A simulation study of the parallax effect of gaseous detectors using the Garfield program is reported. A method that mainly uses non-uniform cathode potentials to reduce the parallax error of planar type gas detectors is described. By applying it to MWPC and Micro-pattern gas detectors, the method reduces the parallax broadening with very good results. For a 13° incidence track, the width (FWHM) of the parallax broadening is reduced to less than 20% of the normal one after using the special cathode potentials.