Anode current saturation of ALD-coated Planacon® MCP-PMTs

, , , , , , , , , and

Published 3 September 2018 © 2018 IOP Publishing Ltd and Sissa Medialab
, , Citation E.V. Antamanova et al 2018 JINST 13 T09001 DOI 10.1088/1748-0221/13/09/T09001

1748-0221/13/09/T09001

Abstract

We have measured and compared the characteristics of ALD-coated Planacon® MCP-PMTs (XP85112/A1-Q-L) with their non-ALD counterparts (XP85012/A1-Q). While the latter show excellent performance, the ALD-coated sensors have surprisingly low current saturation levels (∼ two orders of magnitude lower than expected) and extremely high gain recovery time (more than 7 orders of magnitude higher than expected). We suspect that these problems might be caused by the unexpected side-effects of the ALD process. To make a definite conclusion, more samples need to be tested, preferably from different production runs. If our observation were confirmed, it would mean a serious technological setback for ALD-coated MCP-PMTs.