Abstract
ClearView™ is a single-use radiochromic 3D dosimeter designed by Modus Medical Devices Inc. Some of the dosimetric characteristics of this gel, such as response as a function of dose and as a function of energy were investigated. Moreover, a 400 mL jar was irradiated with a 1x1 cm2 field and PDD and cross-profiles were compared to those obtained with active detectors. Small volumes of gels were found to exhibit a linear response to dose in the dose range 10-80 Gy. Promising PDD and cross profiles were obtained. Artefacts obtained in high resolution for the PDD are suspected to come from small inhomogeneities in the gel.
Export citation and abstract BibTeX RIS
Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.