NOTE

Radiological thickness measurement using a liquid ionization chamber electronic portal imaging device

, , , , and

Published under licence by IOP Publishing Ltd
, , Citation Philip M Evans et al 1999 Phys. Med. Biol. 44 N89 DOI 10.1088/0031-9155/44/6/403

0031-9155/44/6/N89

Abstract

We present a method of calibrating the Portal Vision electronic portal imaging device to obtain radiological thickness maps for compensator design. In this method, coefficients are derived to describe the relationship between intensity and thickness for a set of water-equivalent blocks. The effects of four parameters were studied: (a) The dose response of the system was measured and found to be describable by a square-root function. (b) The calibration data and images were taken with a wedge in situ. The effects of using different wedges and different wedge orientations were investigated. The intrinsic accuracy of the accelerator/imager system was found to be 1.9 mm, for both 15° and 30° wedges. Changing the wedge orientation between calibration and imaging and rotating the calibration coefficients accordingly led to an error of 3.5 mm. (c) The variation in detector response with gantry angle was measured and corrected. The residual error in this process was 2.4 mm. (d) The use of a model to correct the effects of imaging with different field sizes was investigated and found to yield a residual error of 2.9 mm. The overall error in image calibrations was 4 mm or 2% in dose. This is considered to be sufficiently small for the intended use of designing compensators for tangential breast irradiation.

Export citation and abstract BibTeX RIS