The Journal of Automatic Chemistry 
Volume 15 (1993), Issue 3, Pages 79-84
doi:10.1155/S1463924693000136

The role of atomic fluorescence spectrometry in the automatic environmental monitoring of trace element analysis

P. B. Stockwell and W. T. Corns

P S Analytical Ltd, Arthur House, B4 Chaucer Business Park, Watery Lane, Kemsing, Kent, Sevenoaks TN15 6QY, UK

Abstract

Considerable attention has been drawn to the environmental levels of mercury, arsenic, selenium and antimony in the last decade. Legislative and environmental pressure has forced levels to be lowered and this has created an additional burden for analytical chemists. Not only does an analysis have to reach lower detection levels, but it also has to be seen to be correct. Atomic fluorescence detection, especially when coupled to vapour generation techniques, offers both sensitivity and specificity.

Developments in the design of specified atomic fluorescence detectors for mercury, for the hydride-forming elements and also for cadmium, are described in this paper. Each of these systems is capable of analysing samples in the part per trillion (ppt) range reliably and economically. Several analytical applications are described.