Quantitative Analysis of Food and Feed Samples with Droplet Digital PCR

Figure 3

Precision of the duplex ddPCR assay as a function of the target concentration.

MON810 content measured by ddPCR in five series of seven target concentrations. The target MON810 content (3.85%) is indicated by a dotted line. Acceptance criterion for precision is ±25% of the target content (from 2.89% to 4.81%) represented by the dashed lines. Error bars represent the standard deviation of the measured MON810% by ddPCR at each target concentration (five replicates per target concentration).

Figure 3

doi: https://doi.org/10.1371/journal.pone.0062583.g003