processed to collect plasma fractions (n = 100) as previously described [27]. These plasma fractions were aliquoted into cryovials and stored at -80°C until use. The PCDs were eluted at PATH. The eluted material from the triplicate samples was pooled and mixed, and then aliquots of approximately 180 μL were sent to each laboratory on dry ice (see next section), along with matched DBS cards and plasma samples.

thumbnail
Fig 2. Schematic illustrating the plasma collection disc in operation.

The schematic illustrates a single PCD. Photos on the left show the PCD with the the filter disc placed on a manifold for multiple samples and the addition of 35 μL of whole blood to each disc (top), a PDC detail view prior to whole blood being added (middle), and the discs after blood has been absorbed and finally the plasma collection disc exposed after the whole blood treated filter has been removed (bottom). The disc is removed with sterile forceps and is ready for use or can be stored until required.

https://doi.org/10.1371/journal.pone.0277835.g002

Preparation of eluates from DBS and PCDs

The DBS and PCD samples were eluted in the Q-plex assay sample diluent before undergoing analysis with the MN 7-plex assay. DBS processing was carried out in each laboratory the day before assay. Two 6 mm DBS punches, equivalent to approximately 6 μL of serum [23], were placed in 244 μL sample diluent (approx. 1:20 dilution) and left overnight at 4°C. Samples were then shaken at 200 rpm on a microplate shaker for one hour at room temperature before using. PCD processing occurred in one laboratory, with aliquots of elutes then shared with all 3 laboratories. Each PCD containing an estimated equivalent of approximately 8 to 10 μL plasma per PCD was placed in a microcentrifuge tube separation column with a paper filter membrane (Pierce Spin Cups, ThermoFisher Scientific, Waltham, MA, USA) and 200 μL of diluent was added to each PCD, resulting in a dilution of approximately 1 in 20. Samples were incubated at room temperature for two hours before being centrifuged at 12,500 rpm for 15 minutes. Three PCDs per sample were eluted following this procedure, and recovered eluates were pooled for each sample, mixed and aliquoted in ~180 μL volumes for transfer to the laboratories for testing.

Human micronutrient 7-plex array procedure

The MN 7-plex protocol instructs the user to prepare all samples, negative control and calibrator dilutions in sample diluent containing reconstituted competitor. Because the competitor may degrade during the incubation time required for DBS or PCD elution, the standard protocol was modified to add competitor to the mix after sample elution. These protocol alterations were also used for wet plasma specimens included in these analyses.

Competitor was reconstituted in one tenth of the volume of diluent suggested in the MN 7-plex manual to produce a 10X competitor stock and 144 μL of eluate from either DBS or PCD was mixed with 16 μL of 10X assay competitor mix resulting in 1X competitor in a final sample dilution of approximately 1 in 22. Similarly, kit diluent without competitor was used to dilute wet plasma specimens to 1 in 20, and to reconstitute the calibrator provided with the kit and prepare the standard curve; 10X competitor mix was added to prepared solutions as 10% of the final volume. Then, 50 μL of each solution was added into the test wells of the MN 7-plex assay in duplicate, and the remainder of the protocol was carried out as directed in the kit manual.

After addition of the standards and samples, each plate was incubated at room temperature for two hours with shaking at 500 RPM using a flatbed shaker (Titertek Berthold, Huntsville, Alabama, USA). All reactions were aspirated and washed 3 times with 400 μL of wash buffer supplied with the MN 7-plex kit using an automatic plate washer. Next, 50 μL of detection mix was added to each well and the plate was then incubated with shaking at 500 RPM for 1 hour and washed one more time as described above. 50 μL streptavidin horseradish peroxidase solution was added to each well and plates were incubated with shaking for 20 minutes. Following another wash as described above, the chemiluminescent substrate parts A and B were mixed in equal volumes and 50 μL of the mixture was then added to each well.

Each plate was then imaged at 270 seconds of exposure time using a Quansys Q-View Imager LS (Quansys Biosciences). Q-View Software (Quansys Biosciences) was used to overlay a plate map onto the locations of analyte spots in each well and to measure the chemiluminescent signal from each spot in units of pixel intensity. The software applies the calibrator concentration values to the pixel intensities for each spot in the standard curve wells and fits 5 parameter logistic calibration curves for each analyte. The pixel intensities of the spots in each test well are then used to interpolate the concentration of each analyte relative to its calibrator curve. Once the plate image is overlaid with the analysis grid, all of the curve fitting and data reduction steps are automatically applied via the software. The upper and lower limits of quantification determined by Quansys for each kit lot were applied to exclude values beyond the assay concentration ranges. All values were adjusted for dilutions.

Adjusting volume estimates for the PCD

The original concentration estimates for the PCD eluates used the assumption that each disc contained approximately 10 μL of plasma, resulting in a dilution of 1 in 22 when the disc was eluted. However, recovery results suggested that the volume of plasma recovered from each disc is approximately 50% of the original estimate, giving an eluate dilution of 1 in 44. To account for this, the PCD results were doubled to more accurately reflect the eluted plasma volume.

Statistical methods

Reproducibility across the three laboratories was determined using coefficient of variation (CV, standard deviation divided by the mean) of the three measures for each sample type. The CVs were also used to assess the inter-assay and inter-laboratory variation. Concordance between AGP, CRP, ferritin, HRP2, RBP4, sTfR, and Tg results in plasma with both DBS (100 pairs) and PCD (82pairs) was evaluated using scatter plots, Lin’s concordance correlation coefficient (CCC) and Spearman’s correlation coefficient [30]. Agreement in absolute value was evaluated by calculating recovery of plasma values from DBS or PCD (DBS or PCD concentration/plasma concentration, expressed as a percentage) and using Bland Altman plots of the average concentration plotted against the percentage difference in concentrations of each analyte in plasma/DBS or PCD sample pair. Statistical analyses were conducted using Stata 15.1 (StataCorp, College Station, TX, USA). All of the data generated in this study can be publicly accessed at Dataverse (https://doi.org/10.7910/DVN/VY3MDB).

Results

Summary statistics for each analyte and sample type and CVs for the measures across the three participating laboratories are shown in Table 2. The CVs for plasma and DBS were similar, with higher CVs in the PCDs for most analytes. The Eurofins sTfR values were substantially greater than the other datasets and after review, this data was omitted from further analysis. The CVs were also high for HRP2, regardless of sample type, because of the semi-quantitative nature of the assay. Most specimens in the panel contained no detectable HRP2; spiked samples contained high values near or above the upper limit of detection for the assay. For ferritin, the DBS samples saturated the assay and were at the upper limit of detection, leading to high CVs. This was expected given the presence of red blood cell (RBC) associated ferritin present in this sample type. Hemolysis was a concern based on the assumption that these samples may contain RBC ferritin. However, the 14 PCD samples with observed hemolysis after the blood filtration step did not show any evidence of effects on the ferritin concentrations. Concentrations of all samples, spiked, and unspiked, were similar for plasma and DBS, but the PCD values were approximately half the values measured in plasma when dilution was calculated using the original estimate of plasma volume contained in the PCD. Results were adjusted twofold to account for the new plasma volume equivalent estimate. This gave improved agreement with plasma values in terms of absolute values but had minimal impacts on correlation (Fig 3A and 3B).

thumbnail
Fig 3.

A. Lin’s concordance correlation coefficient (CCC) plots comparing the analyte measurements in paired wet plasma (x-axes) and DBS samples (y-axes). AGP, α-1-acid glycoprotein; CRP, C-reactive protein; FER, ferritin; HRP2, histidine rich protein 2; RBP4, retinol binding protein 4; sTfR, soluble transferrin receptor; Tg, thyroglobulin; DBS, dried blood spot; rho_c, CCC. B, Lin’s CCC plots comparing the analyte measurements in paired wet plasma (x-axes) and the adjusted PCD (y-axes). AGP, α-1-acid glycoprotein; CRP, C-reactive protein; FER, ferritin; HRP2, histidine rich protein 2; RBP4, retinol binding protein 4; sTfR, soluble transferrin receptor; Tg, thyroglobulin; DBS, dried blood spot; rho_c, CCC.

https://doi.org/10.1371/journal.pone.0277835.g003

thumbnail
Table 2. Summary statistics and cross-laboratory coefficients of variation (CV) for plasma, DBS and PCD.

https://doi.org/10.1371/journal.pone.0277835.t002

The Spearman correlations between plasma and DBS and between plasma and PCDs are shown in Table 3. The Spearman correlation coefficients (Rs) between the plasma and the DBS samples were high (0.93 or higher), with the exception of ferritin. These results closely match previous observations using DBS samples with the 7-plex assay [22]. As expected, the correlation between plasma ferritin and DBS ferritin was lower (0.64) as DBS eluates contain a mixture of ferritin from serum and from red blood cells. The correlation of plasma ferritin to PCD ferritin was 0.925, highlighting the potential advantage of PCDs for ferritin measurement without the need to centrifuge samples and possibly for dry plasma samples. The correlation for RBP4 from PCD was lower than from DBS (0.993 versus 0.840). The sTfR correlation using the PATH and UW data only data was 0.858.

thumbnail
Table 3. The comparison of results for plasma, DBS and the volume adjusted PCD eluates using Rs and recovery of expected plasma values.

https://doi.org/10.1371/journal.pone.0277835.t003

To complete our comparative analysis, we also assessed the correlation of the DBS and the adjusted PCD samples to the plasma samples using scatter plots with linear regression lines (Fig 3A and 3B) and their agreement using Bland Altman plots (Fig 4A and 4B). We used Lin’s CCC to create scatter plots to compare the measurements [31]. AGP and Tg were both very close to the line of perfect concordance and sTfR is correlative though the reduced major axis is consistently greater than the line of perfect correlation. HRP2, CRP, and RBP4 are less concordant while bias increases showing greater variance at the higher analyte concentrations. The plot for ferritin shows the lowest correlation as is expected since DBS samples have ferritin contributed from RBCs. With the exception of the sTfR plot, all other DBS analytes tended to be lower than plasma values. For most analytes, the correlation of adjusted PCD and plasma was similar to that of DBS and plasma.

thumbnail
Fig 4.

A. Bland Altman plots of the difference in results derived from wet plasma versus DBS types measured using the 7-plex assay. Average of the wet plasma and dry sample type (x-axes) are plotted against the difference between measurements from the 7-Plex assay for paired wet plasma and dry sample types (y-axes). B. Bland Altman plots of the difference in results derived from wet plasma versus the adjusted PCD. For both Figures, the horizontal lines indicate line of perfect agreement (green), mean (purple), and ± 2standard deviations of the difference (red). AGP, α-1-acid glycoprotein; CRP, C-reactive protein; FER, ferritin; HRP2, histidine rich protein 2; RBP4, retinol binding protein 4; sTfR, soluble transferrin receptor; Tg, thyroglobulin; DBS, dried blood spot.

https://doi.org/10.1371/journal.pone.0277835.g004

Discussion

Serum ferritin is a good biomarker for the immunologic determination of iron status from blood. However, RBC ferritin can confound analysis via quantitative ELISA, so blood samples must be fractionated into serum or plasma soon after collection, which is labor intensive and challenging in austere settings. The primary goal of this study was to assess the potential for the non-centrifugal preparation of plasma using a passive blood fractionation tool for its subsequent measurement in a multiplexed immunoassay than can assess MN status, inflammation and malaria antigenemia. We successfully demonstrated that a prototype passive PCD tool could prepare samples of sufficient quality for serum ferritin measurement and that were highly concordant via Rs (0.925) using paired plasma samples.

Over 18% of the PCDs also had visible signs of hemolysis after processing but these did not noticeably affect the ferritin test data. We did observe a clear trend of lower-than-expected recovery values for all of the biomarkers within the PCD dataset. This discrepancy in volume likely arose because while a PCD disc can absorb 10 μL of plasma if directly added to it, the PCD cannot absorb the same volume of plasma via passive diffusion when wicking plasma from the whole blood sample. This was resolved after we adjusted the dilution factor to reflect an assumption that the plasma volume contained in a PCD is approximately 5 μL, and not 10 μL as originally estimated. Further recovery experiments are warranted to confirm the new PCD plasma volume estimate.

When comparing plasma with DBS samples, ferritin failed as expected, but the Rs for the other biomakers were 0.947 or higher. For recovery values, the DBS results were variable with AGP, HRP, and Tg being within <16% of the pooled plasma values. Slight under-recovery was seen with CRP (71%) and RBP4 (74%) and lastly, sTfR demonstrated significant over-recovery at 215%. Therefore, with the exception of ferritin and sTfR, the DBS data matches our previously reported observations [25, 26]. When comparing the PCD data with plasma, the Rs values for the PCD and plasma were not as correlative as seen with DBS with exception of ferritin (0.84). The poorest PCD correlation was for RBP4 at 0.76 whilst the DBS samples gave a near perfect Rs of 0.993. The best correlations from PCD derived analytes were for CRP and Tg at 0.96. It is unsurprising that the correlation between the plasma and DBS results were nearly identical as both have accurate volumes of plasma or blood applied. With the PCDs, the Rs were more discordant than for DBS because while precise volumes of blood were applied to the PCD filters, the exact volumes of plasma retained in the PCDs were unknown. HRP2 was recovered well from both DBS and the PCD eluates, suggesting that the malarial antigen assay is functional with either DBS or PCD samples. For recovery, the PCD data was generally improved versus the DBS after a correction factor was applied. One reason may be that the plasma stored in the PCD can readily diffuse into the buffer whereas with the DBS there is some natural attrition of the biomarkers during drying, retention of analytes in the fibrous dead space within the paper, and some irreversible binding of proteins to the nitrocellulose matrix. While our results are preliminary, they suggest use of PCDs may allow ferritin measurement in plasma filtrates, while also yielding results for other biomarkers similar to those from plasma generated by centrifugation.

It is evident from the mean recoveries that PCDs either contained ~50% less plasma or that a significant amount of material was retained within the disc. Therefore, a rigorous assessment of the volumetric capacity of the PCD when wicking from whole blood is warranted, essentially mimicking the efforts to assess the volume of material retained in DBS relative to sampling volumes [21]. For convenience, the specimens used in this study were derived from a healthy USA-based cohort and so are unlikley to have micronutient deficencies and have no exposure to P. falciparum malaria. Therefore future assessments with more representative populations from LMICs are necessary. While we added experimental rigor by performing the duplicate testing at three sites, we could not reconcile the sTfR values from being grossly elevated from one group wherein all of the other biomarkers remained in range. In general, the sTfR assay in the MN-7-plex array has performed the most poorly with plasma, DBS and PCD sample types [25, 26, 32]. We reported the poor performance of the sTfR immunoassay pair to the manufacturer, who now claim to have successfully improved performance with a new sTfR antibody pair. We are currently assessing the performance of the previous and new version of the MN 7-plex array to independently confirm this.

Our results highlight that there is generally good agreement between the sample types and that the measurement of seven of the analytes in the MN 7-plex array is feasible with a sample collection method that eliminates the need for equipment-based sample processing in the field and that may potentially minimize cold chain requirements with storage as a dried plasma collection card. Future efforts will be required to understand the stability of the biomarkers in the dried PCDs in their intended use for population-based micronutrient surveys in austere settings where access to refrigeration is absent. We aim to validate the performance of these tools in LMIC settings to better understand their performance and limitations. A passive blood fractionation tool to enable HIV viral load testing from dried plasma spots is commercially available (Cobas Plasma Separation card, Roche Diagnostics, Indianapolis, IN, USA) [33]. It remains to be seen if these can be succesfully incorporated with immunoassays such as the MN 7-plex but it highlights that other tools may exist to support effective population-based surveillance of micronutrient status in austere settings.

Acknowledgments

We greatly appreciate the collaborative support of ViveBio LLC (Alpharetta, GA, USA) in supplying prototype PCD devices and for their expert technical input. We would like to thank Dr Emily Smith (The George Washington University, DC, USA) and Dr Ken Brown (University of California Davis, CA, USA) for their critical thinking and advice. We would also like to thank Ms. Olivia Halas for her administrative support in preparing and reviewing the manuscript for submission.

References

  1. 1. World Health Organization. Worldwide prevalence of anaemia 1993–2005: WHO global database on anaemia. World Health Organization; 2008. https://apps.who.int/iris/handle/10665/43894.
  2. 2. Lynch S, Pfeiffer CM, Georgieff MK, Brittenham G, Fairweather-Tait S, Hurrell RF, et al. Biomarkers of Nutrition for Development (BOND)-Iron Review. JNutr. 2018;148: 1001S–1067S. pmid:29878148
  3. 3. Rohner F, Zimmermann M, Jooste P, Pandav C, Caldwell K, Raghavan R, et al. Biomarkers of nutrition for development—iodine review. JNutr. 2014;144: 1322S–1342S. pmid:24966410
  4. 4. Tanumihardjo SA, Russell RM, Stephensen CB, Gannon BM, Craft NE, Haskell MJ, et al. Biomarkers of Nutrition for Development (BOND)-Vitamin A Review. JNutr. 2016;146: 1816S–1848S. pmid:27511929
  5. 5. Bailey RL, West KP Jr, Black RE. The epidemiology of global micronutrient deficiencies. AnnNutrMetab. 2015;66 Suppl 2: 22–33. pmid:26045325
  6. 6. Abe SK, Balogun OO, Ota E, Takahashi K, Mori R. Supplementation with multiple micronutrients for breastfeeding women for improving outcomes for the mother and baby. CochraneDatabaseSystRev. 2016;2: CD010647. pmid:26887903
  7. 7. Cichon B, Fabiansen C, Iuel-Brockdorf AS, Yameogo CW, Ritz C, Christensen VB, et al. Impact of food supplements on hemoglobin, iron status, and inflammation in children with moderate acute malnutrition: a 2 x 2 x 3 factorial randomized trial in Burkina Faso. AmJClinNutr. 2018;107: 278–286.
  8. 8. Wirth JP, Petry N, Tanumihardjo SA, Rogers LM, McLean E, Greig A, et al. Vitamin A Supplementation Programs and Country-Level Evidence of Vitamin A Deficiency. Nutrients. 2017;9. pmid:28245571
  9. 9. Beguin Y. Soluble transferrin receptor for the evaluation of erythropoiesis and iron status. Clin ChimActa. 2003;329: 9–22. pmid:12589962
  10. 10. Almekinder J, Manda W, Soko D, Lan Y, Hoover DR, Semba RD. Evaluation of plasma retinol-binding protein as a surrogate measure for plasma retinol concentrations. ScandJClinLab Invest. 2000;60: 199–203. pmid:10885491
  11. 11. Guagnozzi D, Severi C, Ialongo P, Viscido A, Patrizi F, Testino G, et al. Ferritin as a simple indicator of iron deficiency in anemic IBD patients. InflammBowelDis. 2006;12: 150–151. pmid:16432381
  12. 12. Kifude CM, Rajasekariah HG, Sullivan DJ Jr, Stewart VA, Angov E, Martin SK, et al. Enzyme-linked immunosorbent assay for detection of Plasmodium falciparum histidine-rich protein 2 in blood, plasma, and serum. ClinVaccine Immunol. 2008;15: 1012–1018.
  13. 13. Stinca S, Andersson M, Weibel S, Herter-Aeberli I, Fingerhut R, Gowachirapant S, et al. Dried Blood Spot Thyroglobulin as a Biomarker of Iodine Status in Pregnant Women. JClinEndocrinolMetab. 2017;102: 23–32. pmid:27732337
  14. 14. GALE E, TORRANCE J, BOTHWELL T. The quantitative estimation of total iron stores in human bone marrow. JClin Invest. 1963;42: 1076–1082. pmid:13946227
  15. 15. De Ruyter MG, De Leenheer AP. Determination of serum retinol (vitamin A) by high-speed liquid chromatography. ClinChem. 1976;22: 1593–1595. pmid:975502
  16. 16. Ostler MW, Porter JH, Buxton OM. Dried blood spot collection of health biomarkers to maximize participation in population studies. JVisExp. 2014; e50973. pmid:24513728
  17. 17. Daneau G, Gous N, Scott L, Potgieter J, Kestens L, Stevens W. Human Immunodeficiency Virus (HIV)-Infected Patients Accept Finger Stick Blood Collection for Point-Of-Care CD4 Testing. PLoSOne. 2016;11: e0161891. pmid:27556894
  18. 18. Bahadori M, Sadigh G. Occupational exposure to blood and body fluids. IntJOccupEnvironMed. 2010;1: 1–10. pmid:23022776
  19. 19. Mukherjee S, Kang TG, Chen Y, Kim S. Plasma separation from blood: the “lab-on-a-chip” approach. Crit RevBiomedEng. 2009;37: 517–529. pmid:20565382
  20. 20. Brindle E, Fujita M, Shofer J, O’Connor KA. Serum, plasma, and dried blood spot high-sensitivity C-reactive protein enzyme immunoassay for population research. JImmunolMethods. 2010;362: 112–120.
  21. 21. Mei JV, Alexander JR, Adam BW, Hannon WH. Use of filter paper for the collection and analysis of human whole blood specimens. JNutr. 2001;131: 1631S–1636S. pmid:11340130
  22. 22. Wang S, Sarenac D, Chen MH, Huang SH, Giguel FF, Kuritzkes DR, et al. Simple filter microchip for rapid separation of plasma and viruses from whole blood. IntJNanomedicine. 2012;7: 5019–5028. pmid:23055720
  23. 23. Hall EM, Flores SR, De JV. Influence of Hematocrit and Total-Spot Volume on Performance Characteristics of Dried Blood Spots for Newborn Screening. IntJNeonatal Screen. 2015;1: 69–78. pmid:28868497
  24. 24. Arredondo FX, Craft NE, Fleshman MK. Screening of Dried Matrix Spot Collection Devices to be Used for Nutritional Biomarker Analysis. The FASEB Journal. 2017;31_Supplment. Experimental Biology 2017 Meeting Abstracts.
  25. 25. Brindle E, Lillis L, Barney R, Hess SY, Wessells KR, Ouedraogo CT, et al. Simultaneous assessment of iodine, iron, vitamin A, malarial antigenemia, and inflammation status biomarkers via a multiplex immunoassay method on a population of pregnant women from Niger. PLoSOne. 2017;12: e0185868.
  26. 26. Brindle E, Lillis L, Barney R, Bansil P, Lyman C, Boyle DS. Measurement of micronutrient deficiency associated biomarkers in dried blood spots using a multiplexed immunoarray. PLoSOne. 2019;14: e0210212. pmid:30620768
  27. 27. Brindle E, Stevens D, Crudder C, Levin CE, Garrett D, Lyman C, et al. A multiplex immunoassay method for simultaneous quantification of iron, vitamin A and inflammation status markers. PLoSOne. 2014;9: e115164. pmid:25525806
  28. 28. Thorpe SJ, Walker D, Arosio P, Heath A, Cook JD, Worwood M. International collaborative study to evaluate a recombinant L ferritin preparation as an International Standard. Clin Chem. 1997;43: 1582–1587. pmid:9299937
  29. 29. Wians FH Jr, Urban JE, Kroft SH, Keffer JH. Soluble transferrin receptor (sTfR) concentration quantified using two sTfR kits: analytical and clinical performance characteristics. Clin ChimActa. 2001;303: 75–81. pmid:11163026
  30. 30. Lin L, Hedayat AS, Wu W. A unified approach for assessing agreement for continuous and categorical data. J Biopharm Stat. 2007;17: 629–652. pmid:17613645
  31. 31. Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45: 255–268. pmid:2720055
  32. 32. Brindle E, Lillis L, Barney R, Bansil P, Hess SY, Wessells KR, et al. A multicenter analytical performance evaluation of a multiplexed immunoarray for the simultaneous measurement of biomarkers of micronutrient deficiency, inflammation and malarial antigenemia. PLoS One. 2021;16: e0259509. pmid:34735520
  33. 33. Carmona S, Seiverth B, Magubane D, Hans L, Hoppler M. Separation of Plasma from Whole Blood by Use of the cobas Plasma Separation Card: a Compelling Alternative to Dried Blood Spots for Quantification of HIV-1 Viral Load. J Clin Microbiol. 2019;57: e01336–18. pmid:30728197