End-organ saturations correlate with aortic blood flow estimates by echocardiography in the extremely premature newborn – an observational cohort study
BMC Pediatrics volume 21, Article number: 312 (2021)
Abstract
Background
Near-infrared spectroscopy (NIRS) measures of cerebral saturation (Csat) and renal saturation (Rsat) in extreme premature newborns may be affected by systemic blood flow fluctuations. Despite increasing clinical use of NIRS to monitor tissue saturation in the premature infant, validation of NIRS measures as a correlate of blood flow is still needed. We compared echocardiography (ECHO) derived markers of ascending aorta (AscAo) and descending aorta (DesAo) blood flow with NIRS measurements obtained during the ECHO.
Methods
Newborns < 29 weeks’ gestation (2013–2017) underwent routine NIRS monitoring. Csat, Rsat and systemic saturation at the time of ECHO were retrospectively analyzed and compared with Doppler markers of aortic flow. Renal and cerebral fractional tissue oxygen extraction (rFTOE and cFTOE, respectively) were calculated. Mixed effects models evaluated the association between NIRS and Doppler markers.
Results
Forty-nine neonates with 75 Csat-ECHO and 62 Rsat-ECHO observations were studied. Mean post-menstrual age was 28.3 ± 3.8 weeks during the ECHO. Preductal measures including AscAo velocity time integral (VTI) and AscAo output were correlated with Csat or cFTOE, while postductal measures including DesAo VTI, DesAo peak systolic velocity, and estimated DesAo output were more closely correlated with Rsat or rFTOE.
Conclusions
NIRS measures are associated with aortic blood flow measurements by ECHO in the extremely premature population. NIRS is a tool to consider when following end organ perfusion in the preterm infant.
Background
Extremely premature newborns undergo significant physiologic changes in their immediate post-natal life, with potential hemodynamic instability in the context of persistent fetal shunts. Neonatal care in this vulnerable population aims at maintaining adequate organ perfusion in the setting of changing physiology. Near infrared spectroscopy (NIRS) is a promising technology allowing for real-time monitoring of tissue oxygenation as a surrogate for end-organ perfusion. However, validation of NIRS measures in the premature population as a correlate of blood flow is needed, especially in the context of postnatal delays in transition, such as presence of a patent ductus arteriosus (PDA). Specifically, premature newborns are at risk of diastolic steal effect, which may adversely alter the post-ductal aortic blood flow [1].
NIRS monitoring estimates venous-weighted percentage of saturated hemoglobin in the monitored tissue [2,3,4]. Regional saturations are affected by organ perfusion, hemoglobin concentration, arterial oxygen content, oxygen extraction and oxygen consumption. Perfusion is a reflection of local blood flow, which is dependent on cardiac output, as well as vascular distribution, filling and tone. As such, cerebral oxygen saturation (Csat) and renal oxygen saturation (Rsat) monitored by NIRS may indicate underlying perfusion when all other factors remain stable. Csat is primarily affected by pre-ductal systemic perfusion and cerebral oxygen extraction, while the Rsat is impacted by post-ductal diastolic steal, underlying renal oxygen extraction and altered systemic output. Csat correlates with cerebral perfusion [2, 5] and decreased values have been associated with neurological injury in premature newborns [6,7,8]. We recently published that the velocity time integral (VTI) of the descending aortic flow by echocardiography (ECHO), a marker of stroke distance, correlated with Rsat at the time of the measurement in a population of single ventricle infants [9]. However, limited data exist on the correlation between ultrasound markers of perfusion and cerebral/renal NIRS values in the premature neonate. We sought to compare ECHO derived markers of aortic blood flow with end-organ saturation measurements obtained at the time of ECHO acquisition in extremely premature neonates. We hypothesized that in preterm infants, especially in first week of life when the duct is usually patent, resulting in diastolic runoff, Csat would reflect ECHO-derived markers of aortic pre-ductal blood flow and Rsat would reflect those of post-ductal flow.
Methods
Subjects
We conducted a retrospective review of all premature newborns < 29 weeks of estimated gestational age at birth with available NIRS recordings and admitted from May 2013 to August 2017 at a single institution (Lucile Packard Children’s Hospital, Stanford). Extremely premature newborns were routinely monitored with NIRS in the first 7 days of life per clinical guidelines beginning in May 2013. Csat and Rsat were recorded and stored in a secure database. Patients were excluded if ECHO was not performed during the NIRS monitoring period or if a congenital heart defect was present (except patent ductus arteriosus [PDA] or atrial or ventricular septal defects). Study data were extracted from the electronic medical record and from the neonatal NIRS database, and managed using REDCap electronic data capture tools hosted at the Stanford Center for Clinical Informatics [10]. This study was approved by the Stanford University Institutional Review Board. The study has been granted an exemption from requiring informed consent. All methods were performed in accordance with the Declaration of Helsinki.
NIRS measures and SpO2
Csat and Rsat levels were measured using the INVOS 5100 NIRS monitor (Medtronic, Minneapolis, MN). As previously described by our group [9, 11], neonatal sensors were routinely applied shortly after birth on the central forehead and on either the right or left posterior flank above the iliac crest and below the costal margin (T10-L2) [11]. Csat and Rsat were extracted from our NIRS database, which contains prospectively collected NIRS values recorded at 30 s intervals. Systemic saturation (SpO2) values at the beginning of the ECHO were collected from nursing flowsheets. ECHO scans were stored on the picture archiving and communication system (PACS). Time at the start of the ECHO was retrieved from the images stored on the Syngo Dynamics workstation (Siemens Medical Solutions, USA). The Csat and Rsat values in the hour before and after the start of the ECHO were averaged. Cerebral and renal fractional tissue oxygen extraction (cFTOE and rFTOE, respectively), calculated as (SpO2 – [Csat or Rsat])/SpO2, was similarly collected as a measure of oxygen extraction by the brain or kidneys [3]. Clinicians were not blinded to Csat and Rsat levels, but no management guidelines based on NIRS values were in place other than notification of a medical provider if Csats were sustained < 50%.
Echocardiographic measures
Every ECHO available during NIRS monitoring was analyzed for each patient. Methodology was similar to previous reports by our group [9, 12]. Briefly, stored images on the institutional image server were acquired according to the American Society of Echocardiography guidelines [13]. Images were acquired using Philips iE33 and Philips EPIQ 7 (Philips Medical Systems, Bothell, USA). Our standard neonatal echocardiography protocol does not involve sedation. Each ECHO parameter was prospectively measured from stored images on a Syngo Dynamics workstation (Siemens Medical Solutions, USA) by an investigator trained in pediatric ECHO (GA) and masked to NIRS saturation values. Pulsed-wave (PW) Doppler envelope at various levels in the aorta were analyzed when available (Fig. 1). Data extracted from the supra-sternal view included: ascending aorta PW-Doppler and descending (post-ductal) aorta PW-Doppler. Data extracted from the subcostal view included: descending aorta PW-Doppler. From these PW-Doppler envelopes, peak systolic velocity and velocity time integral (VTI) of the entire cardiac cycle (systolic and diastolic phase) were extracted. The diastolic VTI was considered to be zero in the absence of diastolic flow and considered to be negative when diastolic flow was retrograde. The VTI is an estimated measure of the stroke distance at the sampled area [14]. The diameter of the ascending (suprasternal view) and post-ductal descending aorta (suprasternal and subcostal views) were measured from bidimensional gray-scale images. Aortic outputs were calculated based on the respective VTI and heart rate at PW-Doppler sampling using the same formula: (Aorta diameter)/2)2 x π x heart rate / weight [15, 16].
Clinical outcomes
Urine output was measured for the 8 h before the beginning of the ECHO study and expressed as mL/kg/hour (based on the daily weight). Hemoglobin, hematocrit and blood gas values (pH, pCO2, base excess) were recorded when collected within 24 h of the ECHO. Blood pressure was recorded prior to start of the ECHO.
Statistics
Descriptive statistics were used, including mean (with standard deviations [SD]) and median (interquartile range [IQR]) for continuous variables, and counts (proportions) for categorical variables. Student t-test and Wilcoxon-Mann-Whitney test compared continuous variables for parametric and non-parametric variables, respectively. Average Csat and Rsat in the hour prior to the ECHO were compared to average Csat and Rsat in the hour after the start of the ECHO using paired Student t-test. Csat and Rsat data were correlated with ECHO parameters. Analysis accounted for the non-independent data structure (multiple ECHO measurements for some subjects at various dates) with linear mixed effect models with random intercepts/slope (for continuous variables) and generalized estimating equations (for binary categorical outcomes). The level of significance was set at 0.05 for all comparisons. Statistical analysis was done with Stata/SE 14.2 (Texas, USA).
Results
From May 2013 to August 2017, 97 premature patients < 29 weeks gestation at birth were identified in the NIRS database with either Csat or Rsat monitoring results available. Of these, 50 infants had an available ECHO at the time of NIRS monitoring, of which 1 infant was excluded due to concomitant diagnosis of Tetralogy of Fallot. The final cohort included 49 neonates with 75 Csat-ECHO and 62 Rsat-ECHO observations available for analysis. During NIRS monitoring, a single ECHO was available in 30 patients, two ECHOs in 11 patients, three ECHOs in 7 patients and four ECHOs in 1 patient. Mean gestational age at birth was 25.8 ± 1.3 weeks, with a mean post-menstrual age of 28.3 ± 3.8 weeks at time of ECHO.
Demographic variables are shown in Table 1 and clinical variables pertaining to the ECHO-NIRS dyads in Table 2. There were no significant differences in clinical variables between those with Csat-ECHO measures and those with Rsat-ECHO measures. For those with available Rsat measurements, the average urine output was of 3.9. ± 1.7 mL/kg/min prior to ECHO. The average Rsat was of 58.6 ± 12.6% and the average rFTOE was 37.7 ± 13.3%. There was no difference between the Rsat in the hour following the onset of the ECHO compared to the Rsat in the hour preceding the ECHO (58.6 ± 12.6% vs 58.5 ± 12.3% vs, p = 0.90). For those with available Csat measurements, the average Csat during ECHO was 65.8 ± 8.6% and the average cFTOE was 30.1 ± 8.7%. Similarly, there was no difference between the Csat in the hour following the onset of the ECHO with the Csat in the hour preceding the ECHO (65.8 ± 8.6% vs 66.1 ± 9.1%, p = 0.43). The presence of an umbilical arterial line (UAL), all placed in a high-position (T6-T10), was not associated with a difference in Rsat at the start of echocardiography (54.1 ± 12.3% vs 58.6 ± 19.5%, p = 0.31).
Table 3 illustrates the results of the correlation between NIRS-derived measurements and ECHO-derived doppler flow measurements of the aorta. Although Rsat, Csat, rFTOE, and cFTOE did not correlate with markers of blood flow (VTI, cardiac output, and peak systolic velocity) obtained at the descending aorta and sampled with the subcostal view, the majority of these NIRS measurements did correlate with blood flow markers sampled with the suprasternal view. The unadjusted linear regressions correlating Rsat, Csat, rFTOE and cFTOE with the descending aortic VTI obtained in the suprasternal view are represented in Fig. 2. Rsat and rFTOE also were associated with the peak systolic velocity in that view. Furthermore, rFTOE, Csat and cFTOE correlated with the estimated descending aortic output measured with the suprasternal view. Csat was associated with the stroke distance estimated at the ascending aorta (suprasternal view) and estimated ascending aortic output correlated with cFTOE. Although the vast majority of the cohort had a patent ductus at evaluation, there was an association between PDA size at ECHO and Rsat/Csat values; and when stratifying for PDA presence at ECHO, we found similar results despite the decreasing number of observations (Supplemental Table A). Finally, none of the NIRS measurements were associated with the closest measurements of systolic/diastolic blood pressure, pH, hemoglobin, hematocrit, creatinine and urine output by mixed effect model.
Discussion
In this retrospective study of NIRS measures in extremely premature neonates, we found an association between end organ saturations measured during ECHO with estimates of aortic flow by Doppler. As physiologically expected, markers of pre-ductal aortic flow correlated with Csat and those of post-ductal aortic flow with Rsat.
We previously described that descending abdominal aortic flow by subcostal view correlated with Rsat measures in a congenital heart disease term population [9]. In this current extreme premature population, markers of aortic blood flow in the suprasternal view were correlated with NIRS measures of end-organ saturations. Csat correlated with both ascending aortic measurements (pre-ductal) and descending aortic measurements while Rsat correlated with descending aortic measurements only. Other authors have described a correlation between cerebral saturation and estimated cardiac output in the extreme premature neonatal population [17]. In our study, measurements in the subcostal view were not associated with NIRS measurements. This may be explained by the increased angle of insonation in premature newborns, when evaluating the descending aorta in the subcostal window, compared to a term newborn where one may have more amplitude of manipulation of the probe. During Doppler acquisition, angle of insonation between the sampled vessel and the ultrasound beam must be minimized to avoid imprecision in estimating the Doppler shift [18, 19]. As such, the beam should interrogate the corresponding flow in a parallel fashion, as much as feasible [20]. Suprasternal view minimizes the angle of insonation for Doppler acquisition in the descending aorta, since it is aligned with the flow. Indeed, a previous report indicated that flow velocities for the descending aorta are best sampled from the suprasternal window [21]. Finally, estimation of stroke distance by VTI has been advocated as a simple and reproducible measurement of blood flow [22], which can be used to monitor the hemodynamic changes in critically ill patients [23] and, in select populations, has been associated with patient outcomes [24, 25]. It has previously been described as a marker of lower body perfusion in the pediatric and neonatal population [26]. As such, our data suggest that a suprasternal approach be considered when interrogating the post-ductal aorta to assess VTI in the premature neonate.
NIRS is a promising tool to monitor changes in aortic blood flow, as may be seen with a significant PDA in the extremely premature neonate. Other clinical markers of perfusion commonly used in the clinical setting, including blood pressure, may not detect subtle changes [27, 28] and have not been correlated with echocardiographic markers of blood flow in preterm infants [28]. While indicators of aortic blood flow were associated with Rsat and Csat measurements in our study, they were not associated with other routinely used markers of end-organ hypoperfusion in clinical practice: blood pressure, urine output, pH and hemoglobin. While these traditional markers may be disturbed when tissue ischemia is ongoing [29, 30], they may be imperfect as early detectors of hypoperfusion. Low Csat values have been associated with increased likelihood for severe intraventricular hemorrhage or periventricular leukomalacia in premature newborns [6,7,8]. Altered blood flow patterns may be associated with abnormal Csat values and contribute to the hemodynamic changes involved in the pathophysiology of neurological injury. Rsat values are usually higher than Csat values in term infants. However, in our population, Rsat values were lower, possibly reflecting the high incidence of a left-to-right PDA on the ECHO scans that were included. In the Rsat-ECHO dyads, 92% had a PDA with a significant proportion showing a diastolic steal (56% retrograde flow in the descending aorta), possibly compromising the renal perfusion and explaining the lower Rsat. Our group has previously described an association between a hemodynamically significant PDA and a tendency for a low Rsat [11]. In our cohort, we also found an association between PDA size and Rsat values (Supplemental Table A). Our data suggest that Csat and Rsat, as measured with NIRS, can be utilized as a correlate of pre- and post-ductal systemic blood flow. This adds to the small body of literature associating NIRS measures to markers of perfusion in the pediatric setting [9, 31,32,33].
Several methodologic issues regarding the stability of NIRS measures during concurrent ECHO image acquisition and infusion through an indwelling UAL were investigated. The regional oxygen saturation/extraction values were similar in the hour prior and after the ECHO, suggesting that the image acquisition did not affect oxygen utilization. Furthermore, in our cohort, we did not detect any association between NIRS values and the presence of an UAL. However, the granularity of the NIRS measurements may not have detected subtle blood flow differences secondary to the presence of the line, and time periods of blood sampling from the UAL were not specifically captured. The stability of NIRS measures as described in this study may help to inform future design of simultaneous NIRS and ECHO studies in the preterm infant.
Our study was retrospective, single center and descriptive with a significant risk of being underpowered to detect meaningful associations. Additional limitations include lack of NIRS data for some infants with prematurity and the fact that clinicians were not blinded to the NIRS values. Also, not all infants had an ECHO during NIRS monitoring, and as such, were excluded from the analysis. Hence, the current cohort most likely represents the sickest premature newborns from a clinical perspective, potentially introducing selection bias that would limit generalizability. Arterial oxygen tension (PaO2) at the time of NIRS monitoring could have provided additional information and added accuracy compared to the use of the systemic oxygen saturation but were not consistently available in all patients. Creatinine measurements were only available for some infants and timing of collection was variable since retrospectively extracted from the chart. Precise position of renal parenchyma by ultrasound was not confirmed prior to sensor placement. However, previous studies have evaluated by ultrasound the position of the kidney in newborns and described its location as between the costal margin and the iliac crest [34,35,36]. ECHO was limited by the use of multiple scanning platforms, multiple professionals acquiring the images, absence of some ECHO markers, or limited views in several patients. Indeed, as outlined in Table 3, some ECHO markers were missing in the included ECHOs (potentially introducing a measurement bias) and many newborns had only some of the ECHO markers available for analysis (number of observations available were presented). ECHO measures were not analyzed for intra-reader or inter-reader variability, although our methodology for VTI reproducibility has been validated in a previous publication by our group [24], and by other authors [22, 37, 38]. Furthermore, output estimation by ECHO has limitations regarding the assumption that the cross-sectional area has a circular shape and is calculated solely from the linear diameter measurement of the corresponding structure. However, a MRI study has indicated a strong correlation with ECHO-derived measurements of output, based on this methodology [39]. Finally, the ECHOs were not performed at standardized time-points.
Conclusions
Regional NIRS values are associated with aortic blood flow measurements by ECHO in the extremely premature population. Future studies should investigate if therapeutic approaches to maintain adequate end-organ saturations using NIRS monitoring may also favorably impact blood flow distribution and improve outcomes in the extremely premature population.
Availability of data and materials
The derived data generated in this research will be shared on reasonable request to the corresponding author.
Abbreviations
- Csat:
-
Cerebral regional oxygen saturation level
- ECHO:
-
Echocardiography
- FTOE:
-
Fractional tissue oxygen extraction
- NIRS:
-
Near-infrared spectroscopy
- PDA:
-
Patent ductus arteriosus
- PVR:
-
Pulmonary vascular resistance
- Rsat:
-
Renal regional oxygen saturation level
- SpO2:
-
Systemic oxygen saturation
- UAL:
-
Umbilical arterial line
- VTI:
-
Velocity time integral
References
Broadhouse KM, Price AN, Durighel G, Cox DJ, Finnemore AE, Edwards AD, et al. Assessment of PDA shunt and systemic blood flow in newborns using cardiac MRI. NMR Biomed. 2013;26(9):1135–41. https://doi.org/10.1002/nbm.2927.
Pellicer A, Bravo Mdel C. Near-infrared spectroscopy: a methodology-focused review. Semin Fetal Neonatal Med. 2011;16(1):42–9. https://doi.org/10.1016/j.siny.2010.05.003. Epub 2010 Jun 26.
Chock VY, Davis AS. Bedside cerebral monitoring to predict neurodevelopmental outcomes. NeoReviews. 2009;10(3):e121–9. https://doi.org/10.1542/neo.10-3-e121.
Bailey SM, Hendricks-Munoz KD, Mally P. Cerebral, renal, and splanchnic tissue oxygen saturation values in healthy term newborns. Am J Perinatol. 2014;31(04):339–44.
Naulaers G, Meyns B, Miserez M, Leunens V, Van Huffel S, Casaer P, et al. Use of tissue oxygenation index and fractional tissue oxygen extraction as non-invasive parameters for cerebral oxygenation. Neonatology. 2007;92(2):120–6. https://doi.org/10.1159/000101063.
Tsuji M, Saul JP, du Plessis A, Eichenwald E, Sobh J, Crocker R, et al. Cerebral intravascular oxygenation correlates with mean arterial pressure in critically ill premature infants. Pediatrics. 2000;106(4):625–32. https://doi.org/10.1542/peds.106.4.625.
Katheria AC, Harbert MJ, Nagaraj SB, Arnell K, Poeltler DM, Brown MK, et al. The Neu-Prem Trial: Neuromonitoring of Brains of Infants Born Preterm During Resuscitation-A Prospective Observational Cohort Study. J Pediatr. 2018;198:209–13 e203.
Chock VY, Kwon SH, Ambalavanan N, Batton B, Nelin LD, Chalak LF, et al. Cerebral oxygenation and autoregulation in preterm infants (early NIRS study). J Pediatr. 2020;227:94–100.e1. https://doi.org/10.1016/j.jpeds.2020.08.036.
Altit G, Bhombal S, Tacy TA, Chock VY. End-organ saturation differences in early neonatal transition for left-versus right-sided congenital heart disease. Neonatology. 2018;114(1):53–61. https://doi.org/10.1159/000487472.
Harris PA, Taylor R, Thielke R, Payne J, Gonzalez N, Conde JG. Research electronic data capture (REDCap)—a metadata-driven methodology and workflow process for providing translational research informatics support. J Biomed Inform. 2009;42(2):377–81. https://doi.org/10.1016/j.jbi.2008.08.010.
Chock VY, Rose LA, Mante JV, Punn R. Near-infrared spectroscopy for detection of a significant patent ductus arteriosus. Pediatr Res. 2016;80(5):675–80. https://doi.org/10.1038/pr.2016.148.
Altit G, Bhombal S, Feinstein J, Hopper RK, Tacy TA. Diminished right ventricular function at diagnosis of pulmonary hypertension is associated with mortality in bronchopulmonary dysplasia. Pulmon Circ. 2019;9(3):2045894019878598. https://doi.org/10.1177/2045894019878598.
Lai WW, Geva T, Shirali GS, Frommelt PC, Humes RA, Brook MM, et al. Guidelines and standards for performance of a pediatric echocardiogram: a report from the task force of the pediatric Council of the American Society of echocardiography. J Am Soc Echocardiogr. 2006;19(12):1413–30. https://doi.org/10.1016/j.echo.2006.09.001.
Punn R, Axelrod DM, Sherman-Levine S, Roth SJ, Tacy TA. Predictors of mortality in pediatric patients on venoarterial extracorporeal membrane oxygenation. Pediatr Crit Care Med. 2014;15(9):870–7. https://doi.org/10.1097/PCC.0000000000000236.
Kim KM, Kim HS, Yoon JH, Lee E-J, Yum SK, Moon C-J, et al. Descending aorta blood flow characteristics before the development of necrotizing enterocolitis in preterm neonates. Neonatal Med. 2018;25(2):78–84. https://doi.org/10.5385/nm.2018.25.2.78.
Alverson DC, Eldridge MW, Johnson JD, Burstein R, Papile L, Dillon T, et al. Effect of patent ductus arteriosus on left ventricular output in premature infants. J Pediatr. 1983;102(5):754–7. https://doi.org/10.1016/S0022-3476(83)80252-2.
Janaillac M, Beausoleil TP, Barrington KJ, Raboisson MJ, Karam O, Dehaes M, et al. Correlations between near-infrared spectroscopy, perfusion index, and cardiac outputs in extremely preterm infants in the first 72 h of life. Eur J Pediatr. 2018;177(4):541–50. https://doi.org/10.1007/s00431-018-3096-z.
Camfferman FA, de Goederen R, Govaert P, Dudink J, van Bel F, Pellicer A, et al. Diagnostic and predictive value of Doppler ultrasound for evaluation of the brain circulation in preterm infants: a systematic review. Pediatr Res. 2020;87(Suppl 1):50–8. https://doi.org/10.1038/s41390-020-0777-x.
de Waal KA. The methodology of Doppler-derived central blood flow measurements in newborn infants. Int J Pediatr. 2012;2012:680162.
Mitchell C, Rahko PS, Blauwet LA, Canaday B, Finstuen JA, Foster MC, et al. Guidelines for performing a comprehensive transthoracic echocardiographic examination in adults: recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr. 2019;32(1):1–64. https://doi.org/10.1016/j.echo.2018.06.004.
Hatle L. Assessment of aortic blood flow velocities with continuous wave Doppler ultrasound in the neonate and young child. J Am Coll Cardiol. 1985;5(1 Suppl):113s–9s.
Pees C, Glagau E, Hauser J, Michel-Behnke I. Reference values of aortic flow velocity integral in 1193 healthy infants, children, and adolescents to quickly estimate cardiac stroke volume. Pediatr Cardiol. 2013;34(5):1194–200. https://doi.org/10.1007/s00246-012-0628-6.
Blanco P. Rationale for using the velocity-time integral and the minute distance for assessing the stroke volume and cardiac output in point-of-care settings. Ultrasound J. 2020;12(1):21. https://doi.org/10.1186/s13089-020-00170-x.
Altit G, Bhombal S, Chock VY, Tacy TA. Immediate postnatal ventricular performance is associated with mortality in Hypoplastic left heart syndrome. Pediatr Cardiol. 2019;40(1):168–76. https://doi.org/10.1007/s00246-018-1974-9.
Tan C, Rubenson D, Srivastava A, Mohan R, Smith MR, Billick K, et al. Left ventricular outflow tract velocity time integral outperforms ejection fraction and Doppler-derived cardiac output for predicting outcomes in a select advanced heart failure cohort. Cardiovasc Ultrasound. 2017;15(1):18. https://doi.org/10.1186/s12947-017-0109-4.
Groves AM, Kuschel CA, Knight DB, Skinner J. Echocardiographic assessment of blood flow volume in the superior vena cava and descending aorta in the newborn infant. Arch Dis Child Fetal Neonatal Ed. 2008;93(1):F24–8. https://doi.org/10.1136/adc.2006.109512.
Gale C. Question 2. Is capillary refill time a useful marker of haemodynamic status in neonates? Arch Dis Child. 2010;95(5):395–7. https://doi.org/10.1136/adc.2010.186411.
Groves AM, Kuschel CA, Knight DB, Skinner JR. Relationship between blood pressure and blood flow in newborn preterm infants. Arch Dis Child Fetal Neonatal Ed. 2008;93(1):F29–32. https://doi.org/10.1136/adc.2006.109520.
El-Khuffash A, McNamara PJ. Hemodynamic assessment and monitoring of premature infants. Clin Perinatol. 2017;44(2):377–93. https://doi.org/10.1016/j.clp.2017.02.001.
Rios DR, Bhattacharya S, Levy PT, McNamara PJ. Circulatory insufficiency and hypotension related to the ductus arteriosus in neonates. Front Pediatr. 2018;6:62. https://doi.org/10.3389/fped.2018.00062.
Ortmann LA, Fontenot EE, Seib PM, Eble BK, Brown R, Bhutta AT. Use of near-infrared spectroscopy for estimation of renal oxygenation in children with heart disease. Pediatr Cardiol. 2011;32(6):748–53. https://doi.org/10.1007/s00246-011-9960-5.
Vidal E, Amigoni A, Brugnolaro V, Ghirardo G, Gamba P, Pettenazzo A, et al. Near-infrared spectroscopy as continuous real-time monitoring for kidney graft perfusion. Pediatr Nephrol. 2014;29(5):909–14. https://doi.org/10.1007/s00467-013-2698-y.
Elsayed Y, Amer R, Seshia M. The impact of integrated evaluation of hemodynamics using targeted neonatal echocardiography with indices of tissue oxygenation: a new approach. J Perinatol. 2017;37(5):527–35. https://doi.org/10.1038/jp.2016.257.
Dani C, Pratesi S, Fontanelli G, Barp J, Bertini G. Blood transfusions increase cerebral, splanchnic, and renal oxygenation in anemic preterm infants. Transfusion. 2010;50(6):1220–6. https://doi.org/10.1111/j.1537-2995.2009.02575.x.
Underwood MA, Milstein JM, Sherman MP. Near-infrared spectroscopy as a screening tool for patent ductus arteriosus in extremely low birth weight infants. Neonatology. 2007;91(2):134–9. https://doi.org/10.1159/000097131.
Petrova A, Bhatt M, Mehta R. Regional tissue oxygenation in preterm born infants in association with echocardiographically significant patent ductus arteriosus. J Perinatol. 2011;31(7):460–4. https://doi.org/10.1038/jp.2010.200.
Chaiyakulsil C, Chantra M, Katanyuwong P, Khositseth A, Anantasit N. Comparison of three non-invasive hemodynamic monitoring methods in critically ill children. PLoS One. 2018;13(6):e0199203. https://doi.org/10.1371/journal.pone.0199203.
Villavicencio C, Leache J, Marin J, Oliva I, Rodriguez A, Bodí M, et al. Basic critical care echocardiography training of intensivists allows reproducible and reliable measurements of cardiac output. Ultrasound J. 2019;11(1):5. https://doi.org/10.1186/s13089-019-0120-0.
Ficial B, Finnemore AE, Cox DJ, Broadhouse KM, Price AN, Durighel G, et al. Validation study of the accuracy of echocardiographic measurements of systemic blood flow volume in newborn infants. J Am Soc Echocardiogr. 2013;26(12):1365–71. https://doi.org/10.1016/j.echo.2013.08.019.
Acknowledgements
No particular acknowledgements. Preliminary results were presented in the format of a poster at PAS 2018.
Funding
There was no payment, grant or honorarium given to anyone to produce the research/manuscript.
Author information
Authors and Affiliations
Contributions
G. Altit conceived and designed the study, acquired data, analyzed and interpreted the data, as well as drafted the manuscript. G. Altit is the author that wrote the first draft. S. Bhombal conceived and designed the study, interpreted the data, as well as reviewed critically the manuscript for important intellectual content. V. Chock conceived and designed the study, interpreted the data, as well as reviewed critically the manuscript for important intellectual content. The author(s) read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
This study was approved by the institutional review board (IRB) of Stanford University. The study has been granted an exemption from requiring informed consent (by the Stanford University IRB). All methods were performed in accordance with the Declaration of Helsinki.
Consent for publication
Not applicable.
Competing interests
Gabriel Altit obtained a material grant (consisting of NIRS monitors and sensors) by Medtronic via their peer-review grant program for a study for which he is the principal investigator (ClinicalTrials.gov: NCT04106479). The other authors have no conflicts of interest related to the content of this study.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Additional file 1: Supplemental Table A.
Associations in those with and without PDA at ECHO.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.