Abstract
In order to compare their depth effect, the line source response function at right angles to the image plane was measured using two tomographic and one section-imaging device. It was found that the full- width-at-half-maximum and modulation transfer functions do not adequately measure the capability of a device to resolve structures as a function of depth. A new quantity, called specificity, is proposed which measures the degradation of resolution of one image plane relative to an adjacent image plane as a function of frequency.