Development of a multiplex isothermal amplification molecular diagnosis method for on-site diagnosis of influenza

  • Woong Sik Jang ,

    Contributed equally to this work with: Woong Sik Jang, Da Hye Lim

    Roles Formal analysis, Writing – original draft

    Affiliation Department of Laboratory Medicine, Korea University Guro Hospital, Seoul, Republic of Korea

  • Da Hye Lim ,

    Contributed equally to this work with: Woong Sik Jang, Da Hye Lim

    Roles Data curation, Formal analysis

    Affiliation Department of Laboratory Medicine, Korea University Guro Hospital, Seoul, Republic of Korea

  • Jeonghun Nam,

    Roles Data curation, Formal analysis

    Affiliation Department of Laboratory Medicine, Korea University Guro Hospital, Seoul, Republic of Korea

  • Do-CiC Mihn,

    Roles Supervision

    Affiliation Department of Diagnostic Immunology, Seegene Medical Foundation, Seoul, Republic of Korea

  • Haan Woo Sung,

    Roles Supervision

    Affiliation Department of Veterinary Microbiology, College of Veterinary Medicine, Kangwon National University, Chuncheon, Republic of Korea

  • Chae Seung Lim,

    Roles Conceptualization, Project administration, Resources, Supervision, Writing – review & editing

    Affiliation Department of Laboratory Medicine, Korea University Guro Hospital, Seoul, Republic of Korea

  • Jeeyong Kim

    Roles Supervision, Writing – review & editing

    emperorjy@naver.com

    Affiliation Department of Laboratory Medicine, Korea University Guro Hospital, Seoul, Republic of Korea

Abstract

Influenza, which is an acute respiratory disease caused by the influenza virus, represents a worldwide public health and economic problem owing to the significant morbidity and mortality caused by its seasonal epidemics and pandemics. Sensitive and convenient methodologies for the detection of influenza viruses are important for clinical care and infection control as well as epidemiological investigations. Here, we developed a multiplex reverse transcription loop-mediated isothermal amplification (RT-LAMP) with quencher/fluorescence oligonucleotides connected by a 5′ backward loop (LF or LB) primer for the detection of two subtypes of influenza viruses: Influenza A (A/H1 and A/H3) and influenza B. The detection limits of the multiplex RT-LAMP assay were 103 copies and 102 copies of RNA for influenza A and influenza B, respectively. The sensitivities of the multiplex influenza A/B/IC RT-LAMP assay were 94.62% and 97.50% for influenza A and influenza B clinical samples, respectively. The specificities of the multiplex influenza A/B/IC RT-LAMP assay were 100% for influenza A, influenza B, and healthy clinical samples. In addition, the multiplex influenza A/B/IC RT-LAMP assay had no cross-reactivity with other respiratory viruses.

Introduction

Influenza, which is caused by the influenza virus, is a significant cause of morbidity and mortality that has major social and economic impacts throughout the world [1, 2]. Iuliano et al. reported that 291,243–645,832 seasonal influenza-associated respiratory deaths occur annually (4·0–8·8 per 100,000 individuals) [3]. Influenza viruses belong to the Orthomyxoviridae family and have a single-stranded segmented RNA genome consisting of 7–8 segments encoding 10–11 proteins [4]. The influenza viruses are classified into types A, B, C, and D on the basis of their core proteins [5]. Among the four influenza types, influenza A viruses cause most of the global flu epidemics, influenza B viruses cause smaller localized outbreaks, influenza C viruses generally cause mild illness, and influenza D viruses are not known to cause illness in people [68]. Thus, influenza A and B viruses comprise the major respiratory pathogens in humans that cause seasonal flu epidemics [9, 10].

The three kinds of therapeutic treatments for influenza A and B viruses include M2 inhibitors (amantadine and rimantadine), neuraminidase (NA) inhibitors (zanamivir and oseltamivir) and cap-dependent endonuclease (CEN) inhibitor (baloxavir) [1114]. NA inhibitors and CEN inhibitors have both inhibitory effects on influenza A and influenza B, but M2 inhibitors act specifically on influenza A [15, 16]. Since antiviral drugs can reduce the severity of illness when administered within 48 h of the first symptoms, distinguishing between influenza A and B is critical for choosing the appropriate antiviral drug [17, 18].

Currently, diagnosis methods for influenza viruses include the antigen antibody test, the hemagglutination inhibition test, enzyme immunoassay, microscopic diagnosis, and molecular diagnosis [1921]. However, serological tests have low sensitivity and specificity in comparison with PCR and PCR-based assays requiring expensive instruments, specialized technicians, and complicated procedures; thus, all these methods are unsuitable for rapid diagnosis in field situations [22, 23]. In 2000, loop-mediated isothermal amplification (LAMP) was developed to amplify genes at constant temperatures. The LAMP assay is a rapid, highly sensitive isothermal nucleic acid amplification through chain substitution reaction. LAMP amplifies the target gene at 60–65°C with six primers, including four primers selected by combining six parts of a target DNA strand and two additional loop primers. Bst DNA polymerase, which is a strand displacement DNA polymerase, is used in the LAMP assay to enable loop structure formation of the inner primers, producing LAMP’s unique rapid self-priming amplification [24, 25]. The LAMP assay has been widely applied for the detection of various pathogens [2628]. In particular, the reverse transcription LAMP (RT-LAMP) assay for RNA viruses is widely used for point-of-care testing because it does not require the standard RT reaction time [29].

For multiple detection using the LAMP assay, an assimilating probe consisting of the fluorescently tagged probe and its complementary sequence probe tagged with quencher was developed. This probe works by separating fluorescently-tagged oligonucleotides from the quencher-tagged probe. As a result, fluorescence is observed in real-time and measured from the fluorescently-tagged probe that has been incorporated into RT-LAMP products [30, 31].

In this study, we present a rapid multiplex RT-LAMP diagnosis method for influenza A, influenza B, and an internal control using a newly designed target-specific assimilating probe and fluorescently-tagged strand displaceable probes. This multiplex influenza A/B/IC RT-LAMP assay had high sensitivities for influenza A and influenza B without interference from one another. Moreover, the assay showed no cross-reactivity against other respiratory and hemorrhagic fever viruses.

Materials and methods

Clinical samples and RNA extraction

Nasopharyngeal (NP) swabs were collected from patients presenting flu-like symptoms at Korea University Guro Hospital from January 2018 to December 2018. A total of 314 NP specimens were used in this study, including 100 negative and 214 positive specimens of the following viruses: 11 influenza A/H1, 82 influenza A/H3, 80 influenza B, 4 RSV A, 4 RSV B, 4 adenovirus, 4 parainfluenzavirus (PIV1-4), 9 coronavirus (OC43, NL63, and 229E), 4 human bocavirus (HBoV), 4 human enterovirus (HEV), 4 human rhinovirus (HRV), and 4 metapneumovirus (MPV). All virus specimens were confirmed by polymerase chain reaction (PCR) using an Anyplex II RV16 Detection Kit (Seegene, Inc., Seoul, South Korea). In addition, H5, H7 and H9 subtypes of avian influenza viruses were provided by Department of Veterinary Microbiology, College of Veterinary Medicine, Kangwon National University School of Medicine, Korea. RNA extraction was performed with a QIAamp Viral RNA Mini Kit (Qiagen, Hilden, Germany) according to the manufacturer's manual. RNA was extracted from 140 μL of samples and stored at -50°C. All LAMP assays were performed blindly with the operator unaware of any previous test results. This study was approved by the Medical Ethics Committee of Korea University Guro Hospital (2019GR0055).

Primer design

Influenza A and B LAMP primer sets were designed within conserved regions of segment 7 of influenza A and the nucleoprotein gene of influenza B. For internal control, actin beta LAMP primer set was newly designed within of conserved regions of human actin beta mRNA (NM_001101.5:c.287-c.498), which is commonly used as internal control [32]. All LAMP primers, including two outer primers (forward primer F3 and backward primer B3), two inner primers (forward inner primer FIP and backward inner primer BIP), and two loop primers (forward loop primer LF and backward loop primer LB), were designed using Primer Explorer software (Version 4; Eiken Chemical Co., Tokyo, Japan). For the multiplex LAMP assay, we designed the fluorophore probe oligomer (32 mer) at the 5′ LF or LB primer and the quencher oligonucleotide (30 mer), which is the complementary sequence of the fluorophore probe oligomer, using Random DNA Sequence Generator (https://faculty.ucr.edu/~mmaduro/random.htm). All primers were assessed for specificity before use in the LAMP assays via a BLAST search of sequences in GenBank (National Center for Biotechnology Information [NCBI], Bethesda, MD). All LAMP primers and probes were synthesized by Macrogen, Inc. (Seoul, South Korea; Table 1).

thumbnail
Table 1. Loop-mediated isothermal amplification (LAMP) and Reverse Transcription Polymerase Chain Reaction (RT-PCR) primer sets used in this study.

https://doi.org/10.1371/journal.pone.0238615.t001

Multiplex influenza A/B/IC RT-LAMP assay

The influenza A/B/IC multiplex RT-LAMP assay was performed with a Mmiso RNA amplification kit (Mmonitor, South Korea). The RT-LAMP reaction was prepared with 12.5 μL of 2x reaction buffer, 1.25 μL of influenza A LAMP primer mix, 0.625 μL of influenza B LAMP primer mix, 0.625 μL of internal control LAMP primer mix, 720 nM quencher solution, 2 μL of enzyme mix, and 2.5 μL of sample RNA (final reaction volume: 25 μL). The composition of the LAMP primer mix (influenza A and influenza B) included 4 μM of two outer primers (F3 and B3), 32 μM of two inner primers (FIP and BIP), 10 μM of loopB primer, 4 μM of loopF primer, and 6 μM of loopF probe primer. The composition of the internal control LAMP primer mix included 4 μM of two outer primers (F3 and B3), 32 μM of two inner primers (FIP and BIP), 10 μM of loopF primer, 4 μM of loopB primer, and 6 μM of loopB probe primer. The RT-LAMP assay was run on the CFX 96 Touch Real-Time PCR Detection System (Bio-Rad Laboratories, Hercules, CA, USA) at 60°C for 60 min. In LAMP assay, negative control (human serum RNA and distilled water) were used for setting baseline. Positive signal was determined by checking whether signal is steep or gradual considering the baseline, because baseline of LAMP assay is not stable compared to qPCR or RT-PCR. The FAM, HEX and Texas red fluorescence channels was used for detecting Influenza A, influenza B and internal control (actin beta), respectively.

Real-time RT-PCR

To evaluate the performance of the multiplex influenza A/B/IC RT-LAMP assay, two real-time RT-PCRs, the commercial RealStar® Influenza RT-PCR Kit 2.0 (Altona Diagnostics, Hamburg, Germany) and the World Health Organization (WHO) influenza A/B primer set [33, 34] with the DiaStar OneStep Multiplex qRT-PCR Kit (SolGent Co., Ltd., Daejeon, Korea), were performed using the CFX96 Touch Real-Time PCR Detection System (Bio-Rad Laboratories). The PCR cycling conditions of the WHO influenza A/B primer set were as follows: reverse transcription at 50°C for 20 min, inactivation at 95°C for 15 min, 39 cycles of denaturation at 95°C for 30 s, and annealing with fluorescence detection at 60°C for 40 s. For the RealStar® Influenza RT-PCR Kit 2.0, the thermocycling parameters were as follows: reverse transcription at 55°C for 20 min, inactivation at 95°C for 2 min, 44 cycles of denaturation at 95°C for 15 s, annealing with fluorescence detection at 55°C for 45 s, and extension at 72°C for 15 s.

Limit of Detection (LOD) tests of the multiplex influenza A/B/IC RT-LAMP assay

pTOP Blunt V2 plasmids, including the segment 7 partial sequences of Influenza A, were used to test the LOD of the influenza A LAMP assay. For the LOD of the influenza B LAMP assay, pTOP Blunt V2 plasmids, including the nucleoprotein gene partial sequences of influenza B, were used. pTOP Blunt V2 plasmids, including the beta actin partial sequences of human, were used to test the LOD of the Internal control LAMP assay. All plasmids were constructed by Macrogen, Inc. The plasmids were serially diluted 10-fold from 1.0 × 108 copies/μL to 1.0 × 101 copies/μL to determine the LOD of the multiplex influenza A/B/IC RT-LAMP assay. In addition, the LOD of the multiplex influenza A/B/IC LAMP analysis was tested using clinical samples of influenza A/H1, A/H3, and B. Clinical specimens were diluted 10 times (as much as 10−1 to 10−6 times) based on the original samples. The LOD of the multiplex influenza A/B/IC RT-LAMP analysis for clinical influenza samples was compared to those of the WHO influenza A/B primer set and RealStar® Influenza RT-PCR Kit 2.0.

Results

Optimization of the multiplex influenza A/B/IC LAMP primer set

Before optimization of multiplex LAMP assay, each Influenza A, B and internal control (IC, actin beta) LAMP primer set were tested with clinical samples (Influenza A H1, H1N1, H3N2, Influenza B, human serum RNA and distilled water). All three LAMP primer sets (Influenza A, B and IC) showed the no cross-reactivity (S1 Fig).

For optimization of the multiplex influenza A/B/IC LAMP primer set, different concentration ratios of the influenza A, influenza B, and internal control primer sets (1:1:1, 1:1:0.5, and 1:0.5:0.5, respectively) were tested using human serum RNA samples spiked with influenza A and B plasmids (107 copy, ratio of 1:1) (Fig 1A). As a result, three signals (influenza A, influenza B, and internal control) were detected in the ratios of 1:0.5:0.5, whereas two signals (influenza B and internal control) were detected in the ratio of 1:1:1 and 1:1:0.5. Among three ratio of LAMP primer set, the ratio of 1:0.5:0.5 showed the lowest Ct values of influenza A and influenza B were 12.58 and 10.41, respectively. Therefore, the ratios of influenza A, influenza B, and internal control LAMP primer set (1:0.5:0.5) was determined as optimum ratio for multiplex influenza A/B/IC LAMP assay. Next, temperature gradient tests (58–65°C) were performed to determine the optimal temperature of the multiplex influenza A/B/IC LAMP assay (Fig 1B). Among 8 temperatures (65, 64.6, 62.5, 60.8, 59.4, 58.5 and 58.0°C), the LAMP assay showed lowest Ct values and RFU of all fluorescence channels at 60.8°C (Ct/RFU, influenza A: 15.13/3919, influenza B: 11.02/5237 and internal control: 16.51/2745) and 59.4°C (Ct/RFU, influenza A: 14.36/6996, influenza B: 11.08/5053 and internal control: 16.73/2707). Therefore, we decided the optimal temperature for this assay to be 60°C between 60.8°C and 59.4°C. Fig 1C showed the performance of the multiplex influenza A/B/IC LAMP assay against influenza A RNA samples, influenza B RNA samples, and influenza A/B RNA mixture samples at the optimum conditions (ratio of the influenza A, influenza B, and internal control primer sets was 1:0.5:0.5 at 60°C).

thumbnail
Fig 1. Optimization of the multiplex influenza A/B LAMP primer set.

(A) Different concentration ratios of the influenza A, influenza B, and internal control primer sets (1:1:1, 1:1:0.5, and 1:0.5:0.5, respectively) for human serum RNA (HS RNA) samples spiked with influenza A and B plasmids. (B) Temperature gradient tests (58–65°C) of the multiplex influenza A/B/IC LAMP assay. (C) Performance of the multiplex influenza A/B/IC LAMP assay against HS RNA spiked with influenza A plasmid, HS RNA spiked with influenza B plasmid, and HS RNA spiked with influenza A/B plasmid (1:1).

https://doi.org/10.1371/journal.pone.0238615.g001

LOD tests of the multiplex influenza A/B/IC multiplex LAMP assay

The analytical sensitivity of the multiplex influenza A/B/IC RT-LAMP assay was compared with monoplex LAMP primer sets [In A, In B and Internal control (IC, actin beta)] by testing synthetic RNA plasmids ranging from 108 to 101 RNA copies/μL (Fig 2, Table 2). In monoplex In A, In B and IC LAMP test showed all the detection limits of 1 × 102 copies/μL. In multiplex influenza A/B/IC RT-LAMP assay, both influenza A and B plasmids were detected up to 1 × 102 copies/μL and actin beta plasmid was detected up to 1 × 103 copies/μL. As a result, multiplex influenza A/B/IC showed comparable detection limits with monoplex LAMP assay, although detection limit of IC in multiplex was higher than that of IC monoplex LAMP. Furthermore, the LOD of the RT-LAMP assay was compared with the WHO RT-PCR primer set and commercial RealStar® Influenza RT-PCR Kit 2.0 using serially diluted clinical influenza A/H1, A/H3, and B samples (range of 10−1 to 10−6; Table 3). As a result, the LOD of both the WHO RT-PCR primer set and RealStar® Influenza RT-PCR Kit 2.0 for influenza A/H1 was 10−3 while that of the multiplex influenza A/B/IC RT-LAMP assay was 10−2. For influenza A/H3, the RealStar® Influenza RT-PCR Kit 2.0 assay showed the highest sensitivity (10−5), and the LOD of the other two assays was 10−4. For influenza B, the RealStar® Influenza RT-PCR Kit 2.0 assay showed the highest sensitivity (10−4), and the LOD of the other two assays was 10−3. Overall, the detection limits tested with clinical sample dilutions were the lowest in the analysis of the RealStar® Influenza RT-PCR Kit 2.0, followed by the WHO RT-PCR primer set, and finally the multiplex influenza A/B/IC RT-LAMP assay.

thumbnail
Fig 2. Limit of Detection (LOD) test for the monoplex and multiplex influenza A/B/IC LAMP primer set.

(A) The monoplex influenza A, influenza B and IC LAMP primer sets (left, middle and right panel, respectively). (B) Multiplex influenza A/B/IC LAMP primer set (left, middle and right panel, respectively). The monoplex and multiplex influenza LAMP assays were tested with synthetic influenza A, B or beta actin plasmids ranging from 108 to 101 RNA copies/μL. Colors (red, orange, green, blue, purple, pink, black, olive, sky blue) indicate plasmid copy numbers/μL (1.0 × 108 to 1.0 × 101 copies/μL) and DW (negative control). The LOD test was repeated three times.

https://doi.org/10.1371/journal.pone.0238615.g002

thumbnail
Table 2. Limit of Detection (LOD) test for the monoplex and multiplex influenza A/B/IC LAMP primer set.

https://doi.org/10.1371/journal.pone.0238615.t002

thumbnail
Table 3. Limit of Detection (LOD) tests of the multiplex influenza A/B/IC LAMP assay, World Health Organization (WHO) Influenza RT-PCR, and RealStar® Influenza RT-PCR Kit 2.0 for the clinical influenza A/B samples.

https://doi.org/10.1371/journal.pone.0238615.t003

Comparison of clinical performance between the multiplex influenza A/B/IC RT-LAMP assay, WHO RT-PCR primer set, and RealStar® Influenza RT-PCR Kit 2.0 using clinical samples

To confirm the clinical performance of the multiplex influenza A/B/IC RT-LAMP assay, the sensitivities and specificities of the assay were compared with those of the WHO RT-PCR primer set and RealStar® Influenza RT-PCR Kit 2.0 for 93 influenza A, 80 influenza B, and 100 healthy patient NP specimens (Table 4). For the influenza A H1 clinical samples (n = 11), the sensitivities of the WHO RT-qPCR primer set and RealStar® Influenza RT-PCR Kit 2.0 were 100% and 81.81%, respectively. The sensitivities of the multiplex influenza A/B/IC RT-LAMP assay were 90.90% in the influenza A channel (FAM) and 81.81% in the internal control channel. The specificities of all three assays for influenza A/H1 clinical samples were 100%. For influenza A/H3 clinical samples (n = 82), the sensitivities of the WHO RT-PCR primer set and RealStar® Influenza RT-PCR Kit 2.0 were 97.59% and 89.02%, respectively. The sensitivities of the multiplex influenza A/B/IC RT-LAMP assay were 95.12% in the influenza A channel (FAM) and 36.58% in the internal control channel. The specificities of the WHO RT-PCR primer set and multiplex influenza A/B/IC RT-LAMP assay for influenza A/H3 clinical samples were 100% while that of the RealStar® Influenza RT-PCR Kit 2.0 was 95.12%. Overall, the WHO RT-PCR primer set was found to have the highest sensitivity (97.84%) for all influenza A clinical samples (n = 93), followed by the multiplex influenza A/B/IC RT-LAMP assay (94.62%), and finally the RealStar® Influenza RT-PCR Kit 2.0 (88.17%). The specificities of the former two assays for influenza A clinical samples were 100% while that of the latter was 95.69%. In addition, the multiplex influenza A/B/IC RT-LAMP assay showed 100% of sensitivities and specificities for H5 (n = 10), H7 (n = 10) and H9 (n = 10) subtypes of avian influenza viruses (S1 Table). For influenza B clinical samples (n = 80), the multiplex influenza A/B/IC RT-LAMP assay and WHO RT-PCR primer set showed the highest sensitivities (97.50% and 97.50%, respectively) while that of the RealStar® Influenza RT-PCR Kit 2.0 was 86.25% (Table 4). The internal control channel of the multiplex influenza A/B/IC RT-LAMP assay showed 97.50% sensitivity for influenza B clinical samples. The specificities of all three assays against influenza B clinical samples were 100%. The specificities of the multiplex influenza A/B/IC RT-LAMP assay and WHO RT-PCR primer set for healthy clinical samples (non-infection; n = 100) were 100%, whereas that of the RealStar® Influenza RT-PCR Kit 2.0 was 99% (Table 4). The sensitivity of the internal control channel of the multiplex influenza A/B/IC RT-LAMP assay for healthy clinical samples was 100%.

thumbnail
Table 4. Comparison of clinical performance between the multiplex influenza A/B/IC RT-LAMP assay, WHO Influenza RT-PCR, and RealStar® Influenza RT-PCR Kit 2.0 for influenza A subtypes (H1N1 and H3N2) and influenza B viruses in the clinical samples.

https://doi.org/10.1371/journal.pone.0238615.t004

Cross-reactivity test

To confirm the possibility of cross-reactivity with other infectious viruses, 41 respiratory virus samples, including 4 RSV A, 4 RSV B, 4 adenovirus, 4 PIV(1–4), 9 coronavirus (OC43/HKU1, NL63, and 229E), 4 HBoV, 4 HEV, 4 HRV, and 4 MPV samples, were tested using the multiplex influenza A/B/IC RT-LAMP assay, WHO RT-PCR primer set, and RealStar® Influenza RT-PCR Kit 2.0 (Table 5) [33, 34]. All three molecular diagnostic tests showed no cross-reactivity with other infectious viruses, suggesting that these tests can specifically detect influenza viruses.

thumbnail
Table 5. Cross-reactivity of the multiplex influenza A/B/IC RT-LAMP assay against other human infectious viruses.

https://doi.org/10.1371/journal.pone.0238615.t005

Discussion

Since many respiratory viruses, including influenza viruses, can cause similar symptoms, it is difficult for clinicians to distinguish one virus from another [35]. Given the annual morbidity and mortality caused by influenza viruses, there is an urgent need for sensitive and convenient laboratory methods to identify influenza virus subtypes in clinical care and infection control [36, 37]. There are immunodiagnostic kits that can be utilized quickly, but their sensitivity and specificity are too poor; thus, PCR methods are currently used to make accurate diagnoses [38]. However, PCR-based target gene detection requires bulky and expensive equipment as well as highly skilled technicians [39]. Several isothermal amplification methods, such as HDA, RPA, and LAMP, have been developed for on-site diagnosis of infectious pathogens [4042]. Among them, LAMP is a promising method that has been utilized to detect a variety of pathogens [43, 44]. Recently, a variety of multiplex RT-LAMP methods have been developed to detect influenza viruses by using annealing temperature, nanoparticle hybridization, one-pot colorimetric visualization and immunochromatographic strip etc [4548]. However, their multiplex system amplified each type of influenza individually or consist of two steps, which are lamp assay and rapid test. Thus, it may take a more time when diagnosis a large number of clinical samples. Thus, it can take a more time to diagnose a large number of clinical samples. In addition, it is known that LAMP assay is easy to contaminate [49, 50]. In this study, we developed the one tube-multiplex influenza A/B/IC LAMP assay, including an internal control (actin), for the detection of influenza A/H1, A/H3, and B using newly designed assimilating probes, which has advantages for reducing test time and risk of contamination.

Our results showed that the multiplex influenza A/B/IC RT-LAMP assay had 100% analytical specificity for the identification of influenza A/H1, A/H3, and B viruses, and there was no cross-reaction with other genetically or clinically related control viruses tested in this study. The multiplex influenza A/B/IC RT-LAMP assay for influenza A clinical samples (n = 93) had a sensitivity and specificity of 94.62% and 100%, respectively. However, the multiplex influenza A/B/IC RT-LAMP assay showed a 90% sensitivity for influenza A/H1 because out of the 11 specimens that were positive for influenza A/H1, only 10 were determined by the assay to be positive. Further testing with additional clinical specimens is needed to evaluate the clinical performance characteristics of the multiplex influenza A/B/IC RT-LAMP assay to address the issue of the small sample size used (n = 10). In addition, the internal control signal showed in a very low sensitivity for influenza A clinical samples but not influenza B samples and negative clinical samples. This result might be that LAMP amplification reagents were consumed for the amplification of influenza A or influenza A amplification products interrupt the amplification of the internal control. Interestingly, the multiplex influenza A/B/IC RT-LAMP assay had a lower detection sensitivity for diluted clinical samples than the RealStar® Influenza RT-PCR Kit 2.0 but higher sensitivity for original clinical samples. These results suggest that the multiplex influenza A/B/IC RT-LAMP assay had a higher sensitivity for various influenza genetic sequences while the RealStar® Influenza RT-PCR Kit 2.0 had a higher sensitivity for specific influenza genetic sequences.

RT-LAMP analysis is one of the most promising diagnostic tools for use in the field since it does not require sophisticated and expensive equipment or skilled personnel [51]. The multiplex method developed in this study can diagnose influenza A and B within 60 min using multiple fluorescence. In order to utilize the multiplex influenza A/B/IC RT-LAMP assay in the field, it is necessary to use an isothermal amplification device that detects portable multiple fluorescence. Most of the field isothermal amplifiers currently available have been developed as single channel. However, two-channel isothermal amplifiers (Genie III; OptiGene, West Sussex, UK) have recently been developed and marketed by Chayon Laboratories, Inc. Since there are no field isothermal amplifiers with three channels yet, the influenza A/B and internal control LAMP kits cannot be used. However, by excluding the internal control, influenza A and B can now be diagnosed in the field using commercially available isothermal amplifiers. In addition, conventional RNA extraction methods, which extract RNA using centrifuges from NP swabs or aspirate samples collected from suspected influenza patients, are time-consuming and may be potentially contaminated. Therefore, it is expected that influenza field diagnosis can be performed more effectively by using an RNA extraction chip [52] or the magnetic bead-nucleic acid extraction method [53].

In this study, we developed a multiplex real-time RT-LAMP assay that can diagnose influenza A and influenza B with one step. The multiplex influenza A/B/IC RT-LAMP assay using the probe-quencher that compensates for the disadvantages of LAMP, such as false positive diagnoses, shows similar sensitivity and specificity to the WHO RT-PCR primer set. Since LAMP takes less time (within 60 min) than conventional RT-PCR, the multiplex influenza A/B RT-LAMP assay can be used as an efficient method in on-site molecular diagnostic kits.

Supporting information

S1 Fig. Gel electropherograms of influenza A, B and IC (actin beta) RT-LAMP products.

Influenza A/B virus clinical samples, non-infection human serum RNA and distilled water (DW) were tested by RT-LAMP assays with monoplex influenza A RT-LAMP primer set (A), monoplex influenza B RT-LAMP primer set (B) and monoplex IC (actin beta) RT-LAMP primer set (C). Lane M: DNA ladder marker, Lane 1: Influenza A/H1, Lane 2: Influenza A/H1N1, Lane 3: Influenza A/H3N2, Lane 4: Influenza B, Lane 5: Non-infection human serum RNA and Lane 6: DW (negative control).

https://doi.org/10.1371/journal.pone.0238615.s001

(TIF)

S1 Table. Sensitivities and specificities of the multiplex influenza RT-LAMP assay for H5, H7 and H9 subtypes of avian influenza viruses.

https://doi.org/10.1371/journal.pone.0238615.s002

(DOCX)

References

  1. 1. Bunker D, Ehnis C, Shahbazi M. Managing Influenza Outbreaks Through Social Interaction on Social Media: Research Transformation Through an Engaged Scholarship Approach. Studies in health technology and informatics. 2019;259:39–44. Epub 2019/03/30. pmid:30923270.
  2. 2. Putri W, Muscatello DJ, Stockwell MS, Newall AT. Economic burden of seasonal influenza in the United States. Vaccine. 2018;36(27):3960–6. Epub 2018/05/29. pmid:29801998.
  3. 3. Iuliano AD, Roguski KM, Chang HH, Muscatello DJ, Palekar R, Tempia S, et al. Estimates of global seasonal influenza-associated respiratory mortality: a modelling study. Lancet (London, England). 2018;391(10127):1285–300. Epub 2017/12/19. pmid:29248255; PubMed Central PMCID: PMC5935243.
  4. 4. Coloma R, Valpuesta JM, Arranz R, Carrascosa JL, Ortin J, Martin-Benito J. The structure of a biologically active influenza virus ribonucleoprotein complex. PLoS pathogens. 2009;5(6):e1000491. Epub 2009/06/27. pmid:19557158; PubMed Central PMCID: PMC2695768.
  5. 5. Sreenivasan CC, Thomas M, Antony L, Wormstadt T, Hildreth MB, Wang D, et al. Development and characterization of swine primary respiratory epithelial cells and their susceptibility to infection by four influenza virus types. Virology. 2019;528:152–63. Epub 2019/01/08. pmid:30616205; PubMed Central PMCID: PMC6401229.
  6. 6. Palese P. The genes of influenza virus. Cell. 1977;10(1):1–10. Epub 1977/01/01. pmid:837439.
  7. 7. Henritzi D, Hoffmann B, Wacheck S, Pesch S, Herrler G, Beer M, et al. A newly developed tetraplex real-time RT-PCR for simultaneous screening of influenza virus types A, B, C and D. Influenza and other respiratory viruses. 2019;13(1):71–82. Epub 2018/09/29. pmid:30264926; PubMed Central PMCID: PMC6304318.
  8. 8. Jiang Y, Tan CY, Tan SY, Wong MSF, Chen YF, Zhang L, et al. SAW sensor for Influenza A virus detection enabled with efficient surface functionalization. Sensors and Actuators B: Chemical. 2015;209:78–84.
  9. 9. Petric M, Comanor L, Petti CA. Role of the laboratory in diagnosis of influenza during seasonal epidemics and potential pandemics. The Journal of infectious diseases. 2006;194 Suppl 2:S98–110. Epub 2006/12/14. pmid:17163396.
  10. 10. Pan Q, Wu W, Liao S, Wang S, Zhao C, Li C, et al. Comparison of the detection performance of two different one-step-combined test strips with fluorescent microspheres or colored microspheres as tracers for influenza A and B viruses. Virology journal. 2019;16(1):91. Epub 2019/07/22. pmid:31324259; PubMed Central PMCID: PMC6642511.
  11. 11. Oxford JS, Bossuyt S, Balasingam S, Mann A, Novelli P, Lambkin R. Treatment of epidemic and pandemic influenza with neuraminidase and M2 proton channel inhibitors. Clinical microbiology and infection: the official publication of the European Society of Clinical Microbiology and Infectious Diseases. 2003;9(1):1–14. Epub 2003/04/15. pmid:12691538.
  12. 12. Hayden FG, Treanor JJ, Fritz RS, Lobo M, Betts RF, Miller M, et al. Use of the oral neuraminidase inhibitor oseltamivir in experimental human influenza: randomized controlled trials for prevention and treatment. J Am Med Asso. 1999;282(13):1240–6. Epub 1999/10/12. pmid:10517426.
  13. 13. Aoki FY, Macleod MD, Paggiaro P, Carewicz O, El Sawy A, Wat C, et al. Early administration of oral oseltamivir increases the benefits of influenza treatment. The Journal of antimicrobial chemotherapy. 2003;51(1):123–9. Epub 2002/12/21. pmid:12493796.
  14. 14. Fukao K, Ando Y, Noshi T, Kitano M, Noda T, Kawai M, et al. Baloxavir marboxil, a novel cap-dependent endonuclease inhibitor potently suppresses influenza virus replication and represents therapeutic effects in both immunocompetent and immunocompromised mouse models. PLoS One. 2019;14(5):e0217307. Epub 2019/05/21. pmid:31107922; PubMed Central PMCID: PMC6527232 Research, Co., Ltd, an affiliation of Shionogi. This does not alter our adherence to PLOS ONE policies on sharing data and materials.
  15. 15. Takashita E, Morita H, Ogawa R, Nakamura K, Fujisaki S, Shirakura M, et al. Susceptibility of Influenza Viruses to the Novel Cap-Dependent Endonuclease Inhibitor Baloxavir Marboxil. Frontiers in microbiology. 2018;9:3026. Epub 2018/12/24. pmid:30574137; PubMed Central PMCID: PMC6291754.
  16. 16. Whitley RJ, Hayden FG, Reisinger KS, Young N, Dutkowski R, Ipe D, et al. Oral oseltamivir treatment of influenza in children. The Pediatric infectious disease journal. 2001;20(2):127–33. Epub 2001/02/27. pmid:11224828.
  17. 17. Takayama I, Nakauchi M, Takahashi H, Oba K, Semba S, Kaida A, et al. Development of real-time fluorescent reverse transcription loop-mediated isothermal amplification assay with quenching primer for influenza virus and respiratory syncytial virus. Journal of virological methods. 2019;267:53–8. Epub 2019/03/05. pmid:30831121; PubMed Central PMCID: PMC7113748.
  18. 18. Monto AS, Fleming DM, Henry D, de Groot R, Makela M, Klein T, et al. Efficacy and safety of the neuraminidase inhibitor zanamivirin the treatment of influenza A and B virus infections. The Journal of infectious diseases. 1999;180(2):254–61. Epub 1999/07/09. pmid:10395837.
  19. 19. Vemula SV, Zhao J, Liu J, Wang X, Biswas S, Hewlett I. Current Approaches for Diagnosis of Influenza Virus Infections in Humans. Viruses. 2016;8(4):96. Epub 2016/04/15. pmid:27077877; PubMed Central PMCID: PMC4848591.
  20. 20. Kok J, Blyth CC, Foo H, Patterson J, Taylor J, McPhie K, et al. Comparison of a rapid antigen test with nucleic acid testing during cocirculation of pandemic influenza A/H1N1 2009 and seasonal influenza A/H3N2. Journal of clinical microbiology. 2010;48(1):290–1. Epub 2009/11/06. pmid:19889892; PubMed Central PMCID: PMC2812250.
  21. 21. Templeton KE, Scheltinga SA, Beersma MF, Kroes AC, Claas EC. Rapid and sensitive method using multiplex real-time PCR for diagnosis of infections by influenza a and influenza B viruses, respiratory syncytial virus, and parainfluenza viruses 1, 2, 3, and 4. Journal of clinical microbiology. 2004;42(4):1564–9. Epub 2004/04/09. pmid:15071005; PubMed Central PMCID: PMC387552.
  22. 22. Storch GA. Rapid diagnostic tests for influenza. Current opinion in pediatrics. 2003;15(1):77–84. Epub 2003/01/25. pmid:12544276.
  23. 23. Ma YD, Chang WH, Wang CH, Ma HP, Huang PC, Lee GB. An integrated passive microfluidic device for rapid detection of influenza a (H1N1) virus by reverse transcription loop-mediated isothermal amplification (RT-LAMP). Institute of Electrical and Electronics Engineers. 2017:722–5.
  24. 24. Ghosh DK, Warghane A, Biswas KK. Rapid and Sensitive Detection of Citrus tristeza virus Using Reverse Transcription Loop-Mediated Isothermal Amplification (RT-LAMP) Assay. Methods in molecular biology (Clifton, NJ). 2019;2015:143–50. Epub 2019/06/22. pmid:31222701.
  25. 25. Dhama K, Karthik K, Chakraborty S, Tiwari R, Kapoor S, Kumar A, et al. Loop-mediated isothermal amplification of DNA (LAMP): a new diagnostic tool lights the world of diagnosis of animal and human pathogens: a review. Pakistan journal of biological sciences: PJBS. 2014;17(2):151–66. Epub 2014/05/03. pmid:24783797.
  26. 26. Singh R, Singh DP, Savargaonkar D, Singh OP, Bhatt RM, Valecha N. Evaluation of SYBR green I based visual loop-mediated isothermal amplification (LAMP) assay for genus and species-specific diagnosis of malaria in P. vivax and P. falciparum endemic regions. Journal of vector borne diseases. 2017;54(1):54–60. Epub 2017/03/30. pmid:28352046.
  27. 27. Baek YH, Cheon HS, Park SJ, Lloren KKS, Ahn SJ, Jeong JH, et al. Simple, Rapid and Sensitive Portable Molecular Diagnosis of SFTS Virus Using Reverse Transcriptional Loop-Mediated Isothermal Amplification (RT-LAMP). Journal of microbiology and biotechnology. 2018;28(11):1928–36. Epub 2018/10/03. pmid:30270605.
  28. 28. Edrees MAH, Ali JME, Almansoub HA. Assessment of PCR and lamp tests for the detection of Mycobacterium tuberculosis in sputum samples. European Journal of Pharmaceutical Sciences. 2019;6(5):560–3.
  29. 29. Zhou Y, Wan Z, Yang S, Li Y, Li M, Wang B, et al. A Mismatch-Tolerant Reverse Transcription Loop-Mediated Isothermal Amplification Method and Its Application on Simultaneous Detection of All Four Serotype of Dengue Viruses. Frontiers in microbiology. 2019;10:1056. Epub 2019/05/30. pmid:31139171; PubMed Central PMCID: PMC6518337.
  30. 30. Yaren O, Alto BW, Gangodkar PV, Ranade SR, Patil KN, Bradley KM, et al. Point of sampling detection of Zika virus within a multiplexed kit capable of detecting dengue and chikungunya. BMC infectious diseases. 2017;17(1):293. Epub 2017/04/22. pmid:28427352; PubMed Central PMCID: PMC5399334.
  31. 31. Gadkar VJ, Goldfarb DM, Gantt S, Tilley PAG. Real-time Detection and Monitoring of Loop Mediated Amplification (LAMP) Reaction Using Self-quenching and De-quenching Fluorogenic Probes. Scientific reports. 2018;8(1):5548. Epub 2018/04/05. pmid:29615801; PubMed Central PMCID: PMC5883045.
  32. 32. Poon LL, Leung CS, Chan KH, Lee JH, Yuen KY, Guan Y, et al. Detection of human influenza A viruses by loop-mediated isothermal amplification. Journal of clinical microbiology. 2005;43(1):427–30. Epub 2005/01/07. pmid:15635005; PubMed Central PMCID: PMC540134.
  33. 33. Terrier O, Josset L, Textoris J, Marcel V, Cartet G, Ferraris O, et al. Cellular transcriptional profiling in human lung epithelial cells infected by different subtypes of influenza A viruses reveals an overall down-regulation of the host p53 pathway. Virology journal. 2011;8:285. Epub 2011/06/10. pmid:21651802; PubMed Central PMCID: PMC3127840.
  34. 34. van Elden LJ, Nijhuis M, Schipper P, Schuurman R, van Loon AM. Simultaneous detection of influenza viruses A and B using real-time quantitative PCR. Journal of clinical microbiology. 2001;39(1):196–200. Epub 2001/01/04. pmid:11136770; PubMed Central PMCID: PMC87701.
  35. 35. Coiras MT, Aguilar JC, Garcia ML, Casas I, Perez-Brena P. Simultaneous detection of fourteen respiratory viruses in clinical specimens by two multiplex reverse transcription nested-PCR assays. Journal of medical virology. 2004;72(3):484–95. Epub 2004/01/30. pmid:14748074; PubMed Central PMCID: PMC7166637.
  36. 36. Ding Y, Dou J, Teng Z, Yu J, Wang T, Lu N, et al. Antiviral activity of baicalin against influenza A (H1N1/H3N2) virus in cell culture and in mice and its inhibition of neuraminidase. Archives of virology. 2014;159(12):3269–78. Epub 2014/08/01. pmid:25078390.
  37. 37. Hamilton BS, Gludish DW, Whittaker GR. Cleavage activation of the human-adapted influenza virus subtypes by matriptase reveals both subtype and strain specificities. Journal of virology. 2012;86(19):10579–86. Epub 2012/07/20. pmid:22811538; PubMed Central PMCID: PMC3457293.
  38. 38. Di Trani L, Bedini B, Donatelli I, Campitelli L, Chiappini B, De Marco MA, et al. A sensitive one-step real-time PCR for detection of avian influenza viruses using a MGB probe and an internal positive control. BMC infectious diseases. 2006;6:87. Epub 2006/05/27. pmid:16725022; PubMed Central PMCID: PMC1524785.
  39. 39. Ge Y, Zhou Q, Zhao K, Chi Y, Liu B, Min X, et al. Detection of influenza viruses by coupling multiplex reverse-transcription loop-mediated isothermal amplification with cascade invasive reaction using nanoparticles as a sensor. Int J Nanomedicine. 2017;12:2645–56. Epub 2017/04/25. pmid:28435249; PubMed Central PMCID: PMC5388202.
  40. 40. Chen HT, Zhang J, Sun DH, Ma LN, Liu XT, Quan K, et al. Reverse transcription loop-mediated isothermal amplification for the detection of highly pathogenic porcine reproductive and respiratory syndrome virus. Journal of virological methods. 2008;153(2):266–8. Epub 2008/08/19. pmid:18706931; PubMed Central PMCID: PMC7112790.
  41. 41. Euler M, Wang Y, Otto P, Tomaso H, Escudero R, Anda P, et al. Recombinase polymerase amplification assay for rapid detection of Francisella tularensis. Journal of clinical microbiology. 2012;50(7):2234–8. Epub 2012/04/21. pmid:22518861; PubMed Central PMCID: PMC3405570.
  42. 42. Tong Y, Tang W, Kim HJ, Pan X, Ranalli T, Kong H. Development of isothermal TaqMan assays for detection of biothreat organisms. BioTechniques. 2008;45(5):543–57. Epub 2008/11/15. pmid:19007339.
  43. 43. Yamazaki W, Mioulet V, Murray L, Madi M, Haga T, Misawa N, et al. Development and evaluation of multiplex RT-LAMP assays for rapid and sensitive detection of foot-and-mouth disease virus. Journal of virological methods. 2013;192(1–2):18–24. Epub 2013/04/16. pmid:23583488.
  44. 44. Liu N, Zou D, Dong D, Yang Z, Ao D, Liu W, et al. Development of a multiplex loop-mediated isothermal amplification method for the simultaneous detection of Salmonella spp. and Vibrio parahaemolyticus. Scientific reports. 2017;7:45601. Epub 2017/03/30. pmid:28349967; PubMed Central PMCID: PMC5368564.
  45. 45. Jung JH, Oh SJ, Kim YT, Kim SY, Kim WJ, Jung J, et al. Combination of multiplex reverse-transcription loop-mediated isothermal amplification with an immunochromatographic strip for subtyping influenza A virus. Anal Chim Acta. 2015;853:541–7. Epub 2014/12/04. pmid:25467501; PubMed Central PMCID: PMC7094724.
  46. 46. Ahn SJ, Baek YH, Lloren KKS, Choi WS, Jeong JH, Antigua KJC, et al. Rapid and simple colorimetric detection of multiple influenza viruses infecting humans using a reverse transcriptional loop-mediated isothermal amplification (RT-LAMP) diagnostic platform. BMC infectious diseases. 2019;19(1):676. Epub 2019/08/03. pmid:31370782; PubMed Central PMCID: PMC6669974.
  47. 47. Mahony J, Chong S, Bulir D, Ruyter A, Mwawasi K, Waltho D. Multiplex loop-mediated isothermal amplification (M-LAMP) assay for the detection of influenza A/H1, A/H3 and influenza B can provide a specimen-to-result diagnosis in 40 min with single genome copy sensitivity. Journal of clinical virology: the official publication of the Pan American Society for Clinical Virology. 2013;58(1):127–31. Epub 2013/07/06. pmid:23827787.
  48. 48. Chi Y, Ge Y, Zhao K, Zou B, Liu B, Qi X, et al. Multiplex Reverse-Transcription Loop-Mediated Isothermal Amplification Coupled with Cascade Invasive Reaction and Nanoparticle Hybridization for Subtyping of Influenza A Virus. Scientific reports. 2017;7:44924. Epub 2017/03/23. pmid:28322309; PubMed Central PMCID: PMC5359610.
  49. 49. Hsieh K, Mage PL, Csordas AT, Eisenstein M, Soh HT. Simultaneous elimination of carryover contamination and detection of DNA with uracil-DNA-glycosylase-supplemented loop-mediated isothermal amplification (UDG-LAMP). Chem Commun (Camb). 2014;50(28):3747–9. Epub 2014/03/01. pmid:24577617.
  50. 50. Kil EJ, Kim S, Lee YJ, Kang EH, Lee M, Cho SH, et al. Advanced loop-mediated isothermal amplification method for sensitive and specific detection of Tomato chlorosis virus using a uracil DNA glycosylase to control carry-over contamination. Journal of virological methods. 2015;213:68–74. Epub 2014/12/09. pmid:25483127.
  51. 51. Parida M, Sannarangaiah S, Dash PK, Rao PV, Morita K. Loop mediated isothermal amplification (LAMP): a new generation of innovative gene amplification technique; perspectives in clinical diagnosis of infectious diseases. Reviews in medical virology. 2008;18(6):407–21. Epub 2008/08/22. pmid:18716992; PubMed Central PMCID: PMC7169140.
  52. 52. Yoon J, Yoon Y-J, Lee TY, Park MK, Chung J, Shin Y. A disposable lab-on-a-chip platform for highly efficient RNA isolation. Sensors and Actuators B: Chemical. 2018;255:1491–9.
  53. 53. He H, Li R, Chen Y, Pan P, Tong W, Dong X, et al. Integrated DNA and RNA extraction using magnetic beads from viral pathogens causing acute respiratory infections. Scientific reports. 2017;7:45199. Epub 2017/03/24. pmid:28332631; PubMed Central PMCID: PMC5362898.