In Vivo Intracellular Oxygen Dynamics in Murine Brain Glioma and Immunotherapeutic Response of Cytotoxic T Cells Observed by Fluorine-19 Magnetic Resonance Imaging

Figure 1

In vitro calibration curve for R1 = 1/T1 versus pO2.

Here, solid circles represent average 19F R1 measurements of six oxygen partial pressure values, each in triplicate, ranging from 0% to 100% oxygen in a saturated O2/N2 mixture of PFC emulsion in water. The solid line is the calibration curve fit by the linear least squares method (R2 = 0.99). Data were collected at 11.7 T and 37°C. 100% oxygen is equivalent to 760 mm Hg. (The R1 error bars are smaller than the point size.)

Figure 1

doi: https://doi.org/10.1371/journal.pone.0059479.g001