Paper
20 November 1985 Error Sources In The "Ring Down" Optical Cavity Decay Time Mirror Reflectometer
Thomas M. Crawford
Author Affiliations +
Proceedings Volume 0540, Southwest Conf on Optics '85; (1985) https://doi.org/10.1117/12.976129
Event: 1985 Albuquerque Conferences on Optics, 1985, Albuquerque, United States
Abstract
The "ring down" reflectometer is used to determine mirror reflectances and cavity losses in low loss situations. This is done by measuring the decay time of an optical cavity and relating this decay time to the total loss of the optical cavity. Unfortunately, this reflectometer has several sources of error that may cause problems for the unwary user. Measurement errors may arise from three different sources: (1) not having the source laser and the passive optical cavity tuned to each other properly, (2) having frequency drift in the passive optical cavity, and (3) simultaneously exciting two cavity modes that are close in frequency. In this paper, the theory of the "ring down" reflectometer is developed and analyzed for the purpose of determining its major error sources. The resulting problems are shown and solutions to these problems are suggested.
© (1985) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Thomas M. Crawford "Error Sources In The "Ring Down" Optical Cavity Decay Time Mirror Reflectometer", Proc. SPIE 0540, Southwest Conf on Optics '85, (20 November 1985); https://doi.org/10.1117/12.976129
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical resonators

Reflectometry

Mirrors

Time metrology

Clocks

Pulsed laser operation

Sensors

Back to Top