Evaluation of temperature rise in a tissue mimicking material during HIFU exposure

, , and

Published under licence by IOP Publishing Ltd
, , Citation S Maruvada et al 2011 J. Phys.: Conf. Ser. 279 012010 DOI 10.1088/1742-6596/279/1/012010

1742-6596/279/1/012010

Abstract

In pre-clinical testing it is essential to characterize clinical high intensity focused ultrasound (HIFU) devices using tissue-mimicking materials (TMMs) with well known characteristics, including temperature rise and cavitation properties. The purpose of this study was to monitor cavitation behavior and correlate its effect with temperature rise in a HIFU TMM containing an embedded thermocouple. A 75-μm fine wire thermocouple was embedded in a hydrogel-based TMM previously developed for HIFU. HIFU at 1.1 and 3.3 MHz was focused at the thermocouple junction. Focal pressures from 1-11 MPa were applied and the temperature profiles were recorded. Three hydrophones were used to monitor cavitation activity during sonication. A hydrophone confocal with the HIFU transducer and a cylindrical hydrophone lateral to the HIFU beam were used as passive cavitation detectors for spectral analysis of signals, and a needle hydrophone placed beyond the HIFU focus was used to record changes in the pressure amplitude due to blockage by bubbles at or near the focus. B-mode imaging scans were employed to visualize bubble presence during sonication. In a separate measurement, schlieren imaging was used to monitor the change in field distribution behind the TMM. All hydrophone methods correlated well with cavitation in the TMM.