The Journal of Automatic Chemistry 
Volume 10 (1988), Issue 1, Pages 31-36
doi:10.1155/S1463924688000094

Evaluation of the Coulter Counter S-Plus VI

Giuseppe Banfi, Marina Pontillo, Paola Notti, and Pierangelo Bonini

Istituto Scientifico S. Raffaele, Laboratorio Analisi, Via Olgettina 60, Milano, Italy

Abstract

This article reports an evaluation of the Coulter Counter model S-Plus VI automatic analyser for haematology, and data are presented on linearity, carry-over, precision, accuracy and stability of the instrument, when compared with a model S-Plus IV/D.

The three-part differential count provided by Coulter S-Plus VI has been compared with manual eye counting. The results show a good agreement with only 2.5% of discrepancies in 2271 routine samples.

Advantages of the new instrument include: reduction of running costs, largely due to manpower saving; simple and easy use, and improved operator safety, there being no need for human contact with blood.