Journal of Nippon Medical School
Online ISSN : 1347-3409
Print ISSN : 1345-4676
ISSN-L : 1345-4676
Case Reports
Effectiveness of Endotoxin Scattering Photometry for Determining the Efficacy of Polymyxin B-immobilized Fiber Treatment in Septic Shock: Report of a Case
Yuichiro SakamotoKunihiro MashikoToru ObataHiroyuki Yokota
Author information
JOURNAL FREE ACCESS

2010 Volume 77 Issue 2 Pages 119-122

Details
Abstract

The limulus test, which has been established as a test for endotoxin measurement, is associated with problems, including that posed by the presence of a response inhibitor factor and the longer time needed for the measurement of low concentrations. On the other hand, the technique of direct hemoperfusion with a polymyxin B immobilized fiber column (DHP-PMX) was developed in Japan in 1994 and has been used for the control of endotoxemia in septic shock. The limulus test, which is a common endotoxin measurement test, has several problems with regard to sensitivity. Therefore, this test is no longer used to determine the effectiveness of DHP-PMX. Here, we describe a patient presenting with colonic perforation who recovered from septic shock with DHP-PMX. This treatment effect was reflected by a decrease in plasma endotoxin levels as demonstrated more readily with endotoxin scattering photometry assay than with the standard limulus test. We conclude that endotoxin measurement with endotoxin scattering photometry is superior to nephelometry in patients with endotoxemia.

Content from these authors
© 2010 by the Medical Association of Nippon Medical School
Previous article Next article