Differences in the gray-to-white matter ratio according to different computed tomography scanners for outcome prediction in post-cardiac arrest patients receiving target temperature management

  • Jae Hun Oh,

    Roles Data curation, Formal analysis, Investigation, Writing – original draft, Writing – review & editing

    Affiliation Department of Emergency medicine, College of Medicine, The Catholic University of Korea, Eunpyeong St. Mary’s Hospital, Seoul, Republic of Korea

  • Seung Pill Choi,

    Roles Conceptualization, Data curation, Methodology, Writing – review & editing

    Affiliation Department of Emergency medicine, College of Medicine, The Catholic University of Korea, Eunpyeong St. Mary’s Hospital, Seoul, Republic of Korea

  • Jong Ho Zhu,

    Roles Resources

    Affiliation Department of Emergency medicine, College of Medicine, The Catholic University of Korea, Eunpyeong St. Mary’s Hospital, Seoul, Republic of Korea

  • Soo Hyun Kim ,

    Roles Conceptualization, Resources, Supervision, Writing – review & editing

    unidgirl@catholic.ac.kr

    Affiliation Department of Emergency medicine, College of Medicine, The Catholic University of Korea, Eunpyeong St. Mary’s Hospital, Seoul, Republic of Korea

  • Kyu Nam Park,

    Roles Supervision

    Affiliation Department of Emergency medicine, College of Medicine, The Catholic University of Korea, Seoul St. Mary’s Hospital, Seoul, Republic of Korea

  • Chun Song Youn,

    Roles Formal analysis, Writing – review & editing

    Affiliation Department of Emergency medicine, College of Medicine, The Catholic University of Korea, Seoul St. Mary’s Hospital, Seoul, Republic of Korea

  • Sang Hoon Oh,

    Roles Writing – original draft, Writing – review & editing

    Affiliation Department of Emergency medicine, College of Medicine, The Catholic University of Korea, Seoul St. Mary’s Hospital, Seoul, Republic of Korea

  • Han Joon Kim,

    Roles Formal analysis

    Affiliation Department of Emergency medicine, College of Medicine, The Catholic University of Korea, Seoul St. Mary’s Hospital, Seoul, Republic of Korea

  • Sang Hyun Park

    Roles Data curation, Supervision

    Affiliation Department of Emergency medicine, College of Medicine, The Catholic University of Korea, Yeouido St. Mary’s Hospital, Seoul, Republic of Korea

Abstract

The gray-to-white matter ratio (GWR) has been used to identify brain damage in comatose patients after cardiac arrest. However, Hounsfield units (HUs), the measurement of brain density on computed tomography (CT) images, may vary depending on the machine type or parameter. Therefore, differences in CT scanners may affect the GWR in post-cardiac arrest patients. We performed a retrospective study on comatose post-cardiac arrest patients who visited the hospital from 2007 to 2017. Two CT, Lightspeed and SOMATOM, scanners were used. Two observers independently measured the HUs of the caudate nucleus, putamen, posterior internal capsule, and corpus callosum using regions of interest. We compared the GWR calculated from the HUs measured at different CT scanners. The analysis of different scanners showed statistically significant differences in the measured HUs and GWR. The HUs and GWR of Lightspeed were measured lower than SOMATOM. The difference between the two CT scanners was also evident in groups divided by neurological prognosis. The area under the curve of the receiver operating characteristic curve to predict poor outcomes of Lightspeed was 0.798, and the cut-off value for 100% specificity was 1.172. The SOMATOM was 0.855, and the cut-off value was 1.269. The difference in scanners affects measurements and performance characteristics of the GWR in post-cardiac arrest patients. Therefore, when applying the results of the GWR study to clinical practice, reference values for each device should be presented, and an integrated plan should be prepared.

Introduction

Target temperature management (TTM) is commonly applied to patients who remain comatose following resuscitation after cardiac arrest since it has been shown to improve survival and neurological outcomes [1, 2]. Various tools for evaluating and predicting neurological prognosis following cardiac arrest have been studied, from biomarker to imaging [2]. Electroencephalogram analysis, serum neuron-specific enolase, somatosensory evoked potentials, brain diffusion magnetic resonance imaging (MRI), and the brainstem response have been presented as prognostic predictors in recent studies [2]. Another prognostic factor, the gray-to-white matter ratio (GWR), began to be used as a prognostic factor for comatose post-cardiac arrest patients after Torbey et al. quantitatively measured the difference between gray matter (GM) and white matter (WM) and has since been studied in various ways by other researchers [310].

In the event of cardiac arrest, the patient is placed in a deep hypoxic state. Hypoxia causes lactic acidosis, and a reduced pH causes brain swelling. In computed tomography (CT) images, brain swelling is difficult to distinguish the central sulcus or cisterns, and the ventricle size seems to have decreased [1116]. In addition, this ischemic change appears to reduce the differentiation between GM and WM. The GWR can present patient prognosis prediction criteria by representing qualitative interpretations in quantitative value.

Some studies were conducted on a single scanner with various CT scanner settings, some were conducted with various types of machines, and others provided no details on scanner type or settings [310]. Additionally, brain CT images are affected by several parameters, such as voltage, current, and software version [17]. The values obtained from these different factors are likely to show a clear difference [17, 18]. Therefore, it is necessary to assume that there is no intermachine variability when analyzing the results obtained from different machines and settings. However, to our knowledge, whether these inconsistencies lead to differences in the clinically has never been analyzed.

Researchers have analyzed the brain CT images of normal adults and proved that the Hounsfield units (HUs) values of images obtained from three different machines are different. It was also concluded that the GWR is also affected by inter-scanner variability [19]. The purpose of this study was to analyze differences between the HUs and GWR of two machines in patients who received TTM after recovering from cardiac arrest and understand what these differences could indicate with regards to predicting neurological outcomes.

Materials and methods

Subjects

After deliberation, this study was approved by the Local Ethics Committee of the Catholic University of Korea, Yeouido St. Mary’s Hospital (SC18RESI0068). The medical records of patients who received TTM were reviewed from April 2007 to July 2017. The subjects included in the analysis were adults (≥ 18 years old) patients who received TTM after cardiopulmonary resuscitation and survived until discharge. Patients were excluded if they did not have a brain CT scan performed, had a CT with a contrast agent, died during TTM, suffered structural changes in the brain due to trauma, surgery or cerebrovascular disease, were transferred from other hospitals, or had too many artifacts caused by associated implants or accessories. A total of 86 patients were identified from medical records. Patients were evaluated for age, sex, witnessed the event, initial cardiac rhythm, initial Glasgow coma scale, location of cardiac arrest (in-hospital or out-hospital), the time between CT and arrest, and neurologic outcome at discharge.

This hospital used two CT scanners (CT 1; Lightspeed VCT, GE Healthcare, UK; and CT 2; SOMATOM Definition Flash, Siemens Healthcare, Germany). This study was retrospective, so each CT scanner’s setting channel, current, and voltage were not performed identically. TTM was maintained for 24 hours by lowering the body temperature to 33.0 ± 1°C using an external cooling device with a standardized protocol. Rewarming was performed slowly for more than eight hours until reaching normal body temperature. There were no patients who received extracorporeal membrane oxygenation treatment. Patients who received all treatments were hospitalized in the intensive care unit and received standard intensive care and monitoring, including mechanical ventilation, central venous catheters, and alternative catheters. Patients received brain CT scans as soon as possible when vital signs had stabilized after ROSC. In addition, all patients used a dedicated CT scanner installed in the emergency room. The neurological outcome of the patients was evaluated by the cerebral performance category (CPC) at the time of discharge, and 1 and 2 were defined as good outcomes. The CPC was evaluated by experts at least 72 hours after the ROSC and at discharge.

Density measurements from the image

Two investigators analyzed images of patients. Investigator 1 is an emergency medicine specialist with over twenty years of experience, and investigator 2 is an emergency medicine specialist with eight years of experience. A radiologist trained the two researchers in related research knowledge for this study. Images were analyzed using commercial image viewing software (Marotech; M-view) that had blinded patient information. The level corresponding to the basal ganglia (BG) was determined by investigator 1, and both investigators measured the images of all patients. The two researchers did not share the results of the measurements, and the measurements were taken independently of each other. The density value was measured at 0.1 cm2 wide, and the average HUs value of the area with a circular region of interest (ROI) was used for analysis. The HUs of the caudate nucleus (CN), putamen (PU), posterior internal capsule (PIC), and corpus callosum (CC) on both sides were measured. As in previous studies, to increase the objectivity of the inspection, color mapping can be implemented in the image viewing software, making each area easier to visualize [19]. The area was divided into imaginary lines made using several landmarks, and the average value was defined as the HUs of the area after measurement [19]. We averaged nine values by dividing three sections per area and measuring three ROIs for each section. The mean HUs value was calculated from the CN/PIC, PU/PIC, CN/CC, and PU/CC between single areas. The GWR-AVE, including all values, was defined as (CN+PU)/(PIC+CC).

Statistical analysis

Categorical variables are expressed as numbers and percentages. Continuous variables are expressed as the mean ± standard deviation or median and interquartile range (IQR). Chi-squared and Fisher’s exact tests were used to analyze categorical variables. Continuous variables were tested for normality through the Shapiro-Wilk test. Most continuous variables did not follow a normal distribution, so the average value between the two CT groups was assessed with the nonparametric Mann-Whitney U test. The differences between each group were compared with box and whisker plots. Interclass correlation coefficients (ICCs) were analyzed to evaluate the consistency of measurements between the two investigators. A receiver operating characteristic (ROC) curve was created to obtain a cut-off value for predicting poor outcomes. The cut-off values at 100% specificity were analyzed and compared. Two-sided p-values less than 0.05 were considered statistically significant. Statistical analysis was performed with SPSS ver. 20 (IBM Corp., Armonk, NY, USA) and MedCalc ver. 15.2.2 (MedCalc Software, Mariakerke, Belgium).

Results

Eighty-six patients were enrolled. Forty-three patients were classified into the Lightspeed group, and others were classified into the SOMATOM group (Fig 1). The median time from cardiac arrest to the CT scan was 1.2 hours (IQR, 0.8–2.6 hours), the mean patient age was 56.9 years (SD, ±16.9), and there were twenty-eight (33%) women. Fifty-five patients (64%) experienced witnessed cardiac arrest, and twenty-seven (31%) had a good neurologic outcome. All of the baseline variables showed no statistically significant differences between CT groups (Table 1).

thumbnail
Fig 1. Flow diagram of the study population.

CT, computer tomography; TTM, target temperature management; CVA, cerebrovascular accident; ICH, intracranial hemorrhage; CT 1, LightSpeed VCT; CT 2, SOMATOM definition flash.

https://doi.org/10.1371/journal.pone.0258480.g001

There was a clear difference in the density of GM and WM between the two CT groups. The GWR-AVE, regardless of the CT scanner used, was 1.256 (IQR, 0.938–1.401) that of the Lightspeed scanner was 1.217 (IQR, 0.938–1.311), and that of the SOMATOM scanner was 1.295 (IQR, 1.169–1.401). Similarly, GWR in different areas all showed similar results (Table 2).

thumbnail
Table 2. The density of measured HUs value and GWR according to CT scanner.

https://doi.org/10.1371/journal.pone.0258480.t002

The analysis of neurological outcomes clearly showed a difference between scanners. In the good outcome analysis, the GWR-AVE of the Lightspeed scanner was 1.266 (IQR, 1.240–1.275), and that of the SOMATOM scanner was 1.328 (IQR, 1.312–1.349). In the poor outcome analysis, the Lightspeed scanner was 1.220 (IQR, 1.177–1.242), and the SOMATOM scanner was 1.285 (IQR, 1.236–1.307) (Table 3 and Fig 2). The cut-off means a value with 100% specificity in the ROC curve analysis that predicts the poor neurological outcome of GWR. The cut-off value of the GWR also showed different results depending on the scanner (Table 4). SOMATOM CT scanner showed a higher cut-off value than the Lightspeed CT scanner. The cut-off value for the entire patient matched the cut-off value for the Lightspeed CT scanner. The Cronbach α of the two researchers was 0.904, with ICCs being excellent.

thumbnail
Fig 2. Box-plot is showing the comparison of GWR between each neurologic outcome group.

GWR, gray-to-white ratio; GO, good neurologic outcome; PO, poor neurologic outcome.

https://doi.org/10.1371/journal.pone.0258480.g002

thumbnail
Table 3. Differences in the neurological outcome and GWR for each CT scanner.

https://doi.org/10.1371/journal.pone.0258480.t003

thumbnail
Table 4. AUC value and a cut-off value of GWR for each CT scanner.

https://doi.org/10.1371/journal.pone.0258480.t004

Discussion

The authors used color mapping and sectioning methods as in previous studies to minimize errors in the HUs measurement by ROI [19]. The two methods were used because ROI-based analysis uses the average value of the measurement range. Hence, the value changes even if the measurement of the investigator is slightly off [3, 20]. The color mapping method makes the boundary more clear. Additionally, it is not appropriate to measure the ROI in a single area because the HUs value of the measurement zone is not homogeneous. The sectioning method was used to measure three ROI in each zone after each measurement area was divided into three sections, and the average of the nine values obtained was defined as the HUs value in that area [19]. In this study, two researchers measured the ROI by this method, and the ICCs were excellent (Fig 3).

thumbnail
Fig 3. Bland-altman plots of interclass correlation coefficients between investigators.

https://doi.org/10.1371/journal.pone.0258480.g003

Most previous studies have been retrospective in which only one type of scanner was used. However, there are many kinds of machines used in multi-center studies, and the parameters differ. In a recent study, Gentsch et al. determined a simplified measurement method for the GWR, and the authors commented that three types of CT scanners were used and that the results showed variability [9]. In addition, the radiologist pointed out the variation of scanners and parameters in the commentary for the study analyzing the GWR and neurologic outcome of cardiac arrest patients who performed extracorporeal membrane oxygenation [17]. In this study, different scanners also showed the results of different GWR. In particular, the absolute differences in the GWR-AVE were 0.062 and 0.065 for good and poor neurological outcome, respectively.

These differences can also be misjudged when clinically predicting poor neurologic outcome. When the SOMATOM scanner was applied with a Lightspeed cut-off, the specificity remained 100%, but the sensitivity was reduced. Conversely, applying the cut-off of Somatom to the Lightspeed scanner will predict good neurologic outcome as poor neurologic outcome.

Recently, a multi-center study by Korean Hypothermia Network (KORHN) investigators of PCAS patients receiving TTM showed that the GWR did not correlate with neurologic outcomes [21]. In addition, the poor neurologic outcome prediction sensitivity of the GWRbasal ganglia measured with brain CT obtained within one hour of ROSC showed 3.3% at 100% specificity. In contrast, the prediction sensitivity of poor neurologic outcomes in our study, which showed the GWR at the level of the BG at 100% specificity in the CT1 and CT2 groups, was 24.2% and 42.3%, respectively. Furthermore, regardless of the CT scanner used, the sensitivity decreased to 18.6% when all patients were analyzed. In addition, the cut-off value at 100% specificity also differed by 0.097. It matched the value of the Lightspeed scanner, with a low value for the entire patient population regardless of the scanner used. Analyzing data from a multi-center study without considering the scanner used may result in lower sensitivity because GWR converges to the machine’s cut-off value with the lowest cut-off value.

The recent neurological outcome evaluation of patients resuscitated after cardiac arrest involves multimodal approaches, and the GWR is one of them. However, recent GWR studies have tended to report negative results. However, considering the results of this study, the neurologic outcome prediction power of GWR may have been evaluated less than the actual GWR. In addition, recent imaging techniques, such as MRI or CT, undergo a normalization process to compensate for differences in equipment [18, 2225]. To our knowledge, no study has reported normalization for GWR measurements. This study has several limitations. First, as it was a retrospective study, the same parameters (voltage, thickness, current) of the two CT scans could not be applied. However, all raw data were applied with consistent protocols for each CT. Therefore, the results of this study may reflect the parameter difference. Nevertheless, the results of this study mean that the GWR of the scanner under different conditions may be different. Second, we did not compare the differences between each scanner in the same patient. However, such research design is impossible due to radiation hazards and ethical issues. Third, the purpose of early brain CT scan in cardiac arrest patients is to determine the cause of arrest and determine if it corresponds to an indication of TTM. In addition, recent studies suggest that the GWR of delayed brain CT is suitable for neuroprognostication. In this study, the brain CT scanned after TTM was not analyzed. Fourth, the small sample sizes of single hospitals do not allow the generalization of these results. There is also a potential difference between CT groups in etiology. Therefore, it is necessary to confirm these results in well-designed prospective multi-center studies. Lastly, we did not consider other factors affecting GWR, such as patient age and CT scan time.

Conclusions

This study showed that inter-scanner variability might be observed in post-cardiac arrest patients who received TTM, and the GWR could vary. Therefore, future research on the GWR may need to incorporate a normalization method to correct for differences between machines.

References

  1. 1. Callaway CW, Donnino MW, Fink EL, Geocadin RG, Golan E, Kern KB, et al. Part 8: Post-Cardiac Arrest Care: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2015;132(18 Suppl 2):S465–82. Epub 2015/10/17. pmid:26472996.
  2. 2. Fugate JE, Wijdicks EF, Mandrekar J, Claassen DO, Manno EM, White RD, et al. Predictors of neurologic outcome in hypothermia after cardiac arrest. Ann Neurol. 2010;68(6):907–14. Epub 2010/11/10. pmid:21061401.
  3. 3. Torbey MT, Selim M, Knorr J, Bigelow C, Recht L. Quantitative analysis of the loss of distinction between gray and white matter in comatose patients after cardiac arrest. Stroke. 2000;31(9):2163–7. Epub 2000/09/08. pmid:10978046.
  4. 4. Weinstein MA, Duchesneau PM, MacIntyre WJ. White and gray matter of the brain differentiated by computed tomography. Radiology. 1977;122(3):699–702. Epub 1977/03/01. pmid:841054.
  5. 5. Scheel M, Storm C, Gentsch A, Nee J, Luckenbach F, Ploner CJ, et al. The prognostic value of gray-white-matter ratio in cardiac arrest patients treated with hypothermia. Scand J Trauma Resusc Emerg Med. 2013;21:23. Epub 2013/04/10. pmid:23566292.
  6. 6. Lee YH, Oh YT, Ahn HC, Kim HS, Han SJ, Lee JJ, et al. The prognostic value of the grey-to-white matter ratio in cardiac arrest patients treated with extracorporeal membrane oxygenation. Resuscitation. 2016;99:50–5. Epub 2015/12/23. pmid:26690647.
  7. 7. Lee BK, Kim WY, Shin J, Oh JS, Wee JH, Cha KC, et al. Prognostic value of gray matter to white matter ratio in hypoxic and non-hypoxic cardiac arrest with non-cardiac etiology. Am J Emerg Med. 2016;34(8):1583–8. Epub 2016/06/10. pmid:27278721.
  8. 8. Kim SH, Choi SP, Park KN, Youn CS, Oh SH, Choi SM. Early brain computed tomography findings are associated with outcome in patients treated with therapeutic hypothermia after out-of-hospital cardiac arrest. Scand J Trauma Resusc Emerg Med. 2013;21:57. Epub 2013/07/23. pmid:23870424.
  9. 9. Gentsch A, Storm C, Leithner C, Schroeder T, Ploner CJ, Hamm B, et al. Outcome prediction in patients after cardiac arrest: a simplified method for determination of gray-white matter ratio in cranial computed tomography. Clin Neuroradiol. 2015;25(1):49–54. Epub 2014/03/15. pmid:24626778.
  10. 10. Choi SP, Park HK, Park KN, Kim YM, Ahn KJ, Choi KH, et al. The density ratio of grey to white matter on computed tomography as an early predictor of vegetative state or death after cardiac arrest. Emerg Med J. 2008;25(10):666–9. Epub 2008/10/10. pmid:18843066.
  11. 11. Morimoto Y, Kemmotsu O, Kitami K, Matsubara I, Tedo I. Acute brain swelling after out-of-hospital cardiac arrest: pathogenesis and outcome. Crit Care Med. 1993;21(1):104–10. Epub 1993/01/01. pmid:8420715.
  12. 12. Shirota G, Gonoi W, Ishida M, Okuma H, Shintani Y, Abe H, et al. Brain Swelling and Loss of Gray and White Matter Differentiation in Human Postmortem Cases by Computed Tomography. PLoS One. 2015;10(11):e0143848. Epub 2015/12/01. pmid:26618492.
  13. 13. Yamamura H, Kaga S, Kaneda K, Yamamoto T, Mizobata Y. Head Computed Tomographic measurement as an early predictor of outcome in hypoxic-ischemic brain damage patients treated with hypothermia therapy. Scand J Trauma Resusc Emerg Med. 2013;21:37. Epub 2013/05/16. pmid:23672736.
  14. 14. Metter RB, Rittenberger JC, Guyette FX, Callaway CW. Association between a quantitative CT scan measure of brain edema and outcome after cardiac arrest. Resuscitation. 2011;82(9):1180–5. Epub 2011/05/20. pmid:21592642.
  15. 15. Greer DM, Wu O. Neuroimaging in Cardiac Arrest Prognostication. Semin Neurol. 2017;37(1):66–74. Epub 2017/02/02. pmid:28147420.
  16. 16. Cala LA, Thickbroom GW, Black JL, Collins DW, Mastaglia FL. Brain density and cerebrospinal fluid space size: CT of normal volunteers. AJNR Am J Neuroradiol. 1981;2(1):41–7. Epub 1981/01/01. pmid:6784549.
  17. 17. Chen YF, Chen YS. Is the grey-to-white ratio valuable enough in prediction for cardiac arrest patients rescued with extracorporeal resuscitation? J Thorac Dis. 2016;8(6):E469–70. Epub 2016/06/15. pmid:27294254.
  18. 18. Tsujita N, Kai N, Fujita Y, Hiai Y, Hirai T, Kitajima M, et al. Interimager variability in ADC measurement of the human brain. Magn Reson Med Sci. 2014;13(2):81–7. Epub 2014/04/29. pmid:24769632.
  19. 19. Oh JH, Choi SP, Wee JH, Park JH. Inter-scanner variability in Hounsfield unit measured by CT of the brain and effect on gray-to-white matter ratio. Am J Emerg Med. 2019;37(4):680–4. Epub 2018/07/19. pmid:30017694.
  20. 20. You YH, Park JS, Yoo IS, Min JH, Jeong WJ, Cho YC, et al. Usefulness of a quantitative analysis of the cerebrospinal fluid volume proportion in brain computed tomography for predicting neurological prognosis in cardiac arrest survivors who undergo target temperature management. J Crit Care. 2019;51:170–4. Epub 2019/03/05. pmid:30831551.
  21. 21. Hong JY, Lee DH, Oh JH, Lee SH, Choi YH, Kim SH, et al. Grey-white matter ratio measured using early unenhanced brain computed tomography shows no correlation with neurological outcomes in patients undergoing targeted temperature management after cardiac arrest. Resuscitation. 2019;140:161–9. Epub 2019/04/07. pmid:30953628.
  22. 22. Choi SP, Park KN, Park HK, Kim JY, Youn CS, Ahn KJ, et al. Diffusion-weighted magnetic resonance imaging for predicting the clinical outcome of comatose survivors after cardiac arrest: a cohort study. Crit Care. 2010;14(1):R17. Epub 2010/02/16. pmid:20152021.
  23. 23. Kim J, Kim K, Suh GJ, Kwon WY, Kim KS, Shin J, et al. Prognostication of cardiac arrest survivors using low apparent diffusion coefficient cluster volume. Resuscitation. 2016;100:18–24. Epub 2016/01/18. pmid:26774174.
  24. 24. Wijman CA, Mlynash M, Caulfield AF, Hsia AW, Eyngorn I, Bammer R, et al. Prognostic value of brain diffusion-weighted imaging after cardiac arrest. Ann Neurol. 2009;65(4):394–402. Epub 2009/04/29. pmid:19399889.
  25. 25. Velly L, Perlbarg V, Boulier T, Adam N, Delphine S, Luyt CE, et al. Use of brain diffusion tensor imaging for the prediction of long-term neurological outcomes in patients after cardiac arrest: a multicentre, international, prospective, observational, cohort study. Lancet Neurol. 2018;17(4):317–26. Epub 2018/03/04. pmid:29500154.