Abstract
We investigate three-dimensional transfer functions in order to understand image formation in the recently proposed 4Pi confocal microscope [ J. Opt. Soc. Am. A 9, 2159 ( 1992); Opt. Commun. 93, 277 ( 1992)]. Both nonfluorescence and fluorescence imaging methods are considered, and the effects of apodization functions are taken into account. It is shown that for a 4Pi confocal system of finite-sized aperture the cutoff spatial frequency in the axial direction is larger than that in the transverse direction; this behavior is opposite that of conventional confocal systems, explaining the observation of high axial resolution in 4Pi confocal systems. In addition, the three-dimensional optical transfer function for a conventional confocal fluorescence microscope of high aperture is presented.
© 1994 Optical Society of America
Full Article | PDF ArticleMore Like This
Min Gu, X. Gan, and C. J. R. Sheppard
J. Opt. Soc. Am. A 8(7) 1019-1025 (1991)
Min Gu and C. J. R. Sheppard
J. Opt. Soc. Am. A 9(11) 1991-1999 (1992)
Shigeharu Kimura and Chusuke Munakata
J. Opt. Soc. Am. A 6(7) 1015-1019 (1989)