Paper
8 April 1996 Volumetric analysis of abdominal aortic aneurysm
Kevin M. Baskin, Catherine A. Kusnick M.D., Susanne Shamsolkottabi, Elvira V. Lang, J. D. Corson, William Stanford, Brad H. Thompson, Eric A. Hoffman
Author Affiliations +
Abstract
The purpose of this study was to develop a valid, reliable and accurate system of measurement of abdominal aortic aneurysms, using volumetric analysis of x-ray computed tomographic data. This study evaluates illustrative cases, and compares measurements of AAA phantoms, using standard 2D versus volumetric methods. To validate the volumetric analysis, four phantom aneurysms were constructed in a range of diameters (4.5 - 7.0 cm) which presents the greatest management challenge to the clinician. These phantoms were imaged using a Toshiba Xpress SX helical CT. Separate scans were obtained at conventional (10 mm X 10 mm) and thin slice (5 mm X 5 mm) collimations. The thin slices were reconstructed at 2 mm intervals. Data from each of the 96 scans were interpreted using a standard 2D approach, then analyzed using task-oriented volumetric software. We evaluate patient assessments, and compare greatest outer diameters of phantoms, by standard versus volumetric methods. Qualitative differences between solutions based on standard versus volumetric analysis of illustrative patient cases are substantial. Expert radiologists' standard measurements of phantom aneurysms are highly reliable (r2 equals 0.901 - 0.958; p < 0.001), but biased toward significant overestimation of aneurysm diameters in the range of clinical interest. For the same phantoms, volumetric analysis was both more reliable (r2 equals 0.986 - 0.996; p < 0.001), and more accurate, with no significant bias in the range of interest. Volumetric analysis promotes selection of more valid management strategies, by providing vital information not otherwise available, and allowing more reliable and accurate assessment of abdominal aortic aneurysms. It is particularly valuable in the presence of aortic tortuosity, vessel eccentricity, and uncertain involvement of critical vessels.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Kevin M. Baskin, Catherine A. Kusnick M.D., Susanne Shamsolkottabi, Elvira V. Lang, J. D. Corson, William Stanford, Brad H. Thompson, and Eric A. Hoffman "Volumetric analysis of abdominal aortic aneurysm", Proc. SPIE 2709, Medical Imaging 1996: Physiology and Function from Multidimensional Images, (8 April 1996); https://doi.org/10.1117/12.237875
Lens.org Logo
CITATIONS
Cited by 15 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Error analysis

Tomography

Arteries

Reliability

Analytical research

Computing systems

Surgery

Back to Top